BRESSER 7060200 Tendence Weather Station Instruction Manual

June 9, 2024
BRESSER

BRESSER 7060200 Tendence Weather Station Instruction Manual
BRESSER 7060200 Tendence Weather Station

Imprint

For any warranty claims or service inquiries, please refer to the information on “Warranty” and “Service” in this documentation. We ask for your understanding that unsolicited returns cannot be processed. Errors and technical changes excepted. © 2022 Bresser GmbH All rights reserved. The reproduction of this documentation – even in extracts – in any form (e.g. photocopy, print, etc.) as well as the use and distribution by means of electronic systems (e.g. image file, website, etc.) without the prior written permission of the manufacturer is prohibited. The designations and brand names of the respective companies used in this documentation are generally protected by trade, trademark and/or patent law in Germany, the European Union and/or other countries.

Validity note

This documentation is valid for the products with the following article numbers: 7060200
Manual version: 0922
Manual designation:
Manual_7060200_Tendence-FSX_en-de-fr-esnl_BRESSER_v092022a
Always provide information when requesting service.

About this Instruction Manual

NOTICE
These operating instructions are to be considered a component of the device. Read the safety instructions and the instruction manual carefully before using this device. Keep these instruction manual in a safe place for future reference. If the device is sold or passed on, the instruction manual must be passed on to any subsequent owner/user of the product.

Parts overview and scope of delivery

OVERVIEW

OVERVIEW

Illustration 1: Parts overview for base station (top) and remote sensor (bottom)

  1. Display
  2. Housing
  3. SNOOZE/LIGHT button (snooze function and temporary background lighting)
  4. HISTORY button (retrieve measurements for the past 24 hours)
  5. CH/UP button (sensor channel selection or value change upwards)
  6. 12/24/DOWN button (time mode selection or value change downwards)
  7. MAX/MIN button (switch between highest, lowest or current value display)
  8. BARO button (display of different atmospheric pressure values)
  9. TIME button (manual time setting)
  10. ALARM button (Alarm setting)
  11. Wall mount fixture
  12. °C/°F button (temperature format setting)
  13. RESET button (reset all settings)
  14. Stand, fold-out
  15. Battery compartment cover
  16. Battery compartment
  17. Display
  18. Function indicator (data transmission)
  19. Housing
  20. Battery compartment
  21. RESET button (reset all settings
  22. Wall mount fixture
  23. Channel switch
  24. Battery compartment cover

Scope of delivery
Base station (A), remote sensor (B) Also required (not included): 4 pcs. Mignon batteries (1.5V, AA type)

Screen display

Screen display

Illustration 2: Display of the base unit

  1. AM/PM information in 12- hour time mode
  2. Current time (hours:minutes:seconds)
  3. Alarm symbol (alarm 1 or 2 enabled)
  4. Symbol for active daylight saving time (DST)
  5. Date (month-day or reverse)
  6. Weekday
  7. Transmission symbol (radio-controlled clock CET)
  8. Alarm (AL1 or AL2)
  9. Air pressure (mb/hPa or inHg)
  10. Moon phase
  11. Bar chart for air pressure history
  12. Weather trend (72 hours)
  13. Humidity outdoors
  14. Humidity indoors
  15. Battery status
  16. Temperature (indoor)
  17. Temperature (outdoor)
  18. Channel info (sensors)
  19. Sensor signal status
  20. Weather trend (48 hours)
  21. Weather trend (24 hours)
  22. Singles values for air pressure history (up to 24 hours)

Before commissioning

NOTICE Avoid connection faults!
In order to avoid connection problems between the devices, the following points must be observed during commissioning.

  1. Place the base unit (receiver) and sensor (transmitter) as close together as possible.
  2. Connect the power supply to the base unit and wait until the indoor temperature is displayed.
  3. Establish power supply for the sensor.
  4. Set up/operate the base unit and sensor within the effective transmission range.
  5. Make sure that the base unit and the radio sensor are set to the same channel.

When changing the batteries, always remove the batteries from both the base unit and the sensor and reinsert them in the correct order so that the radio link can be re-established. If one of the two devices is operated via a mains power connection, the power connection must also be briefly disconnected for this device when the batteries are changed. If, for example, only the batteries in the sensor are replaced, the signal may subsequently not be received at all or not be received correctly.

Note that the actual range depends on the respective construction materials used for the buildings as well as the respective position of the base unit and the outdoor sensor. External influences (various radio transmitters and other sources of interference) can greatly reduce the possible range. In such cases, we recommend finding other locations for both the base unit and the outdoor sensor. Sometimes moving the sensor by just a few centimeters is enough! Though the remote unit is weather proof, it should be placed away from direct sunlight, rain or snow.

Setting up power supply

Base unit

  1. Remove the battery compartment cover.

  2. Insert batteries into the battery compartment. Make sure that the batteries are correctly aligned (+/- poles).

  3. Replace the battery compartment cover.

  4. Wait until indoor temperature is displayed on the base unit.
    Wireless sensor

  5. Remove the battery compartment cover.

  6. Move the CH slider to the position for the desired transmission channel (setting CH1, CH2 or CH3 with indication on the display).

  7. Insert batteries into the battery compartment. Make sure that the batteries are correctly aligned (+/- poles).

  8. Replace the battery compartment cover.

  9. Press the SET button for approx. 2 seconds.

NOTICE! When operating only one wireless sensor, channel 1 is recommended as the default setting.

Battery level indicator

  1. When the level of the batteries in the base station or in the wireless sensor reaches a critical level, the battery level symbol appears in the appropriate area on the display.
  2. When replacing one set of batteries, always remove the batteries from the other part of the unit and reinsert them in the correct order (see chapter “Setting up power supply”). Replace the batteries to be changed in the corresponding part of the device with a completely new set with full capacity. This ensures that the connection between the devices will be reestablished again correctly.

Automatic time setting

After the power supply was established, the clock will automatically search for the radio signal. This will take approximately 3 to 8 minutes to finish this process. If the radio signal is received correctly, the date and time will be set automatically and the radio control signal icon turns on.
If the clock fails to receive the time signal, go ahead with the following steps:

  1. Press °C/°F button on the base station for approx. 3 seconds to initate RC signal reception again.
  2. If the device is still not receiving the signal, the time must be set manually.

Manual time setting and other user defined settings

  1. Press the SNOOZE/LIGHT button for about 8 seconds to deactivate the automatic time setting. The reception symbol for the radio-controlled watch disappears from the display.
  2. In normal display mode, press the TIME button for about 2 seconds to enter the settings mode.
  3. Digits to be set are flashing.
  4. Press CH/UP or 12/24/DOWN button to change the value.
  5. Press the TIME button to confirm and switch to the next setting.
  6. Sequence of the settings: Year > Change day/month display > Month > Day > Daylight saving time (DST) AUTO/OFF (off) > Hours > Minutes > Seconds > Language > Time zone (-23 to +23 hours)
  7. Finally press the TIME button to save the settings and exit the settings mode.
  8. NOTICE! To restore the automatic time setting, press the SNOOZE/LIGHT button again for 8 seconds. The reception symbol for the radio-controlled watch reappears in the display.

Alarm settings

  1. In normal display mode, press the TIME button several times to display the Alarm time AL1 or AL2.

  2. Press ALARM button for approx. 3 seconds to enter the alarm time setting mode.

  3. Digits to be set are flashing.

  4. Press CH/UP or 12/24/DOWN button to change the value.

  5. Press ALARM button to confirm and continue to the next setting.

  6. Settings order: hours > minutes > ice alert on/off
    NOTICE! If ice alert is enabled (on), the alarm will sound 30 minutes before the set alarm time if the temperature is -3° or below.

  7. Finally press the ALARM button to save the settings and exit the setting mode. Alarm will be activated automatically. The symbol will be displayed.

  8. In normal display mode, press the ALARM button several times to enable alarm time AL1, AL2 or both. If the alarm is enabled, the corresponding symbol will be displayed ( 1, 2 or 1 2).

Snooze function

  1. When the alarm sounds press the SNOOZE/LIGHT button to  activate the snooze function. The alarm willsound again in 5 minutes.
  2. Press ALARM button when the alarm sounds to interrupt the alarm until the alarm time will be reached again.
  3. The alarm will be turned off automatically if no button is pressed within 2 minutes.

Receiving measurements automatically

Once the power supply is enabled, the base station will display the measurement readings for indoors. Readings from the outdoor sensor will be displayed within 3 minutes after powering it on.
If no signal is received, proceed as follows: Press 12/24 button for approx. 3 seconds to initate reception of measurements again.

Moon phases

This weather station is able to display the moon phases for the northern hemisphere. Here the moon grows from the right. This is because the sunlit side of the moon in the northern hemisphere moves from right to left. The following table shows the representations of the moon phases.

Illustration 3: Moon phases for the northern hemispere.

Weather Trend

The weather station will calculate a weather trend for the next 12 hours on basis of the measured values.

Weather Trend

Illustration 4: Weather trend indicators

  1. Sunny
  2. Partly cloudy
  3. Cloudy
  4. Rain
  5. Storm
  6. Snow

Trend arrow indicators

Trend arrow indicators

  1. Rising
  2. Steady
  3. Falling

The temperature and humidity trend indicator shows the trends of changes in the forthcoming few minutes. Arrows indicate a rising, steady or falling trend.

Barometric / Atmospheric Pressure

Atmospheric Pressure is the pressure at any location on earth, caused by the weight of the column of air above it. One atmospheric pressure refers to the average pressure and gradually decreases as altitude increases. Meteorologists use barometers to measure atmospheric pressure.

Since variation in atmospheric pressure is greatly affected by weather, it is possible to forecast the weather by measuring the changes in pressure.

  1. Press the BARO button to switch between barometric pressure display in inHg or in hPa.

  2. Press the BARO button for 3 seconds to change between absolute and relative atmospheric pressure.

    • ABSOLUTE: the absolute atmospheric pressure of your location.
    • RELATIVE: the relative atmospheric pressure based on the sea level.
      Set relative atmospheric pressure value
  3. Get the atmospheric pressure data of the sea level (it is also the relative atmospheric pressure data of your home area) through the local weather service, internet and other channels.

  4. Press and hold the BARO button for approx. 3 seconds until “abs” or “rel” flashes.

  5. Press CH/UP or 12/24/DOWN button to switch to “rel” mode.

  6. Press the BARO button and the number for “rel” flashes.

  7. Press CH/UP or 12/24/DOWN button to change the value.

  8. Press the BARO button to save and exit the setting mode.

  9. The default relative atmospheric pressure value is 1013 mb/hPa (29.91 inHg), which refers to the average atmospheric pressure.

  10. When you change the relative atmospheric pressure value, the weather indicators will change along with it.

  11. The built-in barometer can notice the environmental absolute atmospheric pressure changes. Based on the collected data a forecast for the weather conditions in the next 12 hours can be made. Therefore, the weather indicators will change according to the detected absolute atmospheric pressure after you operate the clock for 1 hour.

  12. The relative atmospheric pressure is based on the sea level, but it will change with the absolute atmospheric pressure changes after operating the clock for 1 hour.

Connecting remote sensors

The Weather Station can display the readings from up to 3 wireless sensors* of the same type. Each radio sensor must be set to a separate channel. Proceed as follows to set the channel:

  1. Remove the battery compartment cover of the wireless sensor.
  2. Set the channel selection switch to the desired channel (CH1, CH2 or CH3).
  3. Re-attach the battery compartment cover.
  4. NOTICE! Each connected wireless sensor must be set to a different channel. If only one wireless sensor is connected, it should be set to CH1.
  5. Press the CH/UP button on the base station several times to display the measured values for the individual channels. The selected channel is shown in the display
    • one wireless sensor included, others optionally available

Temperature display

Press the °C/°F button to switch between the temperature display in °C or °F. When temperatures of -40°C or below and humidity of 20% or below are reached, the information “LO” is output for the respective range. At temperatures of 70°C or higher and a humidity of 90% or higher, the information “HI” is displayed. For values outside the measurable range “- -” is displayed. When returning to a measurable temperature range, the corresponding temperature is displayed again.

History record for the past 24 hours

The base station automatically records air pressure readings from the last 24 hours. If necessary, press the HISTORY button several times to display the history data for the hourly values (HOUR – up to 24 hours backwards) for the pressure one after the other. When a history date is displayed, press any key (except HISTORY) to return to the normal display mode.

NOTICE! In the history bar graph the values for the pressure of the last 24 hours can be read at any time in compressed form.

MAX/MIN Weather data

The main unit saves highest and lowest value records for indoor and outdoor temperature as well as for humidity for 24 hours:

  1. Press the MAX/MIN button repeatedly to display the stored values of the base station and the currently set remote sensor one after another.
  2. Display order: Highest values > Lowest values > Current values
  3. Press MAX/MIN button for approx. 3 seconds to delete all values of the current recording period.
  4. NOTICE! When the batteries are changed, all values of the current recording period will also be deleted.

 Disposal

Dispose of the packaging materials according to its type. Information on proper disposal can be obtained from the municipal waste disposal service provider or environmental agency. Do not dispose of electronic devices in the household garbage! According to the European Directive 2012/19/EU on Waste Electrical and Electronic Equipment and its transposition into national law, used electrical equipment must be collected separately and recycled in an environmentally sound manner. Batteries and rechargeable batteries must not be disposed of with household waste. You are legally obliged to return used batteries and accumulators and can return them after use either at our sales outlet or in the immediate vicinity (e.g. in the trade or in municipal collection points) free of charge. Batteries and accumulators are marked with a crossed-out dustbin and the chemical symbol of the pollutant, “Cd” stands for cadmium, “Hg” stands for mercury and “Pb” stands for lead.

Technical data

Base station

Batteries 2x AA, 1.5 V
Radio controlled signal DCF
Maximum number of sensors 3
Temperature unit °C/°F
Humidity measuring range 20% to 90%
Humidity resolution 1%
Temperature measuring range -5°C to 50°C (23°F to 122°F)
Barometric pressure unit 540 hPa to 1100 hPa / 15.95 to 32.49 inHg
Time format 12 or 24 hours
Dimensions (WxHxD) 100 x 161 x 21.5 mm

Remote sensor

Batteries 2x AA, 1.5 V
Transmission frequency 433 MHz
Transmission measuring range 30 m
Temperature unit °C
Temperature measuring range -20°C to 60°C (-4°F to 140°F)
Humidity measuring range 1% to 90%
Humidity resolution 1%
Dimensions (W x H x D) 65 x 100 x 35 mm

Warranty

The regular warranty period is 2 years and starts on the day of purchase. To benefit from an extended voluntary warranty period as indicated on the gift box, registration on our website is required. You can consult the full guarantee terms as well as information on extending the guarantee period and details of our services at www.bresser.de/warranty_terms.

EC declaration of conformity

Bresser GmbH hereby declares that the radio equipment type with item number 7060200 is in compliance with Directive 2014/53/EU. The full text of the EC Declaration of Conformity is available at the following web address: www.bresser.de/download/7060200/CE/7060200_CE.pdf

UKCA declaration of conformity

A “Declaration of Conformity” in accordance with the applicable directives and relevant standards has been issued by Bresser GmbH The full text of the UKCA Declaration of Conformity is available at the following web address:
www.bresser.de/download/7060200/UKCA/7060200_UKCA.pdf
Bresser UK Ltd. • Suite 3G, Eden House, Enterprise Way, Edenbridge, Kent TN8 6HF, Great Britain

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

BRESSER User Manuals

Related Manuals