KPS DMM3000 Multifunctional Multimeter User Manual

June 1, 2024
KPS

KPS DMM3000 Multifunctional Multimeter

Specifications

  • Product Name: DMM3000
  • Type: True RMS Multimeter
  • Display: 6000 counts digital and 60 segments analog display
  • Functions: Voltage, Current, Resistance, Continuity, Diode, Capacitance
  • Safety Compliance: Conforms to EU directives

Product Information

The DMM3000 is a True RMS multimeter with a new design for accurate measurements. It features a 6000 counts digital display and a 60-segment analog display. The meter offers various functions including voltage, current, resistance, continuity, diode, and capacitance measurements. It is designed to meet safety standards and directives.

Usage Instructions

Safety Information

Before using the DMM3000, read the user manual carefully and follow the operating instructions. Pay attention to the safety information provided to prevent any accidents or damage.

Operating Guidelines

  • WARNING: Always keep your fingers behind the finger guards when using test leads or probes to avoid bodily harm.
  • CAUTION: Remove test leads from the meter before opening the battery door or meter case.
  • CAUTION: Use the meter only as specified in the manual to prevent any damage.
  • CAUTION: Use proper terminals, switch position, and range for measurements.

Maintenance and Cleaning

Do not attempt to repair the meter yourself as it contains no user-serviceable parts. Only qualified personnel should perform repair or servicing. Periodically wipe the case with a dry cloth and detergent for cleaning. Avoid using abrasives or solvents.

FAQ

Q: How do I know if the battery needs replacement?

A: The meter has a low battery indicator that will alert you when the battery needs replacement. Replace the battery promptly to avoid false readings and ensure accurate measurements.

Q: Can I use the meter in extreme temperatures?

A: It is not recommended to expose the meter to extremes in temperature or high humidity as it may affect its performance. Keep the meter in a suitable environment for reliable measurements.

Read First

Safety Information

Understand and follow operating instructions carefully.

WARNING
Identifies hazardous conditions and actions that could cause BODILY HARM or DEATH.

CAUTION
Identifies conditions and actions that could DAMAGE the meter or equipment under test.

WARNING

  • When using test leads or probes, keep your fingers behind the finger guards.
  • Remove test lead from Meter before opening the battery door or meter case.
  • Use the Meter only as specified in this manual or the protection by the Meter might be impaired.
  • Always use proper terminals, switch position, and range for measurements.
  • Verify the Meter’s operation by measuring a known voltage.
  • If in doubt, have the Meter serviced.
  • Do not apply more than the rated voltage, as marked on Meter, between terminals or between any terminal and earth ground.
  • Only replace the blown fuse with the proper rating as specified in this manual.
  • Use caution with voltages above 30 Vac rms, 42 Vac peak,or 60 Vdc. These voltages pose a shock hazard.
  • To avoid false readings that can lead to electric shock and injury, replace battery as soon as low battery indicator.
  • Disconnect circuit power and discharge all high-voltage capaci-tors before testing resistance, continuity, diodes, or capacitance.
  • Do not use Meter around explosive gas or vapor.
  • To reduce the risk of fire or electric shock do not expose this product to rain or moisture.

CAUTION

  • Disconnect the test leads from the test points before changing the position of the function rotary switch.
  • Do not expose Meter to extremes in temperature or high humidity.
  • Never set the meter in A function to measure the voltage of a power supply circuit in equipment that could result in damage the meter and the equipment under test.

Symbols as marked on the meter and Instruction manual

KPS-DMM3000-Multifunctional-Multimeter-fig-2

Unsafe Voltage
To alert you to the presence of a potentially hazardous voltage, when the Tester detects a voltage ≧30 V or a voltage overload (OL) in V, mV, AutoV. The symbol is displayed.

Maintenance
Do not attempt to repair this Meter.
It contains no user-serviceable parts. Repair or servicing should only be performed by qualified personnel.

Cleaning
Periodically wipe the case with a dry cloth and detergent.
Do not use abrasives or solvents.

Introduction

The Meter Description Front Panel Illustration

  1. VoltSeek Light
  2. 6,000 counts digital and 60 segments analog display
  3. Push-buttons.
  4. Rotary switch for turn the Power On / Off and select the function.
  5. Input Terminal for Multi-function.
  6. Input Terminal for A.
  7. Common (Ground reference) Input Terminal.

KPS-DMM3000-Multifunctional-Multimeter-fig-5

Features

  • 6,000 count digital and 60 segments analog display
  • Large white LED backlight display
  • True RMS measurements on ACV
  • LoZ for prevent false reading from ghost voltage
  • AutoVolt automatic AC/DC voltage selection
  • Auto Ranging
  • 0.5% DCV accuracy
  • Smart Hold
  • Peak Hold for capture peak readings
  • Relative Function
  • Auto Power Off (can be disabled)
  • Capacitance measurements
  • Frequency Counter on ACV
  • Diode Test
  • Frequency Counter on ACA
  • 10Amp ACA/DCA
  • 11A/1000V High Energy Fuse
  • VoltSeekTM for non-contact voltage detection
  • EasyStickTM holster with built in magnetic sticker
  • ContiVisionTM for visible continuity beeper
  • Battery Capacity indication in segments
  • Included Holster with Probe Holder, Tilt Stand and Magnetic sticker
  • CAT IV 600V/ CATIII 1000V standard

Making Basic Measurements

KPS-DMM3000-Multifunctional-Multimeter-fig-6 CAUTION
When connecting the test leads to the DUT (Device Under Test) connect the common test leads before connecting the live test leads ; when removing the test leads, remove the live test leads before removing the common test leads.

Measuring ACV/DCV/AutoV Voltage

Dial the switch to select the measuring function.

Measuring AC/DC Current

Dial the switch to select the measuring function.

Measuring Capacitance / Diode

KPS-DMM3000-Multifunctional-Multimeter-fig-9

Dial the switch and press the Function button to select the measuring function

Measuring Continuity / Resistance

KPS-DMM3000-Multifunctional-Multimeter-fig-10

Dial the switch and press the Function button to select the measuring function.

VoltSeek

Dial the switch and press the RANGE button to select the measuring function and sensitivity.

WARNING
The VoltSeek LED indicates the electric field. If the VoltSeek LED is not on, voltage could still be present.

Using the Function

KPS-DMM3000-Multifunctional-Multimeter-fig-12

Press the Function button to change the function on the same switch position.

Range Button

KPS-DMM3000-Multifunctional-Multimeter-fig-13

Peak Hold (for AC only)

KPS-DMM3000-Multifunctional-Multimeter-fig-14

In the Peak Hold function ,the meter records the peak min. value and the peak max. value when the inputs goes below the recorded peak min. value or above the recorded peak max. value ,the meter records the new value . Press Hold button to pause the recording.

Smart Hold

KPS-DMM3000-Multifunctional-Multimeter-fig-15

The meter will beep continuously and the display will flash if the measured signal is larger than the display reading by 50 counts.(However ,it can not detect across the AC and DC Voltage /Current).

Relative

KPS-DMM3000-Multifunctional-Multimeter-fig-16

Press the Relative button to enable/disenable this function.

Backlight

KPS-DMM3000-Multifunctional-Multimeter-fig-17

Press the Backlight button to turn the backlight on/off.

Auto Power Off

KPS-DMM3000-Multifunctional-Multimeter-fig-18

Wake-up the meter by dialing the switch or pressing any button.

Disable Auto Power Off

KPS-DMM3000-Multifunctional-Multimeter-fig-19

Dial the switch to off position ,then keep the Function button down and turn the meter on.

Testing LCD Monitor

KPS-DMM3000-Multifunctional-Multimeter-fig-20

Dial the switch to off position ,then keep the HOLD button down and turn the meter on.

Fuse Replacement

KPS-DMM3000-Multifunctional-Multimeter-fig-21

Low Battery and Battery Replacement

Replace the battery as soon as the low battery indicator appears, to avoid false reading.
Refer to the following figure to replace the batteries

KPS-DMM3000-Multifunctional-Multimeter-fig-22

CAUTION
Remove test leads from meter before opening the battery cover or meter case.

Specifications

General Specifications

  • Maximum voltage applied to any terminal: 1000 Vac rms or 1000 Vdc. rms
  • Display: 6000 counts.
  • Polarity Indication: Automatic, positive implied, negative indicated.
  • Overrange Indication: OL
  • Batteries Life: 200 hours ALKALINE Battery
  • Low Batteries Indication: Voltage drops below operating voltage will flash.
  • Power Requirement: 9V battery
  • Auto Power Off 20 minutes.
  • Operating Temperature: -10 ~10°C 10°C ~ 30°C (≦80% RH), 30°C ~ 40°C (≦75% RH), 40°C ~ 50°C (≦45%RH)
  • Storage Temperature: -40°C to 60°C , 0 to 80% R.H. (batteries not fitted)
  • Measure: Samples 3 times per second .
  • Altitude: 6561.7 ft (2000m)

CAT Application field

  1. The circuits not connected to mains.
  2. The circuits directly connected to Low-voltage installation.
  3. The building installation.
  4. The source of the Low-voltage installation.
  • Weight: 250g (including battery)
  • Dimensions (W x H x D) : 74mm x 156mm x 44mm.
  • Pollution degree: 2
  • Safety: Complies with EN61010-1,CAT.IV. 600V, CAT.Ⅲ. 1000V
  • EMC: EN 61326-1
  • Shock vibration: Sinusoidal vibration per MIL-PRF- 28800F (5 ~ 55 Hz, 3g max.)
  • Drop Protection: 4 feet drop to hardwood on concrete floor. Indoor Use.

Electrical Specifications

Accuracy is given as ±(% of reading + counts of least significant digit) at 23°C ± 5°C, with relative humidity Less than 80% R.H.

  1. Temperature coefficient
    0.1 x (Specified accuracy) / °C, < 18°C, > 28°C

  2. AC Function
    ACV and ACA specifications are ac coupled, true R.M.S. The crest factor may be up to 3.0 as 4000 counts.
    For square wave, Accuracy is unspecified.
    For non-sinusoidal waveforms, Additional Accuracy by Crest Factor (C.F.) : Add 3.0% for C.F. 1.0 ~ 2.0. Add 5.0% for C.F. 2.0 ~ 2.5. Add 7.0% for C.F. 2.5 ~ 3.0.

  3. DC mV

** Range** OL Reading Resolution Accuracy
600.0mV 660.0mV 0.1mV ±(0.5% + 5D)

Input Impedance: 10M
Overload Protection: AC/DC1000V

  1. DC Voltage
** Range** OL Reading Resolution Accuracy
6.000V 6.600V 0.001V ****


±(0.5% + 2D)

60.00V| 66.00V| 0.01V
600.0V| 660.0V| 0.1V
1000V| 1100V| 1V

Input Impedance: 10M
Overload Protection: AC/DC 1000V

  1. AC Voltage
** Range** OL Reading Resolution Accuracy
600.0mV 660.0mV 0.1mV ±(1.0% + 5D)
6.000V 6.600V 0.001V ****


±(1.0% + 3D)

60.00V| 66.00V| 0.01V
600.0V| 660.0V| 0.1V
1000V| 1100V| 1V

LCD displays 0 counts when the reading < 10 counts.
Input Impedance: 10MQ // less than 100pF
Frequency Response: 45~500Hz(Sine Wave)
Overload Protection: AC/DC 1000V

  1. Auto Voltage
    Range| OL Reading| Resolution| Accuracy
    ---|---|---|---
    600.0V| 660.0V| 0.1V| ****

±(2.0% + 3D)

1000V| 1100V| 1V

LCD displays 0 counts when the reading < 10 counts.
Input Impedance: less than 3k
Frequency Response : 45 ~ 500Hz (Sine Wave)
Overload Protection : AC/DC 1000V

  1. DC Current
    Range| OL Reading| Resolution| Accuracy
    ---|---|---|---
    6.000A| 6.600A| 0.001A| ****

±(1.0% + 3D)

10.00A| 20.00A| 0.01A

Maximum measurement time :

5A for max.3 minutes with at least 20 minutes rest time.
10A for max.30 seconds with at least 10 minutes rest time.
Overload Protection : AC/DC 11A

  1. **AC Current
Range** OL Reading Resolution Accuracy
6.000A 6.600A 0.001A ****

±(1.5% + 3D)

10.00A| 20.00A| 0.01A

In 6A Range,LCD displays 0 counts when the reading < 20 counts.
In10A Range,LCD displays 0 counts when the reading < 10 counts.
Maximum measurement time :

5A for max.3 minutes with at least 20 minutes rest time.
10A for max.30 seconds with at least 10 minutes rest time.
Frequency Response : 45 ~ 500Hz (Sine Wave)
Overload Protection: AC/DC 11A

  1. Resistance
** Range** OL Reading Resolution Accuracy
6OO.O0 66O.O0 O.10 ±(0.9% + 5D)
6.OOOk0 6.6OOk0 O.OO1k0 ±(0.9% + 2D)
6O.OOk0 66.OOk0 O.OOk0 ±(0.9% + 2D)
6OO.Ok0 66O.Ok0 O.1k0 ±(0.9% + 2D)
6.OOOM0 6.6OOM0 O.OO1M0 ±(0.9% + 2D)
4O.OOM0* 44.OOM0 O.O1M0 ±(1.5% + 5D)
  • There is a little rolling less than 50 digits when measuring > 10.00 M.
    Overload Protection: AC/DC 1000V

    1. Continuity
      Range| OL Reading| Resolution| Accuracy
      ---|---|---|---
      6OO.O0| 66O.O0| O.10| ±(0.9% + 5D)

Continuity: Built-in buzzer sounds when measured resistance is less than 200 and sounds off when measured resistance is more than 2000, Between 200 to 2000 the buzzer maybe sound or off either.
Continuity Indicator: 2KHz Tone Buzzer
Response Time of Buzzer: 500 usec
Overload Protection: AC/DC 1000V

  1. Diode
    Range| OL Reading| Resolution| Accuracy
    ---|---|---|---
    1.500V| 1.550V| 0.001V| ±(0.9% + 2D)

Open Circuit Voltage: Approx. 1.8V
Overload Protection: AC/DC 1000V.

  1. Capacitance
** Range** OL Reading Resolution Accuracy
1.000μF 1.100μF 0.001μF ±(1.9% + 5D)
10.00μF 11.00μF 0.01μF ****

±(1.9% + 2D)

100.0μF| 110.0μF| 0.1μF
1.000mF| 1.100mF| 0.001mF
10.00mF| 11.00mF| 0.01mF

Overload Protection : AC/DC 1000V

  1. Frequency
    Range| OL Reading| Resolution| Accuracy
    ---|---|---|---
    100.00Hz| 100.00Hz| 0.01Hz| ****


±(0.1% + 2D)

1000.0Hz| 1000.0Hz| 0.1Hz
10.000kHz| 10.000kHz| 0.001kHz
100.00kHz| 100.00kHz| 0.01kHz

Minimum Sensitivity : > 5V (for ACV 1Hz ~ 10kHz)

20.0V (for ACV 10kHz ~ 50kHz)
unspecified (for ACV 50kHz ~ 100kHz)
0.6A (for ACA)
Minimum Frequency: 1Hz
Overload Protection: AC/DC 1000V or 11A

  1. Peak Hold
    Specified accuracy ± 150 digits.
    Accuracy of Square Wave is unspecified.

  2. VoltSeek
    Voltage Range of High Sensitivity :
    80V ~ 1000V (At the top edge of the meter)
    Voltage Range of Low Sensitivity :
    160V ~ 1000V (At the top edge of the meter)

Limited Warranty

This meter is warranted to the original purchaser against defects in material and workmanship for 2 years from the date of purchase. During this warranty period, Manufacturer will, at its option, replace or repair the defective unit, subject to verification of the defect or malfunction. This warranty does not cover fuses, disposable batteries, or damage from abuse, neglect, accident, unauthorized repair, alteration, contamination, or abnormal conditions of operation or handling. Any implied warranties arising out of the sale of this product, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited to the above. The manufacturer shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expense or economic loss. Some states or countries laws vary, so the above limitations or exclusions may not apply to you.

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals