TSI AM520i SidePak Personal Aerosol Monitor User Guide
- June 9, 2024
- tsi
Table of Contents
TSI AM520i SidePak Personal Aerosol Monitor User Guide
-
Install the battery and secure with screws.
-
Charge the battery (see User Guide for battery charging warnings)
-
Select Impactor and sintered disk (if needed).
-
Power on.
-
If using impactors . . . verify 1.7 L/min flow rate with flow calibrator (sold separately).
-
Zero Cal the instrument
(Menu | Calibration | Zero Cal . . . and follow on-screen instruction.) -
Select Calibration Factor (if needed) (Menu | Calibration | User Cal).
Factory default = 1.0,
Ambient Aerosol = 0.38, or
Custom calibration factor. -
Select Run Mode
(Menu | Survey – Manual – Program).
NOTE: Survey Mode does not log data. Programs 1 to 5 are set up in TrakPro™ Data Analysis Software on a computer and imported into the AM520/AM520i. -
Download logged data to TrakPro™ v5 software (sold separately). AM520/AM520i must be powered “on” to communicate and download data.