MEAN WELL RPS-160 Series 160W Reliable Green Medical Power Supply Owner’s Manual
- June 10, 2024
- MEAN WELL
Table of Contents
**MEAN WELL RPS-160 Series 160W Reliable Green Medical Power Supply
**
Features
- 5″x 3″ compact size
- Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
- Suitable for BF application with appropriate system consideration
- 110W convention, 160W force air
- EMI class B for class Ⅰ configuration
- No load power consumption under 0.5W by PS-ON control (G model)
- 5Vdc standby output, Power Good, Power Fail ; Remote sense for 5~15V
- Protections: Short circuit / Overload / Over voltage / Over temperature
- Operating altitude up to 3000 meters
- 3 years warranty
Applications
- Oral irrigator
- Hemodialysis machine
- Medical monitors
- Sleep apnea devices
- Pumps machine
GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN.aspx
Description
RPS-160 is a 160W highly reliable green PCB type medical power supply with a high power density on a 5” by 3” footprint. It accepts 90~264VAC input and offers various models with the output voltages between 5V and 48V. The working efficiency is up to 88% and the extremely low no load power consumption is down below 0.5W. RPS-160 is able to be used for ClassⅠ(with FG) system design. The extremely low leakage current is less than 160μA. In addition, it conforms to the international medical regulations (2*MOPP) and EMC BS EN/EN55011, perfectly fitting all kinds of BF rated “patient contact” medical system equipment.
Model Encoding
Type | Description | Note |
---|---|---|
Blank | Without 5Vsb | In stock |
G | With 5Vsb & No load power consumption <0.5W | In stock |
SPECIFICATION
MODEL| RPS □ -160-5| RPS □ -160-12| RPS □
-160-15| RPS □ -160-24| RPS □ -160-48
---|---|---|---|---|---
OUTPUT| DC VOLTAGE| 5V| 12V| 15V| 24V| 48V
RATED CURRENT (20.5CFM)| 30A| 12.9A| 10.3A| 6.5A| 3.25A
CURRENT| Convection| 0 ~ 20A| 0 ~ 9.1A| 0 ~ 7.3A| 0 ~ 4.6A| 0 ~
2.3A
20.5CFM| 0 ~ 30A| 0 ~ 12.9A| 0 ~ 10.3A| 0 ~ 6.5A| 0 ~ 3.25A
RATED POWER| Convection Note.2| 103W| 112.2W| 112.5W|
113.4W| 113.4W
20.5CFM| Note.3| 155W| 159.8W| 159.5W| 161W| 161W
RIPPLE & NOISE (max.) Note.4| 80mVp-p| 80mVp-p| 120mVp-p| 120mVp-p|
150mVp-p
VOLTAGE ADJ. RANGE(main output)| 4.5 ~ 5.5V| 10.8 ~ 13.2V|
13.5 ~ 16.5V| 22 ~ 27V| 43.2 ~ 52.8V
VOLTAGE TOLERANCE Note.5| ±4.0%| ±3.0%| ±3.0%| ±2.0%| ±2.0%
LINE REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
LOAD REGULATION| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
SETUP, RISE TIME| 1800ms, 30ms/230VAC 3500ms, 30ms/115VAC at
full load
HOLD UP TIME (Typ.)| 20ms/115VAC 25ms/230VAC at full load
INPUT| VOLTAGE RANGE| Note.6| 90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
POWER FACTOR (Typ.)| PF>0.93/230VAC PF>0.98/115VAC at full load
EFFICIENCY (Typ.)| 86%| 87%| 87%| 87%| 88%
AC CURRENT (Typ.)| 2A/115VAC 1.1A/230VAC
INRUSH CURRENT (Typ.)| COLD START 35A/115VAC 70A/230VAC
LEAKAGE CURRENT **Note.7| Earth leakage current < 160μA/264VAC ,
Touch current < 100μA/264VAC
PROTECTION| OVERLOAD| 105 ~ 135% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is
removed
OVER VOLTAGE| 5.7 ~ 6.8V| 13.8 ~ 16.2V| 17.2 ~ 20.3V| 27.6 ~ 32.4V| 55.2
~ 64.8V
Protection type : Shut down o/p voltage, re-power on to recover
OVER TEMPERATURE| TSW1: Shut down o/p voltage, recovers automatically
after temperature goes down
TSW2: Shut down o/p voltage, re-power on to recover
FUNCTION| 5V STANDBY (G model)| 5Vsb : 5V@0.6A
without fan, 0.8A with fan 20.5CFM ; Tolerance ± 2%, ripple :
50mVp-p(max.)
PS-ON INPUT SIGNAL (G model)| Power on: PS-ON = “Hi” or ” >
2 ~ 5V” ; Power off: PS-ON = “Low” or ” < 0 ~ 0.5V”
POWER GOOD / POWER FAIL| 500ms>PG>10ms PF>1ms
REMOTE SENSE| 5 ~ 15V
ENVIRONMENT| WORKING TEMP.| -20 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 50℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.8| 3000 meters
SAFETY & **EMC
(Note 10)
| SAFETY STANDARDS| IEC60601-1, EAC TP TC 004,ANSI/AAMI ES60601-1, CAN/CSA-C22.2 No. 60601-1:14 – Edition 3 approved,
TUV BS EN/EN60601-1 approved ; Design refer to BS EN/EN60335-1
ISOLATION LEVEL| Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP,
Secondary-Earth:1xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
ISOLATION RESISTANCE| I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/
70% RH
EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY| BS EN/EN55035, BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 4KV/Line-FG ; 2KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25
periods,
100% interruptions 250 periods
OTHERS| MTBF| 2082.3K hrs min. Telcordia SR-332 (Bellcore) ;
234.5K hrs min.| MIL-HDBK-217F (25℃)|
DIMENSION (LWH)| 12776.234.6mm or 5″ 3″ 1.36″ inch
PACKING| 0.33Kg; 36pcs/12.9Kg/0.96CUFT
NOTE| 1. All parameters NOT specially mentioned are measured at 230VAC
input, rated load and 25℃ of ambient temperature.
2. The rated power includes 5Vsb @ 0.6A.
3. The rated power includes 5Vsb @ 0.8A.
4. Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μf & 47μf parallel capacitor.
5. Tolerance : includes set up tolerance, line regulation and load regulation.
6. Derating may be needed under low input voltages. Please check the derating curve for more details.
7. Touch current was measured from primary input to DC output.
8. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
9. HS1,HS2 & HS3 can not be shorted.
10. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re- confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.” (as available on http://www.meanwell.com)
※ Product Liability Disclaimer:For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx
Block Diagram
Derating Curve
Output Derating VS Input Voltage
Mechanical Specification
AC Input Connector (CN1) : JST B3P-VH or equivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1 | AC/N |
JST VHR
or equivalent
|
JST SVH-21T-P1.1
or equivalent
2| No Pin
3| AC/L
Power Good Connector(CN3):JST B4B-XH or equivalent
Pin No. | Status | Mating Housing | Terminal |
---|---|---|---|
1 | PG |
JST XHP
or equivalent
|
JST SXH-001T-P0.6
or equivalent
2| GND
3| -S
4| +S
DC Output Connector (CN2) : JST B8P-VH or equivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1,2,3,4 | +V | JST VHR |
or equivalent
| JST SVH-21T-P1.1
or equivalent
5,6,7,8| -V
5VSB Connector(CN901) : JST B-XH or equivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1 | PS/ON |
JST XHP
or equivalent
|
JST SXH-001T
or equivalent
2,4| GND
3| 5VSB
: Grounding Required
1.HS1,HS2,HS3 cannot be shorted.
2.M1 is safety ground. For better EMC performance, Please secure an electrical
connection between M1,M2, and chassis grounding.