WINNCARE MANGAR Camel Lifting Cushion Instruction Manual
- June 3, 2024
- WINNCARE
Table of Contents
WINNCARE MANGAR Camel Lifting Cushion
Please keep these instructions in a safe location for future reference
WARNINGS
- This equipment MUST only be used by persons familiar in its use.
- The medical state of a person needing to be raised MUST be assessed before commencing a lift.
- Ensure that the surface on which the Camel is to be placed is free from sharp objects. NOTE: Place a blanket, groundsheet, towel or similar under the Camel if using it on a rough surface.
- If a slide sheet has been used to transfer the patient onto the Camel it must be removed from under the patient before commencing a lift.
- Sections 2, 3 and 4 MUST always be fully inflated in sequence from the bottom up, in order to maintain stability. NOTE: The backrest may be partially inflated before any other section is raised. Once the lowest section is fully inflated, the backrest can be raised further as required.
- Do not operate the Airflo 24 compressor in the presence of flammable anaesthetic mixtures with air, oxygen or nitrous oxide, as these mixtures can create a combustible environment leading to the potential for fire or even an explosion in the presence of a spark.
Intended Use
The Mangar Camel is an inflatable lifting device which is designed to sit up
and lift a fallen person from the floor. When inflated, the Camel lifts the
patient into a raised seated position ready to stand with or without
assistance.
This equipment may be used with or without a helper as required.
The Camel is powered by a Mangar Airflo 24 Compressor and controlled using a
4-Way Hand Control.
Patented
PLEASE REFER TO THE SEPARATE AIRFLO 24 COMPRESSOR INSTRUCTIONS BEFORE OPERATING.
Standard Product Contents – fig. 1
- 1 x Camel (A) (including carrying bag).
- 1 x 4-Way Hand Control (B).
- 1 x Airflo 24 Compressor (C) with user instructions.
Connecting the Hand Control and Airflo 24 – fig. 2
- Remove the Camel from the carrying bag, unfasten the straps and unroll.
- Connect the four hoses from the hand control to the corresponding numbered and coloured connectors on the Camel (A).
- Connect the hose (B) from the hand control to the Airflo 24 compressor.
- Position the Airflo 24 compressor so that it will not be in the way while positioning the patient. Ensure that the hoses are not kinked.
Switching the Airflo 24 ‘On’ & ‘Off’ – fig. 3
- To switch on, momentarily press the ‘Power’ button (A fig.3). The bottom battery status indicator light (B) will come on.
- To switch off, momentarily press the ‘Power’ button. The bottom battery status indicator light will go out.
Positioning the patient on the Camel – fig. 4
- Position the Camel at the side of the patient so that the patient’s head is level with the Camel badge (A). NOTE: Place a blanket, groundsheet, towel or similar under the Camel if using it on a rough surface.
- Transfer the patient onto the Camel. The head should lie on the Camel badge (A) and the body should lie centrally.
Operating
The following instructions apply equally when the Camel is being operated by a
person raising themselves without assistance, or by an assistant raising a
patient.
WARNING: If a slide sheet has been used to transfer the patient onto the
Camel it must be removed from under the patient before commencing a lift.
- Switch the Airflo 24 to ‘On’. Press the number 1 button on the hand control to partially inflate the backrest section until the patient is in a comfortable position – fig. 5
- Press the number 2 button to inflate the bottom section until it becomes rigid – fig. 6
- Continue to press buttons 3 and 4 in sequence until the sections become rigid – fig. 7
- Press button 1 to fully raise the backrest and arrive at a sitting position – fig. 8
- • Whether assisted or used alone, the patient should now be in a position to stand or transfer – fig. 9
WARNING: In order to maintain stability, sections must always be fully
inflated from the bottom up, in sequence.
Note: only the backrest (section 1) may be partially inflated before the
lower sections.
After Use
- To deflate, disconnect each of the air hoses from the Camel. NOTE: The initial ‘rush’ of air as the hoses are disconnected from the Camel, can be reduced by first disconnecting the air hose from the Airflo 24 and then pressing each of the buttons on the hand control.
- Roll up the Camel with the hand control in the hollow (A) – fig. 10 secure the straps and place in the carrying bag – fig. 11
Routine Cleaning
Fully inflate all sections of the Camel, sponge clean with a non abrasive
proprietary liquid cleaner or disinfectant and dry thoroughly. Do not dry
using hot air.
Disinfecting
The Camel should be fully inflated and pre-cleaned by washing with detergent
and water, thoroughly rinsed and then disinfected using a 1:10 solution of
household liquid bleach and water (1 part bleach to 10 parts cold water).
Leave the solution on the Camel for approximately 10 minutes before thoroughly
rinsing off with clean cold water. Allow to air dry before storage. Do not dry
using hot air.
Maintenance
Other than routine cleaning, the Camel is maintenance free.
Winncare recommends that your Camel is inspected annually by either your
supplier or other suitably qualified person.
If required, Winncare can provide the above inspection within an annual
service contract
(UK only). Please contact our Customer Service Department for further details
on Freephone 0800 2800 485.
Transportation / Storage
Disconnect the Airflo 24 compressor from the Camel before transporting so as
to avoid accidental inflation of the cushion.
Ensure that the Camel is clean and thoroughly dry before storing. Failure to
thoroughly dry the cushion before storage may result in a build-up of mould on
the surface of the product leading to a weakening of the fabric and
consequential invalidation of the product warranty. Store in a dry
environment.
To maintain the Airflo 24 battery when being stored for long periods, put on
charge for 12 hours every 3 months.
Disposal / Recycling
When the Camel has reached the end of its working life, dispose of it at your
local recycling centre (civic amenity site).
When the Airflo 24 has reached the end of its working life, remove the battery
for safe disposal in accordance with your local council guidelines, and
dispose of the Airflo 24 at your local recycling centre (civic amenity site).
Troubleshooting
Problem| Remedy
Compressor does not run when one or more of the buttons are pressed on the
hand control.| Ensure that all the air hoses from the hand control are
connected to the Camel and Airflo
Compressor keeps stopping when trying to inflate the Camel.|
Check that there are no kinks in the air hoses.
Bottom Battery Status Indicator light flashing ‘Red’.| This happens when the compressor’s internal temperature cut-out has tripped. Leave the Airflo 24 for 60-90 minutes to cool.
Compressor causes electrical interference with another piece of electrical equipment.
| Move either the compressor or other item further away from each other until
the interference stops.
For further assistance with resolving a problem please contact our Customer
Service Department by calling +44(0)1544 267674 or email
sales@winncare.uk
Specifications
- Maximum user weight: 320kg (705lbs)
- Seat height (approx): 56cm (22”)
- Overall length (approx): 112cm (44”)
- Overall width (approx): 70cm (27½”)
- Deflated & rolled (in bag):
- Length : 76cm (30”)
- Diameter: 20cm (8”)
- Weight (Inc. Hand Control): 6.5kg (14.3lbs)
- Maximum air pressure: 0.35bar (5psi)
- Max / Min temperature: +50°C / -15°C
Warranty
The warranty is given against defects in material and workmanship and
comprises piece, parts and labour. The warranty is not valid in the case of
fair wear and tear, wilful damage, common neglect, misuse or unauthorised
alteration or repair by unqualified persons.
The following statements do not affect the consumer’s statutory rights. This
product is sold on the understanding that in the event of any defect of
manufacture or material appearing within the warranty period, the defect will
be rectified free of charge providing that:
- (a) Reasonable evidence is provided (e.g. purchase invoice, delivery note) that the claim is being made within the warranty period.
- (b) The defect is not attributable to accidental damage (either in transit or otherwise), misuse or unauthorised repair.
Winncare is a trading name of Mangar International Ltd.
Mangar International Ltd, Presteigne, Powys, Wales, LD8 2UF
United Kingdom
Tel: +44 (0)1544 267674
Fax: +44 (0)1544 260287
enquiries@winncare.uk
mangarhealth.com
References
- Home Page - Mangar UK | Safe Lifting Products
- Groupe Winncare: Fabricant de lits médicalisés, matelas et coussins thérapeutiques, lèves-personnes et mobilier pour collectivités - Winncare France
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>