ibidi Clamp for Sticky Slides Instructions
- June 12, 2024
- ibidi
Table of Contents
**ibidi Clamp for Sticky Slides Instructions
**
Introductions
The Clamp for sticky-Slides is an instrument to support the assembling of the ibidi stickySlides with a bottom material. The assembled sticky-Slide is inserted into an adapter, which is then compressed between the brackets of the Clamp for sticky-Slides. The applied force is homogeneous on the overall area, to guarantee an optimal adhesive bond between the sticky-Slide and the substrate (e.g. glass coverslip).
Material
The materials of the Clamp for sticky-Slides and the adapters are anodized aluminum and plastic. There are magnets integrated in both parts of the adapters to keep the two parts together.
The Clamp for sticky-Slides and the adapters are not autoclavable since they are temperature stable up to 60°C/140°F only. For sterilization we recommend cleaning all surfaces with an ethanol–soaked tissue.
Specifications
The Clamp for sticky-Slides is available with adapters for several types of sticky-Slides. For detailed information about the handling of the sticky- Slides please refer to the sticky-Slide instructions.
Specifications
Clamping force: 1.2 kN
Material: Anodized aluminium, plastic
Sterilization: Ethanol
Available adapters
sticky-Slide 8 Well / 18 Well
sticky-Slide I Luer
sticky-Slide VI 0.4
sticky-Slide Chemotaxis
Dimensions of Compatible Coverslip
Width: 75 mm
Length: 25 mm
Thickness: 0.1-1.0 mm
Handling of the Clamp for sticky-Slides
The working principle and the order of insertion of the parts of the Clamp for sticky-Slides is shown in the following picture:
For optimal results please follow the proceeding of the protocol below:
-
Take the desired sticky-Slide out of the packaging and remove the protection film of the adhesive surface.
-
Bring the coverslip in position and glue it to the lower side of the sticky-Slide.
-
Put the assembled sticky-Slide in the adapter.
-
Mount the adapter onto the upper bracket of the clamp.
-
Move the lower bracket upwards until it pushes against the adapter. Make sure the lever is opened all the way.
-
Pull the lever to apply pressure on the adapter.
Warnings:
* Always handle with care, because the glass coverslip might break during the procedure!
* Please take care not to trap your hand or fingers between the brackets!
-
Keep the sticky-Slide pressed for approximately 1 min.
-
Open the clamp by pressing the red button and releasing the lever.
-
Remove the adapter from the clamp.
-
Open the adapter and remove the sticky-Slide.
-
For best results, incubate the finished sticky-Slide for 8 hours at 37°C in a dry or humid incubator.
Ordering Information
The Clamp for sticky-Slides is available with adapters for several types of sticky-Slides.
Cat. No. Description
80040 Clamp for sticky-Slides
80042 Adapter for sticky-Slide I Luer
80043 Adapter for sticky-Slide VI 0.4
80044 Adapter for sticky-Slide Chemotaxis
80045 Adapter for sticky-Slide 8 Well / 18 Well
For research use only!
Further information can be found at www.ibidi.com.