Medtronic MiniMed Quick Set Infusion Set User Guide
- June 9, 2024
- Medtronic
Table of Contents
MiniMed™ Quick-set™
Infusion set
User Guide
MiniMed Quick Set Infusion Set
Refer to pump and reservoir user guides for installation and priming.
Fill empty space in cannula with following prime of U-100 Insulin
6mm 0.3u / 9mm 0.5u
Disconnecting
Reconnecting
Indications for use
This product is indicated for the subcutaneous infusion of insulin from an infusion pump.
Contraindications
Infusion sets are indicated for subcutaneous use only and not for intravenous (IV) infusion or the infusion of blood or blood products.
Warnings
Inaccurate medication delivery, infection and/or site irritation may result
from improper insertion and maintenance of the infusion site. Before
insertion, clean the insertion site with isopropyl alcohol. Remove the needle
guard before inserting the infusion set.
If using this infusion set for the first time, do the first set-up in the
presence of your healthcare professional.
Do not leave air in the infusion set. Prime completely.
Check frequently to make sure the soft cannula remains firmly in place as you
may not feel pain if it pulls out. The soft cannula must always be completely
inserted to receive the full amount of medication. If the infusion site
becomes inflamed, replace the set, and use a new site until the first site has
healed. Replace the infusion set if the tape becomes loose, or if the soft
cannula becomes fully or partially dislodged from the skin.
Replace the infusion set every 4872 hours, or per your healthcare
professional’s instructions. Do not reinsert the introducer needle into the
infusion set. Re-insertion may cause tearing of the soft cannula, which may
result in unpredictable medication flow.
Never prime the set or attempt to free a clogged line while the set is
inserted. You may accidentally inject too much medication. Do not put
disinfectants, perfumes, or deodorants on the infusion set as these may affect
the integrity of the set.
Dispose of the infusion set and introducer needle safely, in a sharps
container, after a single use. Do not clean or re-sterilize.
This device is sterile and non-pyrogenic unless the package has been opened or
damaged.
Do not use if the package has been opened or damaged. (For Paradigm users, do
not use the infusion set if the tubing connector needle has been damaged.)
Store infusion sets in a cool, dry place.
Do not leave infusion sets in direct sunlight oinside a vehicle. Use aseptic
techniques when temporarily disconnecting the set and consult your healthcare
provider on how to compensate for missed medication when disconnected. If
infusing insulin, carefully monitor your blood glucose levels when
disconnected and after reconnecting. If infusing insulin, and your blood
glucose level becomes unexplainably high, or an occlusion alarm occurs, check
for clogs and/or leaks.
If in doubt, change the infusion set because the soft cannula may be
dislodged, crimped and/or partially clogged. Should any of these problems
arise, make a plan with your healthcare professional for rapidly replacing
insulin. Test your blood glucose level to make sure the problem is corrected.
If infusing insulin, do not change the infusion set just before bedtime unless
you can check your blood glucose 1-3 hours after insertion.
Reuse of the infusion set may cause damage to the cannula/needle and lead to
infection, site irritation, and/or inaccurate medication delivery. If insulin,
or any liquid, gets inside the tubing connector, it can temporarily block the
vents that allow the pump to properly prime the infusion set.
This may result in the delivery of too little or too much insulin, which can
cause hyperglycemia or hypoglycemia. If this occurs, start over with a new
reservoir and infusion set.
Warranty
For product warranty information, please contact your local help line,
representative, or visit http://www.medtronicdiabetes.com/warranty.
This infusion set is compatible with the Quick-serter® insertion device.
MiniMed, Quick-set and Quick-serter are registered trademarks of Medtronic
MiniMed, Inc.
| Replace every 3 days
---|---
| Manufacturer
| Use by (yearmonth-day)
| Open here
| Do not re-use
| Batch code
| Catalogue number
| Attention: see instructions for use
| Method of sterilization: ethylene oxide
| Non Pyrogenic
| No Latex
Distributed by:
Medtronic MiniMed
18000 Devonshire Street
Northridge, CA 91325
USA 800 646 4633 8
18 576 5555
INS-776-5
Legal Manufacturer:
Unomedical a/s
Aaholmvej 1-3, Osted
4320 Lejre Denmark