PeakTech 6205 Regulated Laboratory Power Supply Instruction Manual
- June 4, 2024
- PeakTech
Table of Contents
Unser Wert ist messbar……
PeakTech® 6205
Regulated Laboratory Power Supply
Operation manual
Safety Precautions
This product complies with the requirements of the following directives of the
European Union for CE conformity: 2014/30/EU (electromagnetic compatibility),
2014/35/EU (low voltage), and 2011/65/EU (RoHS).
To ensure the safe operation of the equipment and eliminate the danger of
serious injury due to shortcircuits (arcing), the following safety precautions
must be observed. Damages resulting from failure to observe these safety
precautions are exempt from any legal claims.
- Do not use this instrument for high-energy industrial installation measurement.
- Prior to the connection of the equipment to the mains, check that the available mains voltage corresponds to the voltage setting of the equipment.
- Connect the mains plug of the equipment only to a mains outlet with an earth connection.
- Do not place the equipment on damp or wet surfaces.
- Check test leads and probes for faulty insulation or bare wires before connection to the equipment.
- Replace a defective fuse only with a fuse of the original rating. Never short-circuit fuse or fuse holding.
- Do not cover the ventilation slots of the cabinet to ensure that air is able to circulate freely inside.
- Do not insert metal objects into the equipment by way of the ventilation slots.
- Do not place water-filled containers on the equipment (danger of short-circuiting in case of knock over of the container)
- Do not operate the equipment near strong magnetic fields (motors, transformers, etc.).
- Do not operate the meter before the cabinet has been closed and screwed safely as the terminal can carry voltage.
- Please use only 4mm-safety test leads to ensure immaculate function.
- To avoid electric shock, do not operate this product in wet or damp conditions. Conduct measuring works only in dry clothing and rubber shoes, i. e. on isolating mats.
- Never touch the tips of the test leads or probe.
- Comply with the warning labels and other info on the equipment.
- The measurement instrument is not to be operated unattended.
- Do not subject the equipment to direct sunlight or extreme temperatures, humidity, or dampness.
- Do not subject the equipment to shocks or strong vibrations.
- Keep hot soldering irons or guns away from the equipment.
- Allow the equipment to stabilize at room temperature before taking up measurements (important for exact measurements).
- Periodically wipe the cabinet with a damp cloth and mild detergent. Do not use abrasives or solvents.
- The meter is suitable for indoor use only
- Do not store the meter in a place of explosive, inflammable substances.
- Opening the equipment and service – and repair work must only be performed by qualified service personnel
- Do not place the equipment face-down on any table or workbench to prevent damaging the controls at the front.
- Do not modify the equipment in any way
- Measuring instruments don’t belong to children’s hands.-
Cleaning the cabinet
Prior to cleaning the cabinet, withdraw the mains plug from the power outlet.
Clean only with a damp, soft cloth and a commercially available mild household
cleanser. Ensure that no water gets inside the equipment to prevent possible
shorts and damage to the equipment.
Introduction
The PeakTech® 6205 is a precision-controlled laboratory power supply with high efficiency. It provides constant voltage mode, constant current mode, overvoltage protection, and overload protection function and can be operated for up to 8 hours continuously at full load. Voltage and current values are adjusted linearly. With the high stability of this power supply, it is suitable for use in schools, training, laboratory, university, and service.
2.1 Main Features
- adjustable Output 0 – 30V
- Fixed output 5V / 1A
- Linear Voltage regulation 0 – 30V
- 4-digit LED displays for Voltage and Current
- Low Ripple and Noise
- Current Output Protection
- CV/CC Mode Automatic Changer
- 8 hours of continuous operation with a full load
- Rugged Metal Cabinet
To extend the operational life span of the power supply, we recommend you limit the working time under full load to eight hours.
Controls and description
- Power-Switch
- CURRENT DISPLAY
- Output ON/OFF-KEY: After power ON, the power supply will be at a protection state (no voltage output at all terminals and OUTPUT indicator [4] is OFF). When the OUTPUT-ON/OFF key is pressed, all terminals will output the voltage as indicated on the voltage display panel and the OUTPUT indicator [4] is ON. Press the key again to cut off the output.
- OUTPUT INDICATOR
- CC – INDICATOR (CONSTANT VOLTAGE)
- CV – INDICATOR (CONSTANT CURRENT)
- VOLTAGE DISPLAY
- Voltage adjustment (COARSE)
- Voltage adjustment (FINE)
- Current adjustment (COARSE)
- Current adjustment (FINE)
- “+” TERMINAL: Positive terminal of 0-30 V adjustable output.
- “-” TERMINAL: Negative terminal of 0-30 V adjustable output.
- GND TERMINAL: This terminal is connecting to the casing and the Earth.
- “+” output terminal: Positive terminal of the 5 V fixed voltage output
- “-” output terminal: Negative terminal of the 5 V fixed voltage output
23. VENTILATION FAN
24. POWER INPUT SOCKET
25. FUSE SOCKET: 115V = 6A/250V ; 230V = 3,15A/250V
26. INPUT VOLTAGE SELECTOR:
To select the required input voltage 115/230 V AC; 50/60 Hz +/- 10%
Technical Specifications
Input Voltage | 115/230 V; 50/60 Hz (switchable); +/-10% |
---|---|
Fuse | 115 V:F 6A/250 V 230 V: F 3,15 A/ 250 V |
Output Voltage | 0 — 30 V |
Output Current | 0— 5 A |
Output Power | 150 W max. |
Display | LED-Display |
Voltage display: +/-2,0% + 2 digit Current display: +/-1,0% + 2 digit
Operating Temperature| 0°C … 40°C; < 80% RH
Storage Temperature| -10°C … + 70°C; < 80% RH
Dimensions (WxHxD)| 130 x 175 x 285 mm
Weight| approx. 6 kg
Accessories| Power cord, operation manual
This power supply needs to warm up for 30 minutes to meet the specifications.
Adjustable Output (0 – 30V)
Stability| Adjustable outputs (V):
1 x 10-4 + 3 mV (+/-10% of nominal voltage)
Adjustable outputs (A):
2 x 10-3 + 3 mA
---|---
Loading Effect
Temperature coefficient (V)
Temperature coefficient (A)| Adjustable voltage output:
<2 x 10-4 + 5 mV (I<3 A)
<2 x 10-4 + 10 mA (I>3 A)
adjustable current output:
<2 x 10-4 + 5 mA (I<3 A)
<2 x 10-4 + 10 mA (I>3 A)
150 ppm/°C
500 ppm/°C
Ripple and Noise| <1 mVrms / < 3 mArms
Overload protection| Current limitation circuit
Fixed Voltage Output
Voltage Range | 5,0 V (+/-8%) |
---|---|
Current Range | 1 A fixed |
Stability | < 5 mV |
Loading Effect | < 15 mV |
Ripple and Noise | < 15 mVrms |
Operation
5.1. Setting the Output Voltage
- Connect the power supply to the power source.
- Press the POWER SWITCH [1] to turn on the power supply.
- Use the VOLTAGE COARSE [8] and VOLTAGE FINE [9] to adjust the voltage to get the desired output voltage.
- Press Output ON/OFF-KEY [3] to activate the output and the OUTPUT INDICATOR [4] will be on.
- The CV INDICATOR [6] lights in green color.
- Connect the circuit to the TERMINALS [12, 14]
- When the CC INDICATOR [5] lights in red color, adjust the CURRENT COARSE [10] and CURRENT FINE [11] to give a suitable current.
Remarks:
- If you want to preset a desired current output before connecting to the circuit, read Section 5.2 first.
Caution:
- Make sure the INPUT VOLTAGE SELECTOR [26] is set to the correct position. Otherwise, it will damage the power supply.
5.2. Setting the Output Current
- Turn the power supply on.
- Refer to Section 5.1 steps 1-6 to give the voltage around 2 – 5V
- For CH1, turn the CURRENT COARSE [10] and CURRENT FINE [11] anticlockwise to reach the minimum current value.
- Short the + and the – TERMINALS [12,14] with a conductor whose cross-section area is not less than 0.5mm².
- Ensure the output indicator is on. Otherwise, press the Output ON/OFF-KEY [3]. Then the CV INDICATOR [6] is off and the CC INDICATOR [5] will turn to orange color.
- Adjust the CURRENT COARSE [10] and the CURRENT FINE [11] to get the desired output current.
- Repress the Output ON/OFF-KEY [3] to cut off the output.
- Then the CV INDICATOR [6] will turn to green color.
- Remove the conductor from the TERMINALS [12,14]
- Set the desired output voltage.
- Connect the circuit to the TERMINALS [12,14]
Remarks:
The conductor is not provided
Caution:
- Ensure the current is set to zero before shorting the TERMINALS. Otherwise, it will damage the power supply.
5.3. Fixed Voltage Output
- Connect the power supply to a power source.
- When the device is turned on there is a 5V fixed voltage at the output.
- Connect the circuit with the output jacks [15, 16].
Note:
The fixed voltage output is not stabilized. If the maximum output current of
1A is exceeded, the output voltage of 5V can not be kept stable and collapses.
Fuse Replacement
Caution:
- Ensure that no power is connected to the power supply; otherwise, electrical shock may occur.
- Do not apply excessive force on the fuse socket, or it may be damaged.
-
Remove all test leads from the output sockets and disconnect the power cord from the input socket.
-
Locate the fuse socket at the rear panel power socket.
-
Replace the fuse with an identical rating.
Fuse: 115 V = 6 A/250 V 5 x 20 mm; 230 V = 3,15 A/250 V 5 x 20 mm -
Reinstate the fuse socket. (Push in the fuse socket to the power socket.)
All rights, also for translation, reprinting, and copy of this manual or parts
are reserved. Reproductions of all kinds (photocopy, microfilm, or other) only
by written permission of the publisher.
This manual is according to the latest technical knowledge. Technical
alterations reserved.
We herewith confirm that the units are calibrated by the factory according to
the specifications as per the technical specifications.
We recommend calibrating the unit again, after 1 year.