Q-Trak 7585 Indoor Air Quality Monitor User Guide

July 20, 2024
Q-Trak

Q-Trak 7585 Indoor Air Quality Monitor

Specifications:

  • Model: Q-TrakTM XP Indoor Air Quality Monitor 7585
  • Includes: Q-TrakTM XP Base Meter 7580, Multi-Sensor IAQ Module 801430, Carbon Dioxide (CO2) sensor 801399
  • Additional Parts : Carrying Case, Lithium Ion Rechargeable Battery Pack, USB Cable, AC Adapter/Power Supply, Gas Sensor Calibration Cap
  • Can accept up to six gas sensors

Product Usage Instructions

Installation of Gas Sensors:

  1. Ensure the instrument is powered OFF.
  2. Remove the sensor module cover.
  3. Position the sensor making good contact with the device. Gas sensors can be installed in any available slot.
  4. Listen for a click to confirm the proper connection.
  5. Attach the cover to the sensor module.

Providing Power:

The Q-TrakTM XP monitor can be powered by a lithium-ion battery or an A/C adapter.

Operating with Lithium-Ion Battery:

  1. Install the battery by connecting it to the handle connector after removing the battery compartment cover.
  2. Charge the battery fully before use.
  3. Calibrate the battery power gauge by fully discharging and then fully charging the battery.

Operating with AC Adapter:

The instrument can be operated with or without a battery using the A/C adapter.

FAQ

  • Q: How many gas sensors can the device accept?
    • A : The device can accept up to six gas sensors in addition to the standard built-in sensors.
  • Q: How long does it take to fully charge the battery?
    • A: Charging the battery may take up to 3 hours. Note the lightning bolt indicator when charging vs fully charged.

Thank you for purchasing a TSI® Incorporated (TS|®)

Q-Trak™ XP Indoor Air Quality (IAQ) Monitor. This guide will help you quickly begin using your monitor.

More detailed information is in the Operation Manual.

Please refer to the manual if you have questions on the operation of your new monitor.

Safety

Refer to the Q-Trak™ XP Indoor Air Quality Monitor Mode 7585 user manual for safety information.

Unpacking and Parts Identification

Q-Trak-7585-Indoo-Air-Quality-Monitor-FIG \(1\)

Q-Trak-7585-Indoo-Air-Quality-Monitor-FIG \(3\)

Carefully unpack the instrument and accessories from the shipping container. Check the individual parts against the list of components below. If anything is missing or damaged, notify TIS/® immediately. All standard equipment can be purchased separately if needed.

Installation of Gas Sensors

In addition to the standard built-in sensors, the device can accept up to six gas sensors.

Install the gas sensors as follows:

  1. Ensure the instrument is powered OFF.
  2. Remove the sensor module cover.Q-Trak-7585-Indoo-Air-Quality-Monitor-FIG \(2\)
  3. Position the sensor as shown, making certain the pins make good contact with the device. Gas sensors can be installed in any available slot. Listen for a click to confirm sensor is connected properly.
  4. Attach the cover to the sensor module.

Providing Power to the Q-Trak ™ XP Monitor

The Q-Trak™ XP monitor can be powered with a lithium-ion battery or with the A/C adapter.

Operating the Instrument with the

Lithium-lon Battery

The lithium-ion battery must be installed by the user and fully charged before use.

Install the battery:

  1. Remove the battery compartment cover.

  2. Connect the battery to the handle connector.
    Q-Trak-7585-Indoo-Air-Quality-Monitor-FIG \(5\),..,.,

  3. Insert the battery as shown in the picture below, keeping the battery wires to the side and clear of the battery cover screw.

Installing the Wi-Fi° Dongle (optional)

If purchased, install the dongle into the USB port located inside the battery compartment.

Attach the battery compartment cover, ensuring battery wires are clear of the screw.

IMPORTANT NOTICE

Calibrate the battery power gauge

Charging the Battery and Calibrating the Power Gauge

  1. With the battery installed, turn on the instrument by pressing the power button and run until the battery is fully discharged. The instrument will automatically turn off when the battery is fully discharged.
  2. Connect the AC adapter to the instrument and fully charge the battery. This may take up to 3 hours. Note the lightning bolt indicator when charging vs fully charged
  3. The instrument is now ready for use with the battery. This Power Gauge calibration procedure does not need to be repeated unless the battery is replaced or disconnected.

Operating the Instrument with the AC Adapter

The instrument can be operated with or without a battery using the A/C adapter.

To operate the instrument with AC adapter:

  1. Plug the AC adapter into an electrical outlet. Plug the other end into the instrument.
  2. Plug the other end into the instrument.

Powering ON the Q-Trak T™ XP IAQ Monitor

  1. Press and release the power button . In 35 seconds the instrument is ready for use.
  2. The first time the Q-Trak™ XP monitor is powered ON the Welcome to Q-Trak™ XP page is presented. Select the appropriate Language and Time Zone then select DONE.
  3. After selecting DONE, the Dashboard page is displayed in Survey mode.

Q-Trak-7585-Indoo-Air-Quality-Monitor-FIG \(14\)

Main Menu

To display the Main Menu, select the icon in the upper left corner of the header on any page of the display.

Select Settings to setup the following:

Manuals, Software, and Other Resources

www.tsi.com

Q-Trak 7585 Indoor Air Quality Monitor

Specifications:

  • Model: Q-TrakTM XP Indoor Air Quality Monitor 7585
  • Includes: Q-TrakTM XP Base Meter 7580, Multi-Sensor IAQ Module 801430, Carbon Dioxide (CO2) sensor 801399
  • Additional Parts : Carrying Case, Lithium Ion Rechargeable Battery Pack, USB Cable, AC Adapter/Power Supply, Gas Sensor Calibration Cap
  • Can accept up to six gas sensors

Product Usage Instructions

Installation of Gas Sensors:

  1. Ensure the instrument is powered OFF.
  2. Remove the sensor module cover.
  3. Position the sensor making good contact with the device. Gas sensors can be installed in any available slot.
  4. Listen for a click to confirm the proper connection.
  5. Attach the cover to the sensor module.

Providing Power:

The Q-TrakTM XP monitor can be powered by a lithium-ion battery or an A/C adapter.

Operating with Lithium-Ion Battery:

  1. Install the battery by connecting it to the handle connector after removing the battery compartment cover.
  2. Charge the battery fully before use.
  3. Calibrate the battery power gauge by fully discharging and then fully charging the battery.

Operating with AC Adapter:

The instrument can be operated with or without a battery using the A/C adapter.

FAQ

  • Q: How many gas sensors can the device accept?
    • A : The device can accept up to six gas sensors in addition to the standard built-in sensors.
  • Q: How long does it take to fully charge the battery?
    • A: Charging the battery may take up to 3 hours. Note the lightning bolt indicator when charging vs fully charged.

Thank you for purchasing a TSI® Incorporated (TS|®)

Q-Trak™ XP Indoor Air Quality (IAQ) Monitor. This guide will help you quickly begin using your monitor.

More detailed information is in the Operation Manual.

Please refer to the manual if you have questions on the operation of your new monitor.

Safety

Refer to the Q-Trak™ XP Indoor Air Quality Monitor Mode 7585 user manual for safety information.

Unpacking and Parts Identification

Q-Trak-7585-Indoo-Air-Quality-Monitor-FIG \(1\)

Q-Trak-7585-Indoo-Air-Quality-Monitor-FIG \(3\)

Carefully unpack the instrument and accessories from the shipping container. Check the individual parts against the list of components below. If anything is missing or damaged, notify TIS/® immediately. All standard equipment can be purchased separately if needed.

Installation of Gas Sensors

In addition to the standard built-in sensors, the device can accept up to six gas sensors.

Install the gas sensors as follows:

  1. Ensure the instrument is powered OFF.
  2. Remove the sensor module cover.Q-Trak-7585-Indoo-Air-Quality-Monitor-FIG \(2\)
  3. Position the sensor as shown, making certain the pins make good contact with the device. Gas sensors can be installed in any available slot. Listen for a click to confirm sensor is connected properly.
  4. Attach the cover to the sensor module.

Providing Power to the Q-Trak ™ XP Monitor

The Q-Trak™ XP monitor can be powered with a lithium-ion battery or with the A/C adapter.

Operating the Instrument with the

Lithium-lon Battery

The lithium-ion battery must be installed by the user and fully charged before use.

Install the battery:

  1. Remove the battery compartment cover.

  2. Connect the battery to the handle connector.
    Q-Trak-7585-Indoo-Air-Quality-Monitor-FIG \(5\),..,.,

  3. Insert the battery as shown in the picture below, keeping the battery wires to the side and clear of the battery cover screw.

Installing the Wi-Fi° Dongle (optional)

If purchased, install the dongle into the USB port located inside the battery compartment.

Attach the battery compartment cover, ensuring battery wires are clear of the screw.

IMPORTANT NOTICE

Calibrate the battery power gauge

Charging the Battery and Calibrating the Power Gauge

  1. With the battery installed, turn on the instrument by pressing the power button and run until the battery is fully discharged. The instrument will automatically turn off when the battery is fully discharged.
  2. Connect the AC adapter to the instrument and fully charge the battery. This may take up to 3 hours. Note the lightning bolt indicator when charging vs fully charged
  3. The instrument is now ready for use with the battery. This Power Gauge calibration procedure does not need to be repeated unless the battery is replaced or disconnected.

Operating the Instrument with the AC Adapter

The instrument can be operated with or without a battery using the A/C adapter.

To operate the instrument with AC adapter:

  1. Plug the AC adapter into an electrical outlet. Plug the other end into the instrument.
  2. Plug the other end into the instrument.

Powering ON the Q-Trak T™ XP IAQ Monitor

  1. Press and release the power button . In 35 seconds the instrument is ready for use.
  2. The first time the Q-Trak™ XP monitor is powered ON the Welcome to Q-Trak™ XP page is presented. Select the appropriate Language and Time Zone then select DONE.
  3. After selecting DONE, the Dashboard page is displayed in Survey mode.

Q-Trak-7585-Indoo-Air-Quality-Monitor-FIG \(14\)

Main Menu

To display the Main Menu, select the icon in the upper left corner of the header on any page of the display.

Select Settings to setup the following:

Manuals, Software, and Other Resources

www.tsi.com

Q-Trak 7585 Indoor Air Quality Monitor

Specifications:

  • Model: Q-TrakTM XP Indoor Air Quality Monitor 7585
  • Includes: Q-TrakTM XP Base Meter 7580, Multi-Sensor IAQ Module 801430, Carbon Dioxide (CO2) sensor 801399
  • Additional Parts : Carrying Case, Lithium Ion Rechargeable Battery Pack, USB Cable, AC Adapter/Power Supply, Gas Sensor Calibration Cap
  • Can accept up to six gas sensors

Product Usage Instructions

Installation of Gas Sensors:

  1. Ensure the instrument is powered OFF.
  2. Remove the sensor module cover.
  3. Position the sensor making good contact with the device. Gas sensors can be installed in any available slot.
  4. Listen for a click to confirm the proper connection.
  5. Attach the cover to the sensor module.

Providing Power:

The Q-TrakTM XP monitor can be powered by a lithium-ion battery or an A/C adapter.

Operating with Lithium-Ion Battery:

  1. Install the battery by connecting it to the handle connector after removing the battery compartment cover.
  2. Charge the battery fully before use.
  3. Calibrate the battery power gauge by fully discharging and then fully charging the battery.

Operating with AC Adapter:

The instrument can be operated with or without a battery using the A/C adapter.

FAQ

  • Q: How many gas sensors can the device accept?
    • A : The device can accept up to six gas sensors in addition to the standard built-in sensors.
  • Q: How long does it take to fully charge the battery?
    • A: Charging the battery may take up to 3 hours. Note the lightning bolt indicator when charging vs fully charged.

Thank you for purchasing a TSI® Incorporated (TS|®)

Q-Trak™ XP Indoor Air Quality (IAQ) Monitor. This guide will help you quickly begin using your monitor.

More detailed information is in the Operation Manual.

Please refer to the manual if you have questions on the operation of your new monitor.

Safety

Refer to the Q-Trak™ XP Indoor Air Quality Monitor Mode 7585 user manual for safety information.

Unpacking and Parts Identification

Q-Trak-7585-Indoo-Air-Quality-Monitor-FIG \(1\)

Q-Trak-7585-Indoo-Air-Quality-Monitor-FIG \(3\)

Carefully unpack the instrument and accessories from the shipping container. Check the individual parts against the list of components below. If anything is missing or damaged, notify TIS/® immediately. All standard equipment can be purchased separately if needed.

Installation of Gas Sensors

In addition to the standard built-in sensors, the device can accept up to six gas sensors.

Install the gas sensors as follows:

  1. Ensure the instrument is powered OFF.
  2. Remove the sensor module cover.Q-Trak-7585-Indoo-Air-Quality-Monitor-FIG \(2\)
  3. Position the sensor as shown, making certain the pins make good contact with the device. Gas sensors can be installed in any available slot. Listen for a click to confirm sensor is connected properly.
  4. Attach the cover to the sensor module.

Providing Power to the Q-Trak ™ XP Monitor

The Q-Trak™ XP monitor can be powered with a lithium-ion battery or with the A/C adapter.

Operating the Instrument with the

Lithium-lon Battery

The lithium-ion battery must be installed by the user and fully charged before use.

Install the battery:

  1. Remove the battery compartment cover.

  2. Connect the battery to the handle connector.
    Q-Trak-7585-Indoo-Air-Quality-Monitor-FIG \(5\),..,.,

  3. Insert the battery as shown in the picture below, keeping the battery wires to the side and clear of the battery cover screw.

Installing the Wi-Fi° Dongle (optional)

If purchased, install the dongle into the USB port located inside the battery compartment.

Attach the battery compartment cover, ensuring battery wires are clear of the screw.

IMPORTANT NOTICE

Calibrate the battery power gauge

Charging the Battery and Calibrating the Power Gauge

  1. With the battery installed, turn on the instrument by pressing the power button and run until the battery is fully discharged. The instrument will automatically turn off when the battery is fully discharged.
  2. Connect the AC adapter to the instrument and fully charge the battery. This may take up to 3 hours. Note the lightning bolt indicator when charging vs fully charged
  3. The instrument is now ready for use with the battery. This Power Gauge calibration procedure does not need to be repeated unless the battery is replaced or disconnected.

Operating the Instrument with the AC Adapter

The instrument can be operated with or without a battery using the A/C adapter.

To operate the instrument with AC adapter:

  1. Plug the AC adapter into an electrical outlet. Plug the other end into the instrument.
  2. Plug the other end into the instrument.

Powering ON the Q-Trak T™ XP IAQ Monitor

  1. Press and release the power button . In 35 seconds the instrument is ready for use.
  2. The first time the Q-Trak™ XP monitor is powered ON the Welcome to Q-Trak™ XP page is presented. Select the appropriate Language and Time Zone then select DONE.
  3. After selecting DONE, the Dashboard page is displayed in Survey mode.

Q-Trak-7585-Indoo-Air-Quality-Monitor-FIG \(14\)

Main Menu

To display the Main Menu, select the icon in the upper left corner of the header on any page of the display.

Select Settings to setup the following:

Manuals, Software, and Other Resources

www.tsi.com

TSI and the TSI logo are registered trademarks of TSI Incorporated in the United States and may be protected under other country’s trademark registrations.

TSI Incorporated – Visit our website www.tsi.com for more information.

USA

Tel: +1 800 680 1220

Germany

Tel: +44 149 4 459200

France

Tel: +33 1 41 19 21 99

UK

Tel: +49 241 523030

India
Tel: +91 80 67877200

China
Tel: +86 10 8219 7688

Singapore
Tel: +65 6595 6388

©2024 TSI Incorporated