Fisher Scientific Isotemp Shaking and General Purpose Water Baths User Manual

June 3, 2024
Fisher Scientific

Isotemp Shaking and General Purpose Water Baths
User Manual
Manual P/N U01318
Rev. 11/05/2015 ****

Isotemp Water Bath Quick Start Guide

This quick start guide is intended for initial start up only. For all other procedures you must refer to the manual. Also, if any of these steps are not clear refer to the manual before proceeding.
Filling (Only approved fluids are Filtered/Single Distilled Water and De- ionized Water)
Ensure the reservoir drain port is closed and that all plumbing connections are securely plumbed or capped. Also ensure any residue is thoroughly removed from the  reservoir before filling.
To avoid spilling, place your samples/trays into the bath before filling.
Slowly fill the reservoir. When adding, point the opening of a container away from yourself.
For Shaking Water Baths, fill the reservoir with a minimum of one inch of water and the maximum level lower than the tray shaft fitting.
Make allowance for water splashing from tray oscillations.
For GPD Baths, fill the reservoir with a minimum of one inch of water and a maximum level one inch lower than the tank upper surface.

Starting

Do not run the bath until fluid is added to the reservoir. Have extra fluid on hand. If the bath does not start refer to the manual.

  • Place the circuit protector located on the rear of the bath to the I position.
  • The screen will momentarily display and then the home screen appears displaying the set point and the audible alarm status. The set point is the desired reservoir fluid temperature.

Viewing/Changing the Shaker Speed (SWB only)
You can view/change the speed value with the bath running or not. From the home screen press the soft key below to display:
Press, and hold, the arrow keys to bring up the desired value. The values are 0 and 30 to 200. Once the desired value is displayed press  or to save the change and return  to the home screen.

Changing the Set Point

When operating without a lid, limit the maximum set point to 60°C.
You can change the set point with the bath running or not. From the home screen press either arrow key to display:
Press, and hold, the arrow keys to bring up the desired set point value. The range is 5°C to 100°C. Once the desired value is displayed, in this case 50.0°C, press the soft key below to save the change or press the soft key below  to not save the change. The display returns to the home screen. Note After a 30 second delay if neither key is pressed  the display returns to the home screen, any change is not saved.

Selecting a Preset
You can select a preset with the bath running or not.
From the home screen press:
Use the arrow keys to select the desired preset. Once the desired preset is highlighted press the soft key below to save the change and return to the home screen. Note After a 30 second delay if neither key is pressed the display returns to the home screen, any change is not saved.

Preface

Compliance
The Declaration of Conformity is available upon request.
Unpacking
The bath is supplied with an electrical power cord. Do not discard the packaging until the cord is located and the bath is operating.
If the bath shows external or internal damage contact the transportation company and file a damage claim. Under ICC regulations, this is your responsibility.
The bath does not have handles. Take into account its weight when unpacking and transporting. We recommend two people lift heavier baths from the bottom.

After-sale Support
Fisher Scientific is committed to customer service both during and after the sale. If you have questions concerning the bath operation, or questions concerning spare parts or  Service Contracts, call our Sales, Service and Customer Support.
Before calling, please obtain the following information:

  • bath model number
  • bath serial number
  • power source voltage

The bath’s model number and serial number are located on the nameplate label on the rear of the bath.

Warranty

The Isotemp® products distributed by Fisher Scientific (“Fisher”), Fisher warrants to the direct purchaser that the product will be free from defects in material or  workmanship for a period of one year from the date of delivery. Fisher will repair or replace the product or provide credit, as its sole option, upon prompt notification and  compliance with its instructions.
The Distributor warrants to Customer that upon prompt notification and compliance with Distributor’s instructions, that the Distributor will repair or replace, at  Distributor’s sole option, any Product which is defective in material or workmanship.
Distributor expressly disclaims all other warranties, whether expressed, implied or statutory, including the warranties of merchantability, and fitness for a particular purpose.  Distributor’s sole responsibility and the Customer’s exclusive remedy for any claim arising out of the purchase of any Product is repair or replacement, as described above. In  no event shall Distributor’s liability exceed the purchase price paid therefore; nor shall Distributor be liable for any claims, losses or damage of any third party or for lost  profits or any special, indirect, incidental, consequential, or exemplary damages, howsoever arising, even if Distributor has been advised of the possibility of such damages.

Chapter 1 Safety

Safety Factors
Make sure you read and understand all instructions and safety precautions listed in this manual before installing or operating your bath. If you have any questions concerning  the operation of your bath or the information in this manual, please contact us.
The are no special personal protective equipment requirements needed to perform normal operation. We do recommend always wearing eye protection and gloves.
DANGER indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It is also be used to alert against unsafe practices.
The lightning flash with arrow symbol, within an equilateral triangle, is intended to alert the user to the presence of non-insulated “dangerous voltage” within the bath’s  enclosure. The voltage magnitude is significant enough to constitute a risk of electrical shock.
This label indicates the presence of hot surfaces.
This label indicates read the manual.
This label indicates a hand crushing hazard. (Shaking Water Bath only)
Observe all warning labels.
Never remove warning labels.

The bath’s construction provides protection against the risk of electrical shock by grounding appropriate metal parts. The protection will not function unless the power cord  is connected to a properly grounded outlet. It is the user’s responsibility to assure a proper ground connection is provided.
Do not use the bath as a sterile or patient connected device. In addition, the bath is not designed for use in Class I, II or III locations as defined by the National Electrical Code.
The circuit protector located on the rear of the bath is not intended to act as a disconnecting means. The bath’s power cord is used as the disconnecting device, it must be  easily accessible at all times.

Never operate the bath with a damaged line cord.
Never place the bath in a location or atmosphere where excessive heat, moisture, or corrosive materials are present.
Never operate the bath without fluid in the reservoir. The only approved fluid is water. General Purpose Baths may also use lab armor beads.
Never operate the bath or add fluid to the reservoir with panels removed.
Operating the baths at high temperature will cause condensation on the underside of the cover. Remove
the cover with care so the water falls back into the reservoir and not on the bath surface.
When operating at high temperatures do not touch the cover surface, always use the cover’s knob/handle.
The user is responsible for any hazard rising from heated material.
Do not clean the bath with solvents, only use a soft cloth and water.
Drain the bath before it is moved. Drain the bath before it is transported and/or stored. Store the bath in the temperature range -25°C to 60°C (with packaging), and <80%  relative humidity.
Always turn the bath off and disconnect the supply voltage from its power source before moving the bath or before performing any service or maintenance procedures.
Transport the bath with care. Sudden jolts or drops can damage its components.
An operating shaking water bath has moving components. Do not insert appendages between the reservoir wall and rack. We recommend stopping the shaker table when  loading/removing samples or vessels.
Refer service and repairs to a qualified technician.
Performance of installation, operation, or maintenance procedures other than the ones described in this manual may result in a hazardous situation and will void the  manufacturer’s warranty.

Chapter 2 General Information

Description
All Fisher Scientific™
Isotemp Water Baths provide constant temperature to applications placed in the reservoir. All have digital displays, four programmable set point temperatures, acoustic and
optical alarms, and offer adjustable high temperature protection.
The interior of the bath is constructed of stainless steel and is designed for operation with water.
The body is made from galvanized steel and is painted for added protection. A drain is located at the far left hand end of the bath. A stainless steel gable cover, required to  maintain optimal temperature sensitivity, is also provided with the bath.
The nameplate label on the bath identifies its electrical requirements.
Intended Use
The baths are intended for use in research and quality control.
They are intended for use by a qualified lab technician trained in basic laboratory procedures and safety protocols. Also, for indoor use in a laboratory environment on a bench top only.

Shaking Water Bath Specifications

| SWB 15| | | SWB 27
---|---|---|---|---
Reservoir Fluid Control
Temperature °C
°F| | Ambient +5 to 100
Ambient +9 to 212|
Set Point Temperature
Range°C
°F| | +5 to 100
+9 to 212|
Ambient Temperature °C
Range °F| | +15 to 45
+59 to 113|
Stability/Unifonnity
IP 37°C| | ±0.1/±0.05|
Heating Output watts| 1200| | | 1500
Bath Volume liters| 15| | | 27
Overall Bath Dimensions
(L x W x H) cm
inches| 39.4 x 63.2 x 24.9
15.5 x 24.9 x 9.8| | | 39.4 x 93.8 x 24.9
15.5 x 36.9 x 9.8
Bath Work Area
Dimensions
(L x W x H) cm
inches| 29.2 x 30.5 x 16.5
11.5 x 12.0 x 6.5| | | 29.2 x 61.0 x 16.5
11.5 x 24.0 x 6.5
Tray Dimensions
(L x W x D) cm
inches| 28.9 x 30.8 x 14.8
11.3 x 12.1 x5.8| | | 28.9 x 61.5 x 14.8
113 x 24.2 x 5.8
Tray Depth cm
inches| | 8 9
3.5|
Shaking Speed rpm| | 0 and 30 – 200|
Stroke Length mm| | 26|
Max Tray Load kg
lb| 9
20| | | 14
30
Bath Weight kg
lb| 22.2
48.0| | | 27.2
62.0
Electrical
Requirements (VAC/Hz)
(Voltage ±10%)| | 100-115/50-60
or
200-230/50-60|
Compliance| | CE RoHS
UL WEEE|
Maximum Relative
Humidity
(Non Condensing)| | 80% (up to 31°C)
80% (up to 88°F)|
Operating Altitude meters
feet| | Sea Level to 2000
Sea Level to 6560|
Overvoltage Category| | II|
Pollution Degree| | 2|
Storage Temperature °C
Range °F| | -25 to +60
-13 to +140|

Low-end temperatures require supplemental cooling.
Specifications obtained at sea level using water.
Bath depth includes bezel, height does not include lid.
Fisher Scientific takes no responsibility for damages caused by the selection of an unapproved fluids.
Fisher Scientific reserves the right to change specifications without notice.

General Purpose Water Bath Specifications

Reservoir Fluid Control
Temperature °C
°F| GPD 02
Ambient to 90
Ambient to 194| GPD 2S 1 GPD °5 1 GPD 10
---|---|---
Ambient to 100
Ambient k 212
Set Point Temperature
Range°C
°F| +5 to 100
+9 to 212
Ambient Temperature °C
Range °F| +15 to 45
+59 to 113
Stability @ 37°C| ±0.1
Uniformity @ 37°C| ±0.2
Heating Output watts| 200| 300| 300| 800
Bath Volume liters| 2| 2| 5| 10
Overall Bath Dimensions
(L xW x H) cm
inches| 23.0 x 19.9 x 23.3
9.1 x 7.8 x 9.2| 24.6 x 35.5 x 23.2
9.7 x 14.0 x 9.1| 24.6 x 35.5 x 23.2
9.7 x 14.0 x 9.1| 39.3 x 38.3 x 23.3
15.5 x 15.1 x9.2
Bath Work Area
Dimensions
(L xW x H) cm
inches| 13.8 x 15.5 x 15.0
5.4 x 6.1 x5.9| 15.3 x30.0 x6.5
6.0 x 11.8 x 2.6| 15.4 x 30.0 x 15.0 6.1 x 11.8 x 5.9| 30.1 x 33.0 x 15.0
11.9 x 13.0 x 5.9
Approximate Weight kg
lb| 3.5
7| 4
9| 4.5
10| 7.5
16
Electrical
Requirements (VAC/Hz)
(Voltage ±10%)| 100-115/50-60
or
200-230/50-60
Compliance| CE RoliS
UL WEEE
Maximum Relative
Humidity
(Non Condensing)| 80% (up to 31°C)
80% (up to 88°F)
Operating Altitude meters
feet| Sea Level to 2000
Sea Level to 6560
Overvdtage Category| II
Pollution Degree| 2
Storage Temperature °C
Range °F| -25 to +60
-13 to +140

Low-end temperatures require supplemental cooling.
Specifications obtained at sea level using water.
Thermal beads may be used instead of water but they will degrade the bath’s uniformity and stability.
Bath depth includes bezel, height does not include lid.
Fisher Scientific takes no responsibility for damages caused by the selection of an unapproved fluids.
Fisher Scientific reserves the right to change specifications without notice.

Chapter 3 Installation

Bath Installation
The bath is designed for continuous operation and for indoor use.
**Never place the bath in a location where excessive heat, moisture or corrosive materials are present.
Ventilation
No special ventilation clearances are required.
Cover Installation**

Electrical Requirements
Refer to the bath’s nameplate for specific electrical requirements.
The bath is intended for use on a dedicated outlet.
The bath construction provides protection against the risk of electrical shock by grounding appropriate metal parts. The protection will not function unless the power cord is  connected to a properly grounded outlet. It is the user’s responsibility to assure a proper ground connection is provided.
The circuit protector on the rear of the bath is designed to protect the bath’s internal components.
Note If the circuit protector activates allow the temperature to cool before resetting. Restart the bath. Contact us if the circuit protector activates again.
The bath’s electrical power cord is the disconnecting device, it must be easily accessible at all times.
****Ensure the cord does not come in contact with the reservoir contents.
Note Before inserting the electrical cord into the bath connection ensure the circuit protector is in the O (off) position.
Once the cord is connected to the bath, connect the other end to the main power source.

Approved Fluids

The only approved fluids are:
Filtered/Single Distilled Water
De-ionized water
For applications requiring resistivity greater than 1 MΩ-cm or maintaining resistivity levels greater than 1 MΩ-cm please call and speak to an applications engineer for  additional information.
Fisher Scientific takes no responsibility for damages caused by the selection of an unapproved fluid.
When using water above 80°C closely monitor the fluid level, frequent top-offs will be required. It will also create steam.
Filtered/Single Distilled Water
Filtered drinking water and single distilled water are good choices for heated water baths. The filtering/distilling process removes microorganisms that are known to cause  biological fouling in water baths. The filtering/distilling process also removes minerals and harmful particulates
from water. Microorganisms, minerals and particulates can lead to deposits and or scaling that eventually promotes corrosion if not removed or treated.
It is also recommended to regularly drain and replace fluid especially when higher temperatures are utilized (above 40°C). Draining and drying the bath after each use is  recommended as leaving standing water for any period of time is known to lead to a buildup of biologic growth.

Tap Water
Usage of tap water may not cause any adverse affects on the equipment in the short term, but in the long term problems may arise due to precipitation of minerals and or  development of algae and bacteria. Please refer to “Process Water Quality and Standard” recommendations in this chapter for guidelines on water usage.
Chlorine
Draining and replacing “used” water is the best preventative maintenance for the cooling/heating equipment. Biological organisms will develop in water that is not replaced  or treated regularly. A mild treatment using an algaecide can help extend the useful life of water.
To help alleviate bacteria or algae growth Fisher Scientific recommends the use of small amounts of chlorine. The usage of chlorine needs to be monitored over time in order  to prevent the formation of corrosion.
The duration of time that chlorine remains in solution depends on factors such as water temperature, pH and availability of direct sunlight. We recommend maintaining  chlorine levels at 1 to 5 ppm free chlorine. PPM levels can be monitored using standard water quality test strips.
For best results, maintain the pH of the fluid between 6.5 and 7.5.
Do not add additional chlorine without first determining the concentration ratio that already exists in the fluid supply. Corrosion and degradation of the circulation  components can result from concentration ratios that are too high. Contact our customer support for additional information.
De-ionized Water
De-ionized water is water that has had its mineral ions removed using ion exchange resins. The purpose of this process is to remove the ions that allow electrical current to  flow more easily through water. This helps to prevent electrical leaks to ground through the recirculating fluid.
De-ionized water is in an unbalanced state and will leach the missing ions from the materials it comes in contact with. The aggressive nature of this leaching can cause pitting  on metal surfaces. Note that the de-ionizing process does not remove microorganisms. Because of this, we recommend de- ionized water only with applications that have it as  a specified requirement. In any case, only de-ionized water with 1 MΩ-cm resistivity, or less, is recommended. Recommended Biocides and Inhibitors
Fisher Scientific offers a biocide and inhibitor package Thermo 200 (Nalco) premixed with five gallons of water or as a kit to be added to water.
****Biocides are corrosive and can cause irreversible eye damage and skin burns.
They are harmful if inhaled, swallowed or absorbed through the skin. Refer to the
manufacturer’s most current MSDS.

