easy home EBP-085 Wrist Blood Pressure Monitor User Manual
- June 9, 2024
- EASY HOME
Table of Contents
- easy home EBP-085 Wrist Blood Pressure Monitor
- General Description
- Indications for Use
- Measurement Principle
- Safety Information
- LCD Display Symbol
- Monitor Components
- Setting Date and Time
- WRIST POSITIONING GUIDE
- Measurement
- Recalling the Records
- Deleting the Records
- Tips for Measurement
- TROUBLESHOOTING
- SPECIFICATIONS
- EMC Guidance
- FCC Statement
- Warranty
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
easy home EBP-085 Wrist Blood Pressure Monitor
General Description
The wrist blood pressure measurement features blood measurement, pulse
measurement and result memory.
This manual contains impcrtant safety and care information, and provides step
by step instructions for using this blood pressure monitor. Read the manual
thoroughly before using this product.
Features:
- Systolic blood pressure
- Diastolic blood pressure
- Pulse rate
- 199 recorded readings for each user
Indications for Use
This Easy@home Wrist Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and heartbeat rate with a wrist circumference ranging from 13.5 cm to 21.5 cm. It is intended for indoor, adult use only.
Contraindications
- The device is not suitable for use on the women who are or may be pregnant.
- The device is not suitable for use on patients with implanted, electrical devices, such as cardiac pacemakers, defibrillators.
Measurement Principle
This product uses the Oscillometric Measuring Method to detect blood pressure. Before every measurement, the unit establishes a “zero point” equivalent to the atmospheric pressure. Then it starts inflating the cuff. Meanwhile, the unit detects pressure oscillation generated by beat-to-beat pulsatile, which is used to determine the systolic pressure and diastolic pressure as well as pulse rate.
Safety Information
The signs below might be in the user manual, labeling or another component. They are the requirement of standard and use.
CAUTION
-
This device is intended for indoor, home use.
-
This device is not intended for public use.
-
This device is portable, but it is not intended for use during patient transport.
-
This device is not suitable for continuous monitoring during medical emergencies or operations.
-
This device is intended for no-invasive measuring and monitoring of arterial blood pressure. It is not intended for use on extremities other than the arm, or for any purpose other than obtaining a blood pressure measurement.
-
This device is for adults. Do not use this device on neonates or infants. Do not use it on children unless otherwise instructed by a medical professional.
-
Do not use on the women in pregnant, including pre-eclamptic, patients .
-
The device is not suitable for use on patients with implanted, electrical devices, such as cardiac pacemakers, defibrillators.
-
The effectiveness of this device has not been established for use:
-
on users with oommon arrhythmias such as atrial or ventricular premature beats or atrial fibrillation,
-
on users with peripheral arterial disease,
-
on users undergoing intravascular therapy, or with arteriovenous (AV) shunt.
Consult a medical professional before use. -
Do not use this device for diagnosis or treatment of any health problem or disease. Contact your physician if you have or suspect any medical problem. Do not change your medications without the advice of your physician or health care professional.
-
If you are taking medication, consult your physician to determine the proper time to measure your blood pressure.
CAUTION
- This device may be used only for the intended use described in this manual, the manufacturer shall have no liability for any incidental, oonsequential, or special damages caused by misuse or abuse.
- Report any unexpected operation or events to the manufacturer.
- Do not apply the cuff on an arm that has an intravenous drip or a blood transfusion attached.
- Warning : Do not kink, fold, stretch, compress, or otherwise deform the tube during measuring, as the cuff pressure might continuously increase, which could prevent blood flow and result injury.
- Warning : Taking blood pressure measurements too frequently could disrupt blood circulation and cause injuries.
- Warning : Do not apply cuff to areas on patient where skin is delicate or damaged. Check cuff site frequently for irritation.
- Warning : Do not place the cuff on the arm of a person whose arteries or veins are undergoing medical treatment, i.e. intra-vascular access or intra-vascular therapy or an arteriovenous (A-V) shunt, which could disrupt blood circulation and cause injuries.
- Do not place the cuff on the arm on the same side of a mastectomy (especially when lymph nodes have been removed). it is recommended to take measurements on the unaffected side.
- Do not wrap the cuff on the same arm to which another monitoring device is applied. One or both devices could temporarily stop functioning if you try to use them at the same time.
- Please check that the operation of the device do not result in prolonged impairment of patient blood circulation.
- Warning : On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, loosen and remove the cuff immediately. Prolonged high pressure applied to the arm (cuff pressure >300 mmHg or constant pressure >15 mmHg for more than 3 minutes) might lead to bruising and discolored skin.
- Warning : Do not use this device with high-frequency (HF) surgical equipment at the same time.
- Warning: This device is not AP/APG equipment. Do not use the device where flammable anesthetic are present, or in environments mixture with air of with oxygen or nitrous oxide.
- The device contains sensitive electronic components.To avoid measurement errors, avoid taking blood pressure measurements near a strong electromagnetic field radiated interference signal or electrical fast transienVburst signal.
- Wireless communication equipment, such as wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies may cause interference that may affect the accuracy of measurements. A minimum distance of 1 foot (30 cm) should be kept from such devices during a measurement.
- You can use this device to take your own measurement, no third-party operator is required.
- Please use the device under the environment which is provided in the user manual. Otherwise, the performance and lifetime of the device will be impacted and reduced.
- The device may require up to 30 minutes to warm up / cool down from the minimum / maximum storage temperature before it is ready for use.
- Warning : Excessive cuff tube lengths could cause strangulation if you don’t manage them properly.
- Warning : Do not touch output of the batteries/adapter and the user simultaneously.
- Adapter is specified as a part of ME EQUIPMENT.
- Warning : The power cord is considered the disconnect device for isolating this equipment from supply mains. Do not position the equipment so that it is difficult to reach or disconnect.
- The blood pressure monitor, its adapter, and the cuff are suitable for use within the patient environment.
- Warning : Do not use this device if you are allergic to polyester, nylon, or plastic.
- Warning : Only use accessories approved by manufacturer. Using unapproved accessories might cause damage to the unit and injure users.
- Warning : If you experience discomfort during a measurement, such as pain in the arm or other complaints, press the Power button immediately to release the air from the cuff.
- No calibration is required within two years of reliable service.
- Do not attempt to repair the unit yourself if it malfunctions. Only have repairs carried out by authorized service centers.
- At the request of authorized service personnel, circuit diagrams, component part lists, descriptions, and calibration procedures wilt be made available by the manufacturer or distributor.
- It is recommended that the performance should be checked after repair, maintenance, and every two years of use, by retesting the requirements in limits of the error of the cuff pressure indication and air leakage (testing at least at 50 mmHg and 200 mmHg).
- Warning : Do not use the device while under maintenance, or being serviced.
- Store your device, cuff and adapter in a clean and dry place, protect it against extreme moisture, heat, lint, dust and direct sunlight. Never place any heavy objects on it.
- Make sure the rubber tube of the cuff is not squeezed, stretched, or kinked during storage.
- Warning : Keep the device, cuff, and batteries away from children as they may pose a risk of choking or strangulation if used improperly.
- Clean both device and cuff with a soft, dry cloth. If necessary use a dampened doth and natural detergent. Do not use alcohol, benzene, or other harsh chemicals.
- Do not wash the cuff in a washing machine or dishwasher!
- The service life of the cuff may vary by the frequency of washing, skin condition, and storage state. The typical service life is 10000 times.
- Dispose of accessories, detachable parts, and the device according to the local guidelines.
LCD Display Symbol
Monitor Components
Component list of pressure measuring system:
- Circuit Board (PCBA);
- Air Pipe;
- Pump;
- Valve;
- Cuff.
