Application
The P18L transducer is a device destined for the continuous conversion of relative humidity or ambient temperature into a standard analog current signal. The transducer is fixed on a wall. Applied sensor shields enable the application of the P18L transducer in various ambient conditions
**** Basic requirements, operational safety
In the security scope, the transducer meets the requirements of the EN 61010 -1 standard.
R emarks concerning the operator safety:
1. General
- All operations concerning transport, installation, and commissioning as well as maintenance must be carried out by qualified, skilled personnel acc. to the section 3.2. of the user’s manual.
- According to this basic safety information, qualified, skilled personnel are persons who are familiar with the installation, assembly, commissioning, and operation of the product and who have qualifications necessary for their occupation.
- In order to avoid any damage, before switching the supply on, it is recommended to ensure that all wires were correctly connected. Before the transducer assembly, one must pass supplying wires through the packing. Screw the packing seal in order to obtain the leaktightness. If the packing seal is not screwed, we cannot ensure the required IP65 leaktightness.
When unpacking the P18L transducer, please check whether the type and version code on the data plate correspond to the order code.
Installation
3.1. Assembly
The P18L transducer is designed to be mounted on a wall by means of a screw
connection or glue without the loss of IP 65 tightness.
The transducer housing is made of a self-extinguishing plastics.
Housing dimensions: 64 x 58 x 35 mm.
The transducer has screw connectors placed inside the transducer, which enable
the connection of external wires of 1 mm² cross-section.
Overall dimensions and the fixing way are shown on the fig. 1
3.2. Electrical connections
The P18L transducer has 2 connecting terminals to which there is access after
removing the cover of the transducer housing.
In case of the transducer work in an environment with high interference, one must apply shielded wires. The shield must be connected to the nearest PE point from the feeder side.
Servicing
After connecting wires, closing and servicing the housing, and switching to the supply on, the transducer is ready to work. The correct measurement result appears only after the time of the measurement initialisation ca 2.5 minutes. After this laps of time, the analog output is correctly steered on. The transducer converts the temperature or relative humidity into a 4-20 mA standard current signal. The choice of the converted quantity is realized by means of the jumper Nr 1 of the 3-position switch placed on the transducer plate. If the jumper Nr 1 is in the position “ON”, the transducer converts the relative humidity. (Fig. 5).
a) temperature,
b) relative humidity
Accessories
As a standard, the P18L transducer is equipped with a metallic shield of the
sensor, destined only for indoors application.
For other applications it is recommended to use additional sensor shields,
depending on the transducer operation conditions.
Sensor protection shields
Item | Order code | Design | Name | Construction | Features | Typical application |
---|---|---|---|---|---|---|
1 | 20-015-00- 00011 | Membra- | ||||
ne filter | Casing made of PC Membrane of teflon lami- |
nated by a film.
Pore size:
1 pm| Mean filtra- tion effect. Maximal temp.: up to 80°C Response time:
t10/90: 15 s| Building automation. In rooms with low pollution.
2| 20-015-00- 00007| | Filter
made of
teflon| Sintered teflon.
Pore size:
50 pm| High chemical resistance..
Maximal temp.: up to 180°C Response time: t10/90: 14 s| Drying process in
chemical applications
3| 20-015-00- 00003| | Filter made of
sintered
bronze| Sintered bronze.
Pore size: 60 pm| High mechanical resistance. To co-opera- to with high
pollution. Applied at small air humidity. Response time:
t10/90: 10 s| Agricultural applications
Technical data
Basic parameters:
– range of relative humidity measurement (RH)
– basic error of relative humidity conversion 1)
– hysteresis of the humidity measurement (T)
– range of temperature
measurement (T)
– basic error of temperature conversion
– additional errors
– temperature influence| 0…100% without condensation
- ± 2% for RH = 10…90%
± 3% for remaining range
± 1% RH
– 20…60°C
± 0.5% of the range
± 25% of the basic error/10°C
---|---
Analog output:
– current
– maximal load resistance of the current output| 4… 20 mA
500 Ω
---|---
Rated operating conditions:
– supply
– consumption
– ambient temperature
– relative air humidity
– rate of air flow| 19…30 V d.c.
< 1.5 VA
– 30…23…85°C
< 95% (admissible condensation)
2 m/s
---|---
- In case of water vapour condensation on the sensor surface, the maximal error can increase to 3% RH
– pre-heating time
– protection degree ensured
by the housing
– fixing
– weight
– dimensions
– work position:
• in applications non-exposed to a direct contact with water
• in applications exposed to a direct contact with water| 15 minutes
IP 65
on a wall
125 g
(35 ´ 58 ´ 118) mm
any
with the sensor chamber
directed towards the ground.
---|---
Electromagnetic compatibility:
– immunity noise
– emission noise| acc. to EN 61000-6-2
acc. to EN 61000-6-4
---|---
Security requirements:
– installation category
– pollution grade
– working voltage in relation to the earth| III
2
50V
---|---
Before a failure will be declared
Possibilities of incorrect work of the P18L transducer Table 2.
Symptoms | Procedure |
---|---|
Data inconsistent with our expec- tations appear on the transducer output. | In |
the version with analog outputs, check parameters of the individual characteristic of analog outputs.
Ordering codes
Ordering codes of P18L transducer versions Table 3.
Code | Description |
---|---|
P18L 000 | Temperature and humidity transducer P18L, built-in temperature |
and humidity sensor, supplied from a current loop, 1 x analog output 4 … 20 mA; without any extra requirements
LUMEL S.A.
ul. Słubicka 4, 65-127 Zielona Góra, Poland
tel.: +48 68 45 75 100, fax +48 68 45 75 508
www.lumel.com.pl
Technical support:
tel.: (+48 68) 45 75 143, 45 75 141, 45 75 144, 45 75 140
e-mail: export@lumel.com.pl
Export department:
tel.: (+48 68) 45 75 130, 45 75 132
e-mail: export@lumel.com.pl
Calibration & Attestation:
e-mail: laboratorium@lumel.com.pl
P18L-09_R1
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>