PRECAUTIONS FOR USE
The protection assured by the current sensor can be compromised if it is used in a way that is not recommended by the manufacturer.
- Comply with the rated maximum voltage and current and the measurement category. Do not use your current sensor on networks where the voltage or category exceeds those stipulated.
- Comply with the conditions of use, that is to say temperature, humidity, altitude, degree of pollution and location of use.
- Do not use the current sensor if its housing is open, deteriorated, or incorrectly reassembled. Before each use, check the integrity of the insulation of the core or of the arms of the clamps, of the housing, and of the leads.
- Do not subject the current sensor to sprayed or falling water or other liquids.
- Keep the gap of the clamp absolutely clean.
- Use suitable personal protective equipment when parts at hazardous voltages may be accessible in the installation where the measurement is made.
- Any repairs must be carried out by accredited skilled personnel.
PRESENTATION
CLAMPS MN93 AND MN93A
CLAMP C193
CLAMP PAC93
MINIFLEX® MA193
MINIFLEX® MA194
MINIFLEX® MA196
The connector on this sensor can be used to form a seal with the terminal to
which it is connected, if the instrument has a threaded receptacle. Simply
screw the ring on the connector onto the receptacle.
AMPFLEX® A193
AMPFLEX® A196A
The connector on this sensor can be used to form a seal with the terminal to
which it is connected, if the instrument has a threaded receptacle. Simply
screw the ring on the connector onto the receptacle.
DELIVERY CONDITION
Clamp MN93
Supplied in a cardboard box with:
- a user manual in 5 languages,
- a verification certificate.
Clamp MN93A
Supplied in a cardboard box with:
- a user manual in 5 languages,
- a verification certificate.
Clamp PAC93
Supplied in a cardboard box with:
- a 9V battery,
- a user manual in 5 languages,
- a verification certificate.
Clamp C193
Supplied in a cardboard box with:
- a user manual in 5 languages,
- a verification certificate.
MiniFlex® MA193 250 mm sensor
Supplied in a cardboard box with:
- a user manual in 5 languages,
- a verification certificate.
MiniFlex® MA193 350 mm sensor
Supplied in a cardboard box with:
- a user manual in 5 languages,
- a verification certificate.
MiniFlex® MA194 250 mm sensor
Supplied in a cardboard box with:
- a user manual in 5 languages,
- a verification certificate.
MiniFlex® MA194 350 mm sensor
Supplied in a cardboard box with:
- a user manual in 5 languages,
- a verification certificate.
MiniFlex® MA194 1000 mm sensor
Supplied in a cardboard box with:
- a user manual in 5 languages,
- a verification certificate.
MiniFlex® MA196 350 mm sensor
Supplied in a cardboard box with:
- a user manual in 5 languages,
- a verification certificate.
AmpFlex® A193 450 mm sensor
Supplied in a cardboard box with:
- a user manual in 5 languages,
- a verification certificate.
AmpFlex® A193 800 mm sensor
Supplied in a cardboard box with:
- a user manual in 5 languages,
- a verification certificate.
AmpFlex® A196A 610 mm sensor
Supplied in a cardboard box with:
- a user manual in 5 languages,
- a verification certificate.
DELIVERY CONDITION
- A main adapter for the PAC93.
For the accessories and spares, consult our web site: www.chauvin- arnoux.com
USE
Current clamps, MiniFlex® and AmpFlex® flexible current sensors are used to
measure the current flowing in a cable without opening the circuit. They also
insulate the userfrom dangerous voltages in the circuit.
The choice of current sensor to be used depends on the current to be measured
and the diameter of the cables.
| MN93| MN93A 5 A| MN93A 100 A
---|---|---|---
Measurement range | 0.05 – 240 A| 0.005 – 6 A| 0.1 – 120 A
Clamping diameter| 20 mm| 20 mm| 20 mm
| C193| PAC93
---|---|---
Measurement range | 1 – 1.200 A| 1 – 1.200 A
Clamping diameter| 52 mm| 39 mm
| MiniFlex ® MA193 250 mm| MiniFlex ® MA193
350 mm| MiniFlex ® MA196 350 mm
---|---|---|---
Measurement range | 100 mA – 10 000 A| 100 mA – 10 000 A| 100 mA – 10 000 A
Clamping diameter| 70 mm| 100 mm| 100 mm
| MiniFlex ® MA194 250 mm| MiniFlex ® MA194
350 mm| MiniFlex ® MA194 1000 mm
---|---|---|---
Measurement range | 100 mA – 10 000 A| 100 mA – 10 000 A| 100 mA – 10 000 A
Clamping diameter| 70 mm| 100 mm| 320 mm
| AmpFlex ® A193 450 mm| AmpFlex ® A196A 610
mm| AmpFlex ® A193 800 mm
---|---|---|---
Measurement range *| 100 mA – 10 000 A| 100 mA – 10 000 A| 100 mA – 10 000 A
Clamping diameter| 140 mm| 190 mm| 250 mm
The measurement ranges are those of the sensors. In some cases, they may differ from the ranges that can be measured by the analyser with which they are associated.
- For three-phase measurements, start by marking the accessories using the coloured rings supplied with the device: a different colour for each current terminal.
- Connect the current sensors to the current terminals of the device.
For the PAC93:
-
Set the switch to “1 mV/A”; the “ON” indicator lights.
-
Once the PAC93 is connected to the device, adjust the zero by turning the potentiometer with no conductor in the jaws of the clamp.
-
At the end of the measurement, remember to switch off the PAC93.
For the clamps:
-
Press the trigger on the clamp to open the jaws.
-
Then clamp the cable that is to be measured. Where possible, the cable should be centred in the jaws of the clamp.
-
Release the trigger.
For the MiniFlex® models:
-
Press on the yellow opening device to open the flexible coil.
-
Open it and place it around the conductor carrying the current to be measured (only one conductor in the sensor). Where possible, the cable should be centred in the sensor.
-
Close the coil. You must hear the click.
For the AmpFlex® models:
-
Press simultaneously on both sides of the opening device.
-
Then clamp the cable that is to be measured. Where possible, the cable should be centred in the sensor.
-
Close the sensor by forcing the moving part into the connector until you hear a click
For details of the measurement configuration and technical characteristics, refer to the user manual of the analyser with which you are using the current sensor.
CHARACTERISTICS
ENVIRONMENTAL CONDITIONS
Indoor use.
Operating range -10 °C to +55 °C and 10 % to 85% RH
Storage range -40 °C to +70 °C and 10 % to 90% RH
Degree of pollution 2
Altitude < 2000 m
CONSTRUCTION SPECIFICATIONS
MN93 and MN93A
- Dimensions (L x W x H) 135 x 51 x 30 mm
- Cord 3.50 metres long
- Weight approx. 290 g
C193
- Dimensions (L x W x H) 216 x 111 x 45 mm
- Cord 3.50 metres long
- Weight approx. 690 g
PAC93
- Dimensions (L x W x H) 236.5 x 97 x 44 mm
- Cord 3.50 metres long
- Weight approx. 540 g
- Battery life 120 hours with an alkaline battery
MiniFlex® MA193 250 mm
- Dimensions (L x W x H) approx. 130 x 80 x 12 mm
- Cord 3 metres long
- Weight approx. 110 g
MiniFlex® MA194 250 mm
- Dimensions (L x W x H) approx. 130 x 80 x 18 mm
- Cord 3 metres long
- Weight approx. 110 g
MiniFlex® MA193 350 mm
- Dimensions (L x W x H) approx. 170 x 110 x 12 mm
- Cord 3 metres long
- Weight approx. 120 g
MiniFlex® MA194 and MA196 350 mm
- Dimensions (L x W x H) approx. 170 x 110 x 18 mm
- Cord 3 metres long
- Weight approx. 120 g
MiniFlex® MA194 1000 mm
- Dimensions (L x W x H) approx. 370 x 320 x 18 mm
- Cord 3 metres long
- Weight approx. 160 g
AmpFlex® A193 450 mm
- Dimensions (L x W x H) approx. 170 x 158 x 25 mm
- Cord 3 metres long
- Weight approx. 220 g
AmpFlex® A193 800 mm
- Dimensions (L x W x H) approx. 280 x 265 x 25 mm
- Cord 3 metres long
- Weight approx. 270 g
AmpFlex® A196A 610 mm
- Dimensions (L x W x H) approx. 215 x 200 x 25 mm
- Cord 3 metres long
- Weight approx. 250 g
Protection index
- IP 40 for the clamps and IP 30 jaws open, according to IEC 60529
- IP 50 for the MiniFlex® MA193 and for the AmpFlex® A193 according to IEC 60529 in operation
- IP 67 for the MiniFlex® MA196 and MA194, and the AmpFlex® A196A according to IEC 60529 when the ring is screwed on
- IK 04 according to IEC 50102
Drop test: according to IEC 61010-1
CONFORMITY TO INTERNATIONAL STANDARDS
Electrical safety according to IEC 61010-2-032.
Maximum appliable voltage:
- MN93 and MN93A: 600 V Cat III or 300 V Cat IV
- C193: 1000 V Cat III or 600 V Cat IV
- PAC93: 600 V Cat III or 300 V Cat IV
- MiniFlex® : 1000 V Cat III ou 600 V Cat IV
- AmpFlex®: 1000 V Cat III or 600 V Cat IV
Type of current sensor as per IEC 61010-2-032:
- MN93 et MN93A : type A
- C193 : type A
- PAC93 : type A
- MiniFlex® : type B
- AmpFlex® : type B
MAINTENANCE
Except for the batteries, the instrument contains no parts that can be replaced by personnel who have not been specially trained and accredited. Any unauthorized repair or replacement of a part by an “equivalent” may gravely impair safety.
CLEANING
Disconnect the current sensor completely. Use a soft cloth, dampened with
soapy water. Rinse with a damp cloth and dry rapidly with a dry cloth or
forced air. Do not use alcohol, solvents, or hydrocarbons. Keep the air gaps
of the clamps perfectly clean. See to it that no foreign body interferes with
the operation of the snap lock device of the sensor.
REPLACING THE BATTERY OF THE PAC93
-
Disconnect the PAC93 completely and turn the rotary switch to OFF.
-
Turn the PAC93 over.
-
Use a screwdriver to unscrew the screws and remove the battery compartment cover.
-
Withdraw the battery from its compartment without pulling on the wires.
-
Disconnect the old battery without pulling on the wires and connect the new one in its place, observing the polarity.
Spent batteries must not be treated as ordinary household waste. Take them to the appropriate recycling collection point.
- Put the battery into its compartment.
- Put the cover back in place and screw the screws back in.
WARRANTY
Except as otherwise stated, our warranty is valid for 24 months starting from thedate on which the equipment was sold. Extract from our General Conditions of Sale provided on request.
The warranty does not apply in the following cases:
- Inappropriate use of the equipment or use with incompatible equipment;
- Modifications made to the equipment without the explicit permission of the manufacturer’s technical staff;
- Work done on the device by a person not approved by the manufacturer;
- Adaptation to a particular application not anticipated in the definition of the equipment or not indicated in the user’s manual;
- Damage caused by shocks, falls, or floods
FRANCE
Chauvin Arnoux
12-16 rue Sarah Bernhardt
92600 Asnières-sur-Seine
Tél : +33 1 44 85 44 85
Fax : +33 1 46 27 73 89
info@chauvin-arnoux.com
www.chauvin-arnoux.com
INTERNATIONAL
Chauvin Arnoux
Tél : +33 1 44 85 44 38
Fax : +33 1 46 27 95 69
Our international contacts
www.chauvin-arnoux.com/contacts
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>