neno bene Portable Mesh Nebulizer User Manual

June 8, 2024
neno

bene Portable Mesh Nebulizer
User Manual
neno bene Portable Mesh Nebulizer

bene Portable Mesh Nebulizer

Dear Customer,
thank you for buying our product, the Neno Bene mesh nebulizer. Please read the user manual before use.
DESCRIPTION
Method of operaon: Thanks to the self-excing oscillaon circuit, the ceramic walls together with a metal mesh and a built-in nebulizing head guarantee successful oscillaon under a high frequency. The device structure ensures smooth flow of the medicine soluon into the micromesh of the metal mesh and nebulizaon into an aerosol form, which flows into paent’s respiratory system through the inhalaon mask to successfully inhale medicine.
Symbols:

| Please read the informaon on the signs listed in the table to avoid user mistakes and to become familiar with the safety measures. Manufacturer
---|---
| The device should be ulized in designated municipal waste areas
| Class 2
| User manual
| BF type parts
| Serial number given by the manufacturer

Accessories
Before installaon and configuraon of the device, check if the device is in good shape and if the quality and quanty of the accessories matches the quanty listed in the user manual. Check, if there are any device defects present, if so, do not use the device, but contact the manufacturer or seller instead.neno bene Portable Mesh Nebulizer - fig

1. Medicine Cup 5. Main Unit Power Buon
2. Mask 6. Main Unit Connecon Cable
3. String 7. Type-C Adapter Head
4. Main  Unit

SPECIFICATION

Power : DC 5V 1A
Aerosol efficiency: ≥0.20 ml/min
Parcle size: MMAD approx. 5um
Sound level: ≤35bB
Ultrasonic frequency: Approx. 114kHz
Medicine cup capacity: 8ml
Medicine cup
temperature range: ≤60°C
Medicine cup weight: Approx. 11g
Dimensions: 42mm x 34mm x 40mm
Safety measures : Do not use the device near flammable, intoxicang gases mixed

with air,
or flammable, intoxicang gasses mixed with oxygen or nitric oxide
Applicaon method:| Medicine in nebulized form
 Contraindicaons:| None
Electrical safety|
requirements:| Internal power mechanisms, type BF applicaon parts
Work mode:| Connuous
Work environment:| Temperature: +10~+40°C
Humidity: 30~85%RH
Pressure: 86~106 kPa
Storage and|
transport condions:| Temperature: -20~+50 C
Humidity: 30~85%RH
Pressure: 86~106 kPa

SAFETY

WARNING:

  • When using a certain kind, dose, and paern of medicine intake, make sure to adhere to doctor’s recommendaons. If failed to adhere to the doctor’s recommendaons, your symptoms may worsen.
  • Clean and disinfect the medicine cup and inhalaon mask aer every use.
  • Clean and disinfect device parts and then carefully dry them, and store in a dry, clean place to prevent bacterial growth.
  • Nebulizer is designed for individual use. It is not recommended to use the device by more than one person.
  • Keep the device out of reach of children and infants to avoid choking hazard. If swallowed, consult your doctor immediately.
  • Do not wash or immerse the device under water, especially hot. Risk of electrical installaon damage, short circuit, or electrocuon. Direct water contact can make the device useless.
  • Neno Bene is a professional nebulizer. Adults and children who are unable to use the device on their own should use the device under a guardian supervision.

ATTENTION:

  • Do not inhale clear water from the medicine cup. Your symptoms may worsen.
  • Do not drop medicine on the main unit. If you drop the medicine, immediately wipe it with a gauze.
  • Using the main unit when wet can cause problems or even electrical shock.
  • Do not pierce or pick the metal mesh with a q-p or a hair pin, it can cause damage to the set and make it unusable.
  • Do not drop or press forcefully on the medicine cup. It can cause damage or even electrical shock.
  • To ensure proper use of the device, read the manual carefully and adhere to the instrucons.
  • If you have any quesons, consult the supplier or the manufacturer.
  • If the storage and transport temperature is lower than 5°C, the device should return to room temperature before use.
  • Please read the package leaflet before use to ensure that the medicinal product is suitable for the mesh nebulizer.

