Easy Home EBT-300 Smart Basal Thermometer
INTENDED USE
The EBT-300 Digital Basal Thermometer is designed to accurately measure a person’s body temperature in its regular mode. With its 1/1 Oath Degree High Precision feature, it can assist women of childbearing age in monitoring their basal body temperature (BBT) effectively. The product is used in oral cavity.
WHY EBT-300?
- Upon setting up, data auto sync to the free Premom app.
- Large and backlit LCD display helps you read results easily.
- High Precision of 11100th degree for temperature readings.
CONTENTS
1 Thermometer.
1 User manual.
1 Storage case.
SET UP AND USE
- Refer the quick guide of Premom App to set up the App before BBT measurement.
- Taking your temperature.
Step1: Keep the smart thermometer near your bed. Make sure to take your basal body temperature each morning at the same time for the best results.
Make sure your lips have been closed for at least five minutes without eating, speaking or getting up before taking your temperature.
Step2: Upon waking, turn on the smart thermometer. Put the metal probe under the tongue and close the lips tight. Hold it until you hear 2 short beeps, which indicates the completion of the measurement. This usually takes 90 to 120 seconds.
DATA STORAGE
Memory stores up to 30 measurements on the device. While the thermometer is off double-click the On/Off button to review the stored data. You can exit the memory display mode by holding down the On/Off button for 3 seconds or it will shut off automatically after 30 seconds.
CHANGING THE BATTERY
Please replace the battery when the low-battery symbol appears on the right of the LCD display. Remove the battery cover and replace it with a battery of the same battery type CR2032. Make sure the “+” sign faces up. Then replace the battery cover. Always dispose of used batteries according to local laws and regulations.
TROUBLESHOOTING
-
Problem: Unstable body temperature readings
Possible Solutions: Ensure the thermometer probe is placed firmly underneath your tongue, close your lips, breathe normally and do not speak, eat, drink, or move. -
Problem: Abnormally high BBT
Possible Solutions: Take measurement immediately after waking up, limit physical activities before and during measurement.
Note:
- If the problem persists, please contact customer support or consult your doctor if concerned.
- If you have any problems with this device, such as setting up, maintaining or using, please contact with Easy@Home customer service. Don’t open or repair the device by yourself.
MAINTENANCE & CLEANING
- Clean the tip of the thermometer with a dry cloth dampened slightly with 75% rubbing alcohol and 25% water before and after every use.
- Do not use cleaning agents other than alcohol or water to clean the thermometer as it may damage or decrease the lifetime of the product and / or present safety risks.
- The thermometer is not waterproof. Never submerge the thermometer in water or any other liquid, this can damage the thermometer, and result in incorrect data, and provide a safety risk.
TECHNICAL SPECIFICATIONS
Product name …………………………….. Digital Thermometer
Model . . . . . . . . . . . . . . . . .. . . . .. . . . . .. . . . . . .
. . . . . . . . . . . . . . . . . . . .. . . EBT -300
Size ………………………………….. 5.20 in x 1.15 in x 0.55 in
Transmission distance ……………………………………. 30 feet
Measurement ………………………………… 89.6°F – 109.4 °F
Accuracy … ±0.18°F/ 0.10°c (from 95.0°F to 102.2°F/ 35°C-39°C) ±0.36°F/
0.20°c (from 89.6°F to 95.0°F/ 32°C-35°C, and from 102.2°F to 109.4 °F/
39°C-43°C)
Laboratory accuracy ………… ±0.09°F/ ±0.05°C {from 95.00°F to 102.20°F/
35.00°C-39.00°c) ±0.18°F/ ±0.10°C (the rest temperature)
Battery ………………………………………………….. CR2032
Receiver ……………………. Apple Phone: iPhone5+ iOSB.0+ Android Phone: Android
4.4+
Battery life ……………………………………… about 3 months
Operating conditions ……………. Temperature: 41°F to 104°F
Weight …………………………….. About 20g (without battery)
Service life …………………………………………………………… 5 years (Please refer to the box for
production date and batch information)
Temperature in working environment ………………………… 5-40°C
Humidity in working environment ………………… 15% to 85% RH
Atmospheric pressure in working environment ….. 86Kpa to 106 Kpa
Temperature in storage environment …………………………. -20-55°C
Humidity in storage environment …………………………….. S85% RH
Atmospheric pressure in storage environment : 86Kpa to 106 Kpa
WARRANTY
- This product is warranted from manufacturing defects for one year from the date of retail purchase. It does not cover damages or wear resulting from an accident, misuse, abuse, commercial use, or an unauthorized adjustment or repair of the product.
- Please direct all returns to the place of original receipt, and you may be asked to provide proof of purchase. To find the customer service menu, please visit our official website at www.premom.com
PRECAUTIONS
- Do not bite, bend, drop or take apart the thermometer.
- Keep the device out of the reach of children/pets to avoid inhalation or swallowing of small parts.
- Do not expose it to direct sunlight, high temperature and moisture.
- Not intended to be sterilized. Prevent saliva or cleaning solution from penetrating the display window.
- If LCD shows the symbol or LCD shows unclear it means the battery has run out. Please replace the battery within 10 minutes.
WARNING
- No servicing/maintenance while the thermometer is in use.
- Not for use in an OXYGEN RICH ENVIRONMENT
- Before every use, check the device. Do not use the device or an electrode if it is damaged in any way. The continuous use of a damaged unit may cause injury, improper results, or serious danger.
- If you have any problems with this device, such as setting up, maintaining or using, please contact with our customer service.
- Don’t open or repair the device by yourself.
- Please report to us if any unexpected operation or events occur.
- The main material of the case is ABS. Be careful to the potential allergic reactions to these materials.
- The typical service life of the new and unused batteries is 100 hours for continuous operation.
- Protection against electric shock: Internally powered ME equipment.
- Protection against harmful ingress of water or particulate matter: IP22.
Below improper operations will affect measuring and cause inaccurate readings:
- Operation outside the manufacturer’s staled temperature and humidity range.
- Storage outside the manufacturer’s stated temperature and humidity range.
- Mechanical shock (for example, drop test).
- Patient temperature is below ambient temperature (operating environment see info in Technical Specification part).
CAUTION
Keep new and used batteries out of the reach of children, to seek immediate medical attention if a battery is ingested, and to follow any other consensus medical advice. Choking Hazard-Small parts not for children under 3 years or any individuals who have a tendency to place inedible object in their mouths.
CALIBRATION
The thermometer is initially calibrated at the time of manufacture. If the thermometer is used according to the use instruction, periodic readjustment is not required.
DISCLAIMER
This thermometer is not a medical device and is only used for fertility tracking.
EMC/FCC INFORMATION
- This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment.
- Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
- Caution: This unit has been thoroughly tested and inspected to assure proper performance and operation!
- Caution: this machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be used.
Guidance and manufacture’s declaration electromagnetic emission
The EBT-300 is intended for use in the electromagnetic environment specified
below.
The customer of the user of the EBT-300 should assure that it is used in such
an environment.
EMISSION TEST| COMPLIANCE| ELECTROMAGNETIC ENVIRONMENT
GUIDANCE
RF emissions CISPR11| Group 1| The EBT-300 use RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emission CISPR11| Class B| The EBT-300 is suitable for use in all
establishments, including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies buildings used
for domestic purposes.
Guidance and manufacture’s declaration electromagnetic immunity
The EBT-300 is intended for use in the electromagnetic environment specified
below.
The customer or the user of EBT-300 should assure that it is used in such an
environment.
IMMUNITY TEST| IEC60601 TEST LEVEL| COMPLIANCE LEVEL|
ELECTROMAGNETIC ENVIRONMENT GUIDANCE
Electrostatic discharge (ESD) IEC 61000-4-2| ±6 kV contact
±8 kV air| ±6 kV contact
±8 kV air| Floors should be wood, concrete or ceramic tile. If floor are
covered with synthetic material, the relative humidity should be at least 30%.
Power frequency (50Hz/60Hz) magnetic field IEC 61000-4-8| 3Alm| 3Alm| Power
frequency magnetic fields should be at levels characteristic of a typical
location in a typical commercial or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test
level.
Guidance and manufacturer’s declaration electromagnetic immunity
The EBT-300 is intended for use in the electromagnetic environment specified
below.
The customer or the user of the EBT-300 should assure that it is used in such
an environment.
IMMUNITY TEST| IEC60601 TEST LEVEL| COMPLIANCE LEVEL|
ELECTROMAGNETIC ENVIRONMENT GUIDANCE
Conducted RFIEC 61000-4-6| 3Vrms 150 kHz to80 MHz| Not Applicant| Portable and
mobile RF communications equipment should be used no closer to any part of the
EBT-300, including cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended separation distance
80 MHz to 800 MHz
800MHz to 2,5 GHz
Radiated RFIEC 61000-4-3| 3Vlm80 MHzto2.5 GHz| 3Vlm| Where P is the maximum
output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in
metres (m). Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, a should be less than the compliance level in
each frequency range. b Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1:At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2:
These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a: Field strengths from fixed transmitters, such as base stations for
radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the EBT-300 is used exceeds
the applicable RF compliance level above, the EBT-300 should be observed to
verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the EBT-300.
b : Over the frequency range 150 kHz to 80 MHz, field strengths should be
less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equipment and the EBT-300 .
The EBT-300 is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the
EBT-300 can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment
(transmitters) and the EBT-300 as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output power of transmitter(W)| Separation distance
according to frequency of transmitter (m)
150 KHz to 80 MHz
**| 80 MHz to
**| 800 MHz 800 MHz to
0.01| 0.12| 0.12| 0.23
0.1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distanced in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where Pis the maximum
output power rating of the transmitter in watts 0N) according to the
transmitter manufacturer.
NOTE 1 : At BO MHz and BOO MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
FCC ID: 2ADNQBTA41 CNBT
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions:
(1) This device may not cause harmful interference,
(2) This device must accept any interference received, including interference
that may cause undesired operation.
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation.
If this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the
following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
EXPLAINATION OF SYMBOLS
| Refer to instruction manual
---|---
| Type BF applied part
| Ingress protection rating
| **** Radio frequency transmitters
| Manufacturing date
| Balch code
| Temperature limit of -4°F-131°F (-20°c -55°CJ
| Humidity limitation less than 85%
| Atmospheric pressure limitation of 86 Kpa -106 Kpa
| WEEE (Waste Electrical and Electronic Equipment)
Customer Support
Download Free Premom APP to log and track your BBT each day for ovulation
prediction.
www.premom.com
Questions or comments?
Please call toll-free:
1-855-822-6999 M-F 9 a.m.-5 p.m.
CST
E-mai l: support@premom.com
Manufactured for Easy Healthcare Corporation
360 Shore Dr., Burr Ridge, IL USA 60527
Made in China
Documents / Resources
| Easy
Home EBT-300 Smart Basal
Thermometer
[pdf] User Manual
EBT-300 Smart Basal Thermometer, EBT-300, Smart Basal Thermometer, Basal
Thermometer, Thermometer
---|---
| easy
home EBT-300 Smart Basal
Thermometer
[pdf] User Manual
EBT-300 Smart Basal Thermometer, EBT-300, Smart Basal Thermometer, Basal
Thermometer, Thermometer
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>