ROPOX TF 200.10.0005_ENG Toilet Lifter User Manual
- October 30, 2023
- ROPOX
Table of Contents
ROPOX TF 200.10.0005_ENG Toilet Lifter User Manual
Symbols used in this manual
Warning! Indication of potentially hazardous situation. If not
avoided, it can result in serious injury or death.|
---|---
Caution! Indication of potentially hazardous situation which may
result in minor or moderate injury. It may also be used to alert against
unsafe practices.|
Notification! This symbol is used to notify correct use and handling
of the product.|
General safety
| This manual must be read and understood before use and installation.
Always keep this manual in close proximity of the product. The use,
installation and service of this product must be in compliance with this
manual to avoid accidents and serious personal injury. Never use or handle
this product in other ways specified in this manual as it can result in
personal safety hazards and/or cause damage to the product. Persons installing
and/or using this product either as operator or user should have the necessary
safety information and access to this manual.
---|---
| Do not perform repairs, disassembly or assembly operations, add-ons,
re-adjustments, or modification of the product beyond what is described in
this manual. These must be carried out by Ropox or Ropox authorized personnel.
Do not perform service while in use. This appliance can be used by children
aged from 8 years and above and persons with reduced physical, sensory, or
mental capabilities or lack of experience and knowledge if they have been
given supervision or instruction concerning use of the appliance in a safe way
and understand the hazards involved. Children shall not play with the
appliance. Do not use the product if it has defects or have become damaged
before being repaired or replaced. Make sure cables are undamaged.
****| The information in this manual is based on correct installation in
accordance with installation instructions for this product. Ropox cannot be
held liable if the product is used in any way that differs from stated in this
manual and/or installation instruction. Ropox reserve the right to amend this
manual and reference documents without prior notice.
Product Unit label
__| This product is CE-marked in accordance with: __ European
Medical Device Regulation (EU)2017/745
---|---
__| Manufacturer name and address
__| Date of production
__| Stock number
__| Serial number
__| Consult manual for important safety related information, warnings,
and safety precautions
__| __ Do not dispose as unsorted municipal waste. Product must be
returned to a designated recycling station
__| __ Operating and storing temperatures may not exceed 5-40°C
__| __ Consult the instructions for use
General requirements
Product information
Manufacturer| Ropox A/S Ringstedgade 221 DK-4700 Naestved+45 55 75
05 00Info@ropox.com
---|---
Product models| Part number| Model| Configuration
40-4502640-46026| Toilet lifter Manual Toilet lifter Manual| Floor drain Wall
drain
Expected service life| 5 years
MDR Class 2017/745| Class I
Intended environments| This product is to be used only in: __
Professional Healthcare Environment Homecare Environment
Maximum user weight according to ISO 17966:2016| 400kg
Height adjustment| 400-620mm
Ambient temperature range| -10 °C to +50 °C Transport and storage+5
°C to +40 °C Operation
Relative humidity| 20% to 80% – non-condensing
Materials in contact with patient| Component __ Covers| Material
description __ ABS/PMMA plastic
Accessories
Item nr.| Model
40-45011| Height adjustable toilet supports, Short straight
40-45012| Height adjustable toilet supports, Long straight
40-45013| Height adjustable toilet supports, Short Wave
40-45014| Height adjustable toilet supports, Long Wave
40-45015| Height adjustable toilet supports, Bariatric Short straight
40-45016| Height adjustable toilet supports, Bariatric Long straight
40-45018| Height adjustable toilet supports, Bariatric Long Wave
40-45008| Clamps, for height adjustment without tools
40-45100| Backrest long (70cm Toilet)
40-45101| Backrest short (56cm Toilet)
40-44070| Wall hung toilet, 56cm
40-44071| Wall hung toilet, 70cm
40-44074| Toilet seat w/o lid
40-44075| Toilet seat w/ lid
40-44276| Svancare Bidet Elegant
Product description
The ROPOX manual toilet lifter can be set in heights ranging from
400-620mm by the use of a crank handle. This can be done on either side of the
toilet lifter if space is limited. Height adjustment will meet different
sitting height needs from various patient heights.
Intended use
The ROPOX height adjustable toilet lifter is intended to be used in both
homecare and professional
healthcare environments. Adjusting the height will meet ergonomic demands
emerging from various patient heights. The toilet lifter is not intended for
playing, or as a training device.
Intended operator
The intended operator may include all that have read and understood this
manual exempt kids.
Essential performance
The device does not have any function related to basic safety or essential
performance. In case of loss or degradation of performance, the user can get
of the toilet.
Non-clinical functions
Height adjusting offers better ergonomics to user and caretaker.
Clinical functions
This product does not offer clinical benefits for patients, when used as
intended by the manufacturer.
Product dimensions
Instructions for use
Installation of product
The following instructions describe how to assemble and mount the toilet lifter, as well as optimal placement of water and drain, based on whether the toilet lifter with floor drain or wall drain has been purchased.
Installation interface for water
It is recommended to always place the water connection as shown below. In
case of deviations, it’s recommended to place the water within the hatched
area.
For 40-45026
For the toilet lifter with a floor drain, it’s recommended to place the drain
within the hatched area, not deviating further than the shown tolerances. This
ensures the best flushing. The drain can be raised up to 40mm, so as to live
up to Säker Vatten.
Installation interface for wall drain
For 40-46026
For the toilet lifter with a wall drain, it’s recommended to place the drain within the hatched area, not deviating further than the shown tolerances. This ensures the best flushing.
NB! All dimensions in mm
Ropox supplied toilets| Shortening of ferrule| Shortening of
flushpipe
---|---|---
**40-44070
| |
**40-44071 **
| | **
Construction
Note!
The toilet lifter is supplied with a set of 6 screws and rawlplugs suitable
for mounting in a concrete wall. The fitter should always consider the
material, condition, and strength of the wall and use screws and rawlplugs
suitable for the specific wall type.
Caution!
Maximum pulling force per screw is 44kg, at 400kg max user weight according to
ISO 17966: 2016
Mounting the product
a. Line up frame with adjustment screws. Mark out 6 holes and drill holes
for appropriate rawlplugs. Ensure that the center of the bottom 2 holes is at
least 64mm above the floor
b. Connect water hose on wall outlet and electric outlet before securing frame to the wall
c. Secure frame to wall
d1. For 40-45026
If drain is in floor, mount the drain bellow as instructed below. The examples
given below to the left, are configurations that will work, but as a general
rule, the installation interface in 4.1.2 should be followed.
Make sure the bellow is not twisted
Mount the bellow as shown below for correct flow
d2. For 40-46026
If drain is in wall, mount the drain bellow in wall drain, mount the
accompanying brackets on the wall, so
that the toilet flange is held in place
Make sure the bellow is not twisted
e. Cut threaded rods to length and secure them in the frame
f. Mount inner covers 1 and 2
g. Mount outer cover
h. Shorten ferrule and flush pipe to length and mount them
i. Mount the toilet and press pad
j. Finish the installation by checking the following:
- Rigidity test, sit on the toilet and check screw connections are firm.
- Check that full travel of the product goes unhindered.
- Check flushing and leaks.
Operating the product
Adjusting the height is done by turning the crank.
Warning!
The product should always be able to travel the full range of actuation
without colliding with objects. Failing to do so will compromise the product
stability. Special care should be taken in ensuring that no children or adults
are sited under the product as severe injury from entrapment may occur.
Residual risks
Warning!
Be aware that there may be no person or body parts under the toilet when it
is driven down, as it is a crushing hazard.| __
---|---
Caution!
Do not place your hand/fingers between the wall and the movable cover, as
you risk crushing your hand/fingers.| __
Warning!
In case of incorrect installation of plumbing parts, water leaks can occur
on the floor, which may increase the risk of falling.| __
Installation of accessories
Installation of height adjustable toilet support arms
a. Attach housing to the profile with 3x M6x12 and 1x M6x16. Use a
locking nut on M6x16.
b. Make sure that the plastic cover is pressed as far onto the frame as
possible. Mount 2x M6 buttonhead screws in the back holes on both sides. Leave
5mm of clearance between underside of screw head to the plastic cover.
c. Slide the assembly onto the two screw heads and attach the assembly with the two M6x30 as shown.
If need be, the two M6x30 can be loosened, and the support arms can be
adjusted in height to better fit the user. The screws must be fastened after
adjustment.
d. Test the toilet support arms in the horizontal direction. If they can
twist significantly, dismount the assembly and leave <5mm between the
underside of the screw head to the plastic cover. There should be no clearance
between the aluminum bracket and plastic cover. Adjust until the support arms
are properly fastened.
e. Slide in the cover and end rubber caps.
Installation of backrest
If Ropox Back Support long is ordered together with Ropox Toilet Lifter, there
will be pre-drilled holes in the cover cap. If the back support is ordered
later for subsequent mounting, new holes must be drilled. Please see the
dimension drawing for the location of holes.
Two M5x16 are used.
a.
b.
Installation of bidet
To mount Svancare Bidet Elegant, see TF 200.01.0025.
Cleaning
Cleaning can be done with lukewarm water and a suitable all-purpose cleaning agent. Use a cloth or soft sponge to wash the product. Dry the product with a wrung cloth.
Notification!
Do not use cleaning agents containing abrasive e.g. scouring powder, steel wool, scouring sponge.
This product is not designed to be sterilized. Autoclaving and sterilization beyond normal means may change the product safety and function
Preapproved cleaning products
Tradename | Solution |
---|---|
Acticlor plus | 1000ppm |
S90 sanirens | 1-3% peracid |
FUTUR, Alkalisk rens | 1% Solution [pH 9,5] |
Maintenance
Notification!
Failing to comply with periodic maintenance may result in degradation of
the product function and safety.
Periodic maintenance
Component| Maintenance activity| Maintenance interval|
Registration date of maintenance
---|---|---|---
Geberit Sigma 12 cistern| Descaling / change of worn components| 6-12 months
dependent on calcium content in water|
Maintenance of pipes and hoses
If need be, it’s possible to open the lower cover plate, by taking the lifter
to the highest position, and sliding the service hatch up. Never attempt to
service the toilet lifter while adjusting the height.
Component parts list
Notification!
Ropox will make parts list available for extraordinary maintenance and repair
beyond what is described in this manual. Repair beyond what is described in
this manual may only be done by Ropox designated service personnel.
Environmental protection
The product is not intended to be disposed as municipal waste. Proper disassembly, sorting and disposal of components must be done by waste disposal professionals.
Part | Disposal recommendation |
---|---|
Frame | Steel |
Cistern | Hard plastic Polystyrene |
Covers | Hard plastic (ABS/PMMA) |
Height adjustable toilet support arms | Steel |
Accident reporting
The manufacturer of this product recommends any accident with this product that meets all the following criteria reported to the appropriate authority as well as to the manufacturer.
- There has been or could have been an accident
- The medical device is considered to be a contributing factor to the case, and
- The event resulted or could have resulted in one of the following outcomes:
- Death – a patient, a user or a third party’s death
- A serious deterioration of a patient’s health, a user’s health or a third party’s health status in the form of:
- a. Life-threatening disease
- b. Lasting damage or injury
- c. Necessary medical or surgical treatment to divert life-threatening illness, permanent injury
- d. Fetal death, fetal injury, or congenital abnormalities