USE GUIDE

A. CONNECT THE POWER CABLE:

  • The device uses an external power system (5V 1A). The main unit has an USB input and a power system (5V 1A). Insert the main unit USB plug into the power input. If you have a type C adapter head purchased, you can use it with the main unit.

The main unit uses an external power supply with baery case. Aenon! Rechargeable baeries cannot be used in the baery case!

  1. Open the baery cover

  2. Insert baeries correctly by aligning the polaries (+ and -) with baery indicaon mark.

  3. Turn on the baery case using the switch, and the indicator on the case will light up.

  4. Connect the main unit plug with the baery case and make sure that it is placed ghtly.

B. BATTERY LIFE AND REPLACEMENT:

  1. When using alkaline baeries:
  • The unit can be used for approximately 3 days (if used 30 minutes a day)
  • When the indicator light changes to red, the baeries are running low. Replace the baeries.
  • If you are planning not to use the device for a long me (3 months or more) take the baeries out.
  • If you leave the baeries inside the device, there is a risk of baery leakage and device damage.
  • Do not insert the baeries inside the device using the wrong polaries.

C. FILL THE MEDICINE CUP :
Open the medicine cup, fill it with medicine and close the lid ghtly. Before using high density or strong medicine consult your doctor. D. CONNECT THE MEDICINE CUP:
Insert the other end of the power cable into the port in the medicine cup, and make sure that it is mounted properly. Do not abruptly shake the medicine cup when filled with liquid. E. MOUNT THE INHALATION MASK: F. PUT THE MASK ON:
G. START THE INHALATION:

  • If the baery case is used to power the device, make sure that it is turned on and that the indicator light is on. If an internal power supply is used, make sure, that the power supply is turned on and that it is compable with the output voltage 5V 1A. Press the power buon to start the inhalaon.

  • Take a deep breath to properly inhale the medicine.

  • When using high density medicine, the nebulizaon can take longer than usual.

  • The sound coming out of metal sheet during the nebulizaon process is a normal phenomenon.

  • If the liquid has high surface acvity, a small quanty of the mixture can flow out of the metal sheet.

  • If the medicine cup runs out of liquid, press the button on the device to avoid potenal damage of the metal mesh.

  • If the medicine cup is running out of medicine, it is recommended to slightly lt t he cup, so the liquid will flow towards the metal mesh.

  • During mask inhalaon, do not cover the air vents of the mask with your hands or other objects.

  • During aerosol inhalaon, do not cover the air vents on the lid with your hands or other objects to not influence the aerosol stream.

  • If the inhalaon is causing you discomfort, immediately turn off the device and consult with your doctor.

H. STOP THE INHALATION
Press the power buon on the device to stop inhalaon. If the power source is the baery case, press the power buon to turn off the baery case. Wait unl the indicator lamp will turn off and disconnect the main unit from the baery case. If the device is powered by an external power source, turn it off, and unplug the main unit from the external power source.neno bene Portable Mesh
Nebulizer - fig 9 CLEANING AND DISINFECTION
CLEANING THE MASK AND MEDICINE CUP

  1. Take out the inhalaon mask and the medicine cup out of the device.
  2. Open the medicine cup and pour out the contents.
  3. Add a small amount (2~5ml) of water into the medicine cup and cover it with the lid. Shake the cup lightly so that the medicine dissolves fully in the water. Open the medicine cup and pour out the contents.
  4. Wash the mask with water.
  5. Aer washing, dry the mask thoroughly.
  6. Store all the nebulizer parts in a dry, clean place.

neno bene Portable Mesh Nebulizer - fig 10 DISINFECTING THE MASK AND MEDICINE CUP:

  1. Immerse the medicine cup in medical alcohol (75%) for no less than 3 minutes, then take the cup out of the alcohol and dry it thoroughly. Store the cup in a clean place.
  2. Immerse the mask in medical alcohol (75%) for no less than 3 minutes. Aer cleaning, dry the mask thoroughly and store it in a clean place.
  3. Aer disinfecng, rinse the device with water to avoid accumulaon of medical alcohol residue. Make sure to disinfect the medicine cup aer each use. In case of permanent staining change the device parts.

MASK AND MEDICINE CUP REPLACEMENT:
Medicine cup should be replaced aer around 20 hours of use. In any other case your health and safety, funconal capacity and safe inhalaon will not be guaranteed. If the medicine cup is damaged or is working improperly, it must be replaced.

 MAINTENANCE

  • Check the device before every use. Do not use the nebulizer if it is damaged. Using the device when damaged can cause harm or danger.
  • Store the device in a cold, dry and venlated room. Keep away from fire and hot surfaces to avoid risk of baery explosion.
  • If you have any problems with the device, like funcon sengs, maintenance or using the device, contact the manufacturer. Do not deconstruct or fix the device on your own.
  • In case of any occurring problems, contact the manufacturer.

TROUBLESHOOTING

Problem Possible cause How to fix
Nebulization power is too weak Membrane is dirty Clean the medicine

container. If the problem can not be fixed by the method above, replace the container with a new one.
The device is not nebulizing| The main unit is not connected properly| Unplug the main unit from the power source and
connect it back again.
The main unit light is on, but the device is not working| The main unit cable is not working properly| Replace the main unit with a new one.
Device membrane is very dirty| Clean the medicine cup according to the user manual. If there are problems after cleaning, replace the medicine cup with a new one.
Fluid leak| Medicine cup is damaged| Replace the medicine cup with a new one.

DISPOSAL
DO NOT throw out the baeries with municipal waste. Throw out the baeries according to the local law on waste ulizaon (for example designated waste collecon points).
The device is made from plasc and stainless steel. Make sure to ulize materials according to the binding waste ulizaon law.
ELECTROMAGNETIC COMPATIBILITY
Guidance and manufacturer’s declaraon

  1. Device parts should be installed and aligned according to the informaon included in the ATTACHED DOCUMENTS.
  2. IMMUNITY LEVEL TESTS for elementary rules of working of medical equipment and their soware should be picked based on a big probability of maintaining security and overall producvity, and according to recommendaons of professional health facilies, house environment and other special designated places of use.
  3. HOUSE ENVIRONMENT is a place in which the paent is living, or a place in which the paent is residing, except professional care facilies with trained medical professionals.

FOR EXAMPLE: As indicated in table 6 IEC 60601-1-2: 2014 for medical equipment, a standard cellphone with a maximum output voltage of 2W stops at d=3.3m with
an immunity level of 3V/m.
A1. Electromagnec emissions for all devices and systems
Guidance and manufacturer’s declaraon – Electromagnec emissions
The device is designed to use in electromagnec environments listed belowpaent using the nebulizer should make sure that they are in the correct environment.

Emission test| Realized standard| Electromagnetic environment – tips
---|---|---
RF CISPR 11 Emissions| Group 1| Nebulizer uses RF energy only for internal functions. Its RF emissions are very low and do not disrupt other devices from working.
RF CISPR 11 Emissions| Class B| The device is optimal to use at any place different than home and public environments, with low power or a mains power supply built in a domestic house grid.
Harmonic emissions IEC 61000-3-2| Class A
Harmonic emissions IEC 61000-3-2| Compatibility

A2. Electromagnec immunity – household environments. ACCESSORIES AND SYSTEMS
Guidance and manufacturer’s declaraon – Electromagnec immunity
The device is designed to use in an electromagnec environment listed below; paent using the nebulizer should make sure that they are in the correct environment.

Immunity test| Measured level IEC 60601| Compatibility level| Electromagnetic environment – tips
---|---|---|---
Electrostatic emissions (ESD)
IEC 61000-4-2| ±8 kV contact
±2 kV, ±4 kV,
±8 kV, ±15 kV air| ±8 kV contact
±2 kV, ±4 kV,
±8 kV, ±15 kV air| The floors should be wooden, concrete of ceramic. If the floor is covered with a synthetic material, the relative humidity should be at lest 30%
Electrical fast transient/burst 61000 44

| ±2kV for power lines| ±2kV for power lines| Power quality network it should be like thissame as typical environment hospital or
commercial.
Surge IEC 61000-4-5| ±1kV line to line ±2kV line
to ground| ±1kV line to line ±2kV line to ground| Mains power quality should be the same as in a standard hospital or commercial environment.
Power frequency (50/60Hz) magnetic field IEC 61000-4-8| 30A/rn| 30A/m| Power frequency of magnetic fields should be at levels characteristic for typical hospital and commercial environments.
Voltage dips, short interruptions, and voltage variations on power supply
input lines IEC 61000-4-11| 0% UT; 0,5 cycle at 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°| 0% UT; 0,5 cycle at 0°, 45°, 900, 135°, 180°, 225°, 270° and 315°| Quality of the mains power supply should be typical for a hospital and commercial
environments. If the user requires continued operation during power
outages, it is recommended to power the device from an uninterrupted power supply.
---|---|---|---
Conducted RF IEC 61000-4-6| 3 Vrms 1501<z to 80MHz| 3 Vrms| The manufacturer should consider
reducing the
minimum separation distance based on risk mangnagement. and usi higher
immunity test levels, that are appropriate for the reduced inimum separation distance. Minimum separation distances for higher immunity test levels shall be calculated using the following equation:
Where “P” is the maximum power in watts. “d” is the minimum separation distance in meters and “E” is the mmunity test level in V/m
Radiated RF IEC 61000-4-3| 10 V/m 80 MHz to 2,7 GHz| 10V/m

ATTENTION: Check the mains power supply before using the device
WARRANTY
The product is covered by a 24-month warranty. The terms of the guarantee can be found at: https://neno.pl/gwarancja
Complaints should be reported using the complaint form located at: https://neno.pl/kontakt
Specificaons and contents of the kit are subject to change without noce. We apologise for any inconvenience.
KGK Trade declares that this device Neno Bene is compliant with significant requirements of direcve 2014/53/EU. The text of this declaraon may be found on the website: https://neno.pl/download/DOC/deklaracja-CE-Neno-Bene.pdf

neno bene Portable Mesh Nebulizer - icon 4Last update of the user’s manual: 01.02.2021
Importer:
KGK Trade sp. z o.o. sp. k.
Ujastek 5b
31-752 Cracow
Poland
Made in P.R.C.

The crossed out trash can symbol indicates that unusable electrical or electronic devices, its accessories (such as power supplies, cords) or components (for example batteries, if included) cannot be disposed of alongside with household waste. In order to dispose of the devices or its components (for example, batteries) deliver the device to the collection point, where it will be accepted free of charge. Disposal is subject to the recast version of the WEEE Directive (2012/19 / EU) and the Directive on batteries and accumulators (2006/66 / EC). Proper disposal of the device prevents degradation of the natural environment. Information about the collection points of the facilities is issued by the competent local authorities. Incorrect disposal of waste is subject to penalties provided for by the law in force in the given area.

Manufacturer’s data:
Vapo Healthcare Co., Ltd.
Southern unit of third floor, building B, No.99 Yudai West Rd, High tech district, KunShan,
Suzhou, Jiangsu Province, China.
EU Authorized Representae:
Share Info GmbH
Heerdter Lohweg 83, 40549 Düsseldorfwww.neno.pl
inspired by Children, designed by Parents

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals