TSI AM520i SidePak Personal Aerosol Monitor User Guide

June 9, 2024
tsi

TSI AM520i SidePak Personal Aerosol Monitor User Guide
TSI AM520i SidePak Personal Aerosol Monitor

  1. Install the battery and secure with screws.

  2. Charge the battery (see User Guide for battery charging warnings)
    Instructions

  3. Select Impactor and sintered disk (if needed).

  4. Power on.
    Instructions

  5. If using impactors . . . verify 1.7 L/min flow rate with flow calibrator (sold separately).
    Instructions
    Display Instructions

  6. Zero Cal the instrument
    (Menu | Calibration | Zero Cal . . . and follow on-screen instruction.)

  7. Select Calibration Factor (if needed) (Menu | Calibration | User Cal).
    Factory default = 1.0,
    Ambient Aerosol = 0.38, or
    Custom calibration factor.

  8. Select Run Mode
    (Menu | Survey – Manual – Program).
    NOTE: Survey Mode does not log data. Programs 1 to 5 are set up in TrakPro™ Data Analysis Software on a computer and imported into the AM520/AM520i.

  9. Download logged data to TrakPro™ v5 software (sold separately). AM520/AM520i must be powered “on” to communicate and download data.
    Instructions

TSI and TSI logo are registered trademarks of TSI Incorporated in the United States and may be protected under other country’s trademark registrations.
SidePak and TrakPro are trademarks of TSI Incorporated.

Support

TSI Incorporated – Visit our websitewww.tsi.com for more information.
USA Tel: +1 800 680 1220
UK Tel: +44 149 4 459200
France Tel: +33 1 41 19 21 99
Germany Tel: +49 241 523030
India Tel: +91 80 67877200
China Tel: +86 10 8219 7688
Singapore Tel: +65 6595 6388
P/N 6009830 Rev. B ©2022 TSI Incorporated

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

tsi User Manuals

Related Manuals