ARROW PR-35052-HPSM Pressure Injectable Two-Lumen PICC with 80 cm Wire Instructions
- June 4, 2024
- Arrow
Table of Contents
PR-35052-HPSM
Pressure Injectable Two-Lumen PICC with 80 cm Wire
Instructions Manual
PR-35052-HPSM Pressure Injectable Two-Lumen PICC with 80 cm Wire
Contents:
1: Two Lumen TaperFree® Catheter: 5Fr.(1.81 mm OD) x50cm, Pressure
Injectable, BlueFlexTip®, and Contamination Guard
1: GlideThru™ Peel-Away Sheath: 5Fr.x4″ (10 cm) Radiopaque over 5 Fr. Dilator
1: Spring-WireGuide: .018″(0.46 mm)dia. x 17-3/4″ (45 cm) (StraightSoft Tip on
both ends)
1: Spring-Wire Guide, Marked: .018″ (0.46mm) dia. x 31-1/2″(80 cm)(Straight
Soft Tip on One End- Straight Stiff Tip on Other)
1: Introducer Needle: Echogenic 21 Ga. x 2-3/4″ (7 cm) TW
1: Syringe: 5 mL Luer-Lock
2: Dust Cap: Non-Vented
1: SecondSite™ Adjustable Hub: Fastener
1: SecondSite™ Adjustable Hub: Catheter Clamp
1: Chart Sticker
1: Patient Information Booklet
1: Patient ID Card
1: Checklist/CLIP Sheet
1: Paper Tape Measure
1: Dressing: STATLOCK®¹ PICC Plus Catheter Stabilization Device
¹A registered trademark of C.R.Bard, Inc.
All components are CE 2797 unless otherwise noted.
Rx only Warning: Read all package insert warnings, precautions, and
instructions prior to use. Failure to do so may result in severe patient
injury or death. www.teleflex.com/IFU
California Prop. 65
⚠WARNING: Cancer and Reproductive Harm.
www.P65Warnings.ca.gov
Not made with natural rubber latex.
Fluid path components are non-pyrogenic.
Contraindications: The Pressure Injectable PICC is contraindicated
wherever there is the presence of device-related infections or the presence of
thrombosis in the intended insertion vessel or catheter pathway. A clinical
assessment of the patient must be completed to ensure no contraindications
exist.
| **Priming Volume*
(mL)| Gravity Flow Rate†
(mL/hr)| Pump Flow Rate††
(mL/h)| MAX Pressure Injection
Flow Rate (mL/sec)
---|---|---|---|---
Distal (18 Ga.)| 0.48| 488| 3860| 5
Proximal (18 Ga.)| 0.5| 488| 3870| 5
- Priming volumes are approximate and are done without accessories.
† Flow rate values are approximate and are determined using deionized water at 100 cm head height.
† † Pump flow rates are determined at a maximum pump pressure of 10 PSIG and represent approximate flow capabilities.
** Pressure injection flow rates are determined at the injector pressure setting of 300 psi maximum using media of 11.8 centipoise viscosity, with 152 cm pressure tubing.
EU Authorized
Representative and Importer:
Teleflex Medical
IDA Business and Technology
Park Dublin Road, Athlone
Co. Westmeath, Ireland
** Arrow International LLC
A subsidiary of Teleflex Incorporated
3015 Carrington Mill Blvd.
Morrisville, NC 27560 USA
www.teleflex.com
LBL062740 R00 (2021-11) Pressure Injectable Two-Lumen PICC with 80 cm Wire
REF: PR-35052-HPSM
5Fr. /2Lumen
50 cm catheter length
.018 inch dia.
spring-wire guide**
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>