QSG-MGC-S Multi Gas Clip Simple User Guide
- June 9, 2024
- gas clip
Table of Contents
MGC-S Multi Gas Clip Simple
User Guide
QSG-MGC-S Multi Gas Clip Simple
READ FIRST BEFORE OPERATION
The Multi Gas Clip Simple is a personal safety device designed to detect the
presence of hydrogen sulfide (H2S), carbon monoxide (CO), oxygen (O2) and
combustible gases (LEL). Before operation, ensure that you have been proper
trained on the use of the detector and the appropriate actions to take in the
event of an alarm condition.
Display Components
Icon | Description |
---|---|
1 | Alarm Condition |
2 | Life Remaining |
3 | Calibration/Test Mode |
4 | Battery Indicator |
5 | Gas Readings |
6 | Gas Identifiers |
When a Gas Reading (5) exceeds its alarm threshold, the detector will go into
visual, vibrating and audible alarm. On the display, the icon will appear
beside the Alarm Condition (1) and the associated Gas Identifier (6) will
flash off and on continuously.
Warning – Users must familiarize themselves with the icons in both alarm
and non-alarm states.
Warning – If the display is missing icons, or cannot be clearly read,
discontinue use and contact Gas Clip Technologies.
Warning – The IR LEL sensor will not detect hydrogen (H2) or acetylene
(C2H2) gases however, due to their cross-sensitive nature, the CO sensor will
go into alarm well below 10% LEL.
Factory Default Alarms
Sensor | LOW | HIGH | TWA | STEL |
---|---|---|---|---|
H2S | 10 ppm | 15 ppm | 10 ppm | 15 ppm |
CO | 35 ppm | 200 ppm | 35 ppm | 50 ppm |
O2 | 19.50% | 23.50% | ||
LEL | 10% | 20% |
Turning On the Detector
Once activated, the detector will run continuously for 24 months. To activate,
press and briefly hold down the power/menu button. The detector will begin to
display Gas Readings (5) immediately. Each Gas Identifier (6) will show a
chasing “0” for the Gas Reading (5) while it is stabilizing and self-testing.
Once all Gas Identifiers (6) have completed the warm up and stabilization
sequence, <65 seconds, the detector is ready to detect all applicable gases.
Display will show:
Warning – Before each use, check that all sensor and alarm ports are clear
of any debris or obstructions.
Bump Testing and Calibrating
These operations can be performed manually or automatically using the MGC
Simple Clip Dock. Refer to the Multi Gas Clip Simple User’s Manual for
instructions.
Auto Zeroing
Continuously hold down the power/menu button to auto zero the detector. Display will show “ZERO” and count down from 3 to 1, followed by “ZERO OK”. Release button when display shows “ZERO OK”. The display will change to “AUTO ZERO”. Wait a few seconds, then press the button briefly when display shows “APPLY GAS”. Detector has been successfully auto zeroed.
Warning – DO NOT auto zero the detector in a combustible atmosphere.
Warning – DO NOT auto zero in temperatures above 104°F (40°C) or below
32°F (0°C).
Battery Information
This detector is designed so that the battery won’t need charging or replacing
during the life of the detector.
Detector Life Countdown
Once activated, the detector will display 24 months of life remaining.
Detector will continue to countdown and display months remaining until 30 days
remain. Display will then show days remaining until end of life.
Additional Questions?
For additional information on the Multi Gas Clip Simple gas detector regarding
programming, bump testing, calibrating, etc., please refer to the Multi Gas
Clip Simple User’s Manual. The manual and other helpful videos & information
are available for viewing or download at
www.gascliptech.com.
Contact Information
Gas Clip Technologies, Inc.
305 W. FM 1382, Suite 540
Cedar Hill, TX 75104 USA
Phone: +1.972.775.7577
Fax: +1.972.775.2483
Toll Free: 1.877.525.0808
Website: www.gascliptech.com
Contact Us:https://www.gascliptech.com/contactus.phpMGC-S QSG v1.04
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>