iFUGE D06
PRODUCT USER MANUAL
INTRODUCTION
This manual provides important safety information for this centrifuge. It should be kept near the centrifuge for quick & easy reference. This centrifuge is equipped with maintenance free brushless DC motor drive and a safety feature of imbalance detection. The centrifuge can be used for multiple application diagnostics & will work with varity of blood collection tubes.
INTENDED USE
This centrifuge is designed for blood centrifugation & other clinical applications like PRP, medical practitioners, pathological laboratories and hospitals.
FEATURES
- Maximum volume of 6 / 8 x 15 ml.
- Imbalance detector with auto cutoff for safe operation.
- Brushless DC motor for maintenance free performance.
- Lid lock safety-the lid opens automatically on run completion.
- Digital display with speed and time indicator.
- RPM setting from 500-6500 RPM with a count of 100.
- One press changeover from RPM to RCF functionality.
- Digital countdown Timer.
- Reduction adaptors for tubes with volume less than 10 ml.
- Emergency lid release during power failure.
- Automatic internal diagnosis & error display.
- Silent operation.
NOTE: Before using the instrument, please read this user manual carefully. This user manual is intended to assist with the operation and care of the unit only and not its repair. For repair please contact the supplier.
TECHNICAL SPECIFICATION
Option to choose 6 or 8 tube rotor that comes with Autoclavable and Tube holders
Autoclavable & Non-Autoclavable| 6 Tube Rotor| 8 Tube
Rotor
---|---|---
Motor Type| Brushless Motor
Maximum Volume| 6×15 ml| 8×15 ml
Speed Setting| 500 to 6500 RPM
Maximum RCF| 3684 x g| 3873 x g
Speed Accuracy| ± 100 rpm
TECHNICAL SPECIFICATION
Run time | 1 min to 30 mins & infinite mode |
---|---|
Acceleration Time | 43 ± 2 seconds |
Deceleration Time | 40 ± 2 seconds |
Size (L x W x H) | 260 x 244 x 203 mm |
Weight | 4 Kg (with rotor) |
Input Voltage | 24 V 4A |
Power Consumption | 100 W |
STANDARD ACCESSORIES
- Power Adaptor: 100-230 VAC, 50/60 Hz (Input), 24 ****4/6.25A (Output).
- 6 / 8 – slots rotor (Pre-installed).
- T-Allen Key.
- Reduction Adaptors (set of 6 / 8).
- Tube Holder & Rubber cap. (6 / 8 nos.)
- Product user manual & warranty card.
INSTALLATION
Gently remove the upper packaging and take out the centrifuge by holding it
from the bottom. When the centrifuge is used for the first time, ensure that
all the packaging
and accessories are removed from the rotor chamber. The rotor is pre-
installed.
Please make sure that rotor is firmly tightened. Accessories and user manual
are kept inside the box with the centrifuge. Please keep all the packaging in
safe storage for at
least 2 years for warranty purpose. The following points are important for
proper installation: ****
6. INSTALLATION
6.1 Connecting Power Adaptor
-
Connect one side of power adaptor to rear side of centrifuge and other to supply as shown in the figure below.
-
Rotate clockwise the adaptor nut to tighten the adaptor with the centrifuge.
Ensure the power switch is OFF while connecting the power adaptor. -
The sticker on the body & jack are indicating the position of the rib on the jack. While fitting the adaptor to the unit-please match the sticker position as shown in the figure below & gently push it in.
-
If matched correctly the plug will go in easily without much force (do not try to force the plug in if it is not matched properly). Once the plug is inserted into position, turn the loose upper ring clockwise to tighten & to secure the fitment.
STANDARD PARTS LISTING
7. STANDARD PARTS LISTING
- View Window
- Groove to open the lid
- Operation and setting Buttons
- Display
- Emergency Lid Release
1. Power Switch
2. Adaptor Port
TUBE HOLDER & ADAPTORS
This centrifuge features centrifugation for wide range of volumes from 1ml to 15ml with the combination of its adaptors. The accessories included are:
“Tube Holders” are always to be used with the rotor. All of the 6 / 8 “Tube Holders” are mandatory to be placed in the rotor whether 2,3,4 or 6 / 8 tubes are used in balanced condition. By combining the above two adaptors and rubber caps, three different
8. TUBE HOLDER & ADAPTORS
reduced adaptor can be designed to hold different types of blood collection
tubes.
The different adaptor combinations along with tube holder, for different blood
collection tubes are illustrated on pages 5,6 and 7 .
NOTE: 1) Maximum RCF value changes for different length of tubes. 2) “Tube
Holders” are the part of Rotor. All 6 / 8 “Tube Holders” are mandatory to
place whether 2,3,4 or 6 / 8 tubes are used in balanced condition.
COMBINATION A:
COMBINATION B:
COMBINATION C:
8. TUBE HOLDER & ADAPTORS
ADAPTOR COMBINATION & TUBE SIZES
-
NO ADAPTOR COMBINATION
The maximum diameter and proposed length for such combination are:
Maximum Tube Diameter : 18 mm
Proposed Tube Length : 90 to 120 mm -
COMBINATION “A” :
The maximum diameter and proposed length for such combination are:
Maximum Tube diameter : 13 mm
Proposed Tube Length : 75 to 100 mm -
COMBINATION “B” OR “C” :
The maximum diameter and proposed length for such combination are:
Maximum Tube Diameter : 13 mm
Proposed Tube Length : 55 to 75 mm
Check both the criteria (Max. Diameter & Proposed Length) of the blood collection tubes and use the appropriate adaptor-tube holder combination or only tube holder to accommodate the tubes.
ROTOR INSTALLATION
ROTOR INSTALLATION
The rotor of the centrifuge is pre-installed. In case of any spillage, remove
the rotor as
shown in the above figure to clean the rotor and its chamber thoroughly.
9.1 REMOVING ROTOR
- To remove rotor, first remove the Tube Holders.
- Using T-Allen key, loosen the rotor nut by rotating rotor nut counter clockwise and take the out rotor vertically
9.2 INSTALLING THE ROTOR
- Make sure the rotor shaft and the rotor mounting hole are clean. Load the rotor vertically into the motor shaft.
- Put the T-Allen key in the rotor nut & turn clockwise to tighten the rotor.
- After properly fastening the rotor, place the Tube Holders into the rotor.
BALANCING THE ROTOR
- Always balance the rotor before centrifugation Symmetrical loading of tubes in the rotor is necessary. Otherwise vibration or imbalance can occur which can cause serious damage to the centrifuge. The following are process of symmetrical loading of centrifuge tubes inside the rotor.
- The samples in the tubes should be of equal volume. To achieve a balanced condition of the 6 “Tube Holders are mandatory to place inside the rotor whether 2, 3, 4 or 6 / 8 tubes are being used
- If the tubes are not loaded symmetrically then the imbalance detector will cut off the running centrifuge for safety reasons. This will stop the centrifuge and Err 55 will be seen indicating the tubes are not loaded symmetrically. To resume operation, load the tubes symmetrically & restart the centrifuge.
- Below are examples of incorrect method of loading tubes into the centrifuge rotor.
OPERATION
After properly installing the centrifuge, switch ON the centrifuge from the
rear side by pressing the power switch. The lid status will be seen in the
display. If the lid is open,
the display will show “Lid” and if it is closed, the set RPM will be
displayed. Please ensure that the Tube Holders are inside the rotor.
11.1 SETTING THE SPEED AND TIME
- To set the speed press the RPM/RCF button first, then with the use of increment & decrement buttons to set the desired speed.
- The increase and decrease of speed will take place in the steps of 10 RPM. The Display window at the bottom will show the values.
- The minimum and maximum speed of the centrifuge is 500 RPM and 6500 RPM respectively. Speed is to be read as value x 10 in the display window.
- The speed can be changed during the operation of the centrifuge, upon changing the speed, centrifuge will start running on the updated speed for the remaining time left on the centrifuge cycle.
- To set time press the TIME button first, then use INCREMENT and DECREMENT buttons time can be set as desired time. The set time will be displayed in the display window at the bottom.
- The time range in this centrifuge allows user to choose between 1 min & 30 minutes & infinite mode. Infinity mode displayed as ][.
- The centrifuge incorporates a count down timer, after each minute of operation the value of the timer is reduced by 1 minute.
NOTE: Centrifuge will not run with the lid open.
11.2 SWITCHING TO RCF DISPLAY
Press the RPM/RCF to change the mode from RPM to RCF, while in RCF mode the
speed value is to be read as value x 10 in the display window. The maximum RCF
of the
centrifuge is 3684 / 3873 xg. Press the RPM/RCF to see corresponding RCF
value.
11.3 START & STOP OPERATION
Disconnect the centrifuge from the main supply. Wait until the rotor comes to
a standstill (this may take some time). Once the rotor has stopped, pull out
the emergency lid
release as shown in the figure. This will open the centrifuge lid.
After the centrifuge lid opens, push the thread back in the hole with your
hand and fit the emergency release cap back in the hole.
11.4 OPENING THE CENTRIFUGE LID IN CASE OF POWER FAILURE Once all the
parameters like Speed & Time are set, press the “Start”button to START
operation.
- Make sure that the lid is closed before starting the centrifuge.
- The display will keep on blinking until the set parameters are (RPM/RCF & Time) reached. Once the set parameters are reached, the display will blink multiple times till the set value would reach.
- When centrifugation is completed the lid of the centrifuge opens automatically and the timer value will again start blinking.
Press the STOP/LID button any time the centrifuge needs to be stopped. After every run the lid opens by itself, In any other case, when the lid needs to be
OPERATION
opened push the STOP/LID button to open the lid.
11.5 IMBALANCE DETECTION
This centrifuge is equipped with an imbalance detection safety feature. When
the rotor is not loaded symmetrically, imbalance detector gets activated. This
will cut off the centrifugation & error “Err55” will be shown on the display.
To correct the imbalance, open the centrifuge lid and correct the imbalance
load using methods described in the balancing the rotor section (page no.8) of
this manual. After correctly loading the tubes into the rotor turn off the
centrifuge & turn it back on.
SAFETY PRECAUTION
- Never use the centrifuge in any manner not specified in the manual.
- Never move the centrifuge while it’s operating .
- The rotor must always be securely fastened. If the centrifuge makes an unusual noise during the operation, the rotor fit needs to be checked.
- The rotor must be loaded symmetrically.
- Do not use the centrifuge or rotor, if it is not installed correctly or it is damaged.
- Do not use the centrifuge or rotor if it shows any sign of external mechanical or electrical damages.
- The centrifuge must be used only for the specified application. It must not be operated in hazardous or flammable environment & no hazardous or flammable content should be centrifuged.
- Do not fill tubes while they are in the rotor. Liquid spillage may harm the device.
- If liquid is spilled on the rotor or rotor chamber, the centrifuge must be cleaned immediately.
- “Tube Holders” are a part of the Rotor. They should be placed in the rotor whether 2,3,4,6 or 8 tubes are being used.
TROUBLESHOOTING
This centrifuge has a self – diagnostic function. If a problem occurs, an
error/warning code will be displayed on the display screen and the operator
can determine the
malfunction with the warning code below.
ERROR | PROBLEM | SOLUTION |
---|---|---|
No display | No main power connection. | Check power & properly plugin power |
adaptor at both ends.
Power failure| Check the main fuse of the lab.
13. TROUBLESHOOTING
No display | Improper connection. | Connect adaptor properly. |
---|---|---|
Err 55 | Rotor not loaded symmetrically. | Load the rotor symmetrically & |
restart the centrifuge.
Centrifuge lid cannot be opened| Rotor is still spinning.| Wait for the rotor
to come to a stop.
Power failure| Press emergency lid release after rotor stops
Centrifuge shakes during acceleration & makes loud running noises while
running| Rotor is not loaded symmetrically.| Load rotor symmetrically &
restart operation
Either a broken tube, damage to the rotor or motor is the cause for run
noise.| Replace broken tube. For damaged rotor/motor contact a service
representative.
Rotor is damaged.| Remove & change rotor
Err 52| Rotor is Stuck| Turn OFF the centrifuge, Check & fit the rotor
properly & turn ON the centrifuge
IMPORTANT NOTE:
- If the centrifuge is working slow or hot due to overdraw current, restart the centrifuge and check it again.
- Do not use liquid with density higher than 1.2g/ml for full load operation.
WARRANTY STATEMENT
This product is warranted to be free from defects in material and workmanship
for a period of two (2) years from date of purchase. Your product will be duly
repaired upon
prompt notification in compliance with the following conditions :
This warranty is valid only if the product is used for its intended purpose
and within the guidelines specified in this instruction manual. This warranty
does not cover
damage caused by accident, neglect, misuse, improper service, natural forces
or other causes not arising from defects in original material or workmanship.
This warranty does not cover any incidental or consequential damages,
commercial loss or any other damages from the use of this product.
The warranty is invalidated by any non-factory modification, which will
immediately terminate all liabilities on us for the products or damages caused
by its use. The buyer
WARRANTY STATEMENT
and its customer shall be responsible for the product or use of products as
well as any supervision required for safety. If requested the products must be
returned to the distributor in well packed and insured manner and all shipping
charges must be paid. Some states do not allow limitation on the length of
implied warranties or the exclusion or limitation of incidental or
consequential damages. This warranty gives you specific legal rights. This
warranty is given expressly in lieu of all other warranties, expressed or
implied. The purchaser agrees that there is no warranty of merchantability or
of fitness for any intended purpose and that there are no other remedies or
warranties, expressed or implied, which extend beyond the description on the
face of the agreement. This warranty is only applicable to the original
purchaser.
Products received without proper authorization will not be entertained. All
items returned for service should be sent postage prepaid in the original
packaging or other
suitable carton, padded to avoid damage. We will not be responsible for damage
incurred by improper packaging.
All items returned for service should be set postage prepaid in the original
packaging or other suitable carton, added to avoid damage.
This warranty is valid only if the warranty is registered with the supplier
within 30 days from the date of purchase.
For you reference, make a note of serial number, date of purchase and supplier
here.
Serial No.:…………
Purchase Date:………….
Supplier:…….
PRODUCT DISPOSAL
In case the product is to be disposed of, the relevant legal regulations are
to be observed.
Information on the disposal of electrical and electronic devices in the
European Community
The disposal of electrical devices is regulated within the European Community
by national regulations based on EU Directive 2012/19/EU on waste electrical
and electronic equipment (WEEE). According to these regulations, any devices
supplied after 13.06.05 in the business to business sphere, to which this
product is assigned, may no longer be disposed off in municipal or domestic
waste. They are marked with the following symbol to indicate this.
15. PRODUCT DISPOSAL
As disposal regulations within the EU may vary from country to country, please
contact your supplier if necessary.