Ardo Medical Carum, Ardo Bellis and Calypso Owner’s Manual

June 12, 2024
Ardo Medical

Medical
Reprocessing the Ardo breastpumps for rental centers
Carum, Ardo Bellis and CalypsoArdo Medical Carum, Ardo Bellis and
Calypso

Carum, Ardo Bellis and Calypso

To safeguard the hygiene of the Ardo breastpumps Ardo Bellis, Carum and Calypso, Ardo medical AG recommends that the surface of the breastpumps and the Carum packaging box be reprocessed immediately after return. The pump surface is reprocessed using a non-phenol disinfectant.
Abrasive agents should not be used. If pumpsets or other accessories have been returned, they must be disposed of because they are intended for personal use.

Important notes

Every time the device is cleaned, it must be switched off, and the power cable must be disconnected from the power supply. Failure to do so may result in an electric shock!
If the pump is contaminated by overflow or another cause, the contaminated parts must be disposed of by the user according to local guidelines.
Do not use cleaners or disinfectants based on phenol! Failure to follow this instruction may destroy the plastic materials due to the formation of fine cracks.

Reprocessing

The device must be cleaned and disinfected each time before it is supplied to a customer. This also applies for the first use after delivery because none of the Ardo breastpumps are supplied disinfected.
Please also comply with the legal provisions and hygiene regulations that apply in your country.

Work station

Unpack and reprocess the Ardo breastpumps at a work station that can be disinfected by wiping down. Keep processed Ardo breastpumps in a separate location to prevent contact with pumps that have not been cleaned and disinfected.

Preparation

  • Prepare a ready-to-use disinfectant that is compatible with the material (see below). Read the product information on the product and check the expiry date before use.
  • For your own protection, we recommend that you wear gloves. Please note the safety data sheet for the disinfectant used.
  • Open the packaging and take out the Ardo breastpump. Dispose of the accessories according to country-specific guidelines.

Operation

Disconnect the Ardo breastpump from the power supply.

  1. Before disinfecting the breastpump, remove any visible soiling from the surface of the Ardo breastpump and its power cord using a damp cloth and a commercially available mild detergent.
  2. Do not allow cleaning fluid to penetrate the connection openings for the pumpset or the electrical connections.
  3. Do not clean the pump under running water or in a water bath.
  4. After cleaning, dry the surfaces with a clean, dry cloth.
  5. Reprocess the surface of the breastpump using a disinfectant that is compatible with the materials (see below). Allow the disinfectant to take effect as described in the manufacturer’s instructions.
  6. Disinfect your work surface after reprocessing the Ardo breastpumps.

Reprocessing the Ardo Bellis breastpump Additional information for Ardo Bellis
The Memory function must be reset before the next leasing.

Possible cleaners

Product name Manufacturer / supplier Ingredients

Surface disinfection / aldehyde-free
Biguanid Roche N| Schumacher / Dengg| Quaternary compound
Biguanid Roche NR| Schumacher / Dengg| Quaternary compound
Surface disinfection / oxidising agent
Perform| SchOtke+| Peroxide compound
Surface disinfection / alcohol-free
mikrozid® sensitive liquid| SchCilke+| Biocide

Reprocessing the Ardo Carum breastpump Additional information for the Carum container
The Carum container can be cleaned with a conventional detergent.
Possible cleaners

Product name Manufacturer / supplier Ingredients

Surface disinfection / aldehyde-free
Biguan d Flache| Schumacher / Dengg| Quaternary compound
Incidin Plus| Ecolab| Alkylamine
TPH protect| Schilke+| Quaternary compound
Surface disinfection / aldehyde-containing
Antiseptica Combi Surface| Antiseptics| Quaternary compound / aldehyde
Surface disinfection / ready-to-use / alcohol-resistant surfaces
Incidin liquid| Ecolab| Alcohols
Mikrozid AF liquid| SchUlke+| AIcohols
Surface disinfection / ready-to-use / non-alcohol-resistant surfac
Acryl Des| SchUlke+| Quaternary compound
Bacillol 30 Foam| Hartmann / Bode| Alcohols / alkylamine derivative
Surface disinfection / disinfectant wipes
Kohrsolin extra tissues| Hartmann / Bode| Aldehyde / quaternary compound
Mikrobac Tissues| ortmann / Bode| Quaternary compound
Sanicloth active wipes| Ecolab| Quaternary compound
Surface disinfection / oxidising agent
Descogen liquid| Antiseptics| Peroxide compound
Incidin Active| Ecolab| Peroxide compound
Perform| Schilke+| Peroxide compound

Reprocessing the Ardo Calypso breastpump

Ardo Medical Carum, Ardo Bellis and Calypso - Additional
2 Additional information for Calypso
For the Calypso, check that there are no batteries in the compartment. If there are batteries, remove them and dispose of them according to standard guidelines.
Possible cleaners

Product name Manufacturer / supplier Ingredients

Surface disinfection / aldehyde-free
Biguan d Flache| Schumacher / Dengg| Quaternary compound
Incidin Plus| Ecolab| Alkylamine
TPH protect| SchCilke+| Quaternary compound
Surface disinfection / aldehyde-containing
Antiseptica Combi Surface| Antiseptics| Quaternary compound / aldehyde
Surface disinfection / ready-to-use / alcohol-resistant surfaces
Incidin liquid| :icolab| Alcohols
Mikrozid AF liquid| 13chUlke+| Alcohols
Surface disinfection / ready-to-usc / non-alcohol-resistant surfaces
Acryl Des| SchUlke+| Quaternary compound
Bacillol 30 Foam| Hartmann / Bode| Alcohols / alkylamine derivative
Surface disinfection / disinfectant wipes
Kohrsolin extra tissues| Hartmann / Bode| Aldehyde / quaternary compound
Mikrobac Tissues| dartmann / Bode| Quaternary compound
Sanicloth active wipes| Icolab| Quaternary compound
Surface disinfection / oxidising agent
Descogen liquid| Antiseptics| Peroxide compound
Incidin Active| Ecolab| Peroxide compound
Perform| SchUlke+| Peroxide compound

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals