APERA INSTRUMENTS EC700 Benchtop Conductivity Meter Installation Guide

June 3, 2024
APERA INSTRUMENTS

EC700 Benchtop Conductivity Meter

EC700 Benchtop Conductivity Meter Instruction Manual
APERA INSTRUMENTS (Europe) GmbH
www.aperainst.de
v2.0

Table of Contents
1 Brief Introduction ………………………………………………………………………………………………- 1 1.1 Main Features ………………………………………………………………………………………………………. – 1 –
2 Technical Specifications……………………………………………………………………………………….- 1 –
3 Instrument Description ……………………………………………………………………………………….- 2 3.1 LCD Display ………………………………………………………………………………………………………….. – 2 3.2 Keypad ……………………………………………………………………………………………………………….. – 3 3.3 Connectors ………………………………………………………………………………………………………….. – 4 3.4 Stable reading display mode……………………………………………………………………………………. – 4 3.5 Data storage, recall, and deletion …………………………………………………………………………….. – 4 3.6 Max/Min Function ………………………………………………………………………………………………… – 5 –
4 Conductivity Measurement………………………………………………………………………………….- 5 4.1 Preparation………………………………………………………………………………………………………….. – 5 4.2 Conductivity Electrode Information ………………………………………………………………………….. – 5 4.3 Information regarding Calibration…………………………………………………………………………….. – 5 4.4 Conductivity Calibration (take 1413 S as an example) …………………………………………………. – 6 4.5 TDS and Conductivity …………………………………………………………………………………………….. – 6 4.6 Sample Test………………………………………………………………………………………………………….. – 7 4.7 Parameter Setting …………………………………………………………………………………………………. – 8 4.8 Conductivity Electrode’s Maintenance ………………………………………………………………………. – 9 –
5 Complete Kit…………………………………………………………………………………………………… – 10 –
6 Warranty ……………………………………………………………………………………………………….. – 10 –

1 Brief Introduction
Thank you for purchasing Apera Instruments EC700 Benchtop Conductivity Meter. This instrument is an outstanding combination of advanced electronics desin and sensor technology. It’s the most economical choice of a reliable lab benchtop conductivity meter.
Please read this maual carefully to properly use and maintain the meter. Apera Instruments reserves the right to update the contents of this manual without giving prior notices.
1.1 Main Features
· The meter has a built-in microprocessor chip, which enables intelligent functions such as automatic calibration, automatic temperature compensation, data storage, parameter setup, max/min reading display, etc.
· The advanced automatic calibration mode has calibration guide and self- diganosis functions, automatically recognizing standard calibration solutions.
· The advanced digital processing technology intelligently improves the meter’s response time and measuring accuracy with stable reading display mode available.
· The kit comes with a combination conductivity electrode with temperature sensor, measuring conductivity and temperature simultaneously, and allows for automatic temperature compensation.
· IP54 dustproof and spillproof. Connectors are protected by silicone seal caps, effectively ensuring the reliability and service life of the meter under harsh environment.
· Detachable electrode holder and 3 bottles of conductivity calibration solutions come with the kit for convenient use.
2 Technical Specifications

Range
Resolution Conductivity Accuracy
Temperature Compensation Electrode Constant Reference Temperature Temperature Compensation Coefficient Calibration

Conductivity: 0-200.0 mS/cm, divided into 4 ranges: (0-199.9) S/cm, (200-1999) S/cm,
(2.00-19.99) mS/cm,(20.0-199.9) mS/cm TDS: (0-100) g/lL devided into 4 ranges:
(0-99.9) mg/L, (100-999) mg/L, (1.00-9.99) g/L, (10.0-100.0) g/L
Conductivity: 0.1/1 S/cm; 0.01/0.1 mS/cm TDS: 0.1/1 mg/L, 0.01/0.1 g/L
±1.0% F.S ±1 digit
0 to 50 °C (Auto or Manual)
0.1/1/10 cm-1
25 °C
0.00-9.99%/ °C, default value: 2.00%/ °C
1-4 point Auto Calibration – 1 –

Temp. Others

Range Resolution Accuracy Data Storage Storage Content Power IP Ratings Size and Dimension

0 to 100°C; 32 to 212°F 0.1°C; 0.1/1°F ±0.5°C ±1 digit 50 groups
Numberings, Measurement, Unit, Temperature, Temperature Compensation Status DC9V/300mA
IP54 dustproof and spillproof (240235103) mm/1kg

3 Instrument Description 3.1 LCD Display

(1) — Measurement mode

Figure-1

(2) — Measurement value

(3) — Calibration icons and numberings

(4) — Measurement unit

(5) — Temperature unit

(6) — The unit displayed during calibration

(7) — The value displayed during calibration, numberings for storage and recall, and reminder icons

(8) — Icons for data storage and recall — M+: store measuring data; RM: Recall measuring data

(9) — Temperature and reminder icons

(10) — Self-Diagnosis icons and numberings (11) — Temperature compensation mode icons — ATC: Auto Temperature Compensation;
MTC: Manual Temperature Compensation (12) — Completed calibration icons (13) — Stable reading icon (14) — Icons for maximum and minimum values
– 2 –

3.2 Keypad

Figure-2 3.2.1. Keypad operations Short press —– <1.5 seconds; Long press —–

1.5 seconds.

3.2.2 For keypad operation, please see Table-1. Table ­ 1 Keypad operations and descriptions

Keypad Operation

Functions

Short Press
Short Press Long Press

Power on/off Select measuring mode: CondTDS. Enter parameter settingP1P2…P6.

Short Press

Enter Calibration mode Cancel any operation and go back to measurement mode

Short Press
Short Press Long Press
Short Press or
Long Press
Short Press

In calibration mode, press to conduct calibration In parameter setting, press to confirm selection
Store measuring data
Recall stored measuring data
1. In manual temperature compensation mode, short press to change temperature, long press to change quickly
2. In parameter setting mode, short press to change parameter and settings.
3. In recall mode, short press to change numberings, long press to change quickly.
4. Buttons for max/min values (refere to section 3.6)

– 3 –

3.3 Connectors

Symbol COND TEMP DC9V

Connector Name Socket for conductivity electrode
Socket for temperature sensor DC9V power socket

Connector Type BNC RCA
2.5 direct type

3.4 Stable reading display mode

When the measuring value is stable, smiley face icon

appears on LCD, see

Figure-3. Please do not get the reading value or make calibration until the

icon appears and stays.

3.5 Data storage, recall, and deletion 3.5.1 Storage

When the measurement is stable, press

key, the meter displays

M+ icon and storage numberings on LCD, and stores the measuring data.

In Figure-4, the meter stores the first group of the measuring data. Press

again to keep storing.

3.5.2 Recall stored data

In the measurement mode, long press

key to recall the last stored

Measuring data. The meter displays RM icon and storage numberings.

In Figure-5, the meter is recalling the 28th stored measuring data.

Keep pressing

and

to recall successively the stored

measuring data. Press and hold

and to quickly recall the stored

measuring data. Short press

to go back to measurement mode.

3.5.3 Clear stored data

The meter can store 50 groups of data. Once the storage is full, if user

presses

again, the FUL icon will be displayed on LCD (see Figure-6).

The storage needs to be cleared in parameter setting P4 so as to keep storing

– 4 –

Figure – 3 Figure – 4 Figure – 5 Figure – 6

data. The numberings in M+ indicates the total number of stored data; The numberings in RM indicates the number of stored data in current mode (cond/TDS). For example, if 20 groups of data is stored in conductivity mode, and 10 groups in TDS mode, then the numberings for M+ will be “M+30” in both conductivity and TDS mode; the numberings for RM in conductivity mode would be “RM20”, in TDS mode “RM10”.

3.6 Max/Min Function

Use this function to display the maximum or minimum records out of groups of measuring data. In

measurement mode, press

, LCD will display flashing icons of “MAX/MIN”, indicating the meter has

entered the mode of recording maximum and minimum data. After testing, press

again, LCD will

alternately display the maximum and minimum value in the past measurements. Press

to go back to

regular measurement mode.

4 Conductivity Measurement
4.1 Preparation 4.1.1 Install the electrode holder
The electrode holder is comosed of a metal stand base, a supporting pole, and a electrode clip. First, install the metal stand base to the meter (screw it on). Please note that the metal base can be installed on either side of the meter. Then install the supporting pole on the metal stand base, and then install the electrode clip on the supporting pole.
4.1.2 Connect the power adapter
Plug the power adapter tightly into the DC9V socket. The power’s voltage should meet the requirements listed on the power adapter.
4.2 Conductivity Electrode Information 4.2.1 Conductivity Electrode
The meter comes with the 2301T-F combination conductivity electrode (K=1.0) with built-in temperature sensor, which enables the automatic temperature compensation. The electrode’s housing adopts polycarbonate materials, which is resistant to shocks and corrosion.
4.2.2 Electrode Connectors
The conductivity electrode has two connectors: the BNC connector connects the conductivity sensor, the RCA connector connects the temperature sensor. Plug these two connectors into `COND” and “TEMP” sockets. Please note do not pull the cables in case of poor contact. Please keep the connectors clean and dry. Refer to section 4.8 regarding how to properly maintain the Conductivity electrode.
4.2.3 Use the Electrode
Stir the electrode for a few seconds after it’s dipped into the solution and then let it stand still such that the bubbles inside the probe can be removed and a stable measurement can be quickly obtained.
4.3 Information regarding Calibration 4.3.1 Conductivity standard solutions
The meter automatically recognizes calibration solutions. 1 point or multi- points calibration (up to 4 points) can – 5 –

