Fisher Scientific Isotemp Shaking and General Purpose Water Baths User Manual
- June 3, 2024
- Fisher Scientific
Table of Contents
Isotemp Shaking and General Purpose Water Baths
User Manual
Manual P/N U01318
Rev. 11/05/2015 ****
Isotemp Water Bath Quick Start Guide
This quick start guide is intended for initial start up only. For all other
procedures you must refer to the manual. Also, if any of these steps are not
clear refer to the manual before proceeding.
Filling (Only approved fluids are Filtered/Single Distilled Water and De-
ionized Water)
Ensure the reservoir drain port is closed and that all plumbing connections
are securely plumbed or capped. Also ensure any residue is thoroughly removed
from the reservoir before filling.
To avoid spilling, place your samples/trays into the bath before filling.
Slowly fill the reservoir. When adding, point the opening of a container away
from yourself.
For Shaking Water Baths, fill the reservoir with a minimum of one inch of
water and the maximum level lower than the tray shaft fitting.
Make allowance for water splashing from tray oscillations.
For GPD Baths, fill the reservoir with a minimum of one inch of water and a
maximum level one inch lower than the tank upper surface.
Starting
Do not run the bath until fluid is added to the reservoir. Have extra fluid on hand. If the bath does not start refer to the manual.
- Place the circuit protector located on the rear of the bath to the I position.
- The screen will momentarily display and then the home screen appears displaying the set point and the audible alarm status. The set point is the desired reservoir fluid temperature.
Viewing/Changing the Shaker Speed (SWB only)
You can view/change the speed value with the bath running or not. From the
home screen press the soft key below to display:
Press, and hold, the arrow keys to bring up the desired value. The values are
0 and 30 to 200. Once the desired value is displayed press or to save the
change and return to the home screen.
Changing the Set Point
When operating without a lid, limit the maximum set point to 60°C.
You can change the set point with the bath running or not. From the home
screen press either arrow key to display:
Press, and hold, the arrow keys to bring up the desired set point value. The
range is 5°C to 100°C. Once the desired value is displayed, in this case
50.0°C, press the soft key below to save the change or press the soft key
below to not save the change. The display returns to the home screen. Note
After a 30 second delay if neither key is pressed the display returns to the
home screen, any change is not saved.
Selecting a Preset
You can select a preset with the bath running or not.
From the home screen press:
Use the arrow keys to select the desired preset. Once the desired preset is
highlighted press the soft key below to save the change and return to the home
screen. Note After a 30 second delay if neither key is pressed the display
returns to the home screen, any change is not saved.
Preface
Compliance
The Declaration of Conformity is available upon request.
Unpacking
The bath is supplied with an electrical power cord. Do not discard the
packaging until the cord is located and the bath is operating.
If the bath shows external or internal damage contact the transportation
company and file a damage claim. Under ICC regulations, this is your
responsibility.
The bath does not have handles. Take into account its weight when unpacking
and transporting. We recommend two people lift heavier baths from the bottom.
After-sale Support
Fisher Scientific is committed to customer service both during and after the
sale. If you have questions concerning the bath operation, or questions
concerning spare parts or Service Contracts, call our Sales, Service and
Customer Support.
Before calling, please obtain the following information:
- bath model number
- bath serial number
- power source voltage
The bath’s model number and serial number are located on the nameplate label on the rear of the bath.
Warranty
The Isotemp® products distributed by Fisher Scientific (“Fisher”), Fisher
warrants to the direct purchaser that the product will be free from defects in
material or workmanship for a period of one year from the date of delivery.
Fisher will repair or replace the product or provide credit, as its sole
option, upon prompt notification and compliance with its instructions.
The Distributor warrants to Customer that upon prompt notification and
compliance with Distributor’s instructions, that the Distributor will repair
or replace, at Distributor’s sole option, any Product which is defective in
material or workmanship.
Distributor expressly disclaims all other warranties, whether expressed,
implied or statutory, including the warranties of merchantability, and fitness
for a particular purpose. Distributor’s sole responsibility and the
Customer’s exclusive remedy for any claim arising out of the purchase of any
Product is repair or replacement, as described above. In no event shall
Distributor’s liability exceed the purchase price paid therefore; nor shall
Distributor be liable for any claims, losses or damage of any third party or
for lost profits or any special, indirect, incidental, consequential, or
exemplary damages, howsoever arising, even if Distributor has been advised of
the possibility of such damages.
Chapter 1 Safety
Safety Factors
Make sure you read and understand all instructions and safety precautions
listed in this manual before installing or operating your bath. If you have
any questions concerning the operation of your bath or the information in
this manual, please contact us.
The are no special personal protective equipment requirements needed to
perform normal operation. We do recommend always wearing eye protection and
gloves.
DANGER indicates an imminently hazardous situation which, if not avoided,
will result in death or serious injury.
WARNING indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION indicates a potentially hazardous situation which, if not
avoided, may result in minor or moderate injury. It is also be used to alert
against unsafe practices.
The lightning flash with arrow symbol, within an equilateral triangle, is
intended to alert the user to the presence of non-insulated “dangerous
voltage” within the bath’s enclosure. The voltage magnitude is significant
enough to constitute a risk of electrical shock.
This label indicates the presence of hot surfaces.
This label indicates read the manual.
This label indicates a hand crushing hazard. (Shaking Water Bath only)
Observe all warning labels.
Never remove warning labels.
The bath’s construction provides protection against the risk of electrical
shock by grounding appropriate metal parts. The protection will not function
unless the power cord is connected to a properly grounded outlet. It is the
user’s responsibility to assure a proper ground connection is provided.
Do not use the bath as a sterile or patient connected device. In addition, the
bath is not designed for use in Class I, II or III locations as defined by the
National Electrical Code.
The circuit protector located on the rear of the bath is not intended to act
as a disconnecting means. The bath’s power cord is used as the disconnecting
device, it must be easily accessible at all times.
Never operate the bath with a damaged line cord.
Never place the bath in a location or atmosphere where excessive heat,
moisture, or corrosive materials are present.
Never operate the bath without fluid in the reservoir. The only approved fluid
is water. General Purpose Baths may also use lab armor beads.
Never operate the bath or add fluid to the reservoir with panels removed.
Operating the baths at high temperature will cause condensation on the
underside of the cover. Remove
the cover with care so the water falls back into the reservoir and not on the
bath surface.
When operating at high temperatures do not touch the cover surface, always use
the cover’s knob/handle.
The user is responsible for any hazard rising from heated material.
Do not clean the bath with solvents, only use a soft cloth and water.
Drain the bath before it is moved. Drain the bath before it is transported
and/or stored. Store the bath in the temperature range -25°C to 60°C (with
packaging), and <80% relative humidity.
Always turn the bath off and disconnect the supply voltage from its power
source before moving the bath or before performing any service or maintenance
procedures.
Transport the bath with care. Sudden jolts or drops can damage its components.
An operating shaking water bath has moving components. Do not insert
appendages between the reservoir wall and rack. We recommend stopping the
shaker table when loading/removing samples or vessels.
Refer service and repairs to a qualified technician.
Performance of installation, operation, or maintenance procedures other than
the ones described in this manual may result in a hazardous situation and will
void the manufacturer’s warranty.
Chapter 2 General Information
Description
All Fisher Scientific™
Isotemp Water Baths provide constant temperature to applications placed in the
reservoir. All have digital displays, four programmable set point
temperatures, acoustic and
optical alarms, and offer adjustable high temperature protection.
The interior of the bath is constructed of stainless steel and is designed for
operation with water.
The body is made from galvanized steel and is painted for added protection. A
drain is located at the far left hand end of the bath. A stainless steel gable
cover, required to maintain optimal temperature sensitivity, is also provided
with the bath.
The nameplate label on the bath identifies its electrical requirements.
Intended Use
The baths are intended for use in research and quality control.
They are intended for use by a qualified lab technician trained in basic
laboratory procedures and safety protocols. Also, for indoor use in a
laboratory environment on a bench top only.
Shaking Water Bath Specifications
| SWB 15| | | SWB 27
---|---|---|---|---
Reservoir Fluid Control
Temperature °C
°F| | Ambient +5 to 100
Ambient +9 to 212|
Set Point Temperature
Range°C
°F| | +5 to 100
+9 to 212|
Ambient Temperature °C
Range °F| | +15 to 45
+59 to 113|
Stability/Unifonnity
IP 37°C| | ±0.1/±0.05|
Heating Output watts| 1200| | | 1500
Bath Volume liters| 15| | | 27
Overall Bath Dimensions
(L x W x H) cm
inches| 39.4 x 63.2 x 24.9
15.5 x 24.9 x 9.8| | | 39.4 x 93.8 x 24.9
15.5 x 36.9 x 9.8
Bath Work Area
Dimensions
(L x W x H) cm
inches| 29.2 x 30.5 x 16.5
11.5 x 12.0 x 6.5| | | 29.2 x 61.0 x 16.5
11.5 x 24.0 x 6.5
Tray Dimensions
(L x W x D) cm
inches| 28.9 x 30.8 x 14.8
11.3 x 12.1 x5.8| | | 28.9 x 61.5 x 14.8
113 x 24.2 x 5.8
Tray Depth cm
inches| | 8 9
3.5|
Shaking Speed rpm| | 0 and 30 – 200|
Stroke Length mm| | 26|
Max Tray Load kg
lb| 9
20| | | 14
30
Bath Weight kg
lb| 22.2
48.0| | | 27.2
62.0
Electrical
Requirements (VAC/Hz)
(Voltage ±10%)| | 100-115/50-60
or
200-230/50-60|
Compliance| | CE RoHS
UL WEEE|
Maximum Relative
Humidity
(Non Condensing)| | 80% (up to 31°C)
80% (up to 88°F)|
Operating Altitude meters
feet| | Sea Level to 2000
Sea Level to 6560|
Overvoltage Category| | II|
Pollution Degree| | 2|
Storage Temperature °C
Range °F| | -25 to +60
-13 to +140|
Low-end temperatures require supplemental cooling.
Specifications obtained at sea level using water.
Bath depth includes bezel, height does not include lid.
Fisher Scientific takes no responsibility for damages caused by the selection
of an unapproved fluids.
Fisher Scientific reserves the right to change specifications without notice.
General Purpose Water Bath Specifications
Reservoir Fluid Control
Temperature °C
°F| GPD 02
Ambient to 90
Ambient to 194| GPD 2S 1 GPD °5 1 GPD 10
---|---|---
Ambient to 100
Ambient k 212
Set Point Temperature
Range°C
°F| +5 to 100
+9 to 212
Ambient Temperature °C
Range °F| +15 to 45
+59 to 113
Stability @ 37°C| ±0.1
Uniformity @ 37°C| ±0.2
Heating Output watts| 200| 300| 300| 800
Bath Volume liters| 2| 2| 5| 10
Overall Bath Dimensions
(L xW x H) cm
inches| 23.0 x 19.9 x 23.3
9.1 x 7.8 x 9.2| 24.6 x 35.5 x 23.2
9.7 x 14.0 x 9.1| 24.6 x 35.5 x 23.2
9.7 x 14.0 x 9.1| 39.3 x 38.3 x 23.3
15.5 x 15.1 x9.2
Bath Work Area
Dimensions
(L xW x H) cm
inches| 13.8 x 15.5 x 15.0
5.4 x 6.1 x5.9| 15.3 x30.0 x6.5
6.0 x 11.8 x 2.6| 15.4 x 30.0 x 15.0 6.1 x 11.8 x 5.9| 30.1 x 33.0 x 15.0
11.9 x 13.0 x 5.9
Approximate Weight kg
lb| 3.5
7| 4
9| 4.5
10| 7.5
16
Electrical
Requirements (VAC/Hz)
(Voltage ±10%)| 100-115/50-60
or
200-230/50-60
Compliance| CE RoliS
UL WEEE
Maximum Relative
Humidity
(Non Condensing)| 80% (up to 31°C)
80% (up to 88°F)
Operating Altitude meters
feet| Sea Level to 2000
Sea Level to 6560
Overvdtage Category| II
Pollution Degree| 2
Storage Temperature °C
Range °F| -25 to +60
-13 to +140
Low-end temperatures require supplemental cooling.
Specifications obtained at sea level using water.
Thermal beads may be used instead of water but they will degrade the bath’s
uniformity and stability.
Bath depth includes bezel, height does not include lid.
