APERA INSTRUMENTS PC60 Premium Multi-Parameter Tester Installation Guide

June 3, 2024
APERA INSTRUMENTS

PC60 Premium Multi-Parameter Tester

PC60 Premium Multi-Parameter Tester (pH/EC/TDS/Salinity/Temp.) Instruction Manual
APERA INSTRUMENTS (Europe) GmbH
www.aperainst.de
V6.4

Thank you for purchasing Apera Instruments PC60 Premium Multi-Parameter Tester. Please carefully read this instruction manual before using the product in order to have an accurate and reliable test result, and avoid unnecessary damages to the meter or probe. For video tutorials, please go to www.aperainst.de
Contents
1. Battery Installation…………………………………………………………………………………………………………………. 3 2. Keypad Functions …………………………………………………………………………………………………………………… 3 3. Complete Kit………………………………………………………………………………………………………………………….. 4 4. Preparation Before Use…………………………………………………………………………………………………………… 4 5. pH Calibration ……………………………………………………………………………………………………………………….. 5 6. pH Measurement …………………………………………………………………………………………………………………… 6 7. Conductivity Calibration………………………………………………………………………………………………………….. 7 8. Conductivity Measurement …………………………………………………………………………………………………….. 7 9. Parameter Setting ………………………………………………………………………………………………………………….. 9 10. Technical Specifications ………………………………………………………………………………………………………. 10 11. Icons and Functions ……………………………………………………………………………………………………………. 11 12. Probe Replacement…………………………………………………………………………………………………………….. 11 13. Warranty …………………………………………………………………………………………………………………………… 11
Upgrade Note The new PC60 Tester comes with an upgraded probe structure, which is equipped with a sensor shield that prevents the glass bulb breakage from accidental collision (see picture below). Users can remove the shield when cleaning the sensor and put it back on after cleaning.
Sensor Shield
2

1. Battery Installation
Please install batteries according to the following steps. *Please note direction of batteries: All
POSITIVE SIDES (“+”) FACING UP. (Wrong installation of batteries will cause damage to the tester and potential hazards)

+ +
– –
+ +
– –

Pull the battery cap up Slide the battery cap along to the direction of arrow Open the battery cap Insert the batteries (ALL POSITIVE SIDES FACING UP) (see graph) Close the battery cap Slide and lock the battery cap along to the direction of arrow Fit the tester’s cap while making sure to push all the way down. The tester’s
waterproof design may be compromised if the cap is not fitted correctly.

2.Keypad Functions

Short press—— < 2 seconds ,Long press——- > 2 seconds

Battery

1.Short press to turn on the tester and long press to turn off the tester. 2.When turned off, long press to enter parameter setting. 3. In measurement mode, short press to turn on backlight.
1.In measurement mode, short press to switch parameter pHCONDTDSSAL 2.In mode setting, short press to change parameter (Unidirectional)
1. Long press to enter calibration mode. 2. In calibration mode, short press to confirm calibration. 3.When measured value is locked, short press to
unlock;

LCD Buttons Probe Probe Cap
Probe
pH Sensor BPB Sensor

3

3. Complete Kit
Diagram – 2
4. Preparation Before Use
If it is first-time use or the tester hasn’t been used for a long time, pour some 3M KCL solution in the probe cap (about 1/5 of the probe cap) and soak the probe for 15-30 minutes. When not in use, we recommend store the pH probe in the storage 3M KCL solution in the probe cap to keep the sensor’s accuracy. But even if stored dry, it won’t do any permanent damage to the sensor. It will only temporarily cause the probe to lose its sensitivity, which can always be restored by soaking in the storage solution. The storage solution is 3M KCL (potassium chloride). One bottle of 10mL storage solution comes with the tester kit. If the soaking solution was contaminated, please replace with new ones timely. * DO NOT use any other brand’s storage solutions because different chemicals may be used and potential permanent damage could be caused to the meter.
4

5. pH Calibration

Things needed in addition to what’s in the box: · A clean cup, · distilled water (8-16oz) · and tissue papers for rinsing · drying the probe

5.1 Short press

to turn on the meter; rinse the probe in distilled water, shake the meter in

the air and use tissue paper to dap off excess water (never rub or wipe the sensor). 5.2 Pour certain amount (about half volume of the calibration bottle) of pH 7.00 and pH 4.00
buffer solution in separate calibration bottles;

5.3 Long press

to enter calibration mode; Short press

to exit.

5.4 Dip the probe in pH7.00 buffer solution, stir gently, and allow it to stand still in the buffer solution until a stable reading is reached. When stable

icon is displayed on the LCD screen (as shown in Diagram 3), short

press to complete 1-point calibration and the tester returns to

Diagram – 3

measurement mode. Indication Icon

will appear at the bottom left of

the LCD screen.

5.5 Rinse the probe in distilled water and dry it. Long press to enter calibration mode. Dip

the probe in pH 4.00 buffer solution, stir gently, and allow it to stand still in the buffer solution.

When stable icon is displayed on the LCD screen, short press to complete 2-point

calibration and the tester returns to measurement mode. Indication icon

will appear

at the bottom left of the LCD screen. 5.6 If necessary, rinse the probe in distilled water and dry it, and dip the probe in 10.01 buffer
solution (sold separately) to complete 3rd point of calibration according to the steps in 5.5,

will appear at the bottom left of the LCD.

Notes
a) Tester will automatically recognize pH buffer solution. Users can perform one-point, two-point, or three-point calibration. But for the 1st point calibration, only 7.00 pH solution can be used. Then use other buffer solutions to conduct 2nd or 3rd point calibration. Tester will automatically recognize 5 kinds of pH buffer solutions. Refer to the table below:

5

Calibration

USA Series

1-point 2-point 3-point

  1. 7.00 pH
  2. 7.00 pH 2) 4.00 or 1.68 pH
  3. 7.00 pH 2) 10.01 or 12.45 pH
  4. 7.00 pH 2) 4.00 or 1.68 pH 3) 10.01 or 12.45 pH

NIST Series

  1. 6.86 pH
  2. 6.86 pH, 2) 4.01 pH or
    1.68 pH 1) 6.86 pH, 2) 9.18 pH or 12.45 pH 1) 6.86pH 2) 4.01 or 1.68pH, 3) 9.18 pH or 12.45 pH

Calibration Indication Icon

Recommended Accuracy and
Range
Accuracy 0.1 pH
Measurement Range7.00
pH
Measurement Range7.00pH
Wide Measurement
Range

b) For pH Calibration buffer solutions, we recommend that users replace new buffer solution after 10 to 15 times of use to keep the standard buffer’s accuracy. Do NOT pour the used calibration solutions back into the solution bottles in case of contamination.
c) This pH probe will NOT give accurate and stable readings for distilled or deionized water. This is because distilled and deionized water do not have enough ions present for the electrode to function properly. Specialized pH probes need to be used for distilled/deionized water measurement. Contact us at info@aperainst.de for more details.
d) When testing purified water like spring water or drinking water, it will take longer for the readings to get stabilized (typically 3-5 minutes) because there is very few ions left to be detected by the sensor in those purified water.
e) Do NOT store pH probe in distilled water to prevent permanent damage to the probe. f) For the self-diagnosis information, please refer to the table below:

Symbol

Self-Diagnosis information

Checking and methods to fix

Wrong calibration solution or the range of calibration solution exceeds standard.

a) Check if calibration solution is correct (1st point of pH calibration must be pH 7.00) b) Check if electrode is damaged. c) Check if there is any air bubble in the glass bulb pH sensor

Is pushed before measurement is Wait for the smile icon to come up and stay,

stable ( comes up and stays)

then press

  • If you find any air bubble in the glass bulb of the pH sensor, simply shake the probe for a few times to remove it. The existence of an air bubble in the glass bulb will cause instable measurements. * The 1st point calibration must be 7.00 pH. Perform the 2nd point calibration (4.00 pH) immediately after the 1st point. Do NOT turn off the meter before you conduct 2nd point calibration. If the meter is turned off after 1st point calibration, users will need to restart the calibration process with the 7.00 pH first and the 4.00 pH following after. Calibrating directly in pH 4.00 after turning meter off and back on will cause Er1.
    6

6.pH Measurement
Short press to turn on the tester. Rinse the probe in distilled water and dry it. Dip the probe in sample solution, stir gently, and allow it to stand still in the solution. Get readings after comes up and stays.

7. Conductivity Calibration
7.1 Press key to switch to conductivity measurement mode. Rinse the probe in distilled water and dry it. 7.2 Pour certain amount (about half volume of the calibration bottle) of 1413S/cm and 12.88 mS/cm conductivity calibration solution into accordant calibration bottles. 7.3 Long press key to enter calibration mode, short press to exit. 7.4 Dip the probe in 1413 S/cm conductivity calibration solution, stir gently and
allow it to stand still in the solution until a stable reading is reached. When stable icon appears and stays on the LCD screen, short press key to complete one-point calibration, the tester returns to measurement mode and indication icon will appear at the bottom left of the LCD screen. 7.5 After calibration, dip the probe in 12.88 mS/cm conductivity calibration solution. If the value is
accurate, it is not necessary to conduct 2nd point calibration. If it is inaccurate, follow the steps in 7.3 to 7.4 to complete the 2nd point of calibration using 12.88 mS/cm buffer solution. * 1000 µS/cm = 1 mS/cm

8. Conductivity Measurement

Press

key to turn on the tester. Rinse the probe in distilled water and dry it.

Dip the probe in sample solution, stir gently, and allow it to stand still in the solution until a stable

reading is reached. Get readings after

comes up and stays. Press to switch from

Conductivity to TDS, and Salinity

Notes a) The TDS and Salinity measurements are converted from the conductivity measurements via a certain conversion factor. b) The tester can calibrate 84S, 1413 S/cm and 12.88 mS/cm conductivity calibration solution. User can conduct 1 to 3 points calibration. Refer to the table below. Usually calibrating the tester with 1413 S/cm conductivity buffer solution alone shall meet the testing requirement.

Calibration Indication Icon Calibration Standards

Measuring Range

84 S/cm

0 – 200 S/cm

1413 S/cm

200 – 2000 S/cm

12.88 mS/cm

2 – 20 mS/cm
7

c) The tester has been calibrated before leaving the factory. Generally, users can use the tester directly or users can test conductivity buffer solutions first. If the error is large, then calibration is needed.
d) For conductivity calibration solutions, we recommend that users replace new solutions after 5 to 10 times of use to keep the standard solution’s accuracy. Do NOT pour the used calibration solutions back into the solution bottles in case of contamination.
e) Temperature compensation factor: The default setting of the temp. compensation factor is 2.0%/. User can adjust the factor based on test solution and experimental data in parameter setting P4.

Solution
NaCl 5% NaOH Dilute ammonia

Temperature compensation factor
2.12%/°C 1.72%/°C 1.88%/°C

Solution
10% Hydrochloric acid
5% Sulfuric acid

Temperature compensation factor
1.32%/°C 0.96%/°C

f) 1000 ppm = 1 ppt g) TDS and conductivity is linear related, and its conversion factor is 0.40-1.00. Adjust the factor
in parameter setting P5 based on the requirements in different industries. The factory default setting is 0.71. Salinity and conductivity is linear related, and its conversion factor is 0.5. The tester only needs to be calibrated in Conductivity mode, then after calibration of conductivity, the meter can switch from conductivity to TDS or salinity. h) Conversion Example: if conductivity measurement is 1000µS/cm, then the default TDS measurement will be 710 ppm (under the default 0.71 conversion factor), and the salinity be 0.5 ppt. i) For the self-diagnosis information, please refer to the table below:

Symbol

Self-Diagnosis information

How to fix

Wrong conductivity buffer solution, which 1. Check if buffer solution is correct
exceeds the recognizable range of the 2. Check if electrode is damaged.
meter.

Is pushed before measurement is Wait for the smile icon to come up and then

stable ( comes up and stays)

press

8

9. Parameter Setting
9.1 Setting Chart

Symbol

Parameter Setting Contents

P1

Select pH buffer standards

P2

Select automatic lock

P3

Select backlight

P4

Temperature compensation factor

P5

TDS factor

P6

Salinity unit

P7

Select temperature unit

P8

Back to factory default

Code
USA ­ NIST Off ­ On
Off – 1 – On 0.00 – 4.00% 0.40 – 1.00
ppt – g/L °C – °F No ­ Yes

Factory Default
USA Off 1 2.00% 0.71 ppt °F No

9.2 Parameter Setting

When turned off, long press

to enter parameter setting short press

to switch

P1-P2… P8. Short Press , parameter flashes short press

to choose

parameter, short press

to confirm Long press

9.3 Parameter Setting Instruction

to turn off.

a) Select standard pH buffer solution (P1): There are two options of standard buffer solutions: USA series and NIST series. Refer to following chart:

Icons

pH Standard Buffer Solution Series

USA series

NIST series

Three-point calibration

1.68 pH and 4.00 pH 7.00 pH
10.01 pH and 12.45 pH

1.68 pH and 4.01 pH 6.86 pH
9.18 pH and 12.45 pH

b) Automatic lock (P2):

Select “On” to activate auto lock function. When reading is stable for more than 10 seconds, the

tester will lock the value automatically, and HOLD icon will display on LCD. Press

key to

cancel auto hold.
9

c) Backlight (P3) “Off”-turn off backlight, “On”-turn on backlight, 1- backlight will last for 1 minute. d) Factory default setting(P7)
Select “Yes” to recover instrument calibration to theoretical value (pH value in zero potential is 7.00, slope is 100%), parameter setting return to initial value. This function can be used when instrument does not work well in calibration or measurement. Calibrate and measure again after recover the instrument to factory default status.

10. Technical Specifications

pH
Cond. TDS Salinity Temp.

Range Resolution Accuracy Calibration Points Automatic Temperature Compensation
Range Resolution Accuracy Calibration Points
Range TDS Factor
Range Range Resolution Accuracy

-2.00 to 16.00 pH 0.01 pH
±0.01 pH ±1 digit 1 to 3 points
0 ­ 50°C (32 ­ 122°F) 0 to 200.0 S, 0 to 2000 S,
0 to 20.00 mS/cm 0.1/1 S, 0.01 mS/cm
±1% F.S
1 to 3 points
0.1 ppm to 10.00 ppt
0.40 to 1.00
0 to 10.00 ppt 0 to 50°C (32-122°F)
0.1°C ±0.5°C

10

11. Icons and Functions

Calibration points indication: Stable Measurement:

Reading value Auto. Lock: HOLD Self-Diagnostic Information: Er1, Er2

Low-Voltage warning: Three-Color backlight:

flashes, reminder of battery replacement

Blue–Measurement Mode; Green–Calibration Mode; Red–Alarm; Auto. Power off in 8 minutes if no operation.

12. Probe Replacement
Screw off the probe ring, unplug the probe, plug in the new replacement probe (pay attention to the probe’s position), and screw on the probe ring. The model numbers of replacement probes that are compatible with PC60 are:
· PC60-E (Regular pH/conductivity probe) · PC60-DE (Double-junction pH/conductivity probe) · PH60-DE (Double-junction pH glass bulb probe) · PH60-E (Regular pH glass bulb probe) · PH60S-E (Spear pH probe for solids /semi-solids pH testing) · PH60F-E (Flat pH probe for surface pH testing) · EC60-E (Conductivity probe)

13. Warranty
We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at option of APERA INSTRUMENTS (Europe) GmbH, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS (Europe) GmbH for a period of TWO YEARS (SIX MONTHS for the probe) from the delivery.

This limited warranty does not cover any damages due to: Transportation, storage, improper use, failure to follow the product instructions or to perform any preventive maintenance, modifications, combination or use with any products, materials, processes, systems or other matter not provided or authorized in writing by us, unauthorized repair, normal wear and tear, or external causes such as accidents, abuse, or other actions or events beyond our reasonable control.

APERA INSTRUMENTS (Europe) GmbH Wilhelm-Muthmann-Straße 18, 42329 Wuppertal Germany info@aperainst.de | www.aperainst.de | Tel. +49 202 51988998
11

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

APERA INSTRUMENTS User Manuals

Related Manuals