Process Fluid Permissible (PPM) Desirable (PPM)

Microbiologicals
(algae, bacteria, fungi)| 0| 0
Inorganic Chemicals
Calcium| <25| <0.6
Chloride| <25| <100
Copper| <13| <10
0.020 ppm if fluid in contact with aluminum
Iron| <0.3| <0.1
Lead| <0.015| 0
Magnesium| <12|  <0.1
Manganese| <0.05| <0.03
Nitrates\Nitrites| <10 as N| 0
Potassium| <20| <0.3
Silicate| <25| <1.0
Sodium| <20| <0.3
Sulfate| <25| <1
Hardness| <17| <0.05
Total Dissolved Solids| <50| <10
Other Parameters
pH| 6.5-8.5| 8-Jul
Resistivity| 0.01| 0.05-0.1

* MΩ-cm (compensated to 25°C)
Unfavorably high total ionized solids (TIS) can accelerate the rate of corrosion. These contaminants function as electrolytes which increase the potential for galvanic cell corrosion.
Tap water is not normally recommended because the total ionized solids level may be too high. As an example, tap water in the United States averages 171 ppm (of NaCl).  The recommended level for use in a water system is between 0.5 to 5.0 ppm (of NaCl).

Filling Requirements
Ensure the reservoir drain port is closed and that all plumbing connections are securely plumbed or capped. Also ensure any residue is thoroughly removed from the  reservoir before filling.
To avoid spilling, place your samples/trays into the bath before filling. Also make allowance for splashing of water from tray oscillation movement.
Slowly fill the reservoir. When adding, point the opening of a container away from yourself.
For Shaking Water Baths, fill the reservoir with a minimum of one inch of water and the maximum level lower than the tray shaft fitting.
For General Purpose Baths, fill the reservoir with a minimum of one inch of water and a maximum level one inch lower than the tank upper surface.
Note Monitor the fluid level whenever heating the fluid.
To conserve energy, reduce evaporation, and increase temperature control accuracy, use the supplied gable cover. Do not use aluminum foil as a cover, as it may cause  corrosion due to an electrochemical reaction.

Filling with Thermal Lab Beads (GPD baths only)
Before filling the reservoir with beads ensure the reservoir plug is securely installed into the drain.
Note Using beads derates the bath’s stability and uniformity capability.

Draining
Drain the bath before moving or storing.
Ensure the fluid is at a safe handling temperature, ~40°C or lower. Wear protective clothing and gloves.

  • Place a suitable receptacle underneath the drain.
  • Slowly turn the drain plug until flow is observed.
  • When the flow stops close the drain plug.

Draining (GPD 10, GPD 20, GPD 28 and GPD 15D only)

  • Remove the drain hose from its housing and insert the fitting into the drain quick disconnect.
  • When the flow stops remove the hose and reinsert it back into the housing.

Accessories
Only use the accessories supplied by Fisher Scientific.
Drain Kit Accessory

  • Before filling the bath remove the drain plug from the bath drain.
  • Wrap coupler with Teflon® tape, or equivalent, and install into bath drain.
  • Insert quick disconnect into hose.
  • To start draining snap quick disconnect onto coupler.
  • To remove quick disconnect from coupler and stop draining press down on the disconnect’s grey release.

Chapter 4 Operation

Controller
The Fisher Scientific Isotemp Water Baths have digital controllers that display the bath’s reservoir fluid temperature, shaker speed and other bath features.

Once the circuit protector on the back of the bath is on, press to start/stop the bath.
Press the two navigation arrows to move through the controller displays and to adjust values Press a soft key to select additional displays.

| Indicates the heater is on. The icon flashes when the reservoir fluid temperature is near or at the set point.
---|---
| Indicates the timer is enabled. The timer is used to turn the bath on, off, or both.
| Indicates the audible alarm status, enabled/disabled.
| Indicates the shaker mode is enabled, SWB only.

Start Up
Before starting, double check all electrical connections.
An operating shaking water bath has moving components. Do not insert appendages between the reservoir wall and rack. We recommend stopping the shaker table when loading/removing samples or vessels.
For shaking water baths, ensure all glass vessels are secure. Unrestrained vessels may shift and break when shaking is engaged. Flask clips and trays are available from
Fisher Scientific.
Do not run the bath until fluid is added to the reservoir. Have extra fluid on hand. If the bath does not start refer to Chapter 6 Troubleshooting.

  • Place the circuit protector located on the rear of the bath to the I position.
  • The screen will momentarily display and then the home screen appears displaying the set point and the audible alarm status. The set point is the desired reservoir fluid  temperature.
  • Press to start the bath and display the reservoir fluid temperature. If the fluid temperature is below the set point the heater will start.

appears on Shaking Water Baths only.

Viewing/Changing the Shaker Speed (SWB only)
You can view/change the shaker speed value with the bath running or not. From the home screen press the soft key below to display:

Press, and hold, the arrow keys to bring up the desired value. The values are 0 and 30 to 200. Once the desired value is displayed press or to save the change and  return to the home screen.

Changing the Set Point
The set point is the desired reservoir fluid temperature. You can change the set point with the bath running or not. From the home screen press either arrow key to display:
****When operating without a lid, limit the maximum set point to 60°C.

Press, and hold, the arrow keys to bring up the desired set point value. The range is 5°C to 104°C. Once the desired value is displayed, in this case 50.0°C, press the soft key  below to save the change or press the soft key below to not save the change.
Note After a 30 second delay if neither key is pressed the display returns to the home screen, any change is not saved.

Electrical Power Cord
Ensure any replacement cord is properly rated.
Optical Disc (SWB only)
Check the optical disc monthly and verify it is free of foreign material. If not, wipe the disc clean.

  1. Remove the four screws, two on top and two on the side securing the top right panel.

  2. Swing panel assembly open.
    Do not operate the bath with the panel open.

  3. Rotate disc, by grasping aluminum components of the eccentric only, one turn clockwise and verify that disc does not rub against optical sensor. If required, use the  3/32″ Allen to loosen two set screws on the eccentric and reposition the disc.

Shaker Assembly
Check the shaker assembly for looseness annually. Contact Fisher Scientific if excessive movement is noted.

Chapter 6 Troubleshooting

Error Displays
An error display indicates an unusual condition. With any message, except CAL, the bath will stop heating and, if applicable, the pump will stop. With any message the alarm, if enabled, will sound.

Display Cause/Action
HTC Fault

fixed high temp protection limit exceeded
allow bath to cool down
restart the bath
if the HTC fault cannot be cleared, the bath must be serviced by an authorized Fisher
Scientific Temperature Control Service Technician.
| High Temp Fault
adjustable high temp fault protection limit exceeded
check limit setting
| Low Fluid Level
low level protection limit exceeded
check fluid level
check for leaks
| Open/Shorted Internal Temp Sensor
the bath must be serviced by an authorized Fisher Scientific Temperature Control Service
Technician.
| Bad Calibration
redo calibration
ensure calibration set points are 35.0°C and 70.0°C
| Shaker Malfunction (SWB only)
ensure there is nothing in the reservoir hampering shaker movement.
the bath must be serviced by an authorized Fisher Scientific Temperature Control Service
Technician.

Checklist
Bath will not start or shuts down

  • Check display for error messages.
  • Ensure wasn’t accidently pressed.
  • Ensure the circuit protector is in the on ( I ) position.
  • Check the line cord connection to your power supply and at the bath.
  • Make sure supply voltage is connected and matches the bath’s nameplate rating ±10%.
  • Restart the bath .

No display

  • Pressing should return the display.
  • Cycle the bath’s circuit protector.

Inadequate temperature control

  • Verify the set point.
  • Low-end temperature set points require supplemental cooling.
  • GPD 02 baths require insulation, especially in the lid, for operation above 90°C.
  • Ensure bath installation complies with the site requirements in Chapter 3.
  • Make sure supply voltage matches bath’s nameplate rating ±10%.

Please contact Fisher Scientific Sales Service and Customer Support if you need any additional information.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Fisher Scientific User Manuals

Related Manuals