PeakTech 5305 A 2-In-1 pH Meter User Manual
- June 4, 2024
- PeakTech
Table of Contents
PeakTech 5305 A 2-In-1 pH Meter
Safety precautions
This product complies with the requirements of the following directives of the European Union for CE conformity: 2014/30/EU (electromagnetic compatibility), 2011/65/EU (RoHS). Damage caused by failure to observe the following information is excluded from claims of any kind. For the operational safety of the device, the following safety instructions for operating the device must be observed.
- Read these instructions carefully and make them accessible to subsequent users.
- Use this device only within its scope and specifications.
- Take all necessary safety precautions, especially when using acids.
- Wear personal safety equipment, gloves and safety glasses when handling chemical products.
- Never operate the device if it is not completely closed.
- Do not operate the device near strong magnetic fields (motors, transformers, etc.)
- Avoid strong vibrations of the device Before starting operation, the device should be stabilized to the ambient temperature (Important when transporting cold to warm rooms and vice versa)
- Do not make any technical changes to the device
- Only qualified service technicians may open the device, as well as perform maintenance and repair work
Cleaning the cabinet
Clean only with a damp soft cloth and a commercially available mild household
cleanser. Ensure that no water gets inside the equipment to prevent possible
shorts and damage to the equipment.
Introduction
The PeakTech 5305 A is used to measure the pH of various liquids and liquids.
With the removable pH probe, it is possible after the respective measurement
to thoroughly free the device and the probe from residues of the liquid and to
clean them. An automatic calibration can be carried out via three measuring
points using a key combination.
To protect the measuring electrode, a protective cap is attached to the end of
the P 5305 (this is removed during the measurement).
- Digital pH meter with illuminated display
- Liquid temperature measurement (°C/°F)
- Multi-Line, 3½-digit LC Display
- Can be calibrated with 3 calibration points
- Exchangable measuring electrode (dry stored)
- Waterproofed case
- Electrode protective cap
Controls
Measuring operation
The pH meter is used to determine the pH values of various liquids. Thanks to the display, which has a backlight, and the intuitive handling of the device, the measurement of the pH value can be carried out quickly and in a user- friendly manner.
Carrying out a pH measurement
In order to carry out the pH measurement, the protective cap protecting the
electrode of the probe must be removed. Then the device must be switched on
with the on / off switch. The background lighting is permanently on during
operation in order to highlight the measured value. If the pH meter has not
been used for a long time, has been in operation for a long time or if it has
been used in environments in which stronger external influences can occur, it
is advisable to calibrate the device (see point 4.2).
Before measuring, clean the probe with distilled water in order to exclude
possible falsifications of the measured value. Carefully dry the probe with a
cloth. Now the probe of the pH meter can be immersed in the liquid to be
measured. Make sure that only the bottom 4 cm of the probe is immersed in the
liquid. After completing the measurement, remove the pH meter from the liquid
and clean the probe again with distilled water. The PeakTech 5305 A is able to
measure the temperature of the liquid during the pH measurement. The display
of the temperature unit can be changed by pressing the Temp/Cal key (°C / °F).
The temperature display can also be updated manually by pressing and holding
the Temp / Cal key (5 sec.).
If no further measurements are to be made, put the protective cap of the electrode back onto the end of the probe to protect it from contamination or damage. Proceed as follows:
- Remove the protective cap from the electrode.
- Clean the electrode with distilled water and dry it well.
- Press “ON” to switch on.
- Immerse the electrode in the test solution and stir gently with the electrode.
- Press “Temp / Cal” to toggle between ° C and ° F. Press and hold “Temp / Cal” for 5 seconds to calibrate the electrode.
- Wash the electrode after use. Switch off and put on protective cap.
Calibration of the pH meter.
For calibration, standard buffer solutions pH6.86, pH4.00 and pH 9.18 are
required, which can be obtained from specialist dealers. The measuring device
can identify the standard buffer solution of pH 4.00 and pH 6.86 itself and
automatically calibrates this value.
- Pour the standard buffer solution pH 6.86 and pH 4.00 (at approx. 25 ° C) separately into three different clean beakers.
- Press “ON” to switch on the device.
- Immerse the electrode in a standard buffer solution pH 6.86 and then stir gently until the value is stable. Press the “Temp / Cal” button for 5 seconds. When the display value “686” corresponds to the standard buffer solution, the calibration is finished.
- Clean the electrode.
- Immerse the electrode in a standard buffer solution pH 4.00 and when the measuring value has stabilized, press “Temp / Cal” for 5 seconds.
- When the display shows “401”, the meter changes to the automatic pH4.01 calibration mode. When the display value corresponds to the standard buffer solution, calibration is complete.
- Clean the electrode.
- Immerse the electrode in the standard buffer solution pH 9.18 and stir gently until the display value is stable. If the measured value is within the acceptable error range, the calibration is complete. If not, repeat steps 4 and 6.
Note : For an even more precise calibration, fill the respective buffer solution into two different beakers. One is only used to clean the electrode and the other is only used for calibration. This allows the contamination by foreign matter to be reduced to a minimum.
Note : Perform a calibration if:
- the electrode has been replaced.
- no calibration has been carried out for a long time.
- the device has been used often and for a long time.
- a particularly high measurement accuracy is required.
Battery replacement
The PeakTech 5305 A is a battery operated measuring device. After a certain
period of use of the device, it may happen that the capacity of the batteries
is used up and new batteries have to be inserted. This becomes visible when
the display gets darker or looks a little blurry. To change the batteries,
make sure that the device is switched off. Remove the upper screw cap of the
device and remove the batteries from the battery compartment. Now insert the
new batteries of the appropriate polarity (which is shown in the housing) into
the battery compartment. Make sure to use the correct batteries that the
measuring device requires (4 x 1.5V AG13 button cells).
After changing the battery, put the screw cap on the device and screw it back
on.
Specifications
Display
| ****
3 ½-digit, LCD-display max 1999 counts
---|---
Functions| PH:
0.00 to 14.00 pH
| Temperature: 0.0°C to 50.0°C
Resolution| 0,01 pH| 0,1°C
Accuracy| ± 0,1pH| ± 1 °C
Sampling rate| 1 – 2 x Sec.
Calibration| Automatic (3 points)
Operation temperature| 0°C…+50°C (32°F…122°C); < 80% RH
Power Supply| 4 x 1,5V Batteries (AG-13)
Dimensions (WxHxD)| 188 x 35 x 35 mm
Weight| 80 g
Accessories| Manual, 4 x 1,5V Batteries (AG-13), Screwdriver
Notification about the Battery Regulation
The delivery of many devices includes batteries, which for example serve to
operate the remote control. There also could be batteries or accumulators
built into the device itself. In connection with the sale of these batteries
or accumulators, we are obliged under the Battery Regulations to notify our
customers of the following:
Please dispose of old batteries at a council collection point or return them
to a local shop at no cost. The disposal in domestic refuse is strictly
forbidden according to the Battery Regulations. You can return used batteries
obtained from us at no charge at the address on the last side in this manual
or by posting with sufficient stamps. Contaminated batteries shall be marked
with a symbol consisting of a crossed-out refuse bin and the chemical symbol
(Cd, Hg or Pb) of the heavy metal which is responsible for the classification
as pollutant:
- “Cd” means cadmium.
- “Hg” means mercury.
- “Pb” stands for lead.
All rights, also for translation, reprinting and copy of this manual or parts are reserved. Reproduction of all kinds (photocopy, microfilm or other) only by written permission of the publisher. This manual considers the latest technical knowing. Technical changings which are in the interest of progress reserved. We herewith confirm, that the units are calibrated by the factory according to the specifications as per the technical specifications. We recommend to calibrate the unit again, after 1 year. © PeakTech® 10/2021 Ehr.
PeakTech Prüf- und Messtechnik GmbH – Gerstenstieg 4 – DE-22926 Ahrensburg / Germany
- +49-(0) 4102-97398 80
- +49-(0) 4102-97398 99
- info@peaktech.de
- www.peaktech.de
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>