List of Package Included Items
- Blood Pressure Monitor EBP-085
- 2x AAA Batteries
- User manual
- Storage bag
Installing and Replacing the Batteries
- Lift the battery cover upwards.
- Install the batteries as indicated in the battery compartment. Always use the correct battery type (2x AAA batteries).
- Close the battery cover.
The typical service life of the new and unused batteries is about 180 measurements. Every operation time is 60 seconds.
Replace the batteries whenever the below happen
- The Battery +bAt Lo shows
- The display is dim
- The display does not light up
CAUTION
- Do not use new and used batteries together.
- Do not use different types of batterles together.
- Do not dispose the batteries in fire. Batteries may explode or leak.
- Remove batteries if the device is not likely to be used for some time.
- Worn batteries are harmful to the environment. Do not dispose with daily garbage.
- Remove the old batteries from the device following your local recycling guidelines.
Setting Date and Time
It Is Important to set the Date and Time before using your blood pressure monitor, so that a correct time stamp can be assigned to each record that Is stored In the memory. (Year Range: 2021-2050; Time Format: 12/24 H)
-
When the monitor is off, press and hold the “ST ART/ STOP ” button to display the date format. Press the “MEM” button to switch the date format between [month/day] and [day/month].
-
Press the “START/ STOP” button to confirm the date format, then the year will flash. Press the “MEM”button to change the year.
-
When you get the right year, press the “START/ STOP” button to confirm the year. The screen will then show a blinking number representing the[MONTH).
-
Repeat steps 2 and 3 to set the [MONTH] and [DAY].
-
Repeat steps 2 and 3 to set the time format.
-
Repeat steps 2 and 3 to set the [HOUR) and [MINUTE].
-
Repeat steps 2 and 3 to confirm [Wrist Positioning Guide) on or off.
Note : When the Wrist Positioning
Guide is OFF, the instruction will not appear at the start of the measurement. -
After confirming the measurement unit, the LCD will display “Done” and then the monitor will tum off.
WRIST POSITIONING GUIDE
When using wrist blood pressure monitor, it is important to take the
measurement while relaxing your wrist at heart level. This will help ensure an
accurate reading will be taken. Because it is difficult to find the ideal
location for a wrist measurement, we have included in the device an optional
Wrist Positioning Sensor that will assist in directing your wrist to the ideal
location for a blood pressure measurement.
Each time you take a measurement, the display will illuminate with different
icons that are designed to help you move your wrist. Once the ideal location
is found, the wrist symbol ✓ will flash and after several seconds, the
measurement will begin. Keep your arm still until the measurement is
completed.
- Move your wrist up.
- Move your wrist down.
- Wrist in ideal position, the symbol ✓ flash.
- Do not move, Measurement start.
Note : Due to differences in individual sizes and physique, this feature may not be helpful in all cases and you may wish to tum this feature OFF. If you feel the suggested wrist position does not match your heart level, please tum this feature OFF and follow your own judgment.
Pair a smartphone or a tablet with the monitor
You are the intended operator of this blood pressure monitor. You can measure your blood pressure and then save and send measurement data to a smartphone or tablet with Bluetooth wireless connectivity and application.
- Turn on Bluetooth and the application. Make sure both are ON when pairing.
- When the monitor is OFF, press and hold the “MEM” button to start pairing. The symbol will be shown on the LCD, indicating pairing is working.
If successful, the Bluetooth symbol will not flash any more, and the LCD will
display “donE” and the monitor will shut off after about 1 minute.
If unsuccessful within 60 seconds, it is judged to timeout and the monitor
will shut off automatically.
- RF Frequency Rans-: 2402 MHz to 2480 MHz
- Output Power Range : s -2.5 dBm
- Supply Voltage : 1.8-3.6V
- Transmitting Distance: 10 meters
CAUTION
- interference may occur In Iha vicinity of equipment mar1<ed with the following symbol . «t» And TMB-2085 may interfere vicinity of electrical equipment.
- Sensitive people, including pregnant women pre-eclamptic and those who implanted medical electronic instruments, should avoid using the unit whenever possible.
- Keep the monitor at least 20 centimeters away from the human body (especially the head) when lhe data after the transmission is proceeding m&asurement.
- To enable the data transmission function, lhls product should be paired to Bluetooth end at 2.4 GHz.
How to mitigate possible interference?
- The range between the device and BT end should be reasonably close, from 1 meter to 10 meters. Please ensure no obstacles between the device end BT eod so as to obtain qua!ily connection and to lower the RF output range.
- To avoid interference, other electronic devices (particularly those with wireless transmission f Transmitter) should be kept at least 1 meter away from lhe monitor.
Tie the Cuff
- Remove a11 accessories (watch, bracelet,etc) from your wrisl If your physician has diagnosed you with poor circulation in one of your wrist, use the other wrist
- Roll or push up your sleeve to expose the skin.
- Apply the cuff to your wrist with your palm facing up.
- Position the edge of the cuff about 1cm-2cm from wrist joint.
- Fasten the wrist cuff around your wrist, leaving no extra room between the cuff and your skin. If the cuff is too loose, the measurement will not be accurate.
- Sit comfortably with your wrist being tested resting on a flat surface. Place your elbow on a table so that the cuff Is at the same level as your heart. Tum your palm upwards. Sit upright In a chair and take 5-6 deep breaths.
- Users with Hypertension:
- The middle of the cuff should be at the level of the right atrium of the heart; Before starting measurement, please sit comfortably with legs uncrossed, feet flat on the floor, back and wrist supported.
- Rest for 5 minutes before measuring.
- Wail at least 3 minutes between measurements. This allows your blood circulation to recover.
- Take the measurement in a silent room.
- The user must relax as much as possible and do not move and talk during the measurement procedure.
- The cuff should maintain at the same level as the right atrium of the heart.
- Do not cross your legs and keep your feel on the ground. Keep your back against the backrest of the chair.
- For a meaningful comparison, try to measure under simllar conditions.
- For example, take daily measurements al approximately the same time, on the same wrist, or as directed by a physician
Measurement
Start the Measurement
-
When the monitor is off, press the “START/STOP” button to tum on the monitor, the display will show a blinking number representing the current [USER], at this process, you can switch the user from User 1 to User 2 or User G (guest) by pressing “MEM” button. After choosing the right user, press the “START/STOP” button to confirm, it will display the latest record of the current user, and after then it will complete the whole measurement automatically. (Take user 1 for example.)
A step-by-step guide showing what you’ll see while taking a measurement- Display the current User
- Display the latest record of current User
- Adjust to zero.
- Inflating and measuring.
-
After the measurement was finished, the symbol and §will display, and the data will start transmitting.
- Display and save the measuring result.
-
If the data transmit successfully, the symbol and symbol will disappear, the image on the right will be displayed, and then the monitor will tum off after about 1 minute.
-
If unsuccessful within 60 seconds, it is judged a timeout and the monitor will shut off. ( Note: the symbol § will be displayed in the next measurement)
-
Press the “START/STOP” button to power off, otherwise, it will shut off after about 1 minute.
Notes : If you have set the [Wrist Positioning Guide] on, when you start the measurement, the blood pressure monitor will detect the position first. -
The wrist must be at the angle between 30′ and 45′. If it’s out of this angle range, the LCD won’t start any measurement until you have the ideal position.
-
If you get the correct position, it will display for about 3 seconds, and then finish the whole measurement. For more details, please see page 18-19.
- Display the current User
- Display the latest record of current User
- Ideal position. measurement start
- Adjust to zero.
Recalling the Records
-
When the monitor is off, press “MEM” button and the user ID will be blinking and the memory number of the reading will be shown. You can switch the user from User 1 to User 2 by pressing “MEM” button. After choosing the right user, press “START/STOP” button to confirm. If there are less than 3 records, it will display the latest record of the current user.
If there are more than 3 records, it will show the average of latest three records first. -
Press “MEM” button to get the record you want. If the symbol Bulethooth and appears along with the record, that means the record pending to transmit.
- Date and time will display altemately.
CAUTION
The most recent record (1) is shown first. Each new measurement is assigned to the first (1) record. All other records are pushed back one digit (e.9., 2 becomes 3, and so on), and the last record (199) is dropped from the list. once 199 records is surpassed.
Deleting the Records
If you did not get the correct measurement, you can delete results by following steps below.
A: To delete a single measurement:
- When the monitor is off, press “MEM” button to show the memory recall mode (you could use the “MEM”‘ button to switch between users).
- Several seconds later, press “MEM” button to find the measurement you would like to erase, press and hold “MEM”” button for 3 seconds, and the display will show a blinking “dEL yES” and the memory number of the reading.
- Use the “MEM” buttons to toggle between “dEL yES” and “dEL no”‘ . Press
“START/STOP'” to confirm the selection. lf”dEL yES” is selected, the unit will delete the record and return to last memory record. All the records are pushed forward one digit (e.g., 03 becomes 02, and so on)
B: To delete ALL measurements:
- When the monitor is off, press “MEM” button to show the memory recall mode (you could use the “MEM” button to switch between users).
- Several seconds later, hold pressing “MEM” +”START/STOP” button for 5 seconds, the display will show a blinking “dEL ALL” and the User ID.
- Use the “MEM” buttons to toggle between “dEL ALL” and “dEL no”. Press
“START/STOP” to confirm the selection. If “dEL ALL” is selected, the unit will display “dEL ALL+ do nE”, and delete all the record of the current user. Several seconds later, ii will display”–“.
Tips for Measurement
Measurements may be inaccurate if taken in the following circumstances.
Maintenance
To obtain the best performance, please follow the instructions below.
What are systolic pressure and diastolic pressure?
When ventricles contract and pump blood out of the heart, blood pressure reaches its maximum value, the highest pressure in the cycle is known as systolic pressure. When the heart relaxes between heartbeats. the lowest blood pressure is the diastolic pressure.
What is the standard blood pressure classification?
The blood pressure classification published by World Health Organization (WHO) and International Society of Hypertension (ISH) in 1999 is as follows.
CAUTION-
Only a physician can tell you your normal blood pressure range and the point
at which you are at risk. Consult your physician to obtain these values.
Please contact a physician if your measuring result falls out of the range.
Irregular Heartbeat Detector
An irregular heartbeat is detected when a heartbeat rhythm varies while the device is measuring systolic pressure and diastolic pressure. During each measurement, the blood pressure monitor will keep a record of all the pulse intervals and calculate the average value of them. If there are two or more pulse intervals, the difference between each interval and the average is more than the average value of £25% , or there are four or more pulse intervals ,the difference between each interval and the average is more than the average value of £15%, then the irregular heartbeat symbol will appear on the display with the measurement result.
CAUTION
The appearance of the IHB icon indicates that a pulse irregularity consistent
with an irregular heartbeat was detected during measurement.
Usually this is NOT a cause for concern. However, if the symbol appears often,
we recommend you seek medical advice. Please note that the device does not
replace a cardiac examination on but serves to detectpulse irregularities at
an early stage.
Why does my blood pressure fluctuate throughout the day?
- Individual blood pressure varies multiple times every day. It is also affected by the way you tie your cuff and your measurement position, so please take the measurement under the same conditions each time.
- If the person takes medicine, the pressure can vary more.
- Wait at least 3 minutes for another measurement.
Why do l get a different blood pressure at home compared to the doctors?
The blood pressure measurement can be different even throughout the day due to
weather, emotion, exercise etc, Also, there is the “white coat” effect, which
means blood pressure usualy increases in dinical settings.
Is the result the same if measuring on the right wrist?
It is ok for both wrists, but there will be some different results for different people. We suggest you measure the same wrist every time.
What you need to pay attention to when you measure your blood pressure at home:
- If the cuff is tied properly. If the cuff is too tight or too loose.
- If the cuff is tied on the wrist. If you feel anxious.
- Taking 2-3 deep breaths before beginning Will be better for measuring.
Advice : Relax yourself for 4-5 minutes until you calm down.
TROUBLESHOOTING
If any abnormality arises during use, please check the following points:
PROBLEM | SYMPTOM | CHECK THIS | REMEDY |
---|
Error message
|
EEx,shows on lhe display.
|
A calibration error occurred.
| Retake the measuremeoUf the problem persists.contact the retailer or our
customer service department for further assistance. Refer to the warranty for
contact infonnation and return instructions.
Warning message|
“out· shows
|
Out of measurement range
| Relax for a moment. Refasten the cuff and then measure again. If the problem persists, contact your physician.
NOTE : If the product still does not work, contact Easy@home Customer Service. Under no circumstance should yu disassemble or attempt to repair the unit by yourself.
Pow er supply | Battery powered mode: 2x AAA batteries 3V ::-:-: |
---|---|
Display mode | Digital LCD V.A.43 mm x 46 mm |
Measurement mode | Oscitlographic testing mode |
Measurement range
| Rated cuff pressure:
0 mmHg – 299 mmHg Measurement pressure: SYS: 60 mmHg – 230 mmHg
DIA: 40 mmHg – 130 mmHg
Pulse value: (40-199) beaVminute
Accuracy| Pressure:5″C • 40°C within ±3 mmHg Pulse value: ±5%
Working condition
| A temperature range of: (41″F – 104°F) 5°C • 40″C
A relative humidity range of 15% to 90%, non-condensing, but not requiring a water vapour partial pressure greater than 50hPa
An atmospheric pressure range of: 700 hPa to 1060 hPa
Storage & transportation condition| Temperature: (-4″F • 140″FJ -20°C – 60°C
A relative humidity range of s 93%, non-condensing,
at a water vapour pressure up to 50hPa
Measurement perimeter of the wrist|
About 13.5-21.5 cm
Weight| App,ox.103 g (Excluding the batteries)
External dimensions| Approx. 85.7 mm>< 60.8 mm>< 24,7 mm (Excludingthe cuff)
Attachment| 2x AAA batteries, user manual, storage bag
SPECIFICATIONS
Mode of operation | Continuous operation |
---|---|
Degree of protection | Type BF applied part |
Device Classification | Internally Powered ME Equipment |
IP Classification
| IP22: It means the device could be protected against solid foreign objects
of 12,5mm _4 > _and greater, and against vertically falling waterd rops
whenenciosure titledupt o 15°
Software Version| A01
Service life of the product| 15,000limes use
WARNING : No modification of this equipment is allowed.
Contact Information
For more information about our products, please visit healthcare-manager.com
You can get customer service and ask usual problems. Easy@home will serve you
anytume.
EMC Guidance
The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments.
Warning : Don’t go near the active HF surgical equipment and the RF
shielded room of an ME system for magnetic resonance imaging, where the
intensity of EM disturbances is high.
Warning : Use of this equipment adjacent to or stacked with other
equipment should be avoided because t could result in improper operation. If
such use is necessary, this equipment and the other equipment should be
observed to verity that they are operating normally.
Warning : Use of accessories, transducers, and cables other than those
specified or provided by the manufacturer of this equipment could result in
increased electromagnetic emissions or decreased electromagnetic immunity of
this equipment and result in improper operation.
Warning : Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be used no closer than 30
cm (12 inches) to any part of the equipment EBP-085, including cables
specified by the manufacturer. Otherwise, degradation of theperformance of
this equipment could result.
Technical description:
- All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE regard to electromagnetic disturbances for the excepted service life.
- Guidance and manufacturer’s declaration- electromagnetic emissions and Immunity.
Gu id ance and manufactur er’s dec laratio n – elect rom agn etic emissio ns
Emi ss ion s test
|
Compli ance
RF emissions CISPR 11|
Group 1
RF emissions CISPR 11
|
Class B
Harmonic emissions IEC 61000-3-2
|
Not applicable
Voltage fluctuations/ flicker emissions IEC 61000-3-3
|
Not applicable
Guidance and manufacturer’ s declaration – electromagnetic Immunity
Immunity Test| IEC 60601-1-2
Test level
|
Compli ance level
Electrostatic discharge (ESD) IEC 61000-4-2
|
±8 kV contact
±2 kV, ±4kV , ±8 kV, ±15 kV air
|
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV air
Electrical fast transienUburst IEC 61000-4-4| N ot applicable| Not applicable
Surge IEC61000-4-5
|
Not applicable
|
Not applicable
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4 – 11
| Not applicable| Not applicable
---|---|---
Power frequency magnetic field IEC 61000-4-8
| 30AJm
50Hz/60Hz
| 30A/m
50Hz/60Hz
Conduced RF IEC61000-4-6
|
Not applicable
|
Not applicable
Radiated RF IEC61000-4-3
| 10 V/m
80 MHz- 2,7 GHz
80 % AM at 1 kHz
|
10V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE Ur is the a.c. mains voltage prior to application of the test level.
Gui danc e and manuf ac tur er’ s decl aratio n • elect r omagneti c Imm un ity
Radiated RF IEC61000-4-3
(Test specifications ‘E”N’CLOSURE
PORT
IMMUNITY to
RF wireless communicati-
005
equipment)
| Test Frequency (MH z )| Band (MHz)| Service| Modulation| Maximum Power (WI| Distance
(m)
| IEC
60601-1-2
Test Level (V/m)
| Compliance level
(Vim)
385| 380-300| TETRA
400
| Pulse
modulation 18Hz
| 1.8| 0.3| 27| 27
450| 430-470| GMRS460,
FRS460
| FM t 5kHz
deviation
1 kHz sine
| | 03| 28| 28
_ 710
__ ill
| 704-787| LTE Band
13,
17
| Pulse modulation 217 Hz| 0.2| 0.3| |
780|
810
870
930
| 800-980| GSM 800/900, TETRA 800,
iOEN 820, CDMA850 ,
LTE Band 5
| Pulse moclulalion 18Hz| | 0.3| 28| 28
| 1720
– – –
1845
1970
| 1700-
1990
| GSM 1800;
COMA 1900;
GSM 1900; DECT;
LTE Band 1,
3, 4,25; UMTS
| Pulse modulation 217 Hz| | 0.3| 28| 28
---|---|---|---|---|---|---|---|---
2450| 2400-
2570
| Bluetooth,
WLAN,
802.11
b/g/n,R FID 2450, LTE
Band7
| Pulse
modulation 217Hz
| | 0.3| 28| 28
5240
– – –
5500
5785
| 5100-
5800
| WLAN
802.11
a/n
| Pulse modulation 217 Hz| 0.2| 0.3| |
FCC Statement
FCC ID : OUT9TMB2085-K
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions:
- this device may not cause harmful interference, and
- this device must accept any interference received, including interference that may cause undesired operation.
Caution : The user is cautioned that changes or modifications not
expressly approved by the party responsible for compliance could void the
user’s authority to operate the equipment.
NOTE : This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to Part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference ina residential installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However,there is no guarantee that interference will not occur
in a particular installation.
If this equipment does cause harmful interference to radi0 or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
FCC Radiation Exposure Statement:
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter.
Warranty
The Wrist Blood Pressure Monitor is guaranteed for 1-year from the date of purchase. If the Wrist Blood Pressure Monitor does not function properly due to defective components or p0or workmanship, we will repair or replace for no additional cost. The warranty does not cover damages to your VWrist Blood Pressure Monitor due to improper handling, loss, or theft. Please contact the manufacturer for details.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>