easy home EHP035 Fingertip Pulse Oximeters User Manual

June 12, 2024
EASY HOME

easy-home-LOGO

easy home EHP035 Fingertip Pulse Oximeters

easy-home-EHP035-Fingertip-Pulse-Oximeters-PRODUCT

General Description
The Easy@Home Fingertip Pulse Oximeter EHP035 is a handheld non-invasive device intended for spot-checking of oxygen saturation of hemoglobin (SpO2) and Pulse Rate of adults, adolescents and children in the home setting. Oxygen binds to hemoglobin in red blood cells when moving through the lungs. It is transported through­out the body as arterial blood. A pulse oximeter uses two frequencies of light (red and infrared) to determine the percentage (%) of hemoglobin in the blood that is saturated with oxygen. The percentage is called blood oxygen saturation, or SpO2. A pulse oximeter also measures and displays the pulse rate at the same time it measures the SpO2 level. Follow and read the operating procedure closely as found in these instructions. Otherwise, abnormality, equipment damage, or personal injury may occur. The manufacturer warranty does not include these safety and reliability issues noted above due to user negligence.
NOTE: This oximeter it is for sport or aviation use only and is NOT intended for medical use.

Measurement Principle
The principle of the oximeter is as follows:

  • The pulse oximeter works by applying a sensor to either pointer finger.
  • The sensor contains a dual light source and photodetector.
  • The one wavelength of the light source is 660nm, which is red light; the other is 905nm, which is infrared-red light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time.
  • The photodetector in the finger sensor collects and converts the light into an electronic signal which is proportional to the light intensity.
  • Systolic blood pressure is the first number measuring the pressure in your blood vessels when your heart beats; the second number is diastolic blood pressure measuring the pressure in your blood vessels as well, but when the heart is resting in between beats.
  • The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement.

This measurement is referred to as SpO2. Diagram of Operation Principle:

  1. Red and Infrared-ray Emission Tube
  2. Red and Infrared-ray Receipt Tube

Box Content

  1. Fingertip pulse oximeter
  2. One lanyard
  3. One Carrying Package
  4. Two AAA batteries
  5. One instruction manual

Product Features

  1. Simple operation, easy to carry.
  2. Light weight and low power usage.
  3. Dual color OLEO displays SpO2, PR, Pl, Pulse bar, and waveform.
  4. 7 display modes
  5. Level 1-10 adjustable brightness.
  6. 2pcs AAA-size alkaline batteries; real-time battery status indication.
  7. Multiple users.

Operation Instructions

  1. Install two AAA batteries according to the Battery Installation instructions.
  2. Place one of your fingers into the rubber opening of the pulse oximeter.
  3. Short press the switch button on the front panel to turn the pulse oximeter on.
  4. Keep your hands still for the reading. The hand should be rested on the chest at the level of the heart. Do not shake your finger during the test. It is recommended that you do not move your body while taking a reading either as it will sway results.
  5. Read the data from the display screen when the waveform is steady and forming a consistent pattern. This is the optimal value for measurement. 120/80 is considered normal for blood pressure, but this may vary. For adults, a normal pulse is 60-100 beats per minute, but this may also vary. Speak with your doctor about any concerns.
  6. The display modes are as follows:easy-home-EHP035-Fingertip-Pulse-Oximeters-FIG-2easy-home-EHP035-Fingertip-Pulse-Oximeters-FIG-3

Notes

  1. Short press the power button to switch the display modes.
  2. Long press the power button to adjust the brightness of the oximeter. There are 10 levels of brightness. The default is level four.
  3. Take out your finger, the screen displays “Finger Out”. It means the measurement has ended.

Front Paneleasy-home-EHP035-Fingertip-Pulse-Oximeters-
FIG-4

Notes

  1. The pulse bar less than 30% indicates signal inadequacy and the displayed SpO2 or pulse rate value is potentially incorrect.
  2. If the screen displays”?”, it means the signal is unstable, please take out your finger and put it back in the oximeter then keep your hands still and retry.easy-home-EHP035-Fingertip-Pulse-Oximeters-FIG-5

Battery Installation

  1. Slide the battery door cover horizontally along the arrow shown in the picture.
  2. Install two AAA batteries into the battery compartment. Match the polarity of the compartment. If the polarities are not matched, damage may be caused to the oximeter.
  3. Close the battery door cover.