be adopted. The calibration icons stand for 4 different calibration points (

solutions), see Table-2:

Table-2

stands for two standard

Calibration icon

pH standard buffer series

4-Point calibration

84 S/cm 1413 S/cm 12.88 mS/cm 111.8 mS/cm

4.3.2 How often to calibrate
· We recommend to calibrate the conductivity meter once per month under normal conditions. · If the requirement for accuracy is high or the temperature of the test sample has a big difference to the
reference temperature (25 °C), we recommend to do the calibration once a week. · Users can tell if the meter needs a calibration by testing the standard calibration solutions. If the
measurements you get have a big difference to the standard value of the calibration solutions, then a calibration is necessary. · For new electrodes or the meter that has been restored to factory default setting, we recommend to conduct 3-4 points calibration. For general uses, users can choose the calibration solutions that are close to the test samples to conduct 1 or 2 points calibration.

4.4 Conductivity Calibration (take 1413 S as an example)
4.4.1. Rinse the electrode in distilled or pure water, allow it to dry, wash with a little of standard solution and submerge it in standard solution. Stir the solution briefly and allow it to stay in the solution until a stable reading is reached.

4.4.2. Press

key to enter the calibration mode. The “CAL” icon will flash at

the top right, and the scanning and locking process of calibration solution will appear at the bottom right.

4.4.3. When the meter locks 1413 S, stable icon

will display on LCD. Press

key to complete calibration. “End” icon will show up and the meter will return

to measuring mode and

will show up on the bottom left (see Figure-7).

Note: pressing

before the reading being stable ( shows up and stays),

the Err4 self-diagnosis icon will appear (see section 6.5.2). 4.4.4. For multi-point calibration, please repeat clause 4.3.1-4.3.2 until all the calibration is done. The meter can repeat calibration in the same calibration solution until the stable value is reached.

Figure-7

4.5 TDS and Conductivity
4.5.1. TDS and conductivity is linear related. The conversion factor is 0.40-1.00. Adjust the factor from – 6 –

parameter P2.6. The factory default setting is 0.71 and please refer to Section 7.4. The meter only needs to be calibrated in Conductivity mode, then after calibration of conductivity, the meter can switch from conductivity to TDS. 4.5.2. Adjust TDS conversion factor in parameter setting P4 according to the data collected during testing and experience. Table ­ 10 lists some commonly used Conductivity and TDS conversion factors. This is for your reference only.
Table ­ 10 Conductivity and TDS conversion factors

Conductivity of solution

TDS conversion factor

0-100 S/cm

0.60

100-1,000 S/cm

0.71

1-10 mS/cm

0.81

10-100 mS/cm

0.94

4.6 Sample Test
4.6.1. Rinse conductivity electrode in distilled or pure water, dry it, and submerge it in the sample solution. Stir

the solution briefly and allow it to stay in the sample solution until a stable reading is reached and

icon

appears on LCD, then get the reading value, which is the conductivity value of the solution. 4.6.2. During the process of calibration and measurement, the meter has self-diagnosis functions, indicating the relative information as below: Table ­ 8.
Table ­ 8 Self-diagnosis information of conductivity measurement mode

Display Icons

Contents

Checking

1. Check whether conductivity calibration

Wrong conductivity calibration solution or solution is correct.

the meter recognition of calibration

2. Check whether the meter connects the

solution is out of range.

electrode well.

3. Check whether the electrode is damaged.

Press

key when measuring value is Press

not stable during calibration.

key after

icon appears.

During calibration, the measuring value 1. Shake the electrode to eliminate bubbles in

electrode head.

being unstable for over 3 minutes.

2. Replace with new conductivity electrode.

4.6.3. Factory default setting For factory default setting, please refer to parameter setting P7 (Section 4.7.2). With this function, all calibration data is deleted and the meter restores to the theory value. Some functions restore to the original value. When calibration or measurement fails, please restore the meter to factory default setting and then perform recalibration or measurement. Please note once set the factory default, all the data deleted will be irretrievable.

