iSTIR HP350/HP550 PRIME
PRODUCT USER MANUAL
INTRODUCTION
This manual provides important safety information for this Hot Plate Magnetic Stirrer. It should be kept near the equipment for quick & easy reference. This stirrer is specially designed for precise control of stirring speed and temperature which allows gentle to vigorous mixing with speed ranging from 200 to 2200 RPM with a maximum capacity of 20 Ltr. The multipurpose digital display assists users by displaying various parameters like actual & settemperature, speed, mode of operation, etc.
INTENDED USE
The hotplate stirrer is suitable to use for mixing and/or heating the liquids with the maximum capacity of 20 Ltr. It is designed for use in general laboratories, pharmacies, schools and universities.
NOTE: Before using the instrument, please read this user manual carefully. This user manual is intended to assist with the operation and care of the unit only and not its repair. For repair please contact the supplier.
SYMBOL
| |
---|---|---
Warning| Waste electrical| Fuse
FEATURES
- Powerful motor for constant speed delivery under varying load
- Stirring capacity up to 20 Ltr
- Nano crystal full ceramic top plate ensures chemical resistivity & constant heat transfer
- Highlyresistantand inert body, even to strong chemicals
- Heater temperature setting from 5°Ct0 550°C (HP550)
- Heater temperature setting from 5°Cto350°C (HP350)
- Variable speed setting from 200 to 2200 RPM in steps of 10 RPM & long timer range from 1-999 mins & infinite mode
- Microprocessor controlled hot plate stirrer with variable speed and time setting, along with last run memory function
- Programmable pulse mode with run pause run feature
- 6heatingmodes for rapid, gradual & accurate heating
- Llarge and clear display for better readability across all parameters, set values readily visible
- Safe temperature circuit cut up to 375°C in HP350 / 575°C in HP550 for greater reliability
- Smallfootprintfor saving valuable bench space
- Hotplateindication when temperaturereaches above 50° C
- Compatible with PT-1000 temperature probe to accurately measure temperature of medium
- Keypad lock feature
ACCESSORIES
- PT 1000 Temperature probe and probe attachment stand
- Magnetic stirring bar
- Power cord
- User manual and warranty card
TECHNICAL SPECIFICATIONS
PRODUCT | HP350 | HP550 |
---|---|---|
Motor | DC | |
Number of stirring position | 1 | |
Max Stirring quantity on Equip (H20) | 20 Ltr | |
Speed range | 200 to 2200 RPM | |
Speed control | 10 RPM / Step | |
Heating Temp. range | Room Temp. to 350°C | Room Temp. to 550°C |
Setting Temp. | 5°C to 350°C | 5°C to 550°C |
Safe Temp | 375°C | 575°C |
Temperature setting resolution | 1°C / Step | |
Time range | 0 to 999 minutes & Infinite | |
Heater Control Accuracy (+/-K) | ±10 °C | |
Recommended Stirring bar length | 25 mm | |
Set-up plate material | Ceramic | |
Set-up plate dimensions | 180 x 180 mm | |
Dimensions (L x W x H) | 313 x 205 x 110 mm | |
Protection class | IP 21 | |
Permissible ambient temp. | 5-40°C | |
Permissible relative humidity | s80% | |
Weight | 4.9 Kg | |
Altitude | Use upto an altitude of 2000 m above MSL | |
Pollution Degree | 2 | |
Environment | For indoor use only | |
Input Voltage | 220-240V, 50Hz | |
Motor rating input & Output | 57W | |
Total Power Consumption | 1000 W | |
Safety Fuse Rating | 5Amp |
SAFETY INSTRUCTIONS
Read all safety & usage information provided in this manual carefully before using the device.
-
Place the device on a flat, stable, clean, non-slippery and fire-proof surface.
-
Ensure that only trained staff uses the device. Keep the instructions manualin a place where it can be accessed easily.
-
Beware of the possible effects of magnetic field on pacemakers, data media, etc.
-
Do not touch the hotplate surface when temperature of hotplate is over 50°C, this could result in serious burns or injury. Pay attention to the residual heat after switching off.
-
Properly lift the device with both hands while moving or installing. Also, the device should only be moved from its position once it attains the room temperature.
-
Make sure that power supply cord or temperature sensor cable must not come in contact with the heated mounting plate.
-
Do not place any steel or magnetic material on the top surface except the recommended magnetic stirring bar with beaker or flask in between. Doing this can affect the magnetism of device.
-
Wear your personal protective equipment in accordance with the hazardous category of the media to be processed. Otherwise, there isarisk from:
• Splashing and evaporation of liquids.
• Ejection of parts.
• The release of toxic or combustible gases. -
The top surface must be clean and kept intact. Wear protective gloves while cleaning the device. Use a soft mild cloth to clean.
-
Do not use damaged beaker, flask, stirring bar or any other component for operation. It may affect the efficiency of the device.
-
11. Reduce the speedif:
• Ifthe medium splashes out of the vessel due to high speed.
• Iftheapplianceis notrunning smoothly.
• Ifthe container moves on the base plate. -
Do not move the device when it’s connected to power supply or during its operation.
-
Ensure that the product is used only for specified operation. It should not be used for shaking hazardous or reactive solutions.
-
Give specific attention to the risks associated with:
• Flammable materials.
• Flammable media with low steam pressure.
• Overfilling of medium.
• Incorrect vessel size.
• Unstable vessel.
• Glass breakage. -
Do not use the device in explosive atmosphere, it is not Ex protected with substance capable of forming an explosive mixture. Appropriate safety measure must be taken.
-
Note that, there is a possibility of contamination or unwanted chemical reaction.
-
The RTD (PT 1000) sensor must always be immersed in the medium by at least 20 mm.
-
The rating of power supply must match with the specified rating.
-
The top surface will be hot after any heating operation. Do not touch the top surface till the hot LED indication glows.
-
Process pathogenic materials only in a closed vessel under a suitable extractor hood. Do not operate the appliance in explosive atmospheres, with the hazardous substance or under water.
-
Only that liquid should be used for mixing which does not react dangerously to the extra energy produced through processing. This also applies to any extra energy produced in other ways. For example: Through light irradiation, through surrounding temperature etc.
-
Abrasion of the dispersion equipmentor the rotating accessories can get into the medium you are working on.
-
The chemical reaction of PTFE can occur when in contact with molten or dissolved alkaline earth metals, as well as with fine partied powders of metals of the 2 and 3 group of the periodical system at temperatures above 300-400 °C. Only elementary fluorine, chlorine trifluoride and alkaline metals do attack PTFE, halogen hydrocarbons have a reversible swelling effect. Only glass-coated magnetic rods should be used in combination with solute alkali metals or alkaline earth metals or at the temperature above 250°C.
-
To protect do not cover the device, even partially, with elements such as metallic plates or sheets otherwise it may overheat. Ensure that mounting plate remains clean.
-
The socket must be earthed (protective ground contact).
-
Do not use the device if the ceramic set-up surface is damaged.
e.g. scratches, splinters or corrosion. A damaged set-up surface could break if used. -
Safety Fuse is provided of 5Amp configuration which can be replace by the operator. The same will protect the machine circuit during an electrical fault or overload.
STANDARD PARTS
Front Side
Rear Side
INSTALLATION
The hot plate stirrer is provided in a box. Open the box, then remove the packaging and gently place the device on the firm & leveled surface. Take care while unpacking & removing all accessories. The user manual should be kept with the device for easy access. Please keep all packaging in safe storage for at least two years for warranty purpose.
9.1 LOCATION & MOUNTING
Place the stirrer on a flat and leveled surface & ensure that all the four
legs of this stirrer stands on the surface firmly. Avoid installing on a
slippery surface or surface
prone tovibration.
- Idealambient temperature is 25°C+1°C, avoid placing the unitin direct sunlight.
- Keep clearance of at least 20 cm from all side to increase the cooling efficiency.
- Keep the unit away from heated solution to avoid sample temperature issues.
- Do not place the equipment at a place where it becomes difficult to operate it.
9.2 ASSEMBLING SUPPORT ROD STAND
- Screwinthe support rod manually until it cannot be tightened any further.
- Assemble the accessories mentioned insection 4 usinga clamp.
9.3 CONNECTING POWER CORD
- Connect one end of the power cord to the rear side of the stirrer and another end to the power supply as shown in the figure below.
- Pushthe power cord firmly for proper connection and turn ON the main switch.
- Make sure thatinput power source is according to stirrer requirement.
USER INTERFACE & DISPLAY
| | |
---|---|---|---
Display shows Setand Actual
Temperature value
|
LEDs indicates in which unit temperature is displayed
|
Display shows current speed & time values
|
LEDs indicates temp. displayed is of PROBE
Item | Name | Function |
---|---|---|
1 | TEMP | Press “TEMP” to select Temperature value. Then rotate knob to set |
Temp value in °C.
2| SPEED| Press “SPEED” to select Speed value. Then rotate knob to set Speed
value in RPM.
3| TIME| Press “TIME” to select time value. Then rotate knob to set Time value
in minutes.
4| TEMP. Selection| Press “°C/°F” to change the unit in which Temp. is
displayed.
5| KNOB| Rotate right side to increase values.
Rotate left side to decrease values.
Press to Lock/Unlock the keypad.
6| PROBE| To display temperature of probe at actual temp. display.
7| START/STOP| Press “START/STOP” to start or stop any operation.
8| HOT LED| Glows when top plate of the stirrer is HOT equal or above 50°C.
9| ATS| “Auto Start Mode”. Machine resumes from the left parameter.
10| PULSE| Pulse mode for Run Pause Run operation.
OPERATION
11.1 SWITCHING ON THE DEVICE
Plug the power cord into its socket on the rear panel, now connect the power
cord to.
AC power output. Now switch ON the mains and switch ON the stirrer from the
rear/back side, as shown in the figure. The power LED on its front panel will
glow.
IMPORTANT NOTE: Values of any parameter will get set saved only after the respective display blinks for multiple times.
11.2 SAFE TEMPERATURE
Safe temp range 50 °Cto 375 °C (HP350) / 50 °C to 575 °C (HP550). Upon turning
the machine on, the stirrer screen will display with SAFE temp setting. Set
the safe temperature by rotation knob clockwise for increment & counter
clockwise for decrement. Only “Knob” & “°C/°F” key will work, other keys will
remain deactivate while setting the safe temperature. Safe Temp can be set in
both – Celsius or Fahrenheit. After adjusting temp, value will be save after
blinks and then it will take to home screen.
NOTE: This is a safety feature where if the temperature of heater exceeds
the value saved in safe mode then operation (heating & stirring) will stop.
The safe temperature limit must always be set to atleast 25 °Clower than the
fire point
of the media used.
11.3 CONTROL PANEL OPERATION
SET and ACTUAL LEDs will FLASH as per status of the command. By rotating knob
valuescan be increased and decreased for each parameter. “START/STOP” button
for operation startand stop. “SPEED”, “TIME” and “TEMP” buttons to select it
respective values.
11.4 START/STOP FUNCTION
Press “START/STOP” key to Start operation (once all the parameters are set).
This will be indicated by start/stop LED. If the LED is ON, means the
operation is running or if the machine is turned off the LED will stop
glowing. Press this button again to “STOP” operation. Iftemp is “Abt” and RPM
O then start button doesn’t work.
11.5 TEMPERATURE BUTTON
To set the temperature, press “TEMP” key and change value by rotary knob from
5°C to 350°C (HP350) / 5°C to 550°C (HP550). After setting desired value,
value will save automatically after blinks. After value saved actual LED will
automatically flash on (once the “START/STOP” key is pressed). Temperature
value can be changed during continuous operation. If the operator want to
check set value then, it can be seen by pressing “TEMP” button again during
operation.
