PCE Instruments PCE-2500N Portable Pen-sized Durometer for Metals User Manual

June 12, 2024
PCE Instruments

PCE Instruments PCE-2500N Portable Pen-sized Durometer for Metals

PCE-Instruments-PCE-2500N-Portable-Pen-sized-Durometer-for-Metals-
product

Product Information

The PCE-2500N/PCE-2600N Durometer is a hardness tester designed for measuring the hardness of various materials. It utilizes the LEEB method of measurement, which calculates the hardness value based on the rebound velocity of an impact body. The device features a 128*64 OLED display for easy reading, a micro USB storage device for direct connection, and a memory capacity for 600 average values and 6 files. It also includes a limit alarm and an auto power-off function when the battery is low. The device is powered by a rechargeable Li- Ion battery and comes with a carrying case for portability.

Applications for the PCE-2500N/PCE-2600N Durometer include testing the hardness of ball bearings and other parts, fault analysis of pressure vessels, steam turbine generators and plants, heavy workpieces, mechanical and permanently installed parts, and limited space environments.

Product Usage Instructions

  1. Read the user manual carefully and completely before using the device for the first time.
  2. The device should only be used by qualified personnel and repaired by PCE Instruments personnel.
  3. Ensure that the battery is fully charged before use.
  4. Connect the micro USB storage device directly to the device if needed.
  5. Press the power button to turn on the device.
  6. Use the buttons on the device to navigate through the menu options.
  7. To measure the hardness of a material, select the appropriate impact device based on the material being tested.
  8. Hold the device against the material surface and press the measurement button to take a reading.
  9. The hardness value will be displayed on the OLED display.
  10. If necessary, use the conversion charts provided to convert the hardness value into other standard hardness units.
  11. After use, turn off the device and store it in the carrying case for safekeeping.
  12. Refer to the user manual for calibration instructions, battery replacement, and additional information.

If you have any questions or need further assistance, please contact PCE Instruments using the contact details provided in the user manual.

Safety notes

Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non- observance of the manual are excluded from our liability and not covered by our warranty.

  • The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
  • The instrument may only be used if the environmental conditions (temperature, relative humidity, …) are within the ranges stated in the technical specifications. Do not expose the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
  • Do not expose the device to shocks or strong vibrations.
  • Avoid dust, strong magnetic fields, oil, etc.
  • The case should only be opened by qualified PCE Instruments personnel.
  • Never use the instrument when your hands are wet.
  • You must not make any technical changes to the device.
  • The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner, no abrasives or solvents.
  • The device must only be used with accessories from PCE Instruments or equivalent.
  • Before each use, inspect the case for visible damage. If any damage is visible, do not use the device.
  • Do not use the instrument in explosive atmospheres.
  • The measurement range as stated in the specifications must not be exceeded under any circumstances.
  • Non-observance of the safety notes can cause damage to the device and injuries to the user.

We do not assume liability for printing errors or any other mistakes in this manual. We expressly point to our general guarantee terms which can be found in our general terms of business. If you have any questions please contact PCE Instruments. The contact details can be found at the end of this manual.

Features

  • 128*64 OLED display;
  • Micro USB storage device can be connected directly;
  • Memory for 600 average values and 6 files;
  • Limit alarm;
  • Auto Power Off when battery is low;
  • Easy-to-read display, clear digits
  • Rechargeable Li-Ion battery
  • Applications: ball bearings and other parts, fault analysis of pressure vessels, steam turbine generators and plants, heavy workpieces, mechanical and permanently installed parts, limited space

Delivery contents

  • 1 x hardness tester PCE-2500N or PCE-2600N
  • 1 x charging cable
  • 1 x test block
  • 1 x cleaning brush
  • 1 x user manual
  • 1 x carrying case

Measuring principle

The LEEB method of measurement was first used in 1978. It is defined as the quotient of the rebound velocity of an impact body over its impact velocity, multiplied by 1000. The harder a material surface, the higher the rebound velocity will be. The Leeb hardness value represents the direct relation of a group of materials (e. g. steel, aluminium) to its hardness properties. For the most common metals, conversion charts are available. These enable the user to convert HL hardness values into other standard hardness units (HB, HV, HRC, etc.).

An impact body with a spherical measuring tip made of tungsten carbide is accelerated towards the test surface and then rebounds. The impact and rebound velocities are measured at a distance of 1 mm from the test surface, using the following method: a permanent magnet in the impact body induces an electrical current into the coil when passing through the coil in the coil holder. This electrical current is proportional to the velocities of the magnet. Leeb hardness values can be calculated by using the following formula:

  • HL = 1000 x (VB / VA)
  • HL: Hardness according to Leeb
  • VB: Rebound velocity
  • VA: Impact velocity

The voltage curve of the output signal, when the impact body passes through the induction coil, is shown in the following figure:

If you measure the hardness of a material with the conventional static test method, the pressure you apply will affect the reading, e. g. 720 HLD ≠ 720 HLC. Since different impact devices correlate with different conversion charts when converting Leeb hardness values into other hardness units, the impact device used should be indicated in addition to the converted value.

Device description

PCE-2500N

PCE-Instruments-PCE-2500N-Portable-Pen-sized-Durometer-for-Metals-
fig-3

  1. Display
  2. Keypad
  3. Loading tube
  4. D-type impact body (within guide tube)
  5. Micro USB charging socket
  6. Support ring

PCE-2600N

PCE-Instruments-PCE-2500N-Portable-Pen-sized-Durometer-for-Metals-
fig-4

  1. Display
  2. Loading tube
  3. Keypad
  4. DL-type impact body (within guide tube)
  5. Micro USB charging socket

Display

After switching on the device, it will show the main screen:

  • HL: Hardness scale
  • AVE: Average value
  • : Impact direction

Difference between PCE-2500N and PCE-2600N

  • PCE-2500N has a D probe for even surfaces
  • PCE-2600N has a DL probe for uneven, small surfaces

Keys

  • On/Off

  • Menu/Confirm

  • Up

  • Down

Technical specifications

Hardness scales HL, HV, HRA, HRC, HRB, HB, HV, HS
Measurement ranges 170 … 960 HLD

17.9 … 69.5 HRC 19 … 683 HB

80 … 1042 HV

30.6 … 102.6 HS

59.1 … 88 HRA 13.5 … 101.7 HRB

Testable materials| Steel and cast steel, alloyed tool steel, stainless steel, grey cast iron, spheroidal graphite iron, cast aluminium, copper-zinc alloy (brass), copper-tin alloy, copper

(bronze)

Accuracy| HLD: ±0.5% (800 HLD)
Repeatability| HLD: 0.8% (800 HLD)
Resolution| 128 x 64 OLED display
Operating time| Approx. 10 h
Operating conditions| Temperature: 10 ~ 50 °C Humidity max. 90 °C
Storage conditions| Temperature: -30 ~ 60 °C Humidity max. 90 °C
Power supply| Rechargeable lithium battery
Dimensions| PCE-2500N: 148 x 32 x 25 mm

PCE-2600N: 205 x 32 x 25 mm

Weight| PCE-2500N: 64 g

PCE-2600N: 80 g

Operation

Before using the device, read the information on calibration and about the factors that have an effect on the measuring accuracy.

Preparation of sample surface

When preparing the sample, it is essential to observe the information given in the charts at the end of this manual.

PCE-Instruments-PCE-2500N-Portable-Pen-sized-Durometer-for-Metals-
fig-10

  • When preparing the sample surface, bear in mind that heat, cold, etc. affect the hardness of the surface to be tested.
  • An excessively rough test surface can cause measurement errors. Therefore, the test surface should have a metallic gloss and be even and free from grease.
  • Surface: The test surface should be as smooth as possible. When the radius of curvature R of the surface is smaller than 20 mm, a small support ring or another ring-shaped fixing aid should be used.

Stabilisation of the sample

  • Heavy samples do not have to be stabilised;
  • Medium-sized samples have to be placed on an even and hard surface. Make sure that the sample rests on the surface evenly and does not move.
  • Make sure that the sample is thick enough. For minimum thicknesses of a sample, please refer to the charts at the end of this user manual.
  • For samples with a hardened surface, find the minimum depths of the hardened layer in the chart at the end of this manual.
  • Coupling
    • Lightweight samples must be closely coupled to a solid support. The two coupling surfaces must be even and you must make sure not to use too much coupling gel. The test direction should be vertical to the coupled surface.
    • When the sample is a large plate, bar or bent part, it can deform considerably and become unstable, even if the weight and thickness are as required. This causes incorrect readings. In this case, the test surface of the sample should be reinforced or supported.
  • The magnetic field of the sample should not exceed 30 Gauss.

Prior to first measurement

Before making your first measurement, you should carry out a test measurement, using the test block which is included in the delivery. Make 5 measurements and calculate the average. If the values from the charts at the end of this manual are exceeded, the internal calibration function can be used.

Power on

  • Press the On / Off key. The hardness tester will switch on and automatically enter measurement mode.

Charging

  • To lock the impact body, push down the loading tube completely. Do not release the loading tube before you have reached the bottom. Then slowly bring it back to its original position.
  • Hold the device between your thumb and index finger and place it onto the sample. While doing so, make sure that the support ring lies firmly on the surface and that the impact direction is vertical to the test surface.

Measurement

  • Before applying the trigger, make sure that the sample stands firmly. Also ensure that the direction of the starting force is in line with the axis of the impact device.
  • Make a measurement by slightly pressing the trigger at the top of the device. The reading will be shown in the display.
  • Each portion of the sample must be measured 5 times. The results should not be below or above the average value of ±15 HL.
  • The distance between any two points of impact or between the middle of a point of impact and the edge of a sample should fulfil the requirements in the following chart.

Centre to centre distance of the two points of impact| Distance from centre of point of impact to edge of sample
---|---
|
3 mm| 5 mm

  • At the end of the measurement, the results can be seen on the display. The hardness value is the average value of 5 measurements at different measuring points.
  • To switch off the device, press the On/Off key.

Readings displayed

  • The hardness value is displayed in a certain hardness unit. If, for instance, a value of 700 is measured with a Leeb impact device and a D-type impact body, the hardness value will be displayed as 700 HLD, HL standing for hardness according to Leeb and D for the D-type impact body.
  • When converting the HLD value into a different hardness unit, the test method should be written as part of the unit. 400HVHLD, for example, stands for a hardness value of 400 HV (hardness according to Vickers), tested by the HLD method (Leeb hardness with D-type impact device).

Menu structure

  • Settings and additional functions can be found in the parameter menu. In the main screen, press the Menu key to enter the main menu.

PCE-Instruments-PCE-2500N-Portable-Pen-sized-Durometer-for-Metals-
fig-11

  • Select the corresponding options to set the parameters. You can make a selection using the arrow keys. Confirm your selection with the Menu/Confirm key.
Calibration
  • Before first use and after not using the hardness tester for a period of time, it must be calibrated by means of the Leeb test block.
  • Go to calibration mode via menu item 7.

x/5

000
Calibration

  • Measure 5 measuring points of the Leeb hardness block.
  • The impact direction must be vertically down.

1/5

772          HL
Calibration

  • After the measurement, the hardness tester will show the average value.
  • Press the arrow keys to set the average value to the standard value of the test block. (You can set the value up to 15 HL higher or lower). Confirm with the Menu/Confirm key. “Calibration complete“ will be displayed.
  • The device shows values in HLDL. In order to calculate your HLD value, please refer to the conversion chart in 10.5.

Battery replacement

  • Unscrew the cover on the back side of the display unit.
  • Remove the battery.
  • Insert the new battery. Observe correct polarity.
  • Screw the cover back on.
  • Switch on the device to find out if it works properly.

Additional information

Specifications of impact device

Type of impact device D (PCE-2500N), DL (PCE-2600N)
Impact energy Impact weight 11 mJ

5,5 g

Ball hardness Ball diameter

Ball material

| 1600 HV

3 mm

Tungsten carbide

Diameter of impact device

Length of impact device Weight of impact device

| 20 mm

147 mm

50 g

Sample requirements

Max. sample hardness 940 HV
Max. sample roughness Ra 1.6 μm

Min. sample weight

Measurement without stabilisation Stabilisation

Coupling necessary

| ****

5 kg

2 ~ 5 kg

0.05 ~ 2 kg

Max. sample thickness Measurement without coupling Coupling necessary| ****

5 mm

≤5 mm

Min. depth of hardened layer| 0.8 mm

Ball indentation

Hardness 300 HV Indentation diameter Penetration depth 0.54 mm 24 μm
Hardness 600 HV Indentation diameter Penetration depth 0.54 mm 17 μm
Hardness 800 HV Indentation diameter Penetration depth 0.35 mm 10 μm

Conversion chart materials

Material Hardness scale Value

Steel and cast steel

| HRC| 17.9~68.5
HRB| 59.6~99.6
HRA| 59.1~85.8
HB| 127~651
HV| 83~976
HS| 32.2~99.5
Steel| HB| 143~650


Alloyed tool steel

| HRC| 20.4~67.1
HV| 80~898


Stainless steel

| HRB| 46.5~101.7
HB| 85~65 5
HV| 85~802


Grey cast iron

| HRC|
HB| 93~334
HV|


Spheroidal graphite iron

| HRC|
HB| 131~387
HV|


Cast aluminium

| HB| 19~164
HRB| 23.8~84.6


Brass

| HB| 40~173
HRB| 13.5~95.3
Bronze| HB| 60~290
Copper| HB| 45~315

Conversion chart HLD/HLDL

HLD| 300| 304| 306| 308| 310| 314| 316| 318| 322| 324| 328| 330| 334| 336| 338| 340| 344| 346| 348| 352| 354
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---
HLDL| 560| 562| 564| 566| 568| 570| 572| 574| 576| 578| 580| 582| 584| 586| 588| 590| 592| 594| 596| 598| 600
HLD| 356| 360| 361| 364| 367| 370| 372| 374| 377| 380| 383| 386| 387| 390| 392| 395| 398| 400| 403| 405| 408
HLDL| 602| 604| 606| 608| 610| 612| 614| 616| 618| 620| 622| 624| 626| 628| 630| 632| 634| 636| 638| 640| 642
HLD| 410| 413| 415| 418| 420| 423| 425| 429| 431| 433| 436| 438| 440| 444| 446| 448| 451| 453| 456| 458| 461
HLDL| 644| 646| 648| 650| 652| 654| 656| 658| 660| 662| 664| 666| 668| 670| 672| 674| 676| 678| 680| 682| 684
HLD| 464| 466| 468| 472| 474| 476| 479| 481| 484| 487| 490| 492| 495| 497| 500| 502| 505| 508| 510| 513| 516
HLDL| 686| 688| 690| 692| 694| 696| 698| 700| 702| 704| 706| 708| 710| 712| 714| 716| 718| 720| 722| 724| 726
HLD| 518| 521| 524| 527| 530| 532| 534| 537| 540| 544| 546| 548| 551| 554| 557| 560| 563| 566| 569| 572| 574
HLDL| 728| 730| 732| 734| 736| 738| 740| 742| 744| 746| 748| 750| 752| 754| 756| 758| 760| 762| 764| 766| 768
HLD| 578| 581| 584| 587| 590| 592| 596| 599| 602| 605| 608| 611| 614| 617| 620| 624| 627| 630| 634| 637| 640
HLDL| 770| 772| 774| 776| 778| 780| 782| 784| 786| 788| 790| 792| 794| 796| 798| 800| 802| 804| 806| 808| 810
HLD| 644| 647| 650| 654| 657| 660| 664| 667| 670| 674| 678| 681| 684| 688| 691| 695| 698| 702| 706| 709| 712
HLDL| 812| 814| 816| 818| 820| 822| 824| 826| 828| 830| 832| 834| 836| 838| 840| 842| 844| 846| 848| 850| 852
HLD| 716| 720| 724| 728| 730| 734| 738| 742| 746| 749| 752| 756| 760| 764| 768| 771| 774| 778| 782| 786| 789
HLDL| 854| 856| 858| 860| 862| 864| 866| 868| 870| 872| 874| 876| 878| 880| 882| 884| 886| 888| 890| 892| 894
HLD| 793| 797| 800| 804| 808| 812| 815| 819| 823| 826| 830| 834| 837| 841| 845| 849| 852| 856| 860| 864| 867
HLDL| 896| 898| 900| 902| 904| 906| 908| 910| 912| 914| 916| 918| 920| 922| 924| 926| 928| 930| 932| 934| 936
HLD| 871| 874| 878| 882| 886| 889| 893| –| –| –| –| –| –| –| –| –| –| –| –| –| –
HLDL| 938| 940| 942| 944| 946| 948| 950| –| –| –| –| –| –| –| –| –| –| –| –| –| –

Contact

  • If you have any questions, suggestions or technical problems, please do not hesitate to contact us.
  • You will find the relevant contact information at the end of this user manual.

Disposal

For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose. In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use them or give them to a recycling company that disposes of the devices in line with the law. For countries outside the EU, batteries and devices should be disposed of in accordance with your local waste regulations. If you have any questions, please contact PCE Instruments.

www.pce-instruments.com

PCE Instruments’ contact information

Germany

United States of America

China

United Kingdom

Turkey

Spain

Italy

Scan

© PCE Instruments

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

PCE Instruments User Manuals

Related Manuals