– 7 –

4.7 Parameter Setting

Symbol

Parameter setting content

Table-9 Parameter

P1

Select electrode constant

P2

Select reference temperature

Adjust temperature compensation P3
coefficient

P4

Adjust TDS Factor

0.1, 1.0, 10 15 – 30 0.00 – 9.99% 0.40 ­ 1.00

P5

Select temperature unit

P6

Clear stored data

No ­ Yes

P7

Restore to factory default

No ­ Yes

Factory default setting 1.0 25
2.00% 0.71
No No

4.7.1 How to do the parameter setting:

In measurement mode, long press

to enter parameter setting P1. Then press

or

to change

menu from P1 to P2…P7. For details, please see Table-10.

Table-10 Parameter Setting

P1 — Select electrode’s constant1.0–10.0–0.1

1. Long press

to enter P1

2. Press

, “1.0” flashes, press

to confirm

to choose 1.010.00.1; press

3. Press

to enter P2, or press

to return to measurement mode.

P2–Select reference temperature (25.0–18.0–20.0)

1. Press

, “25.0” flashes, press

or

to adjust temp. from 15

to 30; press to confirm.

2. Press

to enter P3, or press

to return to measurement mode.

P3–Adjust temp. compensation coefficient (0.00-9.99%)

1. Press

, “2.00” flashes, press

or

to adjust from 0.00-9.99;

press

to confirm.

2. Press

to enter P4, or press

to return to measurement mode.

– 8 –

P4–Adjust TDS Factor (0.40-1.00)

1. Press

, “0.71” flashes, press

press

to confirm.

2. Press

to enter P5, or press

or

to adjust from 0.40-1.00;

to return to measurement mode.

P5 –(°C–°F) 1. Press flashes, press

, “°C” flashes, press to confirm.

2. Press

to enter P6, or press

, “°F” flashes; when parameter to return to measurement mode.

P6 — Clear storage (No–Yes)

1. Press

, “No” flashes, press

to choose NoYes; press

to

confirm No– not to clear data storage; Yes–clear all data storage

2. Press to enter P7, or press

to return to measurement mode.

P7 — Return to factory default mode (No–Yes)

Press

, “No” flashes, press

to choose NoYes; Press

to

confirm, the meter returns to measurement mode. No– not return to factory default mode; Yes–return to factory default mode

4.8 Conductivity Electrode’s Maintenance
4.8.1. Always keep the conductivity electrode clean. Before taking a measurement, rinse the electrode in pure water. It is recommended to rinse it again in the sample solution. When submerge the electrode in solution, stir the solution briefly to eliminate air bubbles and allow it to stay until a stable reading is reached. Conductivity electrodes are usually stored dry. For conductivity electrodes that haven’t been used for a long time, users should soak the electrode in 12.88 mS calibration solution for 5-10 minutes, or to soak it in tap water for 1 to 2 hours. Rinse the electrode in pure water after measurement.
4.8.2. The sensing rod of Model 2301T-F conductivity electrode is coated with platinum black to minimize electrode polarization and expand measuring range. The platinum black coating adopted our special processing technology, which improves the electrode performance and the firmness of the coating. If the platinum black electrode is stained, gently clean the electrode with soft brush in warm water containing detergent or alcohol.

– 9 –

5 Complete Kit

Content

5.1

EC700 Benchtop Conductivity Meter

5.2

2301T-F Plastic Combination Conductivity Electrode

5.3

Conductivity standard calibration solutions: (84S/1413S /12.88mS, 50ml for each)

5.4

9V Power Adapter

5.5

Electrode Holder

5.6

Instruction Manual

Quantity 1 1
1 for each
1 1 set 1

6 Warranty
We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at option of APERA INSTRUMENTS (Europe) GmbH, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS (Europe) GmbH for a period of TWO YEARS (SIX MONTHS for the probe) from the delivery.
This limited warranty does not cover any damages due to: Transportation, storage, improper use, failure to follow the product instructions or to perform any preventive maintenance, modifications, combination or use with any products, materials, processes, systems or other matter not provided or authorized in writing by us, unauthorized repair, normal wear and tear, or external causes such as accidents, abuse, or other actions or events beyond our reasonable control.

APERA INSTRUMENTS (Europe) GmbH Wilhelm-Muthmann-Straße 18, 42329 Wuppertal Germany Email: info@aperainst.de | ww.aperainst.de | Tel. +49 202 51988998
– 10 –

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

APERA INSTRUMENTS User Manuals

Related Manuals