Wahl HSICBB-P Portable Calibration Black Body User Manual
- June 13, 2024
- WAHL
Table of Contents
Wahl HSICBB-P Portable Calibration Black Body
Application
- The Wahl HSICBB-P is a portable ambient temperature black body used for single point calibration and checking of Infrared Thermal Imaging Cameras and Point Infrared Thermometers. It is housed in a watertight IP67 carrying case.
User Controls
- Scale switch – front panel switch is used to select between °F and °C scale. Changing the switch position will result in a change of the scale and displayed temperature on the next display update cycle.
- Power switch – front panel switch is used to power on or off the HSICBB-P.
Operation
- Open enclosure by pulling up on the two side latches.
- Set scale switch to desired scale.
- Set power switch to on position, indicated by “I”.
- Unit will go through self-check and momentarily display all display segments.
- Unit will momentarily blank the display followed by displaying the blackbody’s current temperature.
The unit will update the black body temperature at approximately ten-second intervals.
Typical Calibration Technique and procedure
Stabilization – It is important that both the IR measuring device (Unit Under Test – UUT) being calibrated and the HSICBB-P Calibration Black Body are thermally stable. They should be in a stable (± 4°F) ambient temperature for a minimum period of 30 minutes. This will result in a more accurate calibration check.
- Turn on IR imager or thermometer and HSICBB-P and allow them to stabilize 30 minutes minimum.
- Make certain that if the UUT has variable emissivity, it should be set to .95.
- Aim the UUT at the blackbody target at a distance so that it fills the spot size or measurement area of the UUT.
- Compare and record the Blackbody’s display reading to the UUT reading.
- Document readings as required.
Error Codes
- Cbl1, Cbl2, Cbl3 and Cbl4 indicate open cable or open sensor.
- Low Batt – Icon in bottom right corner of display indicates battery voltage is low.
- “ ————— “ Indicates battery below level for reliable reading.
- Replace battery. For S/N prior to 37105.xxx, use DSA3062 “C” size battery. For S/N after 37105.xxx, use battery #12234-03, “AA” size battery.
Battery Replacement
- When Low Batt Icon or “—-” are displayed replace the battery.
- For S/N prior to 37105.xxx use Wahl P/N DSA3062, “C” size, 3.6Volt, Lithium Thionyl Chloride Battery.
- For S/N after 37105.xxx or S/N that are formatted as “##-#####.###” (example: 19-
- 12345.001) or later use Wahl P/N 12234-03, “AA” size, 3.6V, Lithium Thionyl Chloride Battery.
- The proper replacement battery part number is indicated on the inside of the door.
To replace the battery
- Unlatch and open the door.
- Make sure the unit is powered off.
- Remove the 10 screws with plastic washers, setting them aside.
- Lift the panel out of the housing, noting the orientation.
- On the PC Board mounted to the rear of the panel, remove battery from holder.
- Install the new battery, observing the polarity markings on the battery holder.
- Reinstall the panel, in the proper orientation.
- Reinstall the 10 screws with washers.
- Turn the power switch to the on position and verify the low batt warning has gone off.
Specifications
Radiator | 3” diameter concentric rings with integral 4-wire RTD sensor |
---|---|
Meter | 4-wire RTD thermometer |
Ambient Operating Range | -40° to 158°F (-40°to 70°C) |
Recommended Operating Range | 50° to 122°F (10°to 50°C) |
System Accuracy | ± 0.3°F or ± 0.2°C, over Recommended Operating range |
Scale | °F or °C, User Selectable |
Display | 4-digit 1” high LCD |
Display Resolution | 0.1° |
Display Update Rate | 10 seconds |
Repeatability | 0.1° F or C |
Heating Method | None, assumes the ambient temperature |
Emissivity | 0.95 or ± 0.02 |
Power | 1 – 3.6 volt Lithium Thionyl Chloride Battery, size “AA” |
Battery Life | Approximately 3 years, with “Low Bat” indicator (turning |
off unit when not in use will extend battery life)
Case / Case Rating| Pelican® 1150 / IP67 (1 meter submersion for 30
minutes) with case closed
Case Exterior Dimensions /
Weight
| 9.44″ L x 7.80″ W x 4.29 D” (24 x 19.8 x 10.9 cm) / 3.0 lbs. (1.36 kg)
Included| User Manual (NIST Certificate of Conformance, optional)
Service
- For calibration, service or technical support, contact:
- Palmer Wahl
- 234 Old Weaverville Road Asheville, NC 28804
- Ph.: 800-421-2853 (US only) 828-658-3131
- Fax: 828-658-0728
- Web: www.palmerwahl.com.
234 Old Weaverville Road, Asheville, NC 28804 800-421-2853 • 828-658-3131 • 828-658-0728: www.palmerwahl.com. info@palmerwahl.com.
Palmer Wahl
- 234 Old Weaverville Road
- Asheville, NC 28804
- Toll Free 800-421-2853
- Phone 828-658-3131
- Fax 828-658-0728
- www.palmerwahl.com.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>