Brushless DC Motor Centrifuge
Machine iFUGE MlSK
PRODUCT USER MANUAL
INTRODUCTION
This manual provides important safety information for this centrifuge. lt should be kept near the centrifuge for quick & easy reference. This centrifuge is equipped with maintenance free brushless DC motor drive with microprocessor controlled and a safety feature of imbalance detection. The centrifuge is designed for high speed spin downs applications. Multiple rotor option makes it usable for wide range of applications.
INTENDED USE
This centrifuge is designed for separation of solid suspensions in liquids such as cells, macromolecules such as DNA, RNA & proteins. This centrifuge can be used in various research laboratories & institutions.
FEATURES
- Maximum volume of 12x 2 ml
- lmbalance detector with auto cutofffor safe operation
- Brushless DC motor for maintenance free performance
- Lid lock safety-the lid opens automatically on run completion
- Digital display with speed and time indicator
- RPM setting from 500-15000 RPM with a count of 100
- One press changeover from RPM to RCF functionality
- Reduction adaptors for 0.1/0.2 & 0.5ml tu bes
- Emergency lid release du ring power failure
- Automatie internal diagnosis & error display
NOTE: Before using the instrument, please read this user manual carefully. This user manual is intended to assist with the operation and care of the unit only and not its repair. For repair p/ease contact the supplier.
TECHNICAL SPECIFICATION
Motor Type | Brushless DC Motor |
---|---|
Maximum Volume | 12 x 2 ml (microtubes) |
Speed Setting | 500 to 15000 RPM |
Maximum RCF | 15596 xg |
Speed Accuracy | ± 100 rpm |
Run Time | 1 to 99 minutes & infinite mode |
Acceleration time | 60 ± 2 seconds |
Deceleration time | 50 ± 2 seconds |
Size (LX w X H) | 190 x 120 x 270 mm |
TECHNICAL SPECIFICATION
Weight | 2.450 Kg approx {with rotor) |
---|---|
Input Voltage | 24V ****2.3A |
Power Consumption | 55W |
STANDARD ACCESSORIES
- Power Adaptor: 100-240 VAC, 50/60 Hz, lAmp (Input), 24 V= 4A (Output)
- 12 – slot rotor (Pre-installed) with reduction adaptors for 0.1/0.2 & 0.5ml tu bes
- T-Allen Key
- Product user manual & warranty card
- PCR strip rotor (optional)
ROTOR AND REDUCED ADAPTORS
Below table shows the Maximum RCF of rotor and Maximum RCF with different tu bes &adaptors.
12 tubes rotor | 1.5/2.0ml tubes | 0.5ml Adaptor | 0.1/0.2ml Adaptor |
---|---|---|---|
1.5/2 ml | 0.5 ml | 0.4 ml | 0.2ml |
Max. Speed | 15000 | 15000 | 15000 |
Max. RCF | 15596 | 12577 | 12074 |
PCR strip rotor (Optional) | 0.1/0.2ml PCR strip | ||
--- | --- | ||
Max. Speed | 15000 | ||
Max. RCF | 13080 |
NOTE: Always use recommended accessories for best result & product safety.
SAFETY PRECAUTIONS
-
Never use the centrifuge in any manner not specified in this manual.
-
Using equipment in any manner not specified in this manual or by the manufacturer, will void the warranty.
-
Never move the centrifuge while the rotor is spinning.
-
The rotor and the rotor lid must always be securely fastened. lf the centrifuge makes unusual noise during operation, the rotor or rotor lid fit needs to be checked. Switch OFF the device immediately by pressing STOP, check rotor fit & fasten itwell.
-
The rotors must be loaded symmetrically. Each tube should be counter balanced by another tube of the same weight.
-
Do not use the centrifuge or rotor that have not been correctly installed or damaged.
-
Repairs must only be performed by authorized service technician.
-
Using incorrect rotors & wrong spare parts will void the warranty.
-
Centrifuge may be used for the specified applications only. lt must not be operated in a hazardous or flammable environment and must not be used ill to centrifuge explosive or highly reactive substances.
-
lf liquids are spilled on the rotor or rotor chamber, the centrifuge must be clea ned carefully and properly before being used again.
-
Prior to centrifugation, the tubes should be visually inspected for material damage. Damaged tubes may not be centrifuged. This is because broken tubes can, in addition to sample lass, create imbalance which can result in further damage to the centrifuge and accessories.