11.6 °C/°F KEYS
The Temperature can be set in Celsius or Fahrenheit, on home screen. By
pressing “°C/°F” button, the °C or °F will be displayed alternatively.
Whichever will be selected, the Actual Value and set Value of Temp will change
automatically. While changing Temp Unit, the respective value will be rounded
off for the converted unit & will display. Each unit selected will be indicted
by respective LED.
11.7 HEATING MODE
6 different heating modes can be set when on home screen.
By long press “TEMP/MODE” button, setting display will blink and shown as HO3
as default, by rotary Knob one can switch between – H00, H01,H02,H03,H04 and
HO5. All other functions/keys will remain inactive while setting this
parameter. Once the heating mode is selected the parameter will set
automatically.
NOTE: Temperature accuracy is claimed as per water, if liquid medium
changed the accuracy may change.
-
HOO: Slow Heating for Low volume Liquid & high Accuracy (up to 300ml)
This mode is basically for Lower volume i.e. Up to 300ml. The accuracy for this mode is very High, so itis not suggested for higher volumes. After setting up the required parameters, if the user starts the operation the temperature rises Slowly & stabilizes to the set value.
NOTE: This mode is design for zero over shoot of heating liquid, it will take time if the liquid volume increase or liquid type changes. -
HO1 : Slow heating with High Accuracy (300ml to 800ml)
This mode is very useful for the solutions which are very sensitive to change of temperature, it is basically for volume above 300ml to 800ml.
After setting all parameters when user presses “START”, the initial temperature of device will rise quickly toa point below the set value chternp.,E then the temperature will rise slowly with high accuracy to set value & once it attains the set value then the temperature will remain constant for the remaining time with the high accuracy of +/-1°C inwater.
NOTE: If the Liquid media change than accuracy may differ from specified. -
HO2 : Fast heating with High Accuracy (800ml to 2L)
This mode is useful when the user wants to heat the solution quickly, it is basically for volume above 800ml to 2L volume. In this mode after setting all parameters when the user presses “START”, temp. of unit rises quickly to the set value point & then it stabilizes to setvalue. Initially, temperature may vary up to +/-15°C. After stabilizing, temperature of the plate will remain constant with high accuracy of +/-1°C.
-
HO3 : Rapid Heating with Overshoot (1L to 2L)
This mode of heating is useful when user wants to heat the solution rapidly without the concern of accuracy of temperature, it is basically for volume above 1L to 2L . In this mode after setting all parameters when user presses “START”, initial temperature of device will rise very fast to the set value of temperature and once it reaches the set value device will slowly stabilize the temperature to the set value. In this mode, heating will be much more rapid then previous two modes and possible overheating could be up to+/-20°C. -
HO4: Rapid Heating for High volume Liquid (2L to 5L)
This mode is similar to HO3 mode, but it is basically for high volumei.e. 2L to 5L. After setting up the parameters if the user starts the operation, the temperature will rise with high rate and will gradually stabilize to the set value. In this mode the accuracy is not that accurate as attained in the previous modes.
-
HO5: Rapid Heating for High volume Liquid (6L to 20L)
This mode is basically for higher volume i.e. 5L to 20L. The accuracy for this mode is very low, so it is not suggested for lower volumes. After setting up the required parameters, if the user starts the operation the temperature rises rapidly & stabilizes tothe set value.
11.8 SPEED FUNCTION
To set speed, press “SPEED” button on home screen indicated by speed LED
flashing on. Speed can be set by rotating the knob in proper direction for
increment or decrement ranging from 0-2200 RPM in the range of 10 RPM of
increment or decrement to the parameter. Default RPM is always be as last set
RPM. Heater can work even at 0 RPM. After adjusting the speed, will save
automatically once the screen stops blinking.
11.9 TIME FUNCTION
On Initiation of the machine the value on window will display ” X “. The timer
can be set from 1 to 999 minutes & infinite (” 3L “) minutes. By pressing
“TEMP” key once, timer can be set which will be indicated by LED. By rotary
“knob” the time LED will be flash on. It will show the Time command
activation and we can change time parameters in this situation.
To set time value, rotary Knob for increment and decrement of Value from 1 to
999 Minutes. After set desire value in window, the value will blink then stop,
it means the value is save. If operator want to see Set value then he can be
see by pressing time button. Time unit will be in minute. In running
condition if time rechange, timer will reset & count from the beginning.
11.10 HOT LED
Hot surface indication will turn on when the plate surface reaches 50°C &
remain on until the temperature falls down below 50°C.
The Hot surface indication remains active even the machine is turned off but
the main must remain ON this let one avoid accidental usage of the machine
when the surfaceisstill hot.
- HOT PLATE LED will ON even on/off switch off when temperature comes down below 50°C.
- Stirring will continue start while Hot Plate LED not off.
11.11 OPERATING MODE
All parameter settings are saved automatically when the device suddenly
switched off. Once end user switch on the device, the“START/STOP” key starts
the device again with the set parameters.
11.12 ABT TEMPERATURE
By press “TEMP” button, the temp Set value blink and by Rotating “knob”, you
will see “Abt” on screen. By set Abt on screen you can run stirrer without
heating. Temp setting range ambient to 350°C/550°C, You can not set temp below
Abt temp.
11.13 LOCK /UNLOCK KEYPAD FUNCTION
By Long Pressing “knob”, One can Lock buttons after Starting the operation &
the display Shows “LOC” blinking. Only “START/STOP” button can work during
this Function. By Long Press “knob” again, You can unlock display buttons &
the display Shows “Unlo” blinking. After pressing “STOP” button in “LOC”
condition during operation, “START/STOP” button will also belocked. In that
condition, One have to “Unlock” buttons to Start Operation again. In ATS Mode,
the Lock mode will be unlocked by only two ways. Either by pressing the “knob”
or if the unit is restarted after the ATS operation is over. While the ATS
operation is running the Lockmode will be remain ON even there is a power
cutoff & the unit restarts. Otherwise buttons stay “Unlocked” in all
ondition.
11.14 ATS MODE
By pressing “ATS” key, ATS LED will show & ATS mode will be activated. To
deactivate ATS mode again press “ATS” key and you can press “STOP” button to
exit from ATS Mode. After complete timer countdown, at the end of operation
you will automatically exit from ATS mode. After power cutoff time will save
only in minute notinseconds.
Example: If you have set the timer for 10 minutes and after 7:30 minutes there
is power cutoff then it will save 7 minutes only, after power resuming it will
start from 7 minutes.
11.15 PULSE MODE
You can set the PULSE mode, when you will on home screen. By pressing “PULSE”
button, the White LED will be ON and pulse mode will be activated, by pressing
“PULSE” button again, you can deactivate PULSE mode. You can set Pulse time by
long pressing “PULSE” button, You will see S 60 on speed display when you are
initiating pulse mode first time.
Pulse Time can be set in range of 30 sec to 99 sec by rotating the knob.
Default time for pulse mode will be same as last set pulse time. By default,
stirring direction is clockwise. Even In a pulse mode there will be no change
in stirring direction. During pulse mode timer function will work parallelly.
Pulse mode can be set “ON” or “OFF” while stirrer is in running condition.
Pulse mode should run at any SETRPM whichisin range of 200-2200 RPM. Time
counting of pulse time will counting from beginning.
11.16 PROBE SELECTION
Probe LED will flash only when PT1000 prob is connected to the stirrer.
Without probe stirrer will automatically work in plate mode. When probe is
connected make sure stirrer should not be in running condition. once prob is
connected respective LED will glow.
User will not be able to change plate/probe mode using any key. By removing
probe, again device will be on plate mode. End user should remove the probe
after complete power off.
TROUBLESHOOTING
Any type of malfunction during operation can be identified by error an message on display. Proceed as follows in such cases.
1.Problem:
Root Cause:
Solution:| Error4
Over Temperature Error
- Actual temperature is more then safe temperature
- Will occur when thermocouple connection issue
- Temperature controlling fail
- Inappropriate safe temperature selection
---|---
NOTE: After Error 4 safety relay will on for heater cutoff.
2.Problem:
Root Cause:
Solution:| Error5
Temperature Probe Error - Will occur when probe connection issue
- Loose connection with device
3.Problem:
Root Cause:
Solution:| Error 6
Heater Error - Will occur when heater fail
- No increment in temperature even after 6 min after device ON
NOTE: After any Error detection, need to reset unit.
4.Problem:
Root Cause:
Solution:| Error 8
Probe (PT1000) not in Medium - Will occur when probe is outside from medium placed on hot plate.
- To override it, long press button “TEMP”, when error comes.
MAINTENANCE & CLEANING
- Before cleaning the device, allow the instrument to attain room temperature remove the power cord from the mains.
- To clean the housing of the device from dyes, building materials or cosmetic materials use isopropyl alcohol as a cleaning agent.
- To clean the housing of the device from food materials or fuels, use water. containing detergent as cleaning agent.
- Wear proper protective gloves while cleaning the instrument.
- Becareful thatnoliquid enters the device during cleaning.
- Kindly contact suppliers before using any other methods for cleaning.
WARRANTY STATEMENT
This product is warranted to be free from defects in material and workmanship
for period of two (2) years from the date of purchase. Your product will be
duly repaired upon prompt notification in compliance with the following
conditions This warranty is valid only if the product is used for its
intended purpose and within the guidelines specified in this instruction
manual. This warranty does not cover damage caused by accident, neglect,
misuse, improper service, natural forces or other causes not arising from
defects in original material or workmanship. This warranty does not cover any
incidental or consequential damages, commercial loss orany other damages from
the use of this product. The warranty is invalidated by any non-factory
modification, which will immediately terminate all liabilities on us for the
products or damages caused by its use. The buyer and its customer shall be
responsible for the product or use of products as well as any supervision
required for safety. If requested the products must be returned to the
distributor in well packed and insured manner and all shipping charges must be
paid. Some states do not allow limitation on the length of implied warranties
or the exclusion or limitation of incidental or consequential damages. This
warranty gives you specific legal rights. This warranty is given expressly in
lieu of all other warranties, expressed orimplied. The purchaser agrees that
there is no warranty of merchantability or of fitness for any intended
purpose and that there are no other remedies or warranties, expressed or
implied, which extend beyond the description on the face of the agreement.
This warranty is only applicable to the original purchaser. Products received
without proper authorization will not be entertained. All items returned for
service should be sent postage prepaid in the original packaging or other
suitable carton, padded to avoid damage. We will not be responsible for damage
incurred by improper packaging. Allitems returned for service should be set
postage prepaid in the original packaging orother suitable carton, added to
avoid damage.
This warranty is valid only if the warranty is registered with the supplier
within 30 days from the date of purchase.
For you reference, make a note of serial number, date of purchase and supplier here.
Serial No.:| Purchase Date:
Supplier:
PRODUCT DISPOSAL
In case the product is to be disposed of, the relevant legal regulations are
to be observed. Information on the disposal of electrical and electronic
devices in the European Community. The disposal of electrical devices is
regulated within the European Community by national regulations based on EU
Directive 2012/19/EU on waste electrical and electronic equipment (WEEE).
According to these regulations, any devices supplied after 13.06.05 in the
business to the business sphere, to which this product is assigned, may no
longer be disposed of in municipal or domestic waste. They are marked with the
following symbol to indicate this.
As disposal regulations within the EU may vary from country to country, E
please contact your supplier if necessary.