Tektronix 6013 Digital Capacitance Meter Instruction Manual
- June 3, 2024
- Tektronix
Table of Contents
Tektronix 6013 Digital Capacitance Meter
FEATURES
- Easy and correct readout.
- High measuring accuracy.
- Measurements are possible even under a strong magnetic field.
- LSI-circuit provides high reliability and durability.
- Input overload protection is provided.
- LCD display for low power consumption and clear readout even in bright ambient light conditions.
- In-line pushbuttons allow one hand operation.
- Light-weight and compact construction for easy operation.
- Low battery condition is indicated on the LCD display.
SPECIFICATIONS
GENERAL SPECIFICATIONS
- Display: LCD (Liquid Crystal Display) Max. display 1999.
- Range: 9 range, whole range value (from 0.1pF to 20mF)
- Overload protection: Display “1”.
- Calibrate Adjustment: Have two internal adjustment. One is panel Zero adjustment.
- Zero Adjustment: Manual (range:±20pF)
- Over-range dispaly: Display“1”.
- Sampling Time: 0~5second
- Operating Temp: 0℃ to 40℃.
- Operating Humidity: 80% MAX.R.H.
- Battery: 9V (6F22 or the equivalent
- Battery Life: alkaline battery: approx.200 hours. Zinc-Carbon battery approx. 100 hours
- Typical consumption current: 3~4mA (RANGE: 200pF-200uF)
- Standard Accessories:
- Test alligator clips (red & black): 1 pair.
- Instruction manual 1 pc.
ELECTRICAL SPECIFICATION
- Temperature range for the accuracy guaranteed: 23℃±5℃
- Accuracy: range at 200pF~200μF: ±0.5%+1 digit; range at 2000μF: ±2%+1 digit; range at 20mF: ±2%+2 digit
- Excite voltage: 2.8V peak value
- Overload protection: Protected by a 0.2A/250V self- fused. The capacitance will destroy the meter if the DCV is higher than 60V.
- Zero adjust: limit to 20pF
OPERATION
Pannel description: (see the fig.1)
- LCD: display the measuring value, unit and low battery indication,
- POWER switch: turn on/off the power.
- Zero adjust knob: zero adjust knob before the measuring of lower than 20nF range
- Function knob: to select measuring range.
- Input COM: Capacity(Cx)COM
- Battery case
Notice before measuring:
- Observe polarity when connecting polarized capacitors.
- Discharge completely for any capacitance;
- Do not connect any COM with voltage source absolutely, otherwise it will cause damage badly.
- Do not close (black & red) test leads.
- Before removing the battery and fuse compartment cover, ensure that the instrument is disconnected with any circuit and the power switch is in the off position.
- zero adjust won’t work when using the external components alligators if the capacitance is beyond 20nF.
CAPACITANCE(C) MEASURING PROCEDURE
- Press POWER key, turn on the power.
- Select the range switch for the maximum expected capacitance.
- Check “0” indication: If test range is 200pF, 2nF, 20nF, should check “0” indication before test.
- Observe polarity when connecting polarized capacitors.
- Full discharge any capacitors.
- Connect the alligator clips to the capacitors leads.
- Read the display. The value is direct reading in the electrical unit (pF, nF, uF,mF) indicated at the selected range switch. If display show “1”, It indicate on Out-of-Range measurement. If the display indicates one or more leading zeros, shift to the next lower range scale to improve the resolution of the measurement.
NOTE:
- If the capacitance value is unmarked, start with the 200pF range and keep increasing until the over-range indication goes off and a reading is obtained.
- A shorted capacitor will read over-range on all ranges. A capacitance with low voltage leakage will read over range, or a much higher value than normal. An open capacitor will read zero on all ranges.
- Measure of very low capacitance should be performed using extremely short leads in order to avoid introducing any stray inductance.
- When using the optioned test leads, remember that the leads introduce a measurable capacitance to the measurement. As a first approximation, the test lead capacitance may be measured by opening the leads at the trips, recording the open circuit value and subtracting that value.
- Capacitors, especially electrolytic, often have notoriously wide tolerances. Do not be surprised if the measured value is greater than the value marked on the capacitor, unless it is a close tolerance type. However, value is seldom drastically below the rated value.
- If changing range, measured value will be changed; leakage-voltage capacitors will be checked also. Leakage-resistance will be decreased in lower range.
AUTO POWER OFF
- The meter will be into sleeping mode when it stopped work for 20minutes.
- Please restart the power if you need to continue to measure.
MAINTENANCE
Do not try to modify the electric circuit.
- Keep the meter away from water, dust and shock.
- Do not store and operate the meter under the condition of high temperature, high humidity, combustible, explosive and strong magnetic place.
- Wipe the case with a damp cloth and detergent, do not use abrasives and alcohol.
- If do not operate for a long time, should take out the battery to avoid leakage
- When signal displays, should replace the battery following the steps:
- Take off the holster. (see fig. 2)
- Unlock the screw and remove the battery case.
- Take out the old battery and replace the new one. It’s better to use alkaline battery for longer life
- Fit on the battery case and lock the screw.
- Fit on the holster as the opposite way (see fig.2)
* The specifications are subject to change without totice.
* The content of this manual is regarded as correct, error or omits. Please contact with factory.
* We herby will not be responsible for the accident and damage caused by improper operation.
* The function stated for this User Manual cannot be the reason of special usage.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>