-
The maximum capacity of tubes must not be exceed its corresponding rotor capacity. Do not use liquid with density higher than 1.2g/ml for full load operation.
-
Do not lean on the equipment. lt may damage the equipment or the harm the operator.
-
When moving the centrifuge from a cold room to a normal room, run the centrifuge for 30 minutes beforehand in the cold room to avoid condensation.
Alternately, allow it to warm up in the lab for at least 3 hours before use, but do not plug in the centrifuge in order te prevent possible damage by condensation. -
Be sure to close the tu bes lid tightly prior to centrifugation. Open tu bes lid can be tarn off du ring centrifugation and can damage the rotor lid or centrifuge.
-
Rotors and rotor lids are high-graded components which are subject to extreme mechanical strain. Even slight scratches and tears can lead to serious internal material damage. Ensure to check the rotor for any signs of damage before use.
Rotor & rotor lid showing visible signs of corrosion or mechanical damage should not be used, contact your local supplier or mfg for replacement rotor & parts. -
Do not fill tu bes while they are in the rotor. Liquid spillage may harm the device.
-
In the event of contamination caused by aggressive agents, the rotor must be cleaned immediately using a natural cleaning liquid (like water). This is particularly important for the bores of the tu bes. lf any damage is seen, contact the service technician.
-
Before using cleaning or decontamination methods, other than those stipulate by the manufacturer, contact the manufacturer to ensure that the intended method will not damage the centrifuge.
-
The power cord given with centrifuge unit is designed to use for that particular centrifuge. Do not use any other power cord, using any other power cord /adaptor may damage the centrifuge and will void the warranty.
INSTALLATION
Gently remove the upper packaging and take out the centrifuge by holding it
from the bottom. When the centrifuge is used for the first time, ensure that
all the packaging and accessories are removed from the rotor chamber. The
rotor is pre-installed.
Please make sure that rotor is firmly tightened. Accessories and user manual
are kept inside the box with the centrifuge. Please keep all the packaging in
safe storage for at least 2 years for warranty purpose.
The following points are important for proper installation:
8.1 CONNECTING POWER ADAPTOR
-
Connect one side of power adaptor to rear side of centrifuge and other to supply as shown in the figure below.
-
Rotate clockwise the adaptor nut to tighten the adaptor with the centrifuge. Ensure the power switch is OFF while connecting the power adaptor.
-
The sticker on the body & jack are indicating the position of the rib on the jack.
While fitting the adaptor to the unit-please match the sticker position as shown in the figure below & gently push it in. -
lf matched correctly the plug will go in easily without much force (do not try to force the plug in if it is not matched properly). Once the plug is inserted into position, turn the loose upper ring clockwise to tighten & to secure the fitment.
ROTOR INSTALLATION
The rotor of the
centrifuge is pre-installed. In case of any spillage, remove the rotor as
shown in the above figure to clean the rotor and its chamber thoroughly.
9.1 REMOVING ROTOR
- To remove rotor, first remove the Tube Holders.
- Using T-Allen key, loosen the rotor nut by rotating rotor nut counter clockwise and take the out rotor vertically.
9.2 INSTALLING THE ROTOR
- Make sure the rotor shaft and the rotor mounting hole are clean. Load the rotor vertically into the motor shaft.
- Put the T-Allen key in the rotor nut & turn clockwise to tighten the rotor.
- After properly fastening the rotor, place the Tube Holders into the rotor.
BALANCING THE ROTOR
Always balance the rotor before centrifugation Symmetrical loading of tu bes
in the rotor is necessary. Otherwise vibration or imbalance can occur which
can cause serious damage to the centrifuge. The following are process of
symmetrical loading of centrifuge tu bes inside the rotor. The samples in the
tu bes should be of equal volume.
To achieve a balanced condition, “Reduction Adapters” are mandatory to place
inside the rotor.
PROPER BALANCED ROTOR
OPERATION
After properly installing the centrifuge, switch ON the centrifuge from the
rear side by pressing the power switch. The lid status will be seen in the
display. lf the lid is open, the display will show “Lid” and if it is closed,
the set RPM will be displayed. Please ensure that the Tube Holders are inside
the rotor. 11.1
START
After setting parameters such as speed & timer as instructed below press START
button to run the machine.
11.2 STOP/OPEN LID
To stop the machine at any point of time press “STOP” button. The lid will
open automatically. The LID will open on its own once the completion of the
run. Du ringramp no other button will work except for the STOP button.
11.3 SPEED
Upon turning the power on the machine, RPM/RCF display will be of previous set
speed. The speed can be set from 500-15000 RPM. The speed can be set by using
“+” button for increment & “-” button for decrement of the values. The
increment/decrement is in the 100 RPM only. The Shown values will be multiples
of 10 (x10). The values get saved automatically after blinking for 5 tim es.
11.4 RPM-RCF SWITCH
Lang Press the RPM/RCF button to change the mode from RPM to RCF. The switch
over will be indicated by the LED on respective mode. While RCF mode is ON,
speed is to be read as value x10 in the display window. The maximum RCF of the
centrifuge is 15596 xg. Press the speed button to see the corresponding Value.
11.5 TIME
Upon turning the power on the machine, RPM/RCF display will be of previous set
timer. The timer can be set from 1 minute to 99 minutes & infinite mode. The
timer can be set by using increment (+) and decrement (-) button. The set
timer will automatically get saved. The timer works as countdown which means
the timer will run in descending order during the operation of the machine.
11.6 SHORT-SPIN OPERATION
The centrifuge can be operated for a short run by pressing and holding the
SHORT SPIN button. The centrifuge will continue to run as lang as the Short-
Spin button is pressed and the time in seconds will count up for 1 minute
after 1 minute time count up in minute on the time display. After releasing
short spin button the time in the display shows duration of short spin. The
Short-Spin speed is based on the last run settings remaining in the memory or
set rotational speed prior to short spin as required.
11.7 INCREMENT / DECREMENTS
Press”+” and “-” buttons is increase and decrease the speed or time values
normally.
11.8 OPENING THE CENTRIFUGE LID IN CASE OF POWER FAILURE Disconnect
the centrifuge from the main supply. Wait until the rotorcomes to a
standstill (this may take some time). Once the rotor has stopped, pull out the
emergency lid release as shown in the figure. This will open the centrifuge
lid. After the centrifuge lid opens, push the thread back in the hole with
your hand and fit the emergency release cap back in the hole.
11.9 IMBALANCE DETECTION
This centrifuge is equipped with an imbalance detection safety feature. When
the rotor is not loaded symmetrically, imbalance detector gets activated. This
will cut off the centrifugation & error “ErrSS” will be shown on the display.
To correct the imbalance, open the centrifuge lid and correct the imbalance
load using methods described in the balancing the rotor section (page no.4) of
this manual. After correctly loading the tubes into the rotor turn offthe
centrifuge & turn it back on.
SAFETY PRECAUTION
- Never use the centrifuge in any manner not specified in the manual.
- Never move the centrifuge while it’s operating.
- The roter must always be securely fastened. lf the centrifuge makes an unusual noise du ring the operation, the rotor fit needs to be checked.
- The roter must be loaded symmetrically.
- Do not use the centrifuge or roter, if it is not installed correctly or it is damaged.
- Do not use the centrifuge or roter if it shows any sign of external mechanical or electrical damages.
- The centrifuge must be used only for the specified application. lt must not be operated in hazardous or flammable environment & no hazardous or flammable content should be centrifuged.
- Do not fill tu bes while they are in the roter. Liquid spillage may harm the device.
- lf liquid is spilled on the roter or roter chamber, the centrifuge must be cleaned immediately.
- “Tube Holders” are a part of the Rotor. They should be placed in the roter whether 2, 3, 4, 6 … 12 tu bes are being used.
TROUBLESHOOTING
The centrifuge has a self-diagnostic function. lf a problem occurs, an error/warning code will be displayed on display screen and the operator can be determine the malfunction with the warning code below.
ERROR | PROBLEM | SOLUTION |
---|---|---|
No display | No main power connection | Check power & properly plugin power |
adaptor at both ends
Power failure| Check the main fuse of the lab.
Improper connection| Connect adaptor properly
Lid open| Lid is closed correctly| Close lid correctly
Error with lid closing and opening mechanism| Contact service representative
Err 55| Rotor not loaded symmetrically| Load the rotor symmetrically & restart
the centrifuge
Centrifuge lid cannot be opened| Rotor is still spinning| Wait for the rotor
to come to a stop
Power failure| Press emergency lid release after rotor stops
Centrifuge shakes during acceleration & makes loud
running noises while running| Rotor is not loaded symmetrically| Load rotor
symmetrically & restart operation
Either a broken tube, damage to the rotor or motor is the cause for run noise|
Replace broken tube. For damaged rotor/motor contact a service representative
Rotor is damaged| Remove & change rotor
Display error
Err 1| Loose connection of display
Latch motor damaged, Latch jammed or any Limit switch of latch got damaged|
Contact service representative
Contact service representative
Err 52| Rotor is Stuck| Turn OFF the centrifuge, Check & fit the rotor
properly & turn ON the centrifuge
Power tripping| Cable not fit properly| Remove cable and connectproperly
Last run memory not displayed| Turing ON centrifuge immediately after turning
it OFF| Maintain 3 seconds gap Between switch OFF and switching ON again
System gets hang| Electronics error| Switch off centrifuge and then switch it
ON again .If the error still shows, contact service representative
MAINTENANCE AND CLEANING
-
The rotor and the outside of the centrifuge should be cleaned regularly with a moist cloth.
-
Ensure that while cleaning the unit it is not plugged in.
-
The brushless motor in the centrifuge does not require routine maintenance. Any required service should be performed by authorized, qualified personnel only.
Repairs performed by unauthorized personnel may void the warranty. -
Always keep the centrifuge housing, rotor chamber, rotor and rotor accessories clean. All parts should be wiped down periodically with a soft cloth. For more thorough cleaning, use a neutral cleaning agent (Ph between 6 and 8) applied with a soft cloth. Excessive amounts of liquid should be avoided. Liquid should not come into contact with the motor.
-
After cleaning, ensure that all parts are dried.
-
Do not place the rotor into the cleaning solution !
WARRANTY STATEMENT
This product is warranted to be free from defects in material and workmanship
for a period of two (2) years from date of purchase. Your product will be duly
repaired upon prompt notification in compliance with the following conditions:
This warranty is valid only if the product is used for its intended purpose
and within the guidelines specified in this instruction manual. This warranty
does not cover damage caused by accident, neglect, misuse, improper service,
natural forces or other causes not arising from defects in original material
or workmanship. This warranty does not cover any incidental or consequential
damages, commercial lass or any other damages from the use of this product.
The warranty is invalidated by any non-factory modification, which will
immediately terminate all liabilities on us for the products or damages caused
by its use. The buyer and its customer shall be responsible for the product or
use of products as weil as any supervision required for safety. lf requested
the products must be returned to the distributor in weil packed and insured
manner and all shipping charges must be paid.
Same states do not allow limitation on the length of implied warranties or the
exclusion or limitation of incidental or consequential damages. This warranty
gives you specific legal rights. This warranty is given expressly in lieu of
all other warranties, expressed or implied.
The purchaser agrees that there is no warranty of merchantability or of
fitness for any intended purpose and that there are no other remedies or
warranties, expressed or implied, which extend beyond the description on the
face of the agreement. This warranty is only applicable to the original
purchaser.
Products received without proper authorization will not be entertained. All
items returned for service should be sent postage prepaid in the original
packaging or other suitable carton, padded to avoid damage. We will not be
responsible for damage incurred by im proper packaging.
All items returned for service should be set postage prepaid in the original
packaging or other suitable carton, added to avoid damage.
This warranty is valid only if the warranty is registered with the supplier
within 30 days from the date of purchase.
For you reference, make a note of serial number, date of purchase and supplier here.
Serial No.:
Purchase Date:
Supplier:
PRODUCT DISPOSAL
In case the product is to be disposed of, the relevant legal regulations are
to be observed.
Information on the disposal of electrical and electronic devices in the
European Community.
The disposal of electrical devices is regulated within the European Community
by national regulations based on EU Directive 2012/19/EU on waste electrical
and electronic equipment (WEEE). According to these regulations, any devices
supplied after 13.06.05 in the business to business sphere, to which this
product is assigned, may no langer be disposed off in municipal or domestic
waste. They are marked with the following symbol to indicate this.
As disposal regulations within the EU may vary from country to country, please
contact your supplier if necessary.
iFuge M15K Microcentrifuge | Lab Devices
(praxisdienst.com)
Neuation Technologies Pvt. Ltd.
Plot No. 15, GIDC Electronic Park SEZ Kolavada Road, Gandhinagar – 382026,
Gujarat, India
Website:www.neuation.com
0115-00-0000-02-003-R1
Documents / Resources
|
NEUATION TECHNOLOGIES M15K Brushless DC Motor Centrifuge
Machine
[pdf] User Manual
M15K Brushless DC Motor Centrifuge Machine, M15K, Brushless DC Motor
Centrifuge Machine, DC Motor Centrifuge Machine, Motor Centrifuge Machine,
Centrifuge Machine
---|---