WINNCARE GRÚA People Hoists User Manual
- June 12, 2024
- WINNCARE
Table of Contents
- GRÚA People Hoists
- INTRODUCTION
- ASSEMBLY
- INITIAL CHECKS
- ELECTRICAL EQUIPMENT
- SAFETY TIPS
- USE AND MAINTENANCE RECOMMENDATIONS.
- ELECTRICAL INFORMATION:
- PRODUCT LIFE EXPECTANCY
- CLEANING:
- FAULTS AND REMEDIES
- GUARANTEE
- ANNUAL INSPECTION OF PATIENT HOISTS
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
GRÚA People Hoists
User Manual Manual and Warranty
102102043 -102102044
UDI: 8436571820426 – UDI: 8436571820433
GRÚA People Hoists
DATOS
FABRICANTE
WINNCARE SLU
B-96221718
Tfno. (+34) 96 156 55 21
Ctra Masía del Juez 37b
46909 – Torrent, Valencia (Spain)
A. Leg
B. Column
C. Electric actuator
D. Base opening pedal
E. Control box + batteries
F. Handset| G. Hoist arm
H. Screw
I. Hanger
J. Front wheel without brake
K. Rear wheel with brake
L. Base
---|---
INTRODUCTION
Please read these operating instructions carefully before using your hoist:
contains useful information on safety, handling and maintenance.
Winncare has wide experience in manufacturing hoists acquired throughout the
years.
The constant improvements introduced have bolstered our development as a
Technical Aids manufacturing company, which is clearly reflected in our wide
range of models and applications.
Winncare hoists comply to the New Regulation (EU) 2017/745 for Medical Devices
and so carry the CE mark. They have been tested at the Bio-mechanic institute
of Valencia (IBV) and fulfil all the requirements laid down in standard EN ISO
10535: “Hoists for the transfer of disabled persons. Requirements and test
methods”.
Winncare has a range of hoists with wheels from 130 to 185 Kg, all of them in
electric versions.
These operating instructions have been written in accordance with the current
designs of the products. All of them are constantly subject to a process of
continuous improvement Therefore, we reserve the right to modify any technical
data that appears in this operating instructions without prior warning.
All the data, figures and measurements included in this operating instructions
are approximate and must not be considered as strict technical specifications.
ASSEMBLY
Every hoist is fully assembled in the factory to check that it works properly
and, depend ing on the version, is dismantled or folded for packaging and
transport.
Winncare hoists are designed to facilitate assembly without the use of any
tool.
Remove all the parts of the hoist from the packaging (some of them are heavy
and you may need someone else to help you).
Put the break of the rear wheels on and spread the legs to increase the
stability of the structure.
1. Check that the handle located on the central tube of the hoist’s column is suffi ciently loose to allow the base to be fitted.| 2. lnsert the column into the base.| 3. Tighten up the handle.
INITIAL CHECKS
Before using your hoist, check that:
- The legs of the hoist open and close correctly. It must provide a certain amount of resistance to prevent them from opening or closing accidentally.
- The wheels turn and revolve normally.
- The brakes on the rear wheels work properly and the column is fully attached in its usage position.
- The hanger turns and is adequately balanced.
- There is no wear or deformation of the hanger hooks.
- The slings are not damaged or frayed.
ELECTRICAL EQUIPMENT
- Check that the connection of the actuator, the handset and the charger are firmly connected to the control box.
- Press the up and down buttons (and those opening and closing the legs) to check that the actuator operates correctly.
- Check that the red emergency stop button is working correctly; if you press it, with the hoist moving up or down, it must stop.
- Before using the hoist for the first time, it is advisable to charge the batteries for at least 5 hours.
SAFETY TIPS
Familiarise yourself with the hoist’s devices and how they work before using
it for real.
Your hoist is designed for lifting patients, not for any other purpose.
Check that the weight of the patient does not exceed the maximum weight that
the hoist can lift or the sling bear.
Check that the end or the actuator is fitted between the plate on the hoist
arm and has its pin correctly inserted. Do not force the controls and devices
of the hoist. All of them are easy to use and do not require excessive force
to be applied.
Manoeuvre the hoist by pushing on the handles and never push the column, the
hoist arm or the patient. Move the patient with the legs of the hoist in
closed position. The hoists must be handled gently when moving patients,
maintaining a speed movement appropriate to the situation. Move the hoist on
flat and straight floors. Their use on surfaces with a slope greater than 5
degrees is not recommended; should you have to pass through a sloping area, we
advise you to have a second person to help you.
Never use an electric hoist in a shower. Do not charge the batteries of a
hoist in a bath or shower.
Never leave a patient alone in a hoist.
Winncare hoists have been designed and manufactured for use with Winncare
slings and accessories; the slings and accessories of other manufacturers have
not been tested by Winncare and so their use cannot be recommended. Do not use
a frayed or worn sling end check that the size and characteristics are suited
to the patient.
Adjust the sling in accordance with the sling usage instructions. Before
lifting, at ways make sure that the belts or the sling are correctly
positioned at the end of the hanger; this must be checked when the belts are
taut, but before beginning to lift the patient.
Before lifting the patient, be sure that the strap of the harness are
correctly positioned in the hanger. It must be checked when the straps are
taut but the patient has not been lifted yet.
USE AND MAINTENANCE RECOMMENDATIONS.
-
The hoist has swivel wheels; the rear wheels have brakes that stop them from both turning and swivelling. The brake is applied and released with the foot by means of a lever.
We recommend not putting the brake on the wheels when the user is being lifted so that the hoist can move towards the centre of gravity when the patient is lifted from a chair, bed or any other stationary object, whilst at the same time holding it by the handles. -
The opening of the base, in order to facilitate, for example, the entrance of the wheel-chair, is done by pressing with the foot on the appropriate side of the pedal.
It is advisable to lubricate all the parts of the hoist where movement takes
place. This will improve how it operates and prevent possible wear and tear of
the parts through daily use.
Prevent dirt accumulating on the axle and the hall bearings of the wheels.
Daily check that the screws, nuts, etc… are tight and check for slack between
the different components of the hoist. It is recommended an annual review
following the parameters of Annex .
It is not advisable to leave the hoist in places with a high degree of
humidity (near the sea, indoor swimming pools, etc…) or in corrosive
atmospheres that may affect their components.
The electrical equipment has an emergency lowering system. Depending on the
type of crane, it can be manually or electrically. In the case of the manual
system (nº1), is activated by turning the mechanism. In the case of the
electrical system is activated by pressing the lower button on the control box
(nº2).
It also has two safety systems (electrical and mechanical) that disengages the
motor if, in the downward movement of the hoist, it comes up against an
obstacle that hinders it from descending (parts of the user’s body,
furniture…).
The electrical equipment has an emergency stop button (nº 3) that disconnects
the motor’s battery automatically halting the movement of the hoist. To
disengage the stop button, it must be turned. When the hoist is NOT in use,
press the emergency stop button to prevent the batteries from running down.
Keep the electrical equipment away from water and other liquids to protect it
from possible splashes that may affect its internal circuits causing serious
damage.
The battery housing has several ventilation holes (nº4) that ensure the
necessary and correct air circulation within this space. This ventilation must
not be blocked or covered.
When undertaking the raise-lower movement of the hoist, keep the appropriate
button pressed in, avoiding pressing the buttons repeatedly, as repeated
start-ups of the motor uses up more battery, discharging it more quickly.
The batteries must be replaced after 4 years and perhaps before, depending on
the type of use. Heavy and frequent unloading reduces the life of the battery.
When the low battery beep is heard. In this case will be a maximum of 8 cycles
of elevation and should be put to charge immediately.
For the optimum life of the battery, recharge it as often as possible.
Recharging time is around 5 to 8 hours. It is convenient to charge the battery
every 20 transfers.
While the battery is being charged, the hoist cannot be used.
Batteries of equipment not habitually used must be recharged at least every 3
months; otherwise they could be damaged as a consequence of their own self
discharge. To charge, you must check that the connections are correctly
attached (nº1) and that the emergency stop button is not pressed in (nº2),
otherwise the batteries will not be charged in an appropriate way.
Remember: the battery is the only component of the electrical equipment that
requires continuous maintenance. If you do not follow the advice above, the
battery may discharge to a level that does not allow it to recover its full
charge or may even render it useless.
Nº 3: Blink: Battery charging
Continuous light: Charged battery
ELECTRICAL INFORMATION:
For all its electrical hoist equipment, Winncare uses the Linak company, a
world leader in manufacturing this type of equipment.
The electrical equipment complies with the following standards:
IEC 60601-1 Medical electrical equipment. Part 1: General requirements for
basic safety and essential performance.
IEC 60601-1-2 Medical electrical equipment. Part 1-2: General requirements for
safety collateral standard: electromagnetic compatibility. Requirements and
tests.
The maximum work cycle of the electrical equipment is 10% or 2 minutes of
continuous use followed by 18 minutes rest.
If this work cycle is exceeded there is a risk that the equipment will
overheat and cause a breakdown. Operational environment of the electrical
equipment: from +5° C to +40° C
Actuator 24V continuous current.
Batteries 2 x 12V / 2.9Ah sealed rechargeable batteries of the lead-acid type.
Noise level: (<50dB)
IP EQUIPMENT AND IP PROTECTION| HOIST ACTUATOR| CONTROL BOX|
HANDSET STEUERUNG ODER TASTER
---|---|---|---
Ref. 102102043/UDI: 8436571820426- Básica E130| LA31 IP54| CBJH IPX4| HB71
IPX6
Ref. 102102044/UDI: 8436571820433 – Hop! E150| LA31 IP54| CBJH IPX4| HB71 IPX6
PRODUCT LIFE EXPECTANCY
All Winncare hoists have been tested according to EN ISO 10535, and have a
minimum lifespan of 6 years. In order to prolong the useful life of the hoist,
it is necessary to continue with the annual revisions by our official
technical service.
The following table indicates the estimated life of the motor / actuator
depending on the number of elevations per day.
Elevation X day | Years life |
---|---|
30 | 1 |
15 | 2 |
10 | 3 |
7-8 | 4 |
6 | 5 |
5 | 6 |
4-5 | 7 |
3-4 | 8 |
2-3 | 9 |
1-2 | 10 |
This table can be indicative depending on the use, load and stroke .
The LINAK specifications mentioned can be found on the actuator label, or if
it is not there, they can be consulted in the corresponding data sheet at
www.linak.com
CHANGING BATTERIES
CLEANING:
The hoist must be cleaned using a damp cloth with a product used for cleaning
furniture or similar that does not contain solvents that could damage the
coating of the hoist.
Do not allow the electrical elements of the hoist to come into contact with
water. To disinfect , apply steam throughout the structure .
DISPOSAL METHOD:
The hoist is manufactured with Steel profiles with an epoxy coating.
Therefore, to dispose of it you must follow recycling regulations or local
recommendations on disposal.
Rechargeable batteries containing lead (Pb) must be recycled so they should be
disposed of in containers destined for recycling cars batteries or at the
appropriate waste management sites.
FAULTS AND REMEDIES
SYMPTOM | CAUSE | SOLUTION |
---|---|---|
The lift arm, with the hoist unloaded, sometimes does not lower. | • The lift | |
arm needs a minimum weight in order to descend. | • Lightly push the lift arm |
downwards at the same time as pressing the down button.
The moving parts of the hoist do not function smoothly.| • Lack of
lubrication.| • Lubricate moving parts.
The actuator does not work but a “click” can be heard inside the control box
when pressing the handset.| 1.- Actuator plug badly connected to the control
box.
2.- Fit batteries.
3.- Damaged actuator cable.
4.- Actuator or control box broken down.| 1.- Connect the actuator correctly.
2.- Charge batteries.
3.- Send actuator to be repaired.
4.- Send equipment to be serviced.
The actuator does not work and there is no “click” sound in the control box
when pressing the handset.| 1.- Red emergency button pressed in.
2.- Equipment with removable battery that has not made a good contact.
3.- Bad connection of the electrical components.
4.- Discharged batteries.
5.- Handset cable damaged.
6.- Damaged equipment.| 1.- Turn it to release it.
2.- Put the battery correctly.
3.- Check the connections.
4.- Charge the batteries
5.- Replace the handset.
6.- Send equipment to be serviced.
The actuator stops.| 1.- Discharged batteries (acoustic warning sounds on
pressing the handset)
2.- The weight that the hoist can lift has been exceeded.| 1.- Charge
batteries.
2.- Reduce the weight.
The actuator only works in one direction.| 1.- Breakdown in handset.
2.- Breakdown in control box.| 1.- Replace handset.
2.- Send equipment to be serviced.
The batteries do not charge.| 1.- Bad connection.
2.- Breakdown in charge circuit.| 1.- Check connections.
2.- Send equipment to be serviced.
If the problems cannot be solved through these suggested recommendations,
contact your supplier.
Technical support service
sat@winncare.es
WINNCARE
Carretera Masía del Juez, 37
46909 Torrent Valencia (Spain)
GUARANTEE
All hoists have been rigorously inspected; on their labels there is a lot and
serial number that lists that hoist with the materials, equipment, and
components employed and the work processes undertaken for their manufacturing,
assembly and inspection.
The Winncare hoists have a 5 year warranty and 1 year for the battery, under
normal usage conditions, respecting the recommendation on use and maintenance
and not changing or manipulating any of their components.
Any manipulation or alteration, especially of the electrical equipment,
renders this guarantee null and void. In the event of some defective or
malfunctioning of the hoist or any of its components, please, contact the
distributor where you purchased it, presenting it together with the invoice or
purchase receipt, so that it may be inspected and according with the
manufacturer, decide whether to repair it or replace it. Winncare does not
accept responsibility for imperfections caused by bad usage or as a
consequence of not having followed the recommendations given in these
operating instructions.
According with Council Directive 85/374 / EEC of July 25, 1985, on the
approximation of the legal, regulatory and administrative provisions of the
Member States regarding liability for damage caused by defective products, the
distributor and/or WINNCARE
will not be the damage caused by the defect that did not exist at the time the
product was put into circulation or that this defect appeared later.
Max.→
Cm.
↓
| 102102043
Basic E130| 102102044
Hop! E150
---|---|---
130 Kg.| 150 Kg.
A| 198| 180.5
B| 137.5| 110.5
C| 25.5| 25.5
D| 46| 46
E| 11.5| 11
F| 5| 4.5
G| 10| 7.5
H| 7.5| 7.5
I| 173.5| 157
J| 113| 84
K| 51.5| 39
L| 110| 99.5
M| 133| 117
N| 80| 77
O| 65.5| 56.5
P| 53.5| 44
Q| 147| 130
Kg | 38.1| 29.5
Kg | 16.1| 13.8
Kg | 22| 15.7
Sonorous Pressure| <65 Db
---|---
Ergonomic requirements test| Handset| 0.8 N
Base opening pedal| 240 N
Starting force| 150 N
Turning force| 155 N
Driving force| 83 N
| 102102043/ UDI: 8436571820426
Basic E130| 102102044/ UDI: 8436571820433
Hop! E150
---|---|---
A1| 74322521| 601011003
A2| 6303029390| 601238001
A3| 74500004| 601010003
B1| 605010019| 605010019
B2| 605012022| 605012022
B3| 6300000023| 6300000023
B4/ C1| 82000011| 82000011
B5| 61300062| 61300062
B6| 605018002| 605018002
C2| 81000028| 81000028
C3| 82000012| 82000012
Hanger cover| L-16655| L-16655
Complete hanger| SEL-16525| SEL-16525
ANNUAL INSPECTION OF PATIENT HOISTS
IDENTIFICATION OF THE MEDICAL EQUIPMENT | ETABLISSEMENT |
---|---|
CATEGORY: |
BRAND / MODEL / TYPE:
SERIAL NUMBER:| INVENTORY NUMBER:| DEPARTMENT/LOCATION:
MANUFACTURING DATE:
VISUAL INSPECTION (According to EN ISO 10535)| NA| C| NC
---|---|---|---
Availability and good condition of the instruction manual, loading plate,
identification plate/labelling/printing.| | |
Good general condition (structure, base, articulations, welds, attachments,
etc…) Absence of rusting.| | |
Good general cleanliness.| | |
Good stability of the hoist and good operation of the base opening/closing.| |
|
Checking of the tightness of screws (suspension pin of the lifting arm,
weighing module attachement, etc..) absence of space.| | |
Good condition of slings and cadle or hammock, including the attachment
system.| | |
Good rotation of the gripping bar and good condition of the hooks.| | |
Good operation of the wheels and brakes (swivelling, running, locking).| | |
Good lubrication of articulations (absence of noise, nuisance, squeaking and
hard points).| | |
Good condition of electric cables, control boxes, connectors and actuators
(jacks, motors…)| | |
Good condition of controls and indicator lights.| | |
SAFETY INSPECTION (According to NF EN ISO 10535)| NA| C|
NC
---|---|---|---
Looking out of operational functions and correct operation of visual and
audible alarms.| | |
Emergency stop operates correctly.| | |
Battery and charger operate correctly.| | |
Obstacle stop operates correctly.| | |
Correct operation of the emergency descent.| | |
Test at maximum load (1 complete cycle).| | |
INSPECTION OF THE AMPLITUDE OF MOVEMENTS| NA| C| NC
---|---|---|---
Satisfactory operation of the actuator (elevation/descent)| | |
Maximum and minimum height (according to the manufacturer specifications)| | |
ELECTRICAL SAFETY (According to EN 60601-1)
Continuity:| Tester model:| Serial number:
Breakdown:| |
CONCLUSION| NA| C| NC
---|---|---|---
Operational: Patient safety, care and technical staff will be in danger?| | |
Action to provide:
Recommended for the next QC Date:
OPERATOR
Name:| Date:| Signature:
Symbols
**** | Read the instructions |
---|---|
Warning | |
Manufacturer information | |
Manufacturing date | |
Do not throw in trash bins | |
Use indoors | |
Do not use if the packaging is broken | |
Class 2 equipment | |
Equipment type B according to EN 60601-1 |
We take care of people with loss of autonomy and their caregivers
Carretera Masía del Juez 37b,
46909, Torrent, Valencia
Tel. 96 156 55 21
Fax. 96 156 03 83
sat@winncare.es
www.winncare.es
References
- Winncare Spain : Fabricante de camas hospitalarias, tratamiento y prevención de úlceras por presión: colchones y cojines terapéuticos, grúas eléctricas y mobiliario para colectividades - Winncare Spain
- LINAK - electric linear actuator technology systems for movement
- Winncare Spain : Fabricante de camas hospitalarias, tratamiento y prevención de úlceras por presión: colchones y cojines terapéuticos, grúas eléctricas y mobiliario para colectividades - Winncare Spain
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>