ardo Amaryll Manual Breast Pump Instruction Manual
- June 12, 2024
- ardo
Table of Contents
ardo Amaryll Manual Breast Pump Instruction Manual
Package contents
Amaryll manual breastpump:
a.) Lever
b.) Casing cover
c.+d.) Membrane pot with membrane con rod
e.) 26 mm breast shell
f.) 1 lip valve (+ 1 spare)
g.) 150 ml bottle
h.) Bottle stand
i.) Cap
j.) Cap liner
k.) Cleaning brush
l.) 31 mm breast shell
m.) 28 mm breast shell insert
n.) 26 mm Optiflow massage insert
Amaryll Start: from a to j
Amar yll Kombi kit: from a to d
Indications
The Amaryll manual breast pump is used for expressing, collecting and storing breast milk.
Instructions for use
Assembly
-
Assemble according to the illustration
-
Place the membrane pot securely over the groove of the membrane con rod. Check that the smooth side of the membrane pot faces outwards.
-
Put on the casing cover in the desired position:
-
Standard position
-
Lever position left side
-
Lever position right side
The mark on the casing cover must lie on top of the mark on the breast shell. Then rotate the casing cover clockwise. -
The lever is mounted on the membrane con rod and by means of light pressure down wards clicks into the casing cover.
Expressing milk
-
Wash your hands, sit down comfortably and relax.
-
Place the casing cover in the desired position and tighten it.
-
Centre the nipple so that when pumping it moves freely in the breast shell/breast shell insert and pump by activating the lever.
-
Adjusting the suction intensity (vacuum): The further you press the lever, the greater the suction intensity . 1
-
Adjusting the speed interval: The faster you move the lever up and down, the higher the pumping frequency. 2
-
Our recommendation: Begin at a high speed interval and at low suction intensity in order to stimulate the milk ejection reflex. Once the milk is flowing, reduce the speed interval and increase the suction intensity
Choice of breast shell and Opti flow massage insert
If your nipple rubs uncomfortably on the inner side of the funnel, use a larger breast shell/breast shell insert. If the areola is also sucked into the funnel, use a smaller breast shell. The Optiflow massage insert can only be used with the 31 mm breast shell.
28 mm breast shell insert
Opti flow
![side
of the funne](https://manuals.plus/wp- content/uploads/2023/08/Screenshot_9-272.png)
Storage of breast milk
Recommended storage time at room temperature: 3–4 hours at 16–29°C (60–85°F); in the refrigerator: 3 days at ≤4°C (39°F); in freezing compartment with door: 6 months to 1 year at <-17°C (0°F)
Cleaning / sterilising
At home: After each use, wash with washing up liquid and rinse well. Wash disassembled once a day in the dishwasher and cover. Do not use a brush to clean the valve. The lever, casing cover and bottle stand may also be washed in the dishwasher, but this is not required.
This product can be washed and re-used 40 times.
In hospitals: Disassemble and sterilise at 121°C (250°F) for 20 minutes /
134°C (273°F) for 5 minutes.
The lever, casing cover and bottle stand can, but need not, be sterilized. Can
be disinfected by wiping.
Possible problems and troubleshooting Weak or no vacuum:
- Ensure that the Amaryll manual breast pump has been correctly assembled and is not damaged.
- Check whether the lip valve is cracked or stuck together.
- Make sure that the casing cover seals well.
Pain when pumping:
- Ensure that the nipple is in the middle of the breast shell.
- Check that the breast shell size fits your nipples.
- If problems continue seek professional help.
Contraindications
None known.
Safety regulations: Only use after cleaning or sterilization. After
cleaning check for any damage.
General information
| Materials: Polypropylene (PP) / silicone
---|---
| Temperature limit: Operation -10°C (14°F) to +40°C (104°F) / Transport -20°C
(-4°F) to +50°C (122°F)
| Disposal: Can be safely disposed of in compliance with local regulations.
| Fragile, handle with care
| Humidity limitation: 20–95%
| Atmospheric pressure limitation: 900–1060 hPa
| Keep away from sunlight
| Keep dry
Preferably store Optiflow in a dark place (silicone).
0123 Class IIa
Ardo medical AG
Gewerbestrasse 19
6314 Unterägeri
Switzerland
T : +41(0)41 754 70 70
F : +41(0)41 754 70 71
Email : info@ardo.ch
Website :www.ardomedical.com