KPS DCM300LEAK Bluetooth Leakage Current Clamp Meter User Manual

June 9, 2024
KPS

KPS logo DCM300LEAK
Bluetooth Leakage Current Clamp Meter
User ManualKPS DCM300LEAK Bluetooth Leakage Current Clamp Meter
-

**** Safety Information

To ensure safe operation and service of the Tester, follow these instructions. Failure to observe warnings can result in severe INJURY or DEATH.
Warning
To prevent possible electrical shock, fire, or personal injury.

  • If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
  • Always use proper range for measurements.
  • To reduce the risk of fire or electric shock, do not use this product around explosive gas or in damp locations.
  • Verify the Meter operation by measuring a known current. If in doubt, have the Meter serviced.
  • To avoid false readings that can lead to electric shock and injury, replace battery as soon as low battery indicator appears.
  • Avoid working alone so that assistance can be rendered.
  • Do not use the Tester if the Tester is not operating properly or if it is wet.
  • Individual protective device must be used if hazardous live parts in the installation where the measurement is to be carried out could be accessible.
  • Use caution with voltages above 30 Vac rms, 42 Vac peak, or 60 Vdc. These voltages pose a shock hazard.

Caution
To avoid damage to the Product.

  • Do not subject the jaw to unreasonably strong shock, vibration, or force.
  • If dust gets into the top of the jaws, remove it immediately. Do not close the jaws when dust is trapped in its joints as the sensor may be damaged.
  • Remove the batteries if the Product is not used for an extended period of time, or if stored in temperatures above 50 °C. If the batteries are not removed, battery leakage can damage the Product.
  • Repair the Product before use if the battery leaks.

Symbols as marked on the meter and instruction manual

| WARNING. HAZARDOUS VOLTAGE. Risk of electric shock.| | Caution
---|---|---|---
| Consult user documentation.| | Application around and removal from uninsulated hazardous live conductors is permitted.
| Measurement Category III is applicable to test and measuring
circuits connected to the distribution part of the building’s
low-voltage MAINS installation.
| Alternating current
| Battery capacity
| Do not operate within external low frequency magnetic fields >
30A/m.| | Wireless transmission
| Conforms to European Union directives
| Equipment protected throughout by DOUBLE INSULATION or REINFORCED INSULATION| ****| Do not discard this product orthrow away
|

Meter Description

  1. Induction coil
  2. Hand-Hold area included barrier
  3. Wireless transmission / Worklight button
  4. HFR button
  5. LCD Display
  6. Worklight
  7. Power / Backlight button
  8. Hold / Max-Min button
  9. Range button

KPS DCM300LEAK Bluetooth Leakage Current Clamp Meter - Fig
1

LCD Description

  1. Meter status symbols
  2. Battery capacity symbols
  3. Measure function symbols
  4. 6,000 digit large scale LCD readout
  5. 50-60 Hz / 1k Hz Filter symbols

KPS DCM300LEAK Bluetooth Leakage Current Clamp Meter - fig
2

Power On / Off

KPS DCM300LEAK Bluetooth Leakage Current Clamp Meter - fig
3 The Meter displays battery capacity when powering up. Please replace the battery when 0% is shown.

Auto Power Off

KPS DCM300LEAK Bluetooth Leakage Current Clamp Meter - fig
4

If there have no any operation in meter, it will auto power off after 20 minutes.

Disable Auto Power Off

KPS DCM300LEAK Bluetooth Leakage Current Clamp Meter - fig
5

Position Error

When measuring current, make the cable in the center of clamp. It can make less the position error.
Caution
If the meter is used near the equipment which generates electromagnetic interference, the display may become unstable or the measurements may generate large errors.

AC Current Measurement

KPS DCM300LEAK Bluetooth Leakage Current Clamp Meter - fig
7

Leakage Cuttent Measurement

KPS DCM300LEAK Bluetooth Leakage Current Clamp Meter - fig
8KPS DCM300LEAK Bluetooth Leakage
Current Clamp Meter - fig 9

Data Hold

KPS DCM300LEAK Bluetooth Leakage Current Clamp Meter - fig
10

MAX-MIN

KPS DCM300LEAK Bluetooth Leakage Current Clamp Meter - fig
11

High Frequency Rejection (50-60 Hz & lk Hz)

KPS DCM300LEAK Bluetooth Leakage Current Clamp Meter - fig
12 Auto / Manual Range

KPS DCM300LEAK Bluetooth Leakage Current Clamp Meter - fig
13

Bluetooth

The meter uses low-power Wireless transmission v4.0 wireless technology to transfer the real-time data. You can use the RF communication to link to android or iOS devices.

Battery Replacement

Maintenance

Do not attempt to repair this Tester. It contains no user-serviceable parts. Repair or serving should only be performed by qualified personnel.

Cleaning

Periodically wipe the case with a dry cloth and detergent do not use abrasives or solvents.

General Specifications

LCD display digits: 6,000 digit large scale LCD readout
Measuring rate: 5 times / sec Overload display: OL
Conductor Size: 40mm diameter
Battery Type: 2 Batteries, IEC LR6, NEDA 15A, Alkaline (AA)
Battery Life: Alkaline 60 hours Low
Battery Voltage: 2.5V ± 0.2V
Auto Power Off: The Meter automatically turns off if the buttons are not pressed for 20 minutes.
Dimensions (W x H x D): 100 x 230 x 24 (mm)
Weight: 500g (including batteries)

Environmental Conditions

Temperature Coefficient: 0.1 x (specified accuracy) / °C (<18 °C or >28 °C)
Operating Temperature: -10°C to 30 °C ( ≤80% R.H.)
30°C to 40 °C ( ≤75% R.H.)
40°C to 50 °C ( ≤45% R.H.)
Storage Temperature: -20°C to 60°C 80% R.H., no batteries)
Max. Operating Altitude: 2000m (6562 ft.)
Vibration: Random Vibration per MIL-PRF-28800F Class 2
Calibration cycle: One year
Indoor Use.
Safety Standards: IEC / EN 61010-1
IEC / EN 61010-2-032
CAT. III 600V, Pollution Degree 2
Electromagnetic Compatibility Standards (EMC): EN 61326-1
Application Standards: IEC 61557-13: Class 2, s 30A/m
@ In: 3.5 to 600 mA, Fn: 40 to 1k Hz
Valid at best resolution in 6 mA, 60 mA and 600 mA range.

Intrinsic uncertainty or influence quantity| Designation code| Additional Spec.
---|---|---
Position| El| ±1% of reading
Supply voltage| E2| No additional error
Temperature| E3| 0.1 x (spec. accuracy) / C (<18 C or >28 °C)
Distorted waveform| E9| No additional error
External low frequency magnetic field 15 Hz to 400 Hz according to IEC 61000 -4 -8| Ell| Add ±10pA error per luT (magnetic field)
Load current| 0.00E+00| Add ±6pA error per lA (load current)
Touch current caused by common mode voltage| 0.00E+00| No additional error
Frequency| 0.00E+00| No additional error
Repeatability| 0.00E+00| No additional error

Electrical Specifications

Accuracy is given as ±(% of reading + counts of least significant digit) at 23°C ± 5°C 80% R.H.)
AC Function

  • ACA specifications are ac coupled, true RMS.

  • For non-sinusoidal waveforms, Additional Accuracy by Crest Factor (C.F.):
    Add 1.0% for C.F. 1.0 to 2.0
    Add 2.5% for C.F. 2.0 to 2.5
    Add 4.0% for C.F. 2.5 to 3.0

  • Max. Crest Factor of Input Signal: 3.0 @ 3000 counts
    2.0 @ 4500 counts
    1.5 @ 6000 counts

  • Frequency Response is specified for sine waveform.

AC Current

Range Accuracy
30 to 50 Hz 50 to 60 Hz
6mA*1 ±(2.0%+5D)*2
60mA ±(2.0`)/o+5D)*2
600mA ±(2.0%+5D)*2
6A ±(2.0%+5D)
60A ±(2.0%+5D)

1 Minimum Reading is 10uA
2 Frequency response is 15 to 50 Hz. When frequency < 30 Hz, add 3.0% to accuracy.
*3 Frequency response is 60 to 10k Hz. When frequency > 1k Hz, add 0.5% to accuracy.
Minimum Resolution : 1uA
Position Error : ±1 % of reading
Overload Protection : 60A
Low-Pass Filter (50-60 Hz)

Range Accuracy
30 to 50 Hz 50 to 60 Hz
6mA*1 ±(2.0%+5D)*2
60mA ±(2.0%+5D)*2
600mA ±(2.0`)/0+5D)*2
6A ±(2.0%+5D)
60A ±(2.0%+5D)

1 Minimum Reading is 10uA
2 Frequency response is 15 to 50 Hz.
When frequency < 30 Hz, add 3.0% to accuracy.
Minimum Resolution : 1uA
Position Error : ±1 % of reading
Overload Protection : 60A Cut-off frequency : 200Hz
Low-Pass Filter (1 k Hz)

Range Accuracy
30 to 50 Hz 50 to 60 Hz
6mA*1 ±(2.0%+5D)*2
60mA ±(2.0°/0+5D)*2
600mA ±(2.0%+5D)*2
6A ±(2.0%+5D)
60A ±(2.0°/0+5D)

1 Minimum Reading is 10uA
2 Frequency response is 15 to 50 Hz.
When frequency < 30 Hz, add 3.0% to accuracy.
Minimum Resolution: luA
Position Error: ±1 % of reading
Overload Protection: 60A
Cut-off frequency: 1kHz

KPS logo

CANADA & USA
MGL America, Inc.
[email protected]
2810 Coliseum Centre Drive, Ste. 100. Charlotte, North Carolina 28217 USA
Tel: +1 833 533-5899
MEXICO & LATAM
MGL LATAM S.A DE CV
[email protected]
Colonia Industrial Vallejo Del. Azcapotzalco 02300, Mexico D.F
Tel: +1 833-533-5899
EMEA
MGL Euman, S.L.
[email protected]
Parque Empresarial Argame, 33163 Morcin. Asturias, Spain.
Tel: +34 985-08-18-70
UNITED KINGDOM
MGL GROUP U.K. LIMITED
[email protected]
14 Weller St, London, SE1 1QU, UK
Tel: +34 985-08-18-70
www.kps-intl.com
Incorporated with MGL SKU Number:
KPSDCM300LEAKCBINT
700019854 JUN 2021 V1
©2021 MGL International Group Limited. All rights reserved. Specifications are subject to change without notification.

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals