GW INSTEK GPT-9600 Series Electrical Safety Tester User Manual

June 3, 2024
GW INSTEK

GPT-9600 Series Electrical Safety Tester

Electrical Safety Tester
GPT-9600 Series
USER MANUAL
GW INSTEK PART NO. 82PT-96030EB1
ISO-9001 CERTIFIED MANUFACTURER

This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.

Table of Contents
Table of Contents
SAFETY INSTRUCTIONS ………………………………………….. 4
GETTING STARTED …………………………………………………. 8 GPT-9600 Series Overview ………………… 9 Appearance …………………………………… 12 Set Up………………………………………….. 18
OPERATION ………………………………………………………… 24 Menu Tree…………………………………….. 25 Test Lead Connection ……………………… 29 ACW, DCW and IR Testing ……………….. 33 Common Utility Settings …………………. 55
EXTERNAL CONTROL ……………………………………………. 59 External Control Overview……………….. 60
FAQ ……………………………………………………………………. 66
APPENDIX …………………………………………………………… 68 Fuse Replacement ………………………….. 68 Error Messages ……………………………… 69 GPT-9600 Specifications …………………. 70 GPT-9600 Dimensions ……………………. 73 Declaration of Conformity ……………….. 74
INDEX ………………………………………………………………… 75
3

SAFETY INSTRUCTIONS

SAFETY INSTRUCTIONS

This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to ensure your safety and to keep the instrument in the best possible condition.

Safety Symbols
These safety symbols may appear in this manual or on the instrument.

Warning: Identifies conditions or practices that WARNING could result in injury or loss of life.

CAUTION

Caution: Identifies conditions or practices that could result in damage to the instrument or to other properties.

DANGER High Voltage

Attention Refer to the Manual

Protective Conductor Terminal

Frame or Chassis Terminal

Earth (ground) Terminal

4

SAFETY INSTRUCTIONS

Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased.

Safety Guidelines

General Guideline
CAUTION
Power Supply WARNING

Do not place any heavy object on the instrument.
Avoid severe impact or rough handling that leads to damaging the instrument.
Do not discharge static electricity to the instrument.
Use only mating connectors, not bare wires, for the terminals.
Do not block the cooling fan opening.
Do not disassemble the GPT-9600 unless you are qualified.
(Measurement categories) EN 61010-1:2010 specifies the measurement categories and their requirements as follows. The GPT-9600 does not fall under category II, III or IV.
Measurement category IV is for measurement performed at the source of low- voltage installation.
Measurement category III is for measurement performed in the building installation.
Measurement category II is for measurement performed on the circuits directly connected to the low voltage installation.
AC Input voltage range: 100-120/220-240VAC ±10%
Frequency: 50Hz/60Hz
Fuse: T 4A 250V
To avoid electrical shock connect the protective grounding conductor of the AC power cord to an earth ground.

5

GPT-9600 Series User Manual

Cleaning the GPT-9600 Operation Environment
Storage environment Disposal

Disconnect the power cord before cleaning.
Use a soft cloth dampened in a solution of mild detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material such as benzene, toluene, xylene, and acetone.
Location: Indoor, no direct sunlight, dust free, almost non-conductive pollution (Note below)
Relative Humidity: 70% (no condensation)
Altitude: < 2000m
Temperature: 0°C~40°C
(Pollution Degree) EN 61010-1:2010 specifies the pollution degrees and their requirements as follows. The GPT-9600 falls under degree 2. Pollution refers to “addition of foreign matter, solid, liquid, or gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, nonconductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
Location: Indoor
Temperature: -10°C to 70°C
Relative Humidity: 85% (no condensation)
Do not dispose this instrument as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Please make sure discarded electrical waste is properly recycled to reduce environmental impact.

6

SAFETY INSTRUCTIONS

Power cord for the United Kingdom
When using the safety tester in the United Kingdom, make sure the power cord meets the following safety instructions.

NOTE: This lead/appliance must only be wired by competent persons

WARNING: THIS APPLIANCE MUST BE EARTHED

IMPORTANT: The wires in this lead are coloured in accordance with the

following code:

Green/ Yellow:

Earth

Blue:

Neutral

Brown:

Live (Phase)

As the colours of the wires in main leads may not correspond with the coloured marking identified in your plug/appliance, proceed as follows:

The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.

The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.

The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.

If in doubt, consult the instructions provided with the equipment or contact the supplier.

This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, a cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.

Any exposed wiring from a cable, plug or connection that is engaged in a live socket is extremely hazardous. If a cable or plug is deemed hazardous, turn off the mains power and remove the cable, any fuses and fuse assemblies. All hazardous wiring must be immediately destroyed and replaced in accordance to the above standard.

7

GPT-9600 Series User Manual

GETTING STARTED

This chapter describes the safety tester in a nutshell, including its main features and front / rear panel introduction. After going through the overview, please read the safety considerations in the Set Up chapter.

GPT-9603

AC / DC Withstanding Voltage / Insulation Resistance

Tester

PASS ESC

FAIL FIELD

READY

TEST

CAUTION HIGH VOLTAGE

5.0 kVAC MAX. 6.0 kVDC MAX.

POWER

START

STOP

REMOTE

EDIT

SAVE

UTILITY

RETURN

GPT-9600 Series Overview ………………………………………… 9
Series lineup……………………………………………………………………………………9 Model Overview……………………………………………………………………………..9 Main Features …………………………………………………………………………………9 Accessories and Package Contents ………………………………………………..10
Appearance …………………………………………………………… 12
GPT-9603/9602/9612/9601 Front Panel ………………………………………..12 GPT-9603/9602/9612/9601 Rear Panel …………………………………………16
Set Up ………………………………………………………………….18
Power Up………………………………………………………………………………………18 Workplace Precautions………………………………………………………………….20 Operating Precautions …………………………………………………………………..21 Basic Safety Checks ……………………………………………………………………….23

8

GETTING STARTED

GPT-9600 Series Overview

Series lineup
The GPT-9600 Series Safety Testers are AC/DC withstanding voltage and insulation resistance safety testers.
The GPT-9603 is an AC/DC withstanding and insulation resistance safety tester. The GPT-9602 is an AC/DC withstanding safety tester. The GPT-9612 is an AC withstanding and insulation resistance safety tester. The GPT-9601 is purely an AC withstanding safety tester.
Note: Throughout this user manual, the terms ACW, DCW and IR refer to AC Withstanding, DC Withstanding and Insulation Resistance, respectively. GPT-9600 refers to any of the GPT-96XX models.

Model Overview

Model name

ACW

DCW

IR

GPT-9603

GPT-9602

GPT-9612

GPT-9601

Main Features

Performance

ACW: 5kVAC DCW: 6kVDC IR: 50V, 100V, 250V, 500V, 1000V

9

GPT-9600 Series User Manual

Features Interface

Timer control Safety discharge Over temperature, voltage and current
protection Pass, Fail, Test, High Voltage and Ready
indicators PWM output (increased reliability) Interlock (configurable)
Remote control start/stop interface terminal Signal I/O port for pass/fail/test monitoring
and start/stop control/interlock

Accessories and Package Contents
Check the package contents to make sure all accessories and optional items are included before using the GPT-9600.

Standard Accessories Part number

Description

N/A

GPT-96XX unit.

N/A

User manual CD

N/A

Quick start guide

GHT-114 x1

Test lead

Region dependent Power cord

N/A

Remote terminal male plug

N/A

Interlock key

N/A

CTC (Calibration Traceable

Certificate)

Optional Accessories Part number

Description

GHT-205

High voltage test probe

GHT-113

High voltage test pistol

10

Note

GETTING STARTED

GRA-417

Rack adapter panel (19″, 4U)

Keep the packaging, including the box, polystyrene foam and plastic envelopes should the need arise to return the unit to GW Instek.

11

GPT-9600 Series User Manual

Appearance

GPT-9603/9602/9612/9601 Front Panel

Display
POWER
POWER button

PASS/FAIL indicators

Function keys

Navigation keys

READY indicator

GPT-9603

AC / DC Withstanding Voltage / Insulation Resistance

Tester

PASS ESC

FAIL FIELD

READY

TEST

CAUTION HIGH VOLTAGE

5.0 kVAC MAX. 6.0 kVDC MAX.

TEST indicator
HIGH VOLTAGE indicator

START

STOP

REMOTE

EDIT

SAVE

UTILITY

RETURN

HIGH VOLTAGE output terminal
RETURN terminal

START button REMOTE terminal

Scroll wheel

STOP button

Configuration keys

Display

240 X 48 dot matrix display (LCD)

Function keys

The function keys correspond to the soft-keys directly above on the main display.

Pass/Fail indicators

PASS FAIL

The PASS and FAIL indicators light up upon a PASS or FAIL test result.

Navigation keys

ESC key

ESC

The ESC key is used to exit out of a menu or cancel a setting.

FIELD key

FIELD

The FIELD key is navigate to the next menu item when the tester is either in the EDIT status or in one of the Utility menus.

12

GETTING STARTED

Directional arrow keys

The directional arrow keys are only used for calibration. They are not used during normal operation by the operator. The directional arrow keys can essentially be ignored by the operator.

READY indicator READY

The READY indicator is lit when the tester is ready to begin testing.

TEST indicator TEST

The TEST indicator is lit when a test is on. The START button is used to put the tester into TEST status from READY status.

HIGH VOLTAGE indicator
HIGH VOLTAGE output terminal

CAUTION
HIGH VOLTAGE

5.0 kVAC MAX. 6.0 kVDC MAX.

The HIGH VOLTAGE indicator will flash when the output terminal is active. Only after the test has finished or stopped will the indicator turn off.

CAUTION
HIGH VOLTAGE

5.0 kVAC MAX. 6.0 kVDC MAX.

The HIGH VOLTAGE terminal output is used for outputting the testing voltage. The terminal is recessed for safety. This terminal is used in conjunction with the RETURN terminal.

WARNING
USE EXTREME CAUTION. Do not touch the HIGH VOLTAGE terminal during testing.

13

RETURN terminal
RETURN

GPT-9600 Series User Manual
The RETURN terminal is used for IR, DCW and ACW tests.

Scroll wheel

The scroll wheel is used to edit parameter values or menu settings.

UTILITY key EDIT key

UTILITY EDIT

SAVE key
REMOTE terminal

SAVE
REMOTE

Used to enter the TEST Utility or Common Utility menu.
Used to enter the EDIT status. The EDIT status allows you to select the test mode and the test parameters.
The SAVE key is used to save test parameters when in the EDIT status or to save utility settings.
The REMOTE terminal is used to connect to a remote controller.

STOP button

STOP

The STOP button is used to

stop/cancel tests. The STOP

button will also put the safety

tester in the READY status to

begin testing.

14

START button

START

POWER switch POWER

GETTING STARTED
The START button is used to start tests.
The START button can be used to start tests when the tester is in the READY status. Pressing the START button will put the tester in the TEST status.
Turns the power on. The safety tester will always start up with the last test setting from when the instrument was last powered down.

15

GPT-9600 Series User Manual

GPT-9603/9602/9612/9601 Rear Panel

Calibration port

Fan

GND

CALIBRATION USE ONLY
SIGNAL I / O

WARNING TO AVOID ELECTRIC SHOCK THE POWER CORD PROTECTIVE GROUNDING CONDUCTOR MUST BE CONNECTED TO GROUND. FOR CONTINUED FIRE PROTECTION. REPLACE ONLY WITH SPECIFIED TYPE AND RATED FUSE. NO OPERATOR SERVICEABLE COMPONENTS INSIDE. DO NOT REMOVE COVERS. REFER SERVICING TO QUALIFIED PERSONNEL.
SER. NO. LB

ENSURE THE POWER IS REMOVED FROM THE INSTRUMENT
BEFORE REPLACING THE FUSE
AC POWER MAX. 400VA

GND

LINE VOLTAGE ( 50 / 60Hz )
100 120V
220 240V

FUSE
T 4A 250V

SIGNAL I/O

Line voltage Fuse holder

SIGNAL I/O port

SIGNAL I / O

Calibration port

CALIBRATION USE ONLY

Fan/Fan Vents

GND

GND

The SIGNAL I/O port is used to monitor the tester status (PASS, FAIL, TEST) and input (START/ STOP signals). It is also used with the Interlock key.
The Calibration port is used for calibration purposes only. Calibration is not supported for end-users. Please see your distributor for details.
Exhaust fan. Allow enough room for the fan to vent. Do not block the fan openings.
Connect the GND (ground) terminal to the earth ground.

16

Line voltage input and fuse holder

GETTING STARTED
Line voltage input: 100-120/220-240VAC Line Voltage Fuse: T 4A 250V

17

GPT-9600 Series User Manual

Set Up
Power Up

Background Steps

The GPT-9600 accepts line voltages of 100-120V and 220-240V at 50Hz or 60Hz.
1. Connect the power cord to the AC voltage input.

Warning

2. If the power cord does not

GND

have an earth ground,

ensure the ground

terminal is connected to an

earth ground.

Ensure the power cord is connected to an earth ground. Failure could be harmful to the operator and instrument.

3. Press the Power button.

POWER

4. When the unit is powering up, all the LED indicators will light. Check to make sure all 5 LED indicators are working.
5. Check to make sure the System Self Test passes without errors.

18

GETTING STARTED

SYS T EM SE L F T ES T S y s t em Ch e c k i n g . . . Ha r dwa r e Che c k i ng . . . F i r mw a r e C h e c k i n g

T I MER= 0 0 T I MER= 0 0 T I MER= 0 0 T I MER= 0 0

After the System Self Test completes, the tester will go into READY status and be ready to operate.

READY Status

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V

m A

A CW D CW

IR

T=174S
READY
MOD E T I ME

19

GPT-9600 Series User Manual

Workplace Precautions

Background

The GPT-9600 Series are a high voltage instruments that output dangerous voltages. The following section describes precautions and procedures that must be followed to ensure a safe work environment.

WARNING

The GPT-96XX generates voltages in excess of 5kVAC or 6kVDC. Follow all safety precautions, warnings and directions given in the following section when using the instrument.

1. Only technically qualified personnel should be allowed to operate the safety tester.

2. The operating workplace must be fully isolated, especially when the instrument is in operation. The instrument should be clearly labeled with appropriate warning signage.

3. The operator should not wear any conductive materials, jewelry, badges, or other items, such wrist watches.

4. The operator should wear insulation gloves for high voltage protection.

5. Ensure the earth ground of the line voltage is properly grounded.

6. Ensure any devices that are adversely affected by magnetic fields are not placed near the tester.

20

GETTING STARTED

Operating Precautions

Background

The GPT-9600 Series are high voltage instruments that output dangerous voltages. The following section describes precautions and procedures that must be followed to ensure that the testers are operated in a safe manner.

WARNING

The GPT-96XX generates voltages of up to 5kVAC or 6kVDC. Follow all safety precautions, warnings and directions given in the following section when using the instrument.

1. Never touch the safety tester, lead wires, terminals, probes and other connected equipment when the tester is testing.

2. Do not turn the safety tester on and off quickly or repeatedly. When turning the power off, please allow a few moments before turning the power back on. This will allow the protection circuits to properly initialize.

Do not turn the power off when a test is running, unless in an emergency.

3. Only use those test leads supplied with the instrument. Leads with inappropriate gauges can be dangerous to both the operator and the instrument.

4. Do not short the HIGH VOLTAGE terminal with ground. Doing so could charge the chassis to dangerously high voltages.

5. Ensure the earth ground of the line voltage is properly grounded.

21

GPT-9600 Series User Manual
6. Only connect the test leads to the HIGH VOLTAGE terminals before the start of a test. Keep the test leads disconnected at all other times.
7. Always press the STOP button when pausing testing.
8. Do not leave the safety tester unattended. Always turn the power off when leaving the testing area.
9. When remotely controlling the safety tester, ensure adequate safety measures are in place to prevent:
Inadvertent output of the test voltage.
Accidental contact with the instrument during testing. Ensure that the instrument and DUT are fully isolated when the instrument is remotely controlled.
10. Ensure an adequate discharge time for the DUT.
When DCW or IR tests are performed, the DUT, test leads and probes become highly charged. The GPT-96XX has discharge circuitry to discharge the DUT after each test. The time required for a DUT to discharge depends on the DUT and test voltage.
Never disconnect the safety tester before a discharge is completed.
22

GETTING STARTED

Basic Safety Checks

Background

The GPT-9600 Series are high voltage devices and as such, daily safety checks should be made to ensure safe operation.

1. Ensure all test leads are not broken and are free from defects such as cracks or splitting.

2. Ensure the safety tester is always connected to an earth ground.

3. Test the safety tester operation with a low voltage/current output: Ensure the safety tester generates a FAIL judgment when the HIGH VOLTAGE and RETURN terminals are shorted (using the lowest voltage/current as the testing parameters).

WARNING

Do not use high voltages/currents when the HIGH VOLTAGE and RETURN terminals are shorted. It may result in damage to the instrument.

23

GPT-9600 Series User Manual
OPERATION
Menu Tree …………………………………………………………….25
Menu Tree Overview…………………………………………………………………….26
Test Lead Connection ………………………………………………29
ACW, DCW, IR Connection ………………………………………………………….29 ACW and DCW Grounding Mode Information …………………………….30 IR Grounding Mode Information ………………………………………………….31
ACW, DCW and IR Testing ………………………………………..33
Setting the Test Function……………………………………………………………….34 Edit Test Settings…………………………………………………………………………..34 Setting the Upper and Lower Limits ……………………………………………..35 Setting the Test Time (Timer) ………………………………………………………..37 Setting the ARC Mode …………………………………………………………………..39 Saving and Exiting EDIT Status …………………………………………………….41 Setting the Test Voltage and Running a Test …………………………………42 PASS / FAIL Test Judgments ………………………………………………………..45 Auto Mode Testing ……………………………………………………………………….49 PASS / FAIL Test Judgments for Auto Mode ………………………………..52
Common Utility Settings ………………………………………….55
LCD Settings …………………………………………………………………………………55 Control Settings …………………………………………………………………………….56
24

OPERATION

Menu Tree
This section describes the overall structure of the operation statuses and modes for the GPT-9600 safety testers. The testers have 6 operation statuses (EDIT, READY, TEST, STOP, PASS and FAIL).

FA I L
Status (buzzer on)
Press STOP
FA I L
Status (buzzer off)
Press SAVE
Common Utility Settings1

Press STOP
Press UTILITY

PASS/FAIL Judgement

TEST
status
Press START

READY
status

Press SAVE

Press EDIT

ED I T
status1

1. Press ESC to cancel and return to the previous screen.

Press START Press STOP
Press UTILTY

PASS
status

S TOP
status

Press STOP

Press STOP

Press SAVE
TEST UTILITY Settings1

25

GPT-9600 Series User Manual

Menu Tree Overview

READY status EDIT status

When the tester is in READY status, it is ready to begin testing. Pressing the START button will begin testing and put the tester into TEST status. Pressing the EDIT key will put the tester into the EDIT status. Pressing the UTILITY key will enter the Common Utility settings.

READY Status

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V

m A

A CW D CW

IR

T=174S
READY
MOD E T I ME

EDIT status is used to edit the test parameters. Pressing the SAVE key will save any changes and return to the READY status. Pressing the ESC key will cancel any changes and return to the READY status. Pressing the UTILITY key will enter the TEST Utility settings.

EDIT Status

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V

m A

A CW D CW

IR

T=174S
ED I T
MOD E T I ME

26

TEST status STOP status FAIL status

OPERATION

TEST status is active when a test is running. Pressing STOP will cancel the test and put the tester into the STOP status. Waiting for the test to complete will result in a PASS or FAIL judgment.
TEST Status

H= 2 2 . 0mA L = 0 . 0 0mA

0 14k V 0 00mA

A CW D CW

IR

T=008S
TEST
MOD E T I ME

STOP status is shown when a test did not finish running and has been stopped by the operator. Pressing STOP will return the tester to READY status.

STOP Status

H= 2 2 . 0mA L = 0 . 0 0mA

0 14k V 0 00mA

A CW D CW

IR

T=008S
S TOP
MOD E T I ME

When a DUT fails a test, the FAIL status is displayed. Pressing STOP twice will return the tester to the READY status.

FAIL Status

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V 23 0mA

A CW D CW

IR

T=060S
FA I L
MOD E T I ME

27

GPT-9600 Series User Manual

PASS Status
Common Utility Settings
TEST Utility Settings

When a DUT passes a test, the PASS status is displayed. Pressing STOP returns the unit to the READY status, pressing START will re-run the test again.

PASS Status

H= 2 2 . 0mA L = 0 . 0 0mA

0 90k V 0 00mA

A CW D CW

IR

T=060S
PASS
MOD E T I ME

This utility controls the LCD and control settings. These settings are system wide.

COMMON U T I L I T Y LCD Co n t r a s t : 7 L CD B r i g h t n e s s : BR I GHT
LCD CTRL

T I MER= 0 0 T I MER= 0 0
R=0 0 T I MER= 0 0

The TEST Utility configures the ARC mode (ACW, DCW) and the ACW test frequency.

TEST UT I L I TY A RC MOD E : O F F FREQUENCY : 6 0 HZ

T I MER= 0 0 T I MER= 0 0 T I MER= 0 0 T I MER= 0 0

28

OPERATION

Test Lead Connection
This section describes how to connect the GPT-9600 to a DUT for withstanding or insulation resistance tests.

ACW, DCW, IR Connection

Background
ACW, DCW, IR Connection

ACW, DCW and IR tests use the HIGH VOLTAGE terminal and RETURN terminal with the GHT-114 test leads.
GPT-9600 High Voltage terminal
DUT
Return terminal

Steps

1. Turn the power off on the safety tester.
2. Connect the high voltage test lead(red) to the HIGH VOLTAGE terminal and screw firmly into place.
3. Connect the return test lead(white) into the RETURN terminal and screw the protector bar into place, as shown below.

HIGH VOLTAGE Terminal

RETURN Terminal
29

GPT-9600 Series User Manual

ACW and DCW Grounding Mode Information

Background Example

For ACW and DCW tests, the GROUND MODE is always set to ON. In the ON setting, the GPT9600 grounds the return terminal to the ground. This mode is best for DUTs that are grounded to an earth ground by their chassis, fixtures or operation environment. This mode measures the potential of the HIGH VOLTAGE terminal with respect to earth ground. This means that any stray capacitance/resistance that leaks to earth ground will also be measured.

GROUND MODE = ON, DUT grounded
GPT-9600

High Voltage terminal Return terminal DUT

stray resistance, capacitance

GROUND MODE = ON, DUT floating
GPT-9600

High Voltage terminal Return terminal DUT

stray resistance, capacitance

Ground Mode Icon for ACW or DCW tests

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V

m A

A CW D CW

IR

T=174S
READY
MOD E T I ME

Ground mode ON

30

OPERATION

IR Grounding Mode Information

Background
Warning Example

For IR tests the GROUND MODE is always set to OFF. Here, the return terminal is floating with respect to the earth ground. This mode is for DUTs that are floating and not directly connected to an earth ground. In this mode any stray capacitance/resistance that leaks to the earth ground from the DUT side of the testing circuit and will not be measured.

When GROUND MODE is set to OFF, the DUT, fixtures or connected instrumentation cannot be grounded. This will short circuit the internal circuitry during a test.

GROUND MODE = OFF, DUT floating
GPT-9600

High Voltage terminal Return terminal DUT

stray resistance, capacitance

GROUND MODE = OFF, DUT grounded
GPT-9600

High Voltage terminal Return terminal DUT

stray resistance, capacitance

31

Ground Mode Icon for IR tests

GPT-9600 Series User Manual

H=

M L = 0 0 0 1 M

0 10k V

M

A CW D CW

IR

T =OF F
READY
MOD E

Ground mode OFF

32

OPERATION
ACW, DCW and IR Testing
This section describes how to create, edit and run a single ACW, DCW or IR safety test. Each setting described in this chapter only applies to the selected test mode ­ For example the test time setting for an ACW test is different to the DCW and IR test time settings. Setting the Test Function from page 34. Editing the Test Settings from page 34. Setting the Upper and Lower Limits from page 35. Setting the Test Time (Timer) from page 37. Setting the ARC Mode from page 39. Saving and Exiting EDIT Status from page 41. Setting the Test Voltage and Running a Test from page 42. PASS/FAIL Test Judgments from page 45. Auto Mode Testing from page 49. PASS/FAIL Test Judgments for Auto Mode from page 52. Before operating the GPT-9600 please read the safety precautions as outlined in the Set Up chapter on page 18.
33

GPT-9600 Series User Manual

Setting the Test Function

Background Steps

There are three test functions, AC Withstand, DC Withstand and Insulation Resistance.

1. To choose the test function, press the ACW, DCW or IR soft-keys when the tester is in the READY or EDIT status.

A CW

D CW

IR

2. The test function soft-key is highlighted.

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V

m A

A CW D CW

IR

Test function

T=174S
READY
MOD E T I ME

Edit Test Settings

Background Steps

To edit any test settings, the tester must be in EDIT status.

Any settings or parameters that are edited only apply to the currently selected test function.

1. Press the EDIT key when in

EDIT

READY status to enter the EDIT

status. This will enter the EDIT

status for the current test function.

34

OPERATION

Note

READY

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V

m A

A CW D CW

IR

T=174S
ED I T
MOD E T I ME

2. The Status changes from READY to EDIT.

Pressing the SAVE key will save the settings for the current test function and return back to READY status.

Setting the Upper and Lower Limits

Background Steps

There is both L (low) and H (high) judgment settings. When the measured value is below the L setting, the test will be judged as FAIL. When the value exceeds the H setting the test will be judged as FAIL. Any measurement between the L and H setting is judged as PASS. The L limit cannot be made greater than the H limit.

1. Press the FIELD soft-key to bring FIELD the cursor to the H setting or the L setting when in the EDIT status.

cursor

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V

m A

A CW D CW

IR

T=174S
ED I T
MOD E T I ME

2. Use the scroll wheel to set the H limit*.

35

Note

GPT-9600 Series User Manual

ACW (H) 0.01mA~22.0mA

DCW (H) 0.01mA~6.00mA

IR (H)

0002M~

3. Press the FIELD soft-key to bring FIELD the cursor to the L setting or the H setting.

cursor

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V

m A

A CW D CW

IR

T=174S
ED I T
MOD E T I ME

4. Use the scroll wheel to set the H/L limit*.

ACW (L) DCW (L) IR (L)

0.00mA~21.9mA 0.00mA~5.99mA 0001M ~ 2000M

The L limit cannot be greater than the H limit.
When setting the current, be aware that a maximum of 100VA can be set for ACW and 25W for DCW.
The word “OVER” will be displayed on unit when the measured resistance is beyond 2,000 ohms.

36

OPERATION

Setting the Test Time (Timer)

Background

The TIMER setting is used to set the test time for the test. The test time determines how long to run the test at the test voltage. This test time does not include the Ramp up time (fixed at 100ms). The test time can be set from 001 seconds to 180 seconds for ACW, DCW and IR tests, with a resolution of 1 second for all modes. The timer can also be turned off. The total discharge time depends on the DUT and is not specified.

Steps

Output V

TEST TIME

time

RAMP 100ms

Discharge time

1. When in the EDIT status, press the FIELD FIELD key to bring the cursor to the T setting.

cursor

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V

m A

A CW D CW

IR

T=174S
ED I T
MOD E T I ME

2. Use the scroll wheel to set the T value*.

Range

001s~180s

37

GPT-9600 Series User Manual

Note
Turning the test time off.

*Note that the T value cannot be edited with the scroll wheel if it is set to OFF. See below to toggle the test time on or off.
When the output current exceeds 15mA, AC withstanding tests cannot be continuously performed without a cool down period between tests. For 15mAI20mA, the cool down period must equal or exceed the test time. See the AC Withstanding specifications on page 71 for details.
The test time (T setting) can be turned on or off. When turned off, a test will run indefinitely until the test has failed or is stopped by the user.
Press the TIME soft-key to toggle T I ME the test time (T setting) on or off.

38

OPERATION

Setting the ARC Mode

Background
Steps Note

ARC detection, otherwise known as flashover detection, detects fast voltage or current transients that are not normally detected. Arcing is usually an indicator of poor withstanding insulation, electrode gaps or other insulating problems that cause temporary spikes in current or voltage during ACW.
There are three ARC detection settings: OFF, ON AND CONTINUE, ON AND STOP. The ON AND CONTINUE setting will detect arcs over the ARC current level and continue the test, the ON AND STOP setting will stop the test when an arc is detected.
ARC mode settings only apply to ACW and DCW tests.
1. Press the UTILITY key on the front UTILITY panel when the tester is in EDIT status. The tester will go to the TEST Utility.

TEST UT I L I TY A RC MOD E : O F F FREQUENCY : 6 0 HZ

T I MER= 0 0 T I MER= 0 0 T I MER= 0 0 T I MER= 0 0

The TEST utility applies to the ACW and DCW test functions only.

2. Use the FIELD key to move to the FIELD ARC MODE setting.

39

Note
Note 40

GPT-9600 Series User Manual

3. Use the scroll wheel to set the ARC mode.

ARC MODES: OFF, ON AND CONTINUE, ON AND STOP
4. Press the SAVE key to save and exit SAVE the TEST Utility and go back to EDIT status.

The ESC key can be pressed at any time in the Utility menu to cancel and exit.

5. If the ARC MODE was set to either ON AND CONTINUE, or ON AND STOP, the ARC current level will now appear on the display and can be edited.

ARC MODE

  • H = 2 . 0 0 m A L = 0 . 0 0 m A

= 1 . 0 0mA T = 0 5 3 S

0 75k V 0 00mA T E S T

A CW D CW

IR

MOD E

6. Use the FIELD key to move the

FIELD

cursor to the ARC setting.

7. Use the scroll wheel to edit the ARC level.

ACW DCW

1.00mA~40.0mA 1.00mA~12.0mA

When the signle “*” symbol appears, it indicates the function of ARC is turned ON without ARC detection. When two “**” symbols appear, it means the function of ARC is turned ON with ARC detection during test.

OPERATION

Saving and Exiting EDIT Status

Background Steps
Note

After all test parameters have been set, the test can be saved.

1. When in EDIT status, press the

SAVE

SAVE key to save the test.

ED I T

H=

M L = 0 0 0 1 M

0 50k V

M

A CW D CW

I R

T =OF F
READY
AC – I R

2. The status changes from EDIT to READY.

Pressing the SAVE key again will return the tester back to EDIT status for the current test.

41

GPT-9600 Series User Manual

Setting the Test Voltage and Running a Test

Background Note
Steps

The test voltage can be set when the tester is in the READY status. After the test voltage has been set, the test can then be run.

The tester cannot start to run a test under the following conditions:
The INTERLOCK function is ON and the Interlock key is not inserted in the signal I/O port (page 56).
The STOP signal has been received remotely.
If Double Action is ON, ensure the START button is pressed immediately after the STOP button (<0.5s). See page 56.

1. Connect the DUT to the tester.

Page 29

2. Ensure the tester is in READY status and that all the testing parameters have been set for the test.

From Page 30

READY Status

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V

m A

A CW D CW

IR

T=174S
READY
MOD E T I ME

3. The READY indicator will be lit READY blue when in the READY status.

42

OPERATION

4. For ACW and DCW tests, turn the scroll wheel to set the desired test voltage.

Range

0.1kV ~ 5.0kV(ACV)/6.0kV(DCV)

5. For IR tests, turn the scroll wheel to set the voltage to one of five settings.

Range

50V, 100V, 250V, 500V, 1000V

Voltage setting

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V

m A

A CW D CW

IR

T=174S
READY
MOD E T I ME

6. Press the START button. The test

START

starts automatically and the tester

goes into the TEST status.

Note

7. The TEST indicator will be lit

TEST

orange when in the TEST status.

8. The test will start by showing the remaining test time. The test will continue until the test is finished or the test is stopped.

ACW and DCW mode can also edit the test voltage when a test is running.

Conversely, IR tests cannot edit the test voltage when a test is running.

43

GPT-9600 Series User Manual

Example Stop the Test Example

Measured current

TEST voltage

TEST Status

H= 2 2 . 0mA L = 0 . 0 0mA

T=008S

0 94k V 9 00mA T E S T

A CW D CW

IR

MOD E T I ME

Remaining test time

9. To stop the test at any time when it

STOP

is running, press the STOP button.

The test will stop immediately.

When the STOP button is pressed,

a judgment is not made on the test.

All panel keys except the STOP button are locked when the tester is in STOP status.

STOP Status

H= 2 2 . 0mA L = 0 . 0 0mA

0 94k V 9 00mA

A CW D CW

IR

T=008S
S TOP
MOD E T I ME

10. To put the tester back into READY

STOP

status, press the STOP button

again.

Note

Do not touch any terminals, test leads or any other connections when the test is on.

44

OPERATION

PASS / FAIL Test Judgments

Background

If the test is allowed to run to completion (the test is not stopped or a protection setting is not tripped) then the tester will judge the test as either PASS or FAIL.

Note

The test will be judged PASS when: The H and L limits (high and low limits)
have not been tripped during the test time.
The test will be judged FAIL when: Either the H or L limit has been tripped
during the test time. The ARC current limit has been tripped
during a test (if enabled).

PASS Judgment 1. When the test is judged as PASS,

PASS

PASS will be displayed, the buzzer

will sound once and the PASS

indicator will be lit green.

H= 2 2 . 0mA L = 0 . 0 0mA

0 90k V 1 00mA

A CW D CW

IR

T=060S
PASS
MOD E T I ME

2. The PASS judgment will be held on the display until the STOP or START button is pressed.

Pressing the STOP button will

STOP

return the tester to the READY

status.

45

GPT-9600 Series User Manual

Pressing the START button will

START

restart the test again.

PASS Timing Diagrams ACW PASS Timing
DCW PASS Timing
IR PASS Timing

The timing diagrams below show the ACW, DCW and IR timing for the START status, TEST status and PASS judgment.

START TEST PASS

Output V

time TEST TIME

RAMP (100ms)

Discharge time

START TEST PASS

Output V

time TEST TIME

RAMP (100ms)

Discharge time

START TEST PASS

Output V

time TEST TIME

RAMP (100ms)

Discharge time

46

OPERATION

FAIL Judgment 1. When the test is judged as FAIL,

FAIL

FAIL will be displayed, the buzzer

will sound and the FAIL indicator

will be lit red.

As soon as a test is judged FAIL, the output is cut from the terminals.

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V 23 0mA

A CW D CW

IR

T=060S
FA I L
MOD E T I ME

2. The FAIL judgment will be held on

STOP

the display until the STOP button

is pressed twice. Pressing the STOP

button once will stop the buzzer,

pressing the STOP button again

will return the tester to the READY

x2

status.

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V

m A

A CW D CW

IR

T=174S
READY
MOD E T I ME

FAIL Timing Diagrams

The timing diagrams below show the ACW, DCW and IR timing for the START status, TEST status and FAIL judgment.

ACW FAIL Timing

START TEST FAIL

Output V

time TEST TIME

RAMP (100ms)

Discharge time

47

GPT-9600 Series User Manual

DCW FAIL Timing

START TEST FAIL

Output V

time TEST TIME

RAMP (100ms)

Discharge time

IR FAIL Timing

START TEST FAIL

Output V

time TEST TIME

RAMP (100ms)

Discharge time

48

OPERATION

Auto Mode Testing

Background Steps

The Auto Mode function allows you to run two test functions back to back automatically. Only after both tests have run will a PASS/FAIL judgment be determined on the tests.

1. Connect the DUT to the tester.

Page 29

2. Ensure the tester is in READY

From Page

status and that all the testing

33

parameters have been set for both the

tests* that you wish to run in Auto

Mode.

3. Press the Mode key to toggle

MOD E

between each of the Auto Modes.

Each mode indicates the order that the tests will execute in. For example: IR- DC indicates that an IR test will be followed by a DCW test.

Range

MODE, AC-IR, IR-AC, DC-IR, IRDC

Note: MODE is the default setting which only runs a single test a time.

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V

m A

A CW D CW

IR

T=174S
READY
AC – I R T I ME

Selected Auto mode

4. The Auto Mode function is now ready to test two tests in succession.

49

GPT-9600 Series User Manual

5. With the tester in READY status,

START

press the START button. The test

starts the first test and goes into

the TEST status for the first test.

6. The TEST indicator will be lit

TEST

orange when in the TEST status.

7. The first test will start by showing the remaining test time.

8. For ACW or DCW tests, the scroll wheel can be used to edit the test voltage when the tester is in the TEST status.

ACW

0.10kV ~ 5.0kV

DCW

0.10kV ~ 6.0kV

9. The test will continue until the test time runs down or the test is stopped. When the test time ends, the tester will automatically go onto the second test.

Like the first test, the second test will show the remaining test time and will continue until the test time runs out or the test is stopped.

If both tests run to completion, each test will be judged as a PASS or a FAIL, see page 52 for details.

50

OPERATION

Example Stop the Test Example

Measured current

TEST voltage

TEST Status

H= 2 2 . 0mA L = 0 . 0 0mA

T=008S

0 94k V 9 00mA T E S T

A CW D CW

IR

AC – I R T I ME

Currently running test

Selected Remaining test

Auto Mode

time

10. To stop the tests at any time when

STOP

they are running, press the STOP

button. Testing will stop

immediately. When the STOP

button is pressed, a judgment is

not made on the current test. If a

test has already been run, its

judgment is still valid.

All panel keys except the STOP button are locked when the tester is in STOP status.

STOP Status

H= 2 2 . 0mA L = 0 . 0 0mA

0 94k V 9 00mA

A CW D CW

IR

T=008S
S TOP
AC – I R T I ME

11. To put the tester back into READY

STOP

status, press the STOP button

again.

Note

Do not touch any terminals, test leads or any other connections when the test is on.

51

GPT-9600 Series User Manual

PASS / FAIL Test Judgments for Auto Mode

Background

The test results for the Auto Mode testing are identical as the results from a single test, however:
If one of the tests fails, then the FAIL judgment for that test will be shown on the display, even if the other test is judged as PASS.
Only one test result can be viewed at any one time.

Note

A test will be judged PASS when: The H and L limits (high and low limits)
have not been tripped during the test time.
A test will be judged FAIL when: Either the H or L limit has been tripped
during the test time. The ARC current limit has been tripped
during a test (if enabled).

View Test

1. When both tests are judged as

PASS

Results~PASS

PASS, PASS will be displayed for

the last test, the buzzer will sound

once and the PASS indicator will

be lit green.

H= 2 2 . 0mA L = 0 . 0 0mA

0 90k V 1 00mA

A CW D CW

IR

T=060S
PASS
I R – AC T I ME

2. Turn the scroll wheel left to view the first test result. Turn the scroll wheel back to the right to view the last test result again.

52

OPERATION

3. The PASS judgment will be held on the display until the STOP or START button is pressed.

Pressing the STOP button will

STOP

return the tester to the READY

status.

Pressing the START button will

START

restart the tests again.

FAIL Judgment

4. After both tests have finished the

FAIL

output will be cut from the

terminals. If any or both of the tests

are judged as FAIL, FAIL will be

displayed on the screen, the buzzer

will sound and the FAIL indicator

will be lit red.

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V 23 0mA

A CW D CW

IR

T=060S
FA I L
AC – I R T I ME

Note

If both tests fail, then the last test will be shown on the display.

5. The FAIL judgment will be held on

STOP

the display until the STOP button

is pressed.

Pressing the STOP button will turn the buzzer sound off and allow you to view the test results.

53

GPT-9600 Series User Manual

6. Turn the scroll wheel left to view the first test result and turn the scroll wheel back to the right to view the last test result again.

7. Pressing the STOP button again

STOP

will return the tester to the READY

status.

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V

m A

A CW D CW

IR

T=174S
READY
AC – I R T I ME

Timing Diagrams

The PASS/FAIL timing diagrams for the Auto Mode are same as those for single tests. See page 45 for PASS/FAIL judgments and timing diagrams.

54

OPERATION

Common Utility Settings
The Common Utility settings are system-wide settings that apply to all function tests.

The Common Utility menu includes the following settings: LCD settings from page 55. Control settings from page 56.
LCD Settings

Description Steps

The LCD settings include contrast and brightness controls.

1. Ensure the tester is in READY status.

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V

m A

A CW D CW

IR

T=174S
READY
MOD E T I ME

2. Press the UTILITY key.

UTILITY

3. Press the LCD soft-key to bring up L C D the LCD Common Utility menu.

COMMON U T I L I T Y LCD Co n t r a s t : 7 L CD B r i g h t n e s s : BR I GHT
LCD CTRL

T I MER= 0 0 T I MER= 0 0
R=0 0 T I MER= 0 0

4. Press the FIELD key to choose a

FIELD

menu item: LCD Contrast, LCD

Brightness.

55

GPT-9600 Series User Manual

Note

5. Use the scroll wheel to select a parameter for the chosen menu item.
LCD Contrast 1(low) ~ 8(high) LCD Brightness BRIGHT, DARK

6. Press SAVE to save the settings

SAVE

and exit to READY status.

The ESC key can be pressed at any time to cancel and exit back to READY status.

Control Settings

Description

The Control settings are accessed in the Common Utility menu. The Control settings include: Start Control, Double Action, Key Lock and Interlock.
Start Control is used to determine how a test is started. Tests can be started via the front panel (START/STOP buttons), from a remote controller or via the SIGNAL I/O port.
The Double Action function is a safety feature used to prevent accidentally starting a test. Normally to start a test, the START button is pressed when the tester is in the READY status. To start a test when Double Action is ON, the STOP button must first be pressed, followed by the START button within 500ms.
Key Lock disables the front panel keys from changing the test function or testing parameters. Only the Utility menu and any keys required for testing are not disabled.

56

Steps

OPERATION

The Interlock function is a safety feature. The interlock function prevents a test from running, unless the interlock pins on the signal I/O port connector are shorted. The included interlock key can be used for this purpose. See page 65 for details.

1. Ensure the tester is in READY status.

H= 2 2 . 0mA L = 0 . 0 0mA

1 00k V

m A

A CW D CW

IR

T=174S
READY
MOD E T I ME

2. Press the UTILITY key.

UTILITY

3. Press the CTRL soft-key to bring C T R L up the Control Common Utility menu.

COMMON U T I L I T Y

S t a r t C t r l : FRONT PANE L

T I MER= 0 0

Do u b l e A c t i o n : OFF . . .

T I MER= 0 0

Ke y L o c k : OFF

I NT ER LOCK : OF F R= 0 0

LCD CTRL

T I MER= 0 0

4. Press the FIELD key to choose a

FIELD

menu item: Start Ctrl, Double

Action, Key Lock or INTERLOCK.

5. Use the scroll wheel to select setting for the chosen menu item.

Start Ctrl
Double Action Key Lock INTERLOCK

FRONT PANEL, REMOTE CONNECT, SIGNAL IO ON, OFF ON, OFF ON, OFF

57

Note

GPT-9600 Series User Manual

6. Press SAVE to save the settings

SAVE

and exit to READY status.

If a test is started with INTERLOCK ON, but the interlock signal I/O pins are not shorted (either with the included interlock key or manually), the INTERLOCK OPEN message will be displayed, preventing the test from starting.

Interlock open message

H= 6 . 0 0mA L = 0 . 0 0mA

= 1 . 0 0mA T =OF F

0 00 R E A D Y I N T E R L O C K O P E N

kV

mA

A CW D CW

IR

MOD E

58

EXTERNAL CONTROL
EXTERNAL CONTROL
The External Control chapter covers the REMOTE terminal and the SIGNAL I/O port.
External Control Overview ………………………………………. 60
Remote Terminal Overview…………………………………………………………..60 Remote Controller Operation ………………………………………………………..61 SIGNAL I/O Overview …………………………………………………………………62 Using the SIGNAL I/O to Start/Stop Tests …………………………………..64 Using the Interlock Key …………………………………………………………………65
59

GPT-9600 Series User Manual

External Control Overview
The External Control section describes the front panel REMOTE terminal connection and the rear panel SIGNAL I/O port.
Remote Terminal Overview

Overview
WARNING Pin Assignment and Connection

The REMOTE terminal connector is a standard 5-pin DIN terminal suitable for a remote controller.

Keep any cables that are connected to the REMOTE terminal away from the HIGH VOLTAGE and RETURN terminals.

RMT_START RMT_STOP

REMOTE

25 4

1

3

Pin Pin name 1 RMT_STOP 2 COM 3 Not used 4 RMT_START 5 Not used Signal Properties High level input voltage Low level input voltage Input period

Description Remote Stop signal Common line
Remote Start signal
2.4V~3.3V 0~0.8V minimum of 1ms

60

EXTERNAL CONTROL

Remote Controller Operation

Description Steps

The GPT-9600 accepts external remote controllers with a START and STOP button. To use the REMOTE terminal, the GPT-9600 must first be configured to accept a remote controller.

Operating a remote controller is the same as operating the START and STOP buttons on the front panel.

1. Insert the lead of remote controller into the REMOTE terminal.

REMOTE

NOTE

2. Configure the Start Ctrl option to REMOTE CONNECT in the Common Utility menu.

Page 56

3. The tester will now only be able to start a test using a remote controller.

Even if the GPT-9600 is configured to use the REMOTE CONNECT option, the STOP button on the front panel can still be used to stop a test.

4. To return the operation control to Page 56 the front panel, configure the Start Ctrl option to FRONT PANEL.

61

GPT-9600 Series User Manual

SIGNAL I/O Overview

Overview Pin Assignment

The SIGNAL I/O port can be used to remotely start/stop tests and monitor the test status of the instrument. The SIGNAL I/O port is also used for the interlock function (page 56).
The SIGNAL I/O port uses a DB-9 pin female connector.
6 78 9

1 234 5

Pin name INTERLOCK1 INTERLOCK2 INPUT_COM INPUT_START INPUT_STOP OUTPUT_TEST OUTPUT_FAIL OUTPUT_PASS OUTPUT_COM

Pin Description 1 When INTERLOCK is ON, a test is only allowed 2 to start when both INTERLOCK pins are shorted. 3 Common input line 4 Start signal input 5 Stop signal input 6 Indicates that a test is in progress 7 Indicates that a test has failed 8 Indicates that a test has passed 9 Common output line

Interlock connection

PIN 1 INTERLOCK1 PIN 2 INTERLOCK2

Input Connection

PIN 3 INPUT_COM PIN 4 INPUT_START PIN 5 INPUT_STOP

62

Output Connection
Signal Properties

EXTERNAL CONTROL

PIN 6 OUTPUT_TEST
PIN 7 OUTPUT_FAIL
PIN 8 OUTPUT_PASS
PIN 9 OUTPUT_COM
Input Signals High level input voltage Low level input voltage Low level input current Input period Output Signals Output Type Output Rated Voltage Maximum output current

5V ~ 32V 0V ~ 1V Maximum of -5mA Minimum of 1ms
Relay form A 30VDC 0.5A

63

GPT-9600 Series User Manual

Using the SIGNAL I/O to Start/Stop Tests

Background

To use the SIGNAL I/O port the Start Ctrl settings have to be set to SIGNAL I/O in the Common Utility menu.

Panel operation 1. Set the Start Ctrl option to SIGNAL Page 56 I/O.

2. Connect the Input/Output signals to the SIGNAL I/O port.

3. To start the testing, short the INPUT_START and INPUT_COM lines for a minimum of 1ms.

4. To stop the testing, temporarily short the INPUT_STOP and INPUT_COM line again.

NOTE

Even if the GPT-9600 is configured to use the SIGNAL I/O interface, the STOP button on the front panel can still be used to stop a test.

NOTE

If Double Action is used, you will need to first short the INPUT_STOP and INPUT_COM followed immediately (<500ms) by INPUT_START and INPUT_COM.

64

EXTERNAL CONTROL

Using the Interlock Key

Background

When the INTERLOCK function is set to ON, tests are only allowed to start when both Interlock pins on the signal I/O port are shorted. Using the Interlock key will short the INTERLOCK1 and INTERLOCK2 pins on the signal I/O port. See page 62 for the Signal I/O pin assignment.

Panel operation

1. Insert the Interlock key into the SIGNAL I/O port on the rear panel.

Note

2. Set the INTERLOCK option to ON Page 56 in the Common Utility.
With INTERLOCK set to ON, the tester can now only start a test when the Interlock key is connected. Do not remove the interlock after starting a test. It must be connected after a test has started or is running.
Set INTERLOCK to OFF to disable this feature.

65

GPT-9600 Series User Manual
FAQ
· The tester will not turn on. · Some/all of the panel keys are not working. · When I press the START button the tester will not start testing. · The accuracy does not match the specification.
The tester will not turn on.
Ensure the power cord is connected. Check to make sure the fuse is not blown. See page 68.
Some/all of the panel keys are not working.
Ensure the tester is not in SIGNAL I/O mode, Remote Connect mode or that the key lock function is not turned on. See page 56 for details.
When I press the START button the tester will not start testing?
The tester must first be in the READY status before a test can be started. Ensure the tester displays READY before pressing the START button, page 42. If “Double Action” is enabled, the START button must be pressed within 0.5 seconds after the STOP button is pressed, otherwise the tester will not start testing. See the Control Settings section on page 56.
66

FAQ If “Interlock” is enabled, the interlock key must be inserted into the signal I/O port on the rear panel before a test can be started. See page 65 for details. Lastly, ensure that the Start Ctrl setting is correctly configured in the Common Utility menu. For example, to enable the START button to start a test, ensure that the Start Ctrl setting is set to FRONT PANEL. See page 56 for details.
The accuracy does not match the specification.
Make sure the tester is powered on for at least 30 minutes, within +15°C~+35°C. This is necessary to stabilize the unit to match the specification. For more information, contact your local dealer or GWInstek at www.gwinstek.com / marketing@goodwill.com.tw.
67

GPT-9600 Series User Manual

APPENDIX

Fuse Replacement

Steps

1. Turn the instrument

POWER

off.

2. Remove the fuse socket using a flat screwdriver.

3. Replace the fuse in the fuse holder.

Fuse Rating

T 4A, 250V

68

APPENDIX

Error Messages
Test Errors
The following error messages or messages may appear on the GPT-9600 screen when configuring or running tests.

Error Messages I ERR SHORT V ERR R ERR TIME ERR
OVER 25W

Description
For ACW, DCW tests. Shown when the current is abnormal. Voltage is too low. Indicates that the DUT could be shorted. For ACW, DCW and IR tests. Indicates that the voltage is abnormal. For IR tests. Resistance=0. Check to see whether the DUT or test lead is shorting. For ACW tests. TIME ERR is displayed when the H setting is 15~20mA if the TIME setting is OFF. For DCW tests. OVER 25W is displayed if the H setting multiplied by the voltage setting is greater than 25W.

69

GPT-9600 Series User Manual

GPT-9600 Specifications
The specifications apply when the GPT-9600 is powered on for at least 30 minutes at 15°C~35°C.

Specifications

Environment Range Warranty Operation Storage Installation Location

Temperature

Humidity

15°C ~ 35°C

70% (No condensation)

0°C ~ 40°C

70% (No condensation)

-10°C ~ 70°C

85% (No condensation)

Indoors at an amplitude of up to 2000m.

AC Withstanding Voltage Output Voltage Range Output Voltage Resolution Maximum Rated Load Maximum Rated Current
Output Voltage Waveform Frequency Voltage Regulation Output Voltage Accuracy Voltmeter Accuracy Current Measurement Range Current Best Resolution
Current Measurement Accuracy Current Judgment Accuracy
Window Comparator Method ARC DETECT RAMP (Ramp Time) TIMER (Test Time) GND Judgment Valid Range
Continued…

0.10kV~ 5.00kV 10V 100 VA (5kV/20mA) 20mA 0.01mA ~ 5mA(0.1kVV0.5kV) 0.01mA ~ 20mA(0.5kV<V5kV) Sine wave 50 Hz / 60 Hz ± 1.5%+2 counts [Maximum rated load no load] ± (1.5% of setting +2 counts)with no load ± (1.5% of reading+ 2 counts) 0.01mA~20.0mA 10uA 0.01mA(0.01mA~9.99mA) 0.1mA(10.0mA~20.0mA) ± 2% of reading + 10 counts when I <1.00mA ± 2% of reading + 3 counts when I 1.00mA ± 3% of setting + 10 counts when I <1.00mA ± 3% of setting + 3 counts when I 1.00mA Yes Yes 0.1S fixed OFF, 1S~180S ON H/L Setting: 0.10mA to 9.99mA
10.0mA to 20.1mA

70

APPENDIX

Output Limitation for

Upper Current

Withstanding Voltage Testing 15mAI20mA

0.01mAI<15mA

Pause

Output Time

At least as long Maximum 180

as the output seconds

time

Not necessary Continuous

output

possible

DC Withstanding Voltage Output Voltage Range Output Voltage Resolution Maximum Rated Load Maximum Rated Current
Voltmeter Accuracy Output Voltage Accuracy Voltage Regulation
Current Measurement Range Current Best Resolution
Current Measurement Accuracy Current Judgment Accuracy
Window Comparator Method ARC DETECT RAMP (Ramp Time) TIMER (Test Time) GND Judgment Valid Range

0.10kV~ 6.00kV 10V 25W (5kV/5mA) 6mA 0.01mA ~ 2mA (0.1kVV0.5kV) 0.01mA ~ 6mA (0.5kV<V6kV) ±(1.5% of reading + 2 counts) ±(1.5% of setting + 2 counts) with no load ± 1.5% +2 counts
[Maximum rated load no load] 0.01mA~6.00mA 10uA 0.01mA(0.01mA~6.00mA) ± 2% of reading + 10 counts when I <1.00mA ± 2% of reading + 3 counts when I 1.00mA ± 3% of setting + 10 counts when I <1.00mA ± 3% of setting + 3 counts when I 1.00mA Yes Yes 0.1S fixed OFF, 1S~180S ON H/L Setting: 0.10mA to 6.00mA

Insulation Resistance Test Output Voltage Output Voltage Accuracy Resistance Measurement Range Test Voltage
50V/100V/250
500V/1000V
Voltmeter accuracy Continued…

50V, 100V, 250V, 500V, 1000V ±(3% of setting+1 count) with no load 1M ~ 2000M

Measurement Range Accuracy

1 ~ 50M

±(5% of reading + 2M)

51~2000M

±(10% of reading + 2M)

1 ~ 500M

±(5% of reading + 2M)

501~2000M

±(10% of reading + 2M)

±(3% of reading + 1 count)

71

GPT-9600 Series User Manual

Voltage Regulation
Window Comparator Method RAMP (Ramp Time) TIMER (Test Time) GND Output Impedance

±1.5% + 2 counts [Maximum rated load no load] Yes 0.1S fixed OFF, 1S~180S OFF 600k

Interface

REMOTE (Remote terminal) Yes

SIGNAL IO

Yes

General DISPLAY AUTO MODE POWER SOURCE
ACCESSORIES
DIMENSIONS & WEIGHT

240 x 48 dot matrix LED back light LCD AC-IR / IR-AC / DC-IR / IR-DC AC 100V- 120V/220V-240V 10% 50Hz/60Hz Power cord x1, Quick Start Guide x1 User Manual x1 (CD) GHT-114×1 Approx. 330(W) x 148(H) x 385(D) mm (Max.), 9kg(Max)

72

GPT-9600 Dimensions

APPENDIX

GPT-9603

AC / DC Withstanding Voltage / Insulation Resistance

Tester

PASS ESC

FAIL FIELD

READY

TEST

CAUTION HIGH VOLTAGE

5.0 kVAC MAX. 6.0 kVDC MAX.

POWER

START

STOP

REMOTE

EDIT

SAVE

UTILITY

RETURN

322.0

134.0 148.2

328.9 385.0

73

GPT-9600 Series User Manual

Declaration of Conformity

We GOOD WILL INSTRUMENT CO., LTD. No. 7-1, Jhongsing Rd, Tucheng Dist., New Taipei City 236, Taiwan

GOOD WILL INSTRUMENT (SUZHOU) CO., LTD. No. 69 Lushan Road, Suzhou New District Jiangsu, China.

declare that the below mentioned product

Type of Product: Electrical Safety Tester

Model Number: GPT-9603, GPT-9602, GPT-9612, GPT-9601

are herewith confirmed to comply with the requirements set out in the

Council Directive on the Approximation of the Law of Member States

relating to Electromagnetic Compatibility (2004/108/EC) and Low Voltage

Directive (2006/95/EC).

For the evaluation regarding the Electromagnetic Compatibility and Low

Voltage Directive, the following standards were applied:

EMC

EN 61326-1

Electrical equipment for measurement, control and

EN 61326-2-1

laboratory use ­­ EMC requirements (2013)

Conducted Emission

Electrostatic Discharge

Radiated Emission

EN 61000-4-2: 2009

EN55011: 2009+A1: 2010

Current Harmonics

Radiated Immunity

EN 61000-3-2: 2006+A1:2009

EN 61000-4-3: 2006 +A1:2008

+A2:2009

+A2:2010

Voltage Fluctuations

Electrical Fast Transients

EN 61000-3-3: 2008

EN 61000-4-4: 2012

————————-

Surge Immunity

EN 61000-4-5: 2006

————————-

Conducted Susceptibility

EN 61000-4-6: 2009

————————-

Power Frequency Magnetic Field

EN 61000-4-8: 2010

————————-

Voltage Dip/ Interruption

EN 61000-4-11: 2004

Low Voltage Equipment Directive 2006/95/EC

Safety Requirements

EN 61010-1: 2010

EN 61010-2-030: 2010

74

INDEX
Accessories …………………………… 10 Auto mode
test results………………………………..52 Auto mode testing ………………… 49 Caution symbol ………………………. 4 Cleaning the instrument …………. 6 Declaration of conformity……… 74 Dimensions…………………………… 73 Disposal instructions ………………. 6 EN61010
measurement category……………….5 pollution degree…………………………6 Environment safety instruction ……………………….6 External control…………………….. 59 Interlock key…………………………….65 overview ………………………………….60 remote operation ……………………..61 remote terminal ……………………….60 signal I/O operation ………………..64 signal I/O overview …………………62 FAQ ……………………………………… 66 Front panel diagram……………… 12 Ground symbol……………………………………….4 Interlock key…………………………. 65 Line voltage selection ……………. 18 List of features………………………… 9 Marketing contact ……………………………………..67 Menu tree……………………………… 25 Operating precautions ………….. 21 Overview ……………………………….. 9 Power on/off safety instruction ……………………….5 Rear panel diagram ………………. 16 Service operation

INDEX
about disassembly…………………….. 5 contact ……………………………………. 67 Setting the test voltage ………….. 42 Specifications ………………………… 70 Starting a test ………………………… 42 Test errors …………………………….. 69 Test results ……………………………. 45 Auto mode ……………………………… 52 Auto mode viewing ………………… 52 fail judgment ………………………….. 47 pass judgment ………………………… 45 timing diagrams ……………………… 46 Testing ARC mode ……………………………… 39 auto mode testing …………………… 49 ground mode
ACW and DCW ………………………… 30 IR mode ……………………………………. 31
overview ………………………………… 33 running a test………………………….. 42 saving …………………………………….. 41 terminal connections……………….. 29 test function ……………………………. 34 test limits………………………………… 35 test settings …………………………….. 34 test time ………………………………….. 37
cool down …………………………………. 38
test voltage settings…………………. 42 UK power cord………………………..7 Utility settings
Control settings ………………………. 56 double action ………………………….. 56 key lock ………………………………….. 56 start control…………………………….. 56 Utility settings LCD………………………………………… 55 Warning symbol………………………4 Workplace precautions …………. 20

75

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals