LUMEL VA28B Smart Digital Multimeter User Manual
- September 28, 2024
- LUMEL
Table of Contents
VA28B Smart Digital Multimeter
“`html
Specifications
- Model: VA28B
- Measurement Category: CAT III 600V
- Overvoltage Category: CAT III 600V
- Pollution Degree: 2
Product Usage Instructions
Safety Information
Measurement category III is for the measurements performed on
circuits directly connected to the low voltage installation. Follow
all safety and operating instructions to ensure safe usage. Always
adhere to the following:
-
Do not exceed the protection limit for each measurement
range. -
Avoid measuring voltages exceeding 600V above earth ground in
category III installations. -
Exercise caution when working with voltages above 60V dc or 30V
ac rms. -
Avoid performing resistance measurements on live circuits.
-
Inspect test leads and probes for any damage before use.
Maintenance
To ensure proper maintenance and safety:
-
Always disconnect test leads before opening the case.
-
Replace fuse only with specified ratings.
-
Do not use the meter if the back cover is not closed
completely. -
Clean the meter with a damp cloth and mild detergent, avoiding
abrasives or solvents.
Operating Instructions
This smart digital multimeter can perform various measurements
including:
-
SCAN (SMART) Measurement Mode: Select this
mode for smart measurements. -
Voltage Measurement: Measure AC and DC
voltage. -
Capacitance Measurement: Measure
capacitance. -
Resistance Measurement: Perform resistance
measurements. -
Frequency Measurement: Measure frequency.
-
Temperature Measurement: Use K-type
thermocouple for temperature measurements. -
Diode Test: Test diodes for
functionality. -
Audible Continuity Test: Check continuity
audibly. -
LIVE Test: Perform live line tests with only
one red test lead plugged in. -
Non-contact Voltage Detect (NCV/EF) Test:
Detect voltage without contact.
FAQ
Q: What should I do if the meter displays an overvoltage
warning?
A: If an overvoltage warning is displayed, immediately
disconnect the meter from the circuit to prevent damage to the
device or its components.
Q: Can I measure resistance on live circuits?
A: It is not recommended to perform resistance measurements on
live circuits to avoid potential hazards and inaccuracies in
readings.
Q: How do I replace the battery or fuse of the meter?
A: Refer to the user manual for detailed instructions on
replacing the battery and fuse of the digital multimeter.
“`
SMART DIGITAL MULTIMETER
VA28B
USER’S MANUAL
1
Spis treci 1. 2. 3. 4. 5. 6. 7.
8. 9.
SAFETY INFORMATION
4
DURING USE
4
SAFETY SYMBOL
4
MAINTENACE
5
GENERAL DESCRIPTION
5
PANNEL AND LCD
6
SPECIFICATION
6
7.2 Non-contact Voltage detect
7
7.3 LIVE test
7
7.4 Temperature Measurement (K-type thermocouple)
7
7.5 Current ((Depending Model))
8
7.6 Resistance
8
7.7 Continuity Check
9
7.8 Diode Test
9
7.9 Capacitance (Depending model)
9
7.10 Linear Frequency
10
7.11 SCAN(SMART) measure mode range
10
GENERAL SPECIFICATIONS
11
OPERATING INSTRUCTION
12
9.1 SCAN (SMART) Measurement Mode
12
9.2 Voltage Measurement
12
9.4 Capacitance Measurement
13
9.5 Resistance Measurement
13
9.6 Frequency Measruement
13
9.7 Temperature Measurement
13
9.8 Diode Test
13
9.9 Audible Continuity Test
14
9.10 LIVE Test (ONLY ONE RED TESTLEAD PLUGIN) 14
9.11 Non-contact Voltage Detect (NCV/EF) Test
14
2
KEY FUNCTION
14
10.1 Hold Key / Backlight Key
14
10.2 FUNC Key / Range Key
15
AUTO POWER OFF FUNCTION
15
CURREN PLUG-IN DETECT
15
DANGER VOLTAGE ALARM
15
BATTERY & FUSE REPLACEMENT
15
14.1 To replace the Meter’s battery:
15
14.2 To replace the Meter’s fuse:
16
WARNING!
16
ACCESSORIES
16
3
1. SAFETY INFORMATION
Measurement category III is for the measurements performed on circuits
directly connected to the low voltage installation. This meter has been
designed according to IEC-61010-1 concerning electronic measuring instruments
with an overvoltage category (CAT III 600V) and pollution degree 2. Follow all
safety and operating instructions to ensure the meter is used safely and is
kept in good condition. With proper use and care, your digital multimeter will
give you years of satisfactory service.
2. DURING USE
Never exceed the protection limit indicated in the specifications for each
range of measurement.
Never use the meter to measure voltages that might exceed 600V above earth
ground in category III installations.
Always be careful when working with voltages above 60V dc or 30V ac rms. Keep
fingers behind the probe barriers while measuring.
Do not perform resistance measurements on live circuits Inspect test leads and
probes for cracks, breaks or crazes in the insulation before using the meter.
3. SAFETY SYMBOL
Caution: refer to the instruction manual. Incorrect use may result in damage
to the device or its components.
~ AC (Alternating Current)
4
DC (Direct Current)
AC or DC
Earth ground
Double insulated
Fuse
Conforms to European Union directives
4. MAINTENACE
Before opening case, always disconnect test leads from all energized circuits.
For continuous protection against fire, replace fuse only with ratings: 1A
range: F 1A /600V 630 (Quick Acting). 10A range: F 10A /600V 630 (Quick
Acting) Never use the meter unless the back cover is closed completely. Do not
use abrasives or solvents on the meter. To clean it use only a damp cloth and
mild detergent.
5. GENERAL DESCRIPTION
This mini smart digital multimeter is designed to measure AC and DC voltage,
AC and DC current, Resistance, Capacitance, Diode, Frequency, Non-contact
voltage, Live line detect and audible continuity checks with accuracy and
easy. Small and light weight device with test leads set, this instrument will
provide you years of satisfactory service
5
6. PANNEL AND LCD
1A model
10A model
No Current model
7. SPECIFICATION
Accuracy is specified for one year after calibration, at operating temperatures of 18 to 28, with relative humidity at 0% to 75%. Accuracy specifications take the form of: ±(% of Reading + Number of Least Significant Digits)
7.1 Voltage
Function
DC Voltage V
Range
4,000 V 40,00 V 400,0 V 600 V
Resolution
1 mV 10 mV 100 mV
1 V
6
Accuracy ±(0.5% of rdg +3 digits)
AC Voltage1,2
V~
4,000 V
40,00 V 400,0 V
1 mV
10 mV 100 mV
600 V
1 V
1. Frequency Range: 40Hz~1kHz RMS. 2. AC minimum measurement: 5% of lowest range; 3. Overload Protection: 600V dc or 600V ac rms.
±(1.0% of rdg + 6 digits) ±(1.0% of rdg + 3digits)
7.2 Non-contact Voltage detect
Voltage
Frequency
Indication
50~1000V
50Hz~400Hz 4Bars display/ Alarm light/Beep
7.3 LIVE test
Voltage 100~1000V
Frequency
Indication
50Hz~400Hz “Hi” display/ Alarm light/Beep
7.4 Temperature Measurement (K-type thermocouple)
Range -200~1000 -328~1832
Resolution 1 1
Accuracy ±(2% of rdg +3 digits) ±(2% of rdg +6 digits)
7
7.5 Current ((Depending Model))
Function
1A model
10A model
Resolution
Accuracy
DC Current mA
AC Current mA ~
40,00 mA 400,0 mA 1,000 A
/ 40,00 mA 400,0 mA
1,000 A
/ / 4000 mV 10,00 A / /
4000 mA
0,01 mA 0,1 mA 1 mA 1 mA 0,01 mA 0,1 mA
1 mA
±(1% of rdg+3 digits) ±(1.5% of rdg+3 digits) ±(1.5% of rdg+3 digits)
/
10,00 A
10 mA
Overload protection: 1A range: Maximum input 1A DC or AC RMS. F 1A/600V fuse. 10A range: Maximum input 10A DC or AC RMS. F 10A/600V fuse. Over Load indication: OL Displayed. >1A for 1min load on then 10min load off. Make sure A terminal socket have a good connect.
±(2% of rdg+3 digits)
7.6 Resistance
Function
Range 400,0
Resolution 0,1
4,000 k
1
Resistance
40,00 k 400,0 k
10 100
4,000 M
1 k
40,00 M
10 k
Overload protection: 600V dc or 600V ac rms.
Accuracy ±(0.5% of rdg+3 digits)
±(0.5% of rdg+2 digits)
±(1.5% of rdg+3 digits)
8
7.7 Continuity Check
Function Continuity Test
Range 200
Resolution 0,1
Overload protection: 600V dc or 600V ac rms. Test Condition: Open circuit voltage: approx. 0.5V
7.8 Diode Test
Description Continuity beeper50
Function Diode test
Range 1V
Resolution 0,001 V
Overload protection: 600V dc or 600V ac rms. Test Condition: Forward DC current approximately 1mA. Reversed DC voltage approximately 1.5V
Accuracy 1.0% uncertainty
7.9 Capacitance (Depending model)
Function
Range
Resolution
Accuracy
4,000 nF
1 pF
±5.0% of rdg+30 digits]
40,00 nF
10 pF
Capacitance
400,0 nF 4,000F 40,00F
0,1 nF 1 nF 10 nF
±(3.0% of rdg+5 digits)
400,0F
0,1 uF
1,000 mF
1 uF
Overload protection: 600V dc or 600V ac rms. (Auto range only.)
9
7.10 Linear Frequency
Range 10.00~40.00Hz
Resolution 0.01Hz
40.0~400.0Hz
0.1Hz
400~4000Hz
1Hz
Overload protection: 600V dc or 600V ac rms. Cannot accept below 10Hz and Over 1KHz.
Accuracy ±(0.5% of rdg+3 digits)
7.11 SCAN(SMART) measure mode range
Function
Range
DC Voltage
0.700V~600.0V
AC Voltage
0.700V~600.0V
Resistance
50.0~40.00M
Continuity
0.0~50.0
DC Current
1mA~10.00A
AC Current
4mA~10.00A
Please check Accuracy at Above function table
10
8. GENERAL SPECIFICATIONS
Environment conditions: 600V CAT III MAX. Voltage between terminals and earth
ground: 600V AC rms or 600V DC. Pollution degree: 2 Altitude < 2000m Operating
temperature: 0~40 (32~122) Storage temperature: -10~60 (14~140) Fuse
Protection: 1A range: F 1A /600V 630 (Quick Acting). 10A range: F 10A /600V
630 (Quick Acting). Sample Rate: 3 times/sec for digital data. Display: 3999
LCD display. Automatic indication of functions and symbols. SMART/SCAN
function: Auto recognize measurement function (AC or DC
Voltage/Resistor/Continuity/AC or DC Current). Range selection: automatic.
Over Range indication: display ,,OL”. Low battery indication: Yes Polarity
indication: ,,” displayed automatically. Hazardous voltage indication: >36V ,,
,, displayed. Current Mis-plug alarm: Alarm light/Beep. Auto Power off: 30
Minutes. Backlight and Alarm light and Flashlight: Yes Battery type: 3V,
AAA*2. (3.7V 700mAh Li changeable battery with TYPEC USB is option)
Dimensions: 130(L)63(W)35(H) mm. Weight: 110g. Approx. (battery included).
11
9. OPERATING INSTRUCTION
9.1 SCAN (SMART) Measurement Mode 1. When LCD display SCAN symbol, unless use
FUNC. Key change. 2. Connect the black and red test leads plug to the COM and
V terminal. (When measure current, RED test lead Must plug-in mA/A terminal).
3. Connect the test leads probe to the target. 4. Device will according the
target type to select measure function by self, target should be AC or DC
Voltage/Resistor/Continuity/AC or DC Current.
9.2 Voltage Measurement 1. Use the FUNC. Key can select DCV or ACV measure
mode. 2. Connect the black and red test leads plug to the COM and V terminal.
3. Connect the test leads probe to the target. 4. Read the displayed value.
Red test lead connection will be positive when making a DCV measurement.
9.3 Current Measurement 1. Turn off power to the circuit. Discharge all high
voltage capacitors. 2. Connect the black and red test leads to the COM and
mA/A terminal. 3. Use the FUNC. Key can select DCA or ACA measure mode. 4.
Break the circuit path to be tested. Touch the black probe to the more
negative side of the break; touch the red probe to the more positive side of
the break. (Reversing the leads will give a negative reading, but will not
damage the Meter.) 6. Turn on power to the circuit; then read the display. 7.
Turn off power to the circuit and discharge all high voltage capacitors.
Remove the Meter and restore the circuit to normal operation. *At current
function, red blacklight will interval flash for remind.
12
9.4 Capacitance Measurement 1. Use the FUNC. Key can select Capacitance
measure mode. 2. Connect the black and red test leads plug to the COM and V
terminal. 3. Connect the test leads probe to the target. 4. Read the displayed
value. Discharge all high voltage capacitors before measure. Do not input a
Voltage source at this mode.
9.5 Resistance Measurement 1. Use the FUNC. Key can select Resistance measure
mode. 2. Connect the black and red test leads plug to the COM and V terminal.
3. Connect the test leads probe to the target. 4. Read the displayed value. Do
not input a Voltage source at this mode.
9.6 Frequency Measruement 1. Use the FUNC. Key can select Frequency measure
mode. 2. Connect the black and red test leads plug to the COM and V terminal.
3. Connect the test leads probe to the target. 4. Read the displayed value.
9.7 Temperature Measurement 1. Use the FUNC. Key can select Temperature
measure mode. 2. Connect the K-type thermocouple sensor to COM and V terminal
and read the displayed value.
9.8 Diode Test 1. Use the FUNC. Key can select Diode measure mode. 2. Connect
the black and red test leads plug to the COM and V terminal. 3. Connect the
test leads probe to the target. Red test lead connection will be positive. 4.
The meter will show the approx. forward voltage of the diode. Do not
13
input a Voltage source at this mode.
9.9 Audible Continuity Test 1. Use the FUNC. Key can select Continuity measure
mode. 2. Connect the black and red test leads plug to the COM and V terminal.
3. Connect the test leads probe to the target. 4. The meter will show the
Resistance for connection. When the circuit is below 50, a continuous beeping
will indicate it. Do not input a Voltage source at this mode.
9.10 LIVE Test (ONLY ONE RED TESTLEAD PLUGIN) 1. Hand hold the meter. Keep
press the FUNC. Key 3second and then press once FUNC. Key to select LIVE test
function. 2. Connect the red test lead plug to the V terminal. (Only need one
RED test lead) 3. Connect the red test leads to the circuit being measured 4.
“Hi” will been shown when connect the red test leads to LIVE wire.
9.11 Non-contact Voltage Detect (NCV/EF) Test 1. Keep press the FUNC. Key
3second to select NCV function. 2. Make upper right corner of device (marking
NCV) close to test wire/ socket. 3. It will showing 4 bars according LIVE
voltage level and distance.
10. KEY FUNCTION
10.1 Hold Key / Backlight Key Data Hold function: Press once (short press)
Data Hold mode makes the meter stop updating the display. Backlight and flash
light on/off: Keep press 3seconds (long press).
14
10.2 FUNC Key / Range Key Function key: Press once (short press) Switches
alternate function for: SCAN function, DCV, ACV, Resistance, Continuity,
Diode, Capacitance, Frequency, Temperature. Into/Exit EF/LIVE test mode: Keep
press 3seconds (long press).
11. AUTO POWER OFF FUNCTION
The Meter enters the ,,sleep mode” and blanks the display if the Meter is on
but not used for 30 minutes. Keep press “HOLD” key then Press “ON/OFF” key to
turn on meter, meter will disable auto power off function, auto power off
symbol will be disappeared.
12. CURREN PLUG-IN DETECT
When test lead plug-in at mA/A terminal, then any measurement function will be
changed to current measurement function.
13. DANGER VOLTAGE ALARM
Device getting a Hazardous voltage >36V, a thunder symbol ,, ” will be
displayed.
14. BATTERY & FUSE REPLACEMENT
14.1 To replace the Meter’s battery: If the sign ,, ” appears on the LCD
display, it indicates that the battery should be replaced. Remove the screw on
the back cover and open the
15
battery case. Replace the exhausted batteries with two new 1.5V batteries of the same type (AAA).
14.2 To replace the Meter’s fuse:
Fuse rarely need replacement and blow almost always as a result of operator’s
error. Open the case and replace the blown fuse with the same rating
specified: 1A Model: F 1A /600V 630. 10A Model: F 10A /600V 630.
15. WARNING!
Before attempting to open the case, always be sure that test leads have been disconnected from measurement circuits. Close case and tighten screws completely before using the meter to avoid electrical shock hazard.
16. ACCESSORIES
Battery Test Lead User manual Temperature sensor
1.5V (AAA) X 2pcs 1set 1 piece 1 piece