INVACARE 1558208 Softform Premier Active 2 Hybrid Mattress and Pump User Guide
- June 4, 2024
- INVACARE
Table of Contents
1558208 Softform Premier Active 2 Hybrid Mattress and Pump
User Guide
Pos. | Description | Function |
---|---|---|
A | Mute button | The relevant audible/visible indicator turns on when low |
pressure, power failure or alternate failure is recognized. To mute the
audible signal, press the Mute button. The visible indicator will flash until
the problem is solved. If the problem indicated has not been solved within 10
minutes, the indicator will sound again. Press the Mute button to mute the
audible indication, the indicator will not sound again. Refer to Identifying
and repairing faults on page 3 of this guide.
B| System in cycle/Setup indicator (Operative LED)| Green LED indicator
flashes whilst system is setting up — Solid green LED indicates normal
operation.
C| Alternating system/Timing failure indicator| Red LED indicator flashes and
audible buzzer sounds if the system recognizes a cycle time problem.
D| Power failure indicator| Red LED indicator flashes and audible buzzer
sounds if the system has no power connected. This could be in the event of a
power failure, for example, caused by a power cord unplugged or power off when
the control unit is working. When power is restored, the control unit will
automatically start working.
E| Pressure failure/Low-pressure indicator| Red LED indicator flashes and
audible buzzer sounds if the system recognizes a low-pressure failure.
Operation
- The power indicator will flash and the control unit will enter static mode initially until pressure reaches 15 mmHg. Then control unit will enter an alternate mode and the power indicator will remain on. The cycle time is set at 10 min, and the pressure is set at 60 ± 3 mmHg.
The first inflation of a new mattress: the indicator function (visible and audible indicators) will activate if the mattress does not inflate completely within 15 minutes.
Operation failures
- During control unit operation, the low-pressure indicator will activate within 1 minute if air hoses are disconnected or air cells are broken.
- In the case of abnormal alternate or no alternate, the audible signal will be activated and the alternate failure indicator will flash.
Installing the system
-
Hang the control unit by means of the 2 built-in hangers A at the end of the bed or place it on a horizontal surface. (Placing the control unit on the floor will not affect the performance, but may expose it to accidental damage.)
-
Connect the CPR hose E to air outlets on the mattress and control unit B, ensuring that the hose is not kinked or twisted. Allow space for bed rails to drop freely.
-
Connect the mains power cable to the power supply socket C of the control unit and a suitable outlet.
-
Switch on mains power of the outlet if present.
-
Move the mechanical switch D on the left side of the Control unit to the On position. A faint single beep sounds and the system in cycle light flashes while the system powers up.
CPR procedure
The Invacare® Softform Active® 2 Rx mattress has been fully tested to comply with the current CPR standard of 5 – 6 cm compression depth. This was achieved at all stages of inflation/deflation.
-
Hold down Red CPR button A.
-
Pull hose connector firmly away from the control unit B.
-
Switch off the control unit.
The mattress will start to deflate. The deflation time is 20 seconds. -
When CPR is complete reactivate the system following the section Installing the system.
Identifying and repairing faults
There are audio and visual alarms present on the control unit.
Problem/Alarm | Cause | Solution |
---|
Mattress not inflating (not alternating properly).
Alternating system / Timing failure indication| Mattress CPR hose
disconnected.| Connect the CPR hose connector, and lock it in place.
The power cable and fuse have been checked, the control unit does not
operate.| Send control unit back to Invacare for repair.
A major leak in air cell or complete air insert.| Replace leaking air cells.
| CPR hose or tube connectors kinked or split.| Unkink or replace split CPR
hose or tube connectors.
Not alternating, rotor failure.| Send control unit back to Invacare for
repair.
No air (control unit failure).| Send control unit back to Invacare for repair.
No power/Power failure indication| Control unit off.| Check power source, turn
unit on.
The power cord disconnected.| Connect the power cord and ensure the power
source is on.
No power in the power outlet.| Switch on the power outlet (if a pole switch is
available).
Has the power outlet been repaired by an electrician?
Power outage.| Wait until the power source has power.
Fuse blown.| Change fuse on power inlet connector with spare fuse or identical
replacement only (consult a trained engineer if you are unsure how to change a
fuse).
Pressure failure/Low-pressure indication| Disconnection of CPR (connection
hose).| Connect hose properly.
Disconnection of connector tubes to air cells in air insert.| Check individual
air cells in insert are correctly connected to connector tubes.
Kinked connection hose| Make sure that there are no kinks or bends in the
hose.
Invacare Limited
Pencoed Technology Park, Pencoed
Bridgend CF35 5AQ
UK
** Yes, you can**
Documents / Resources
|
INVACARE 1558208 Softform Premier Active 2 Hybrid Mattress and
Pump
[pdf] User Guide
1558208, Softform Premier Active 2 Hybrid Mattress and Pump, Active 2 Hybrid
Mattress and Pump, Hybrid Mattress and Pump, 1558208, Mattress and Pump
---|---
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>