Hillrom A-71404 Advance Cervical Traction Bar Instruction Manual
- June 9, 2024
- Hillrom
Table of Contents
- Hillrom A-71404 Advance Cervical Traction Bar
- INSTRUCTIONS FOR USE
- General Information
- Safety Considerations
- Operating the system
- System
- Indication for use
- Equipment Setup and Use
- Safety Precautions and General Information
- Product Specifications
- Cleaning and Disinfection Instruction
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Hillrom A-71404 Advance Cervical Traction Bar
INSTRUCTIONS FOR USE
IMPORTANT NOTICES
WARNING: Prior to using this or any other type of medical apparatus with a patient, it is recommended that you read the Instructions for Use and familiarize yourself with the product.
-
Read and understand all warnings in this manual and on the device itself prior to use with a patient.
-
The symbol is intended to alert the user to important procedures or safety instructions regarding the use of this device.
The symbol on the labels is intended to show when the IFU should be referenced for use. -
The techniques detailed in this manual are only manufacturer’s suggestions. The final responsibility for patient care with respect to this device remains with the attending physician.
-
Device function should be checked prior to each usage.
-
This device should only be operated by trained personnel.
-
All modifications, upgrades, or repairs must be performed by an authorized specialist.
-
Keep this manual available for future reference.
-
Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority listed in this document.
General Information
Allen Medical Systems, Inc. is a subsidiary of Hill-Rom, Inc., (NYSE: HRC), a leading worldwide manufacturer and provider of medical technologies and related services for the health care industry. As an industry leader in patient positioning, our passion is improving patient outcomes and caregiver safety, while enhancing our customers’ efficiency. Our inspiration comes from providing innovative solutions to address our customers’ most pressing needs. We immerse ourselves in our customers’ world, to better address these needs and the daily challenges of their environment. Whether developing a solution to address patient positioning challenges or creating a system to offer safe and effective surgical site access for the surgical team, we are committed to providing products of exceptional value and quality. Allen products are backed by responsive and reliable service and complimentary on-site product demonstrations.
Copyright Notice
Revision
© 2019 Allen Medical Systems Inc. ALL RIGHTS RESERVED.
No part of this text shall be reproduced or transmitted in any form or by any
means, electronic or mechanical, including photocopying, recording, or by any
information or retrieval system without written permission from Allen Medical
Systems, Inc. (Allen Medical).The information in this manual is confidential
and may not be disclosed to third parties without the prior written consent of
Allen Medical.
Trademarks
Trademark information can be found at Allenmedical.com/pages/terms-
conditions.Products may be
covered by one or more patents. Please consult listing at
Hill-rom.com/patents for any patent(s).
Contact Details
- For ordering information please see catalog. Allen Customer Service Contact Information
Safety Considerations
Safety hazard symbol notice
WARNING: DO NOT USE IF PRODUCT SHOWS VISIBLE DAMAGE OR MATERIAL DEGRADATION.
Equipment misuse notice:
Do not use the product if package is damaged or unintentionally opened before
use. All modifications, upgrades, or repairs must be performed by an
authorized specialist.
Notice to users and/or patients:
Any serious incident that has occurred in relation to the device should be
reported to the manufacturer and the competent authority of the Member State
in which the user and/or patient is established.
Note: Refer to the surgical table manufacturer’s user guide for
instructions on use. Always refer to the surgical table manufacturer’s weight
limits.
WARNING: NEVER EXCEED THE WEIGHT CAPACITY OF THE OPERATING ROOM TABLE
Safe disposal:
Customers should adhere to all federal, state, regional, and/or local laws and
regulations as it pertains to the safe disposal of medical devices and
accessories.
If in doubt, the user of the device shall first contact Hill-Rom Technical
Support for guidance on safe disposal protocols.
Operating the system
Applicable Symbols
Symbol used | Description | Reference |
---|
| Indicates the device is a medical device| ****
MDR 2017/745
| Indicates the medical device manufacturer|
EN ISO 15223-1
| Indicates the manufacturer’s serial number. The device serial number is encoded as 1YYWWSSSSSSS.
· YY indicates the year of manufacture. i.e. 1 18 WWSSSSSSS where 18 represents the year 2018.
· WW indicates the number of the manufacturing week per a standard shop calendar. (Leading zeros included.)
· SSSSSSS is a sequential unique
number.
| ****
EN ISO 15223-1
****| Indicates the medical device Global Trade Item Number| 21 CFR 830 MDR 2017/745
| Indicates the manufacturer’s lot code using the Julian Date yyddd, where yy indicates the last two digits of the year and ddd indicates the day of the year,
i.e. April 4th, 2019 would be represented
as 19094.
| ****
EN ISO 15223-1
|
Indicates the date when the medical device was manufactured
| ****
EN ISO 15223-1
****| Indicates the manufacturer’s
catalogue number
| ****
EN ISO 15223-1
****| Indicates the need for the user to consult the instructions for use for important cautionary information such
as warnings and precautions.
| ****
EN ISO 15223-1
---|---|---
| Indicates the device do not contain natural rubber or dry natural
rubber latex|
EN ISO 15223-1
| Indicates the authorized representative in the European Community| ****
EN ISO 15223-1
****| Indicates the Medical Device complies to REGULATION (EU) 2017/745|
MDR 2017/745
|
Indicates a Warning
| ****
IEC 60601-1
****| Indicates the need for the user to consult the instruction for use|
EN ISO 15223-1
Intended User and Patient Population :
Intended User: Surgeons, Nurses, Doctors, Physicians and OR healthcare
professionals involved in the device intended procedure. Not intended for Lay
persons.
Intended Populations:
This device is intended to be used with adults and pediatric patients
Compliance with medical device regulations
This Product is a non-invasive, Class I Medical Device. This system is CE- marked according to Annex VIII, Rule 1, of the Medical Device Regulations (REGULATION (EU) 2017/745).
EMC considerations
This is not an electromechanical device. Therefore, EMC Declarations are
not applicable.
EC authorized representative
- HILL-ROM SAS
- B.P. 14 – Z.I. DU TALHOUET 56330 PLUVIGNER FRANCE
- TEL: +33 (0)2 97 50 92 12
Manufacturing Information
- ALLEN MEDICAL SYSTEMS, INC. 100 DISCOVERY WAY
- ACTON, MA 01720 US
- 800-433-5774 (NORTH AMERICA) 978-266-4200 (INTERNATIONAL)
EU Importer Information
- Baxter Medical Systems GmbH + Co. KG Carl-Zeiss Straße 7-9
- 07318 Saalfeld/Saale
- Germany
Authorized Australian sponsor
- Welch Allyn Australia Pty. Ltd. 1 Baxter Drive
- Old Toongabbie, NSW 2146 Australia
System
System components Identification
Product Code and Description
- A-71404 – Advance Cervical Traction Bar
List of Accessories and Consumable Components Table
The following list are accessories and components that may be used with this device.
Name of Accessory | Product Number |
---|---|
Not Applicable | Not Applicable |
Name of Consumable | Product Number |
--- | --- |
Not Applicable | Not Applicable |
Indication for use
The Advance Cervical Traction Bar is used in a variety of surgical procedures including, but not limited to cervical surgery. These devices are capable of being used with a broad patient population as determined appropriate by the caregiver or institution.
Intended use
The Advance Cervical Traction Bar is designed to be used as a distractor in
a variety of surgical procedures including, but not limited to cervical
surgery. These devices are intended to be used by healthcare professionals
within the Operating Room setting.
Equipment Setup and Use
Prior to use
- a. Inspect the product looking for any visible damage or sharp edges that could be caused by a drop or impact during storage.
- b. Make sure product has been properly cleaned and disinfected and wiped dry prior to each use.
- c. Tools required: None
- d. Parts required: A-71404 (Advance Cervical Traction Bar)
- e. Pre-Use Checks table
Condition | Associated Indicator(s) |
---|
The Table power cord is connected to the Table and to an applicable power outlet.|
The Power indicator is on.
The On/Off switch is on.| The Battery Charge Level indicator is on.
The brakes are locked.
| Locked indicator is green ; Unlocked indicator is
off.
The Table is in Pendant mode.
| The Pendant Control indicator is green ; the Manual Flip indicator is off.
The columns are level (at the same height).
| Near the bottom on the outside of the columns there are numbers to help identify height level.
The support top is horizontal (not tilted).
| Tilt Level indicator is green ; the Tilt indicator is in the center of the Tilt Indicator Area.
Setup
The cervical traction bar gets installed on the support top-side of the head- end H-bracket (see figure System components Identification section 2.1).
- Determine the hole location for the cervical traction bar.
- Put the bar into the H-bracket openings.
- Make sure the drop locks on each end of the bar moves to the side to hold the bar in position.
- For use information, refer to the Allen®1 Advance Table User Manual (D-720686) and to the warnings on 4.1 General Safety Warnings and Cautions section 4.
Device controls and indicators
- Refer to the Allen®1 Advance Table User Manual (D-720686)
Storage, Handling and Removal Instructions
Storage and Handling
The product should be stored in a clean and safe environment to prevent
product damage. See storage Specifications under Product Specification
section.
Removal Instruction
Pull the bar from the H-bracket openings
Troubleshooting Guide
This device does not have a troubleshooting guide. For technical support
user of the device shall first contact Hill-Rom Technical Support.
Device Maintenance
Make sure that all labels are installed and can be read. Replace labels as
necessary by using a plastic scraper to remove the label. Use an alcohol wipe
to remove any adhesive residue.
Contact us if you need to repair or replace the device by using the
information from the contact details section (1.3).
Safety Precautions and General Information
General Safety Warnings and Cautions
WARNING
- a. Do not use if product shows visible damage.
- b. Prior to using this device, please read the instructions for equipment set up and use. Familiarize yourself with the product before application on a patient.
- c. To prevent patient and/or user injury and/or equipment damage, examine the device and surgical-table side rails for potential damage or wear prior to use. Do not use the device if damage is
- visible, if parts are missing or if it does not function as expected.
- d. The maximum diameter of the traction rope should be 0.25″ (6.35 mm). The use of a traction rope that is greater than 0.25″ (6.35 mm) in diameter could cause binding inside the cervical traction assembly. Patient injury could occur as a result of inaccurate traction.
- e. The maximum weight load for the traction assembly is 50 lb (22.6 kg). To exceed this weight could cause patient fall, injury, or equipment damage.
- f. Make sure the traction weight is in a safe position, free from possible unintended movement. Failure to do so could cause patient fall, injury, or equipment damage.
- g. Do not press the Cervical Traction Release button while traction weight is being applied. To do so could cause the assembly to retract and the traction weight to move. Patient fall, injury, or equipment damage could occur.
CAUTION: Do not exceed safe working load shown in the product specification table
Product Specifications
Mechanical Specifications | Description |
---|---|
Product Dimensions | The Cervical Traction Bar is 11.75” long |
(29.8 cm)
Material| Stainless Steel
Safe Working Load on the device| The maximum weight load for the traction
assembly is 50 lbs (22.6 kg)
Overall Weight of Complete Device| Approx. 1lbs (0.45 kg)
Storage Specifications| Description
Storage temperature| -29° C to +60° C
Storage Relative humidity range| 15% to 85%
---|---
Operating temperature| This device is intended to be used in a controlled
Operating Room environment.
Operating Relative humidity range
Electrical Specifications| Description
Not Applicable.| Not Applicable.
Software Specifications| Description
Not Applicable.| Not Applicable.
Compatibility Specifications| Description
The Advance Cervical Traction Bar is
compatible with:
| Allen® Advance Table (A-71100) through the
Advance H-Brackets (A-71402)
Note: Consult the corresponding IFU for the products mentioned in the above table.
Sterilization Instruction:
This device is not intended to be sterilized. Equipment damage may occur.
Cleaning and Disinfection Instruction
WARNING:
- Do not use bleach or products that contain bleach to clean the device. Injury or equipment damage can occur.
- After each use, clean the device with alcohol-based wipes.
- Do not put the device into water. Equipment damage can occur.
- Use a cloth and a quaternary ammonium disinfecting/cleaning solution to clean and disinfect the device.
- Read and follow the manufacturer’s recommendation for low-level disinfection.
- Read and follow the cleaning product’s instructions. Use caution in areas where liquid can get into the mechanism.
- Wipe the device with a clean, dry cloth.
- Make sure that the device is dry before you store it or use it again.
List of Applicable Standards
Sl. no | Standards | Description |
---|
| ****
EN 62366-1
| Medical devices – Part 1: Application of usability engineering to medical
devices
Sl. no| Standards| Description
---|---|---
| ****
EN ISO 14971
| Medical devices- Application of risk management to medical devices.
| ****
EN 1041
| Information supplied by the manufacturer of medical devices
| ****
EN ISO 15223-1
| Medical devices – Symbols to be used with medical device labels, labelling and information to be supplied – Part 1: General requirements
| ****
EN ISO 10993-1
| Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
| ****
IEC 60601-2-46
| Medical electrical equipment – Part 2-46: Particular requirements for the basic safety and essential performance of operating tables
| ****
ISTA
| International Safe Transit Association standards for package testing
Hillrom Technical Support
hillrom.com/en-us/about-us/locations/
Allen Medical Systems, Inc. is a subsidiary of Hill-Rom, Inc.
References
- Advancing Connected Care | Hillrom
- allenmedical.com - allenmedical Resources and Information.
- allenmedical.com - allenmedical Resources and Information.
- Patents | Hillrom
- Patents | Hillrom
- Hillrom Locations by Country | Hillrom
- allenmedical.com - allenmedical Resources and Information.
- Patents | Hillrom
- Hillrom Locations by Country | Hillrom
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>