Fisher Scientific takes no responsibility for damages caused by the selection
of an unapproved fluids.
Fisher Scientific reserves the right to change specifications without notice.
Chapter 3 Installation
Bath Installation
The bath is designed for continuous operation and for indoor use.
**Never place the bath in a location where excessive heat, moisture or
corrosive materials are present.
Ventilation
No special ventilation clearances are required.
Cover Installation**
Electrical Requirements
Refer to the bath’s nameplate for specific electrical requirements.
The bath is intended for use on a dedicated outlet.
The bath construction provides protection against the risk of electrical
shock by grounding appropriate metal parts. The protection will not function
unless the power cord is connected to a properly grounded outlet. It is the
user’s responsibility to assure a proper ground connection is provided.
The circuit protector on the rear of the bath is designed to protect the
bath’s internal components.
Note If the circuit protector activates allow the temperature to cool before
resetting. Restart the bath. Contact us if the circuit protector activates
again.
The bath’s electrical power cord is the disconnecting device, it must be
easily accessible at all times.
****Ensure the cord does not come in contact with the reservoir contents.
Note Before inserting the electrical cord into the bath connection ensure the
circuit protector is in the O (off) position.
Once the cord is connected to the bath, connect the other end to the main
power source.
Approved Fluids
The only approved fluids are:
Filtered/Single Distilled Water
De-ionized water
For applications requiring resistivity greater than 1 MΩ-cm or maintaining
resistivity levels greater than 1 MΩ-cm please call and speak to an
applications engineer for additional information.
Fisher Scientific takes no responsibility for damages caused by the selection
of an unapproved fluid.
When using water above 80°C closely monitor the fluid level, frequent top-offs
will be required. It will also create steam.
Filtered/Single Distilled Water
Filtered drinking water and single distilled water are good choices for heated
water baths. The filtering/distilling process removes microorganisms that are
known to cause biological fouling in water baths. The filtering/distilling
process also removes minerals and harmful particulates
from water. Microorganisms, minerals and particulates can lead to deposits and
or scaling that eventually promotes corrosion if not removed or treated.
It is also recommended to regularly drain and replace fluid especially when
higher temperatures are utilized (above 40°C). Draining and drying the bath
after each use is recommended as leaving standing water for any period of
time is known to lead to a buildup of biologic growth.
Tap Water
Usage of tap water may not cause any adverse affects on the equipment in the
short term, but in the long term problems may arise due to precipitation of
minerals and or development of algae and bacteria. Please refer to “Process
Water Quality and Standard” recommendations in this chapter for guidelines on
water usage.
Chlorine
Draining and replacing “used” water is the best preventative maintenance for
the cooling/heating equipment. Biological organisms will develop in water that
is not replaced or treated regularly. A mild treatment using an algaecide can
help extend the useful life of water.
To help alleviate bacteria or algae growth Fisher Scientific recommends the
use of small amounts of chlorine. The usage of chlorine needs to be monitored
over time in order to prevent the formation of corrosion.
The duration of time that chlorine remains in solution depends on factors such
as water temperature, pH and availability of direct sunlight. We recommend
maintaining chlorine levels at 1 to 5 ppm free chlorine. PPM levels can be
monitored using standard water quality test strips.
For best results, maintain the pH of the fluid between 6.5 and 7.5.
Do not add additional chlorine without first determining the concentration
ratio that already exists in the fluid supply. Corrosion and degradation of
the circulation components can result from concentration ratios that are too
high. Contact our customer support for additional information.
De-ionized Water
De-ionized water is water that has had its mineral ions removed using ion
exchange resins. The purpose of this process is to remove the ions that allow
electrical current to flow more easily through water. This helps to prevent
electrical leaks to ground through the recirculating fluid.
De-ionized water is in an unbalanced state and will leach the missing ions
from the materials it comes in contact with. The aggressive nature of this
leaching can cause pitting on metal surfaces. Note that the de-ionizing
process does not remove microorganisms. Because of this, we recommend de-
ionized water only with applications that have it as a specified requirement.
In any case, only de-ionized water with 1 MΩ-cm resistivity, or less, is
recommended. Recommended Biocides and Inhibitors
Fisher Scientific offers a biocide and inhibitor package Thermo 200 (Nalco)
premixed with five gallons of water or as a kit to be added to water.
****Biocides are corrosive and can cause irreversible eye damage and skin
burns.
They are harmful if inhaled, swallowed or absorbed through the skin. Refer to
the
manufacturer’s most current MSDS.
Process Fluid | Permissible (PPM) | Desirable (PPM) |
---|
Microbiologicals
(algae, bacteria, fungi)| 0| 0
Inorganic Chemicals
Calcium| <25| <0.6
Chloride| <25| <100
Copper| <13| <10
0.020 ppm if fluid in contact with aluminum
Iron| <0.3| <0.1
Lead| <0.015| 0
Magnesium| <12| <0.1
Manganese| <0.05| <0.03
Nitrates\Nitrites| <10 as N| 0
Potassium| <20| <0.3
Silicate| <25| <1.0
Sodium| <20| <0.3
Sulfate| <25| <1
Hardness| <17| <0.05
Total Dissolved Solids| <50| <10
Other Parameters
pH| 6.5-8.5| 8-Jul
Resistivity| 0.01| 0.05-0.1
* MΩ-cm (compensated to 25°C)
Unfavorably high total ionized solids (TIS) can accelerate the rate of
corrosion. These contaminants function as electrolytes which increase the
potential for galvanic cell corrosion.
Tap water is not normally recommended because the total ionized solids level
may be too high. As an example, tap water in the United States averages 171
ppm (of NaCl). The recommended level for use in a water system is between 0.5
to 5.0 ppm (of NaCl).
Filling Requirements
Ensure the reservoir drain port is closed and that all plumbing connections
are securely plumbed or capped. Also ensure any residue is thoroughly removed
from the reservoir before filling.
To avoid spilling, place your samples/trays into the bath before filling. Also
make allowance for splashing of water from tray oscillation movement.
Slowly fill the reservoir. When adding, point the opening of a container away
from yourself.
For Shaking Water Baths, fill the reservoir with a minimum of one inch of
water and the maximum level lower than the tray shaft fitting.
For General Purpose Baths, fill the reservoir with a minimum of one inch of
water and a maximum level one inch lower than the tank upper surface.
Note Monitor the fluid level whenever heating the fluid.
To conserve energy, reduce evaporation, and increase temperature control
accuracy, use the supplied gable cover. Do not use aluminum foil as a cover,
as it may cause corrosion due to an electrochemical reaction.
Filling with Thermal Lab Beads (GPD baths only)
Before filling the reservoir with beads ensure the reservoir plug is securely
installed into the drain.
Note Using beads derates the bath’s stability and uniformity capability.
Draining
Drain the bath before moving or storing.
Ensure the fluid is at a safe handling temperature, ~40°C or lower. Wear
protective clothing and gloves.
- Place a suitable receptacle underneath the drain.
- Slowly turn the drain plug until flow is observed.
- When the flow stops close the drain plug.
Draining (GPD 10, GPD 20, GPD 28 and GPD 15D only)
- Remove the drain hose from its housing and insert the fitting into the drain quick disconnect.
- When the flow stops remove the hose and reinsert it back into the housing.
Accessories
Only use the accessories supplied by Fisher Scientific.
Drain Kit Accessory
- Before filling the bath remove the drain plug from the bath drain.
- Wrap coupler with Teflon® tape, or equivalent, and install into bath drain.
- Insert quick disconnect into hose.
- To start draining snap quick disconnect onto coupler.
- To remove quick disconnect from coupler and stop draining press down on the disconnect’s grey release.
Chapter 4 Operation
Controller
The Fisher Scientific Isotemp Water Baths have digital controllers that
display the bath’s reservoir fluid temperature, shaker speed and other bath
features.
Once the circuit protector on the back of the bath is on, press to start/stop
the bath.
Press the two navigation arrows to move through the controller displays and to
adjust values Press a soft key to select additional displays.
| Indicates the heater is on. The icon flashes when the reservoir fluid
temperature is near or at the set point.
---|---
| Indicates the timer is enabled. The timer is used to turn the bath on, off,
or both.
| Indicates the audible alarm status, enabled/disabled.
| Indicates the shaker mode is enabled, SWB only.
