UV monitor CHS9AM Series Monitoring Radiometer User Manual
- June 12, 2024
- UV monitor
Table of Contents
UV monitorCHS9AM series
Operation manual
V1.12
Appearance of monitor introduction
- UV monitor CHS9AM series
Mode explanation
Basic function description
- Mega-functional MCU built-in controller with high steady
- MIN 48*94 standard contorl plate
- 5 units (0.56 inches )LED display
- Output industrial standard signals & Convenient used for embeded system
- Easily operation for users
- Input voltage range :85–265V AC
- UV lamp lighting time accumulated automatically
- Lighting intensity attenuate 50% off alarming
Technology parameter
Voltage:85-265VAC 50/60Hz| Measurement range: 0~19999uW/cm2
Mesurement accuracy :±(0.2%FS±1bite)| Input voltage:linar 0-5VDC
Output specific(linar):
①Output voltage (isolated):1-5VDC
②Output current (isolated):4-20mA
③Passive delay output (resistor load): 250VAC 3A/30VDC 3A
| Display solution: 5units LED
Size:4896105
Consumption:≤5W
Installation mode:embeded| Weight : 420 g
Working temperature:0~+50℃| Working humidy:35~85%RH
Herein outputs analog signals inside connected same AGND without isolated.
Safety notice
Warning
- Must fixed/debug under expert or professional stuff
- Cut down the power for the unit,then do installation.
Herein outputs analog signals inside connected same AGND without isolated.
Analog signals(4-20mA/ 1~5VDC output value corresponding to the relative value
of UV 0~100%).
Blue cable:Vin-18 /Brown cable:5VDC-19 /Black cable:GND-20
Port No. | Name | Function | Note |
---|---|---|---|
1、2 | POWER |
POWER INPUT PORTS
|
4、6| NC| NORMAL CLOSED|
5、6| NO| NORMAL OPEN|
7、8| 1-5VDC|
VOLTAGE OUTPUT PORTS
| Port7-GND
9、10| 4-20mA|
CURRENT OUTPUT PORTS
| Port9-GND
18| Vin|
SIGNALS INPUT FOR SENSOR
|
19、20| 5VDC/GND| SENSOR POWER CABLE| POWER PROVIDED BY APPARATUS
Symbol description
Symbol | Item | Operation | Adjust Range | Note |
---|---|---|---|---|
C-00 | UV |
relative value
| Enter in the parameters setting mode,keep pressing the “SET” button lasts 3S.| 100%| Set the UV value as whole referance value in order to monitor the UV lamp attenuation.
C-01
| C-00| Press “SET”enter in the parameters indicate mode,repress SET”button
enter in parameters setting mode .Press the “page shift set “button to adjust
the unit value .Press the “shift left” button to change the unit needs to
adjust .
Setting over ,press the “SET”button lasts for 3s,start device memory and back
to standby mode .|
0-100%
| According the actual requirement , set the lamp low limited warming line .In
case the inspected value lower than last value ,warming up and the delay be
actived .
C-02| C-00| Page turned to C-02 parameters setting mode , press the
“SET”button enter in the parameters indicate mode.Then press the “SET”button
enter in the parameters setting mode ,details as same reason as C-01setting .|
0-19999Hours| After set the lamp using time . In actually run time ,when the
totally run time exceeds the original setting time .
Warning up and delay be actived .
OPERATION DESCRIPTION
-
FACE PALTE EXPLANATION
-
Description:
2.1)Press the “page shift set “button ,switchover the display . When switch on the parameters setting plate ,C0-C4 would shown .Then press the “set”button ,then will show nowtime parameters values .Re-press the “set” , enter in the parameters setting mode .The value you setting will falshing ,press the “page set button”to adjust –add or sub.Use the “left shift” button to change the setting unit .After all setting ,press the “SET” button 3s to quit setting mode and return tovalue-shown mode .Press the “page shift set “button return to code-shown mode .Details value names are on the plate .
2.2) Convenient for client to zero clearing setting and reset setting operation .The apparatus can up to those functions .
①zero clearing setting enter in the “run time”mode ,press the “SET” button 3S
②reset setting if you need to set thr “UV relative value “up to 100% .enter in the “C-00″mode ,then press the “SET”button and dis-release into the value- shown mode,then press the “SET” button and dis-release into the 100%value- setting mode, three indicators flashing at this time ,press the “SET”button 3S. After above ,the show value set to 100% well and turn back to normal mode .
Buttons fuctional operation explanation
SET: Under the parameter settingm mode,valid in UV running mode ,keep pressing
3S to zero clearing.
shift left : Under the parameter settingm mode ,change the unit need to adjust
.the flashing unit can be change.”shift left” button is circle run .Invalid in
other modes .
page shift left:
①under the parameter settingm mode,press the button to change the show mode
just as parameters or codes modes .
②under the parameter settingm mode,press the button back to parameters code
mode .
③under the parameter settingm mode,press the button back to parameters value
mode . combination :
under the parameter settingm mode,press the “SET” +”page shift left”buttons
can back to parameters code mode without record the setting parameters .
Display plate and operation processing
A–press the “page shift left”button to circle display the interface step ① to⑨
Parameters adjustment steps (for parameters C-00 ):
B–Step ① to ⑦ for C-00 parameters setting
Parameters adjustment steps (for parameters C-01 ):
C–Step① to ⑦ for C-01 parameters setting
Parameters adjustment steps for C-02 same as above C-01 setting
steps.
Function settings
Symbol | Item | Description | Defects elimination |
---|---|---|---|
E-01 | UV low limited |
warning
| As the UV relative value down to 50% of original seetings , device be
warning .| 1D. etector may be dirty, please do clearing
2.The intensity of illumination may be weak ,please replace the lamp.
E-02| UV lighting period exceeded warning| As the UV run time exceed to
original seetings run-time , device be warning .| **** Replace lamp
Defective codes explanation
Symbol | Item | Operation | Adjust Range | Note |
---|---|---|---|---|
C-00 | UV |
relative value
| Enter in the parameters setting mode,keep pressing the “SET” button lasts 5S.| 100%| Set the UV value as whole referance value in order to monitor the UV lamp attenuation.
C-01
| C-00| Press “SET”enter in the parameters indicate mode,repress SET”button enter in parameters setting mode .Press the “page shift set “button to adjust the unit value .
Press the “shift left” button to change the unit needs to adjust . Setting over ,press the “SET”button lasts for 5s,start device memory and back to standby mode .
| ****
0-100%
| ****
According the actual requirement ,set the lamp low limited warming line .In case the inspected value lower than last value ,warming up and the delay be actived .
C-02| C-00| Page turned to C-02 parameters setting mode , press the “SET”button enter in the parameters indicate mode.Then press the “SET”button enter in the parameters setting mode ,details as same as C-01setting .| 0-19999Hours| After set the lamp using time . In actually run time ,when the totally run time exceeds the original setting time .Warning up and delay be actived .
Dimension of appearance
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>