Notes

  • Please remove the batteries if the pulse oximeter will not be used for long periods of time, as corrosion can occur.
  • Please replace the battery when the power indicator starts flickering.easy-home-EHP035-Fingertip-Pulse-Oximeters-FIG-6easy-home-EHP035-Fingertip-Pulse-Oximeters-FIG-7

Using the lanyard

  1. Thread the thinner end of the lanyard through the loop.
  2. Thread the thicker end of the lanyard through the threaded end before pulling it tightly.

Warnings

◊ Keep the oximeter away from young children. Small items such as the battery door, battery, and lanyard are choking hazards.
◊ Do not hang the lanyard from the device’s electrical wire.
◊ Please notice that the lanyard, which is tied to the oximeter, may cause strangling due to excessive length.easy-home-EHP035-Fingertip-Pulse-
Oximeters-FIG-8easy-home-EHP035-Fingertip-Pulse-
Oximeters-FIG-9

Inaccurate measurements may be  caused by

  1. Significant levels of dysfunctional hemoglobin (such as carbonyl – hemoglobin or methemoglobin).
  2. lntravascular dyes such as indocyanine green or methylene blue.
  3. High ambient light. Shield the sensor area if necessary.
  4. Excessive patient movement.
  5. High-frequency interference and defibrillators.
  6. Venous pulsations.
  7. Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line.
  8. The person has hypotension, severe vasoconstriction, severe anemia, or hypothermia.
  9. he person is in cardiac arrest or is in shock.
  10. Fingernail polish or false fingernails.
  11. Weak pulse quality (also known as low perfusion) .
  12. Low hemoglobin.

Maintenance and Storage

  1. Replace the batteries in a timely manner when low voltage lamp is lighted.
  2. Clean surface of the fingertip oximeter before it is used.
  3. Remove the batteries if the oximeter will not be operated for a long time.
  4. It is best to store the product in -13°F -158°F (-25°C-+ 70°C) and,; 93% humidity.
  5. Keep in a dry place. Extreme moisture may affect oximeter’s lifetime and may cause damage.
  6. Dispose batteries properly; follow any applicable local battery disposal laws.

Cleaning the fingertip pulse oximeter

  • Please use medical alcohol to clean the silicone by touching the finger inside of oximeter with a soft cloth dampened with 70% isopropyl alcohol. Also clean the finger being tested using alcohol before and after each test.
  • Do not pour or spray liquids onto the oximeter, and do not allow any liquid to enter any openings in the device. Allow the oximeter to dry thoroughly before reuse.
  • The fingertip pulse oximeter requires no routine calibration or maintenance other than the replacement of batteries.

NOTE: The use life of the device is five years when it is used for 15 measurements every day and 10 minutes per one measurement. Stop using and contact the manufacturer if one of the following cases occurs:

  • Troubleshooting error notifications are displayed on screen.
  • The oximeter cannot be powered on if you have ruled out any possible battery issues.
  • There is a crack on the oximeter or damage on the display resulting in readings not being able to be identified; the spring is invalid; or the key is unresponsive or unavailable.

Disinfecting

  • The applied parts touching the finger should be disinfected once after each use.
  • The recommended disinfectants include: alcohol 70%, isopropanol 70%, and glutaraldehyde-type 2% liquid disinfectants.
  • Note : Disinfection may cause equipment damage. Therefore only disinfect the rubber cushion of the clip.
  • CAUTION : Never use EtO or formaldehyde for disinfection.

Precautions for Use

  1. Before use, please read the manual carefully.
  2. The fingertip pulse oximeter must be able to measure the pulse properly to obtain an accurate SpO2 measurement. Verify that nothing is hindering the pulse measurement before relying on the SpO2 measurement.
  3. Explosive hazard-do not use the fingertip pulse oximeter in an environment with flammable gas such as ignitable anesthetic agents.
  4. The fingertip pulse oximeter is intended only as an adjunct in assessment. It may be used in conjunct tion with other methods of assessing.
  5. In order to ensure correct sensor alignment and skin integrity, the maximum application time for a single use session of this device should be less than half an hour. It is not for continuous monitoring.
  6. Do not sterilize the device using autoclaving, ethylene oxide sterilizing, or immersing the device in liquid. The device is not intended for sterilization.
  7. Follow local ordinances and recycling instructions regarding disposal or recycling of the device anddevice components, including batteries.
  8. This equipment complies with IEC 60601-1-2:2014 for electromagnetic compatibility for medical electrical equipment and/or systems. However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in healthcare and other environments, it is possible that high levels of such interference due to close proximity or strength of a source might disrupt the performance of this device.
  9. Portable and mobile RF communications equipment can affect electrical equipment.
  10. This equipment is not intended for use during transport.
  11. This equipment should not be used adjacent to or stacked with other equipment.
  12. It may be unsafe to:
    • use accessories, detachable parts and materials not described in the instructions for use
    • interconnect this equipment with other equipment not described in the instructions for use
    • disassemble, repair or modify the equipment
  13. These materials that contact with the patient’s skin contain medical silicone and ABS plastic enclosure are all pass the ISO10993-5 Tests for invitro cytotoxicity and ISO10993-10 Tests for irritation and delayed-type hypersensitivity.
  14. When the signal is not stable, the reading may be inaccurate. Please do not rely on

SPECIFICATION

  1. Display Type
    • OLED display
  2. Sp02
    • Display range : 0%-100%
    • Measurement range : 70%-100%
    • Accuracy : 70%-100%±2%; 0%-69% no definition
    • Resolution : 1%
  3. Pulse Rate
    • Display range : 30bpm-250bpm
    • Measure range : 30bpm-250bpm
    • Accuracy : 30bpm-99bpm, ±2bpm; 100-250bpm, ±2%
    • Resolution : lbpm
  4. Perfusion Index
    • Display range : 0.1%-20.0%
    • Measure range : 0.2%-20.0%
    • Accuracy : 0.2%-1.0%, ±0.2digits; 1.1%-20.0%, ±20%
    • Resolution : 0.1%
  5. Probe LED Specifications easy-home-EHP035-Fingertip-Pulse-Oximeters-FIG-12

Power Requirements
Two AAA alkaline Batteries

  • Power consumption: Less than 40mA
  • Battery Life: Two AAA 1.5V, 1200mAh alkaline batteries could be continuously operated as long as 24 hours.

Environment Requirements

  • Operation Temperature: 41 °F -104°F (5°C-40°C)
  • Storage Temperature: -13°F -+158°F (-25°C-+ 70°C)
  • Ambient Humidity: 15% ~ 93% no condensation in operation; :;;;93% no condensation in storage/transport
  • Atmosphere pressure: 70kPa-106kPa

Equipment Response Time
As shown in the following figure. Response time of slower average is 8s .easy-home-EHP035-Fingertip-Pulse-Oximeters-FIG-11

Classification

  • According to the type of protection against electric shock: INTERNALLY POWERED EQUIPMENT
  • According to the degree of protection against electric shock: TYPE BF APPLIED PART (The application part Accordina to the dedree or protection adains electro ShOCK’ IYPEBRAPPIERARTe abolcation Dart is rubber inside of the Pulse Oximeter)
  • According to the degree of protection against ingress of water : IP22 According to the mode of operation: CONTINUOUS OPERATION

ARMS Value Analysis Statement easy-home-EHP035-Fingertip-Pulse-
Oximeters-FIG-13

Bland-Altman Plot Graphic

easy-home-EHP035-Fingertip-Pulse-Oximeters-FIG-10

Troubleshooting

easy-home-EHP035-Fingertip-Pulse-Oximeters-
FIG-14

Symbol Definitions

easy-home-EHP035-Fingertip-Pulse-Oximeters-FIG-15

HeaIth care-Manager.com
Questions or comments? Please call toll- free:1-855-822-6999 M-F 9 a.m.-5 p.m. CST E-mail: [email protected] Easy Healthcare Corporation 360 Shore Dr. Unit B, Burr Ridge, IL 60527 Made in China

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

EASY HOME User Manuals

Related Manuals