Start Up
Before starting, double check all electrical connections.
An operating shaking water bath has moving components. Do not insert
appendages between the reservoir wall and rack. We recommend stopping the
shaker table when loading/removing samples or vessels.
For shaking water baths, ensure all glass vessels are secure.
Unrestrained vessels may shift and break when shaking is engaged. Flask clips
and trays are available from
Fisher Scientific.
Do not run the bath until fluid is added to the reservoir. Have extra fluid on
hand. If the bath does not start refer to Chapter 6 Troubleshooting.
- Place the circuit protector located on the rear of the bath to the I position.
- The screen will momentarily display and then the home screen appears displaying the set point and the audible alarm status. The set point is the desired reservoir fluid temperature.
- Press to start the bath and display the reservoir fluid temperature. If the fluid temperature is below the set point the heater will start.
appears on Shaking Water Baths only.
Viewing/Changing the Shaker Speed (SWB only)
You can view/change the shaker speed value with the bath running or not. From
the home screen press the soft key below to display:
Press, and hold, the arrow keys to bring up the desired value. The values are 0 and 30 to 200. Once the desired value is displayed press or to save the change and return to the home screen.
Changing the Set Point
The set point is the desired reservoir fluid temperature. You can change the
set point with the bath running or not. From the home screen press either
arrow key to display:
****When operating without a lid, limit the maximum set point to 60°C.
Press, and hold, the arrow keys to bring up the desired set point value. The
range is 5°C to 104°C. Once the desired value is displayed, in this case
50.0°C, press the soft key below to save the change or press the soft key
below to not save the change.
Note After a 30 second delay if neither key is pressed the display
returns to the home screen, any change is not saved.
Electrical Power Cord
Ensure any replacement cord is properly rated.
Optical Disc (SWB only)
Check the optical disc monthly and verify it is free of foreign material. If
not, wipe the disc clean.
-
Remove the four screws, two on top and two on the side securing the top right panel.
-
Swing panel assembly open.
Do not operate the bath with the panel open. -
Rotate disc, by grasping aluminum components of the eccentric only, one turn clockwise and verify that disc does not rub against optical sensor. If required, use the 3/32″ Allen to loosen two set screws on the eccentric and reposition the disc.
Shaker Assembly
Check the shaker assembly for looseness annually. Contact Fisher Scientific if
excessive movement is noted.
Chapter 6 Troubleshooting
Error Displays
An error display indicates an unusual condition. With any message, except CAL,
the bath will stop heating and, if applicable, the pump will stop. With any
message the alarm, if enabled, will sound.
Display | Cause/Action |
---|---|
HTC Fault |
fixed high temp protection limit exceeded
allow bath to cool down
restart the bath
if the HTC fault cannot be cleared, the bath must be serviced by an authorized
Fisher
Scientific Temperature Control Service Technician.
| High Temp Fault
adjustable high temp fault protection limit exceeded
check limit setting
| Low Fluid Level
low level protection limit exceeded
check fluid level
check for leaks
| Open/Shorted Internal Temp Sensor
the bath must be serviced by an authorized Fisher Scientific Temperature
Control Service
Technician.
| Bad Calibration
redo calibration
ensure calibration set points are 35.0°C and 70.0°C
| Shaker Malfunction (SWB only)
ensure there is nothing in the reservoir hampering shaker movement.
the bath must be serviced by an authorized Fisher Scientific Temperature
Control Service
Technician.
Checklist
Bath will not start or shuts down
- Check display for error messages.
- Ensure wasn’t accidently pressed.
- Ensure the circuit protector is in the on ( I ) position.
- Check the line cord connection to your power supply and at the bath.
- Make sure supply voltage is connected and matches the bath’s nameplate rating ±10%.
- Restart the bath .
No display
- Pressing should return the display.
- Cycle the bath’s circuit protector.
Inadequate temperature control
- Verify the set point.
- Low-end temperature set points require supplemental cooling.
- GPD 02 baths require insulation, especially in the lid, for operation above 90°C.
- Ensure bath installation complies with the site requirements in Chapter 3.
- Make sure supply voltage matches bath’s nameplate rating ±10%.
Please contact Fisher Scientific Sales Service and Customer Support if you need any additional information.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>