Fisher Scientific 88861049 Multi-Purpose Tube Rotator Instruction Manual
- June 12, 2024
- Fisher Scientific
Table of Contents
Fisher Scientific 88861049 Multi-Purpose Tube Rotator Instruction Manual
Overview
Connections
Structure Diagram
Specifications
Speed
Speed Range | 5 ~80 rpm |
---|---|
Display Accuracy | 1 rpm |
Speed Accuracy | ±1 rpm |
Display | LED |
Note: Maximum speed may vary with heavy or unbalanced loads.
Angle
Angle Adjustment Range | 95°~180° |
---|---|
Step Angle | 17° |
Adjustment Method | Manual Adjustment |
Load
Maximum Capacity | 64×Φ18.5mm culture tubes (drum rotor) |
---|---|
Maximum Load (Centered on tray) | 5 kg (Clamps Included) |
Time
Range | 0, 1 min. ~ 99h59min. |
---|---|
Display | LED |
Size
Overall Dimensions | 415×293×464mm(16.3×11.5×18.3in) |
---|---|
Standard Match Dimensions | 15×50ml rotary plate (Diameter Φ290mm/11.4 in) |
Package Dimensions | 536×502×505mm(21.1×19.8×19.9in) |
Weight
Net Weight | 9.4Kg(20.7lb) |
---|---|
Gross Weight | 13.2Kg(29.0lbs) |
Power Supply
Power Supply | AC100~240V, 50Hz/60Hz, 0.2A |
---|
Others
Certificate | ROHS, WEEE, cCSAus, CE Mark |
---|
Environmental Conditions
Application Environmental Conditions: indoor use
Temperature | 5 to 40 ℃(41 to 104 ℉) |
---|---|
Humidity | 20% to 85%, |
Altitude | ≤2,000 m |
Voltage Fluctuation | ±10% of the nominal voltage |
Storage Environmental Conditions
Temperature | 0 to 60 ℃(36 to 124 ℉) |
---|---|
Humidity | 20% to 90%, non-condensing |
Safety Instructions
Please read the entire instruction manual before operating the Multi-Purpose Tube Rotator.
WARNING DO NOT use the in a hazardous atmosphere or with hazardous materials for which the unit was not designed. Also, the user should be aware that the protection provided by the equipment may be impaired if accessories used are not provided or recommended by the manufacturer, or are used in a manner not specified by the manufacturer.
CAUTION ! To avoid electrical shock, completely cut off power to the unit by disconnecting the power cord from the unit or unplug from the wall outlet.
Disconnect unit from the power supply prior to maintenance and service. Any spills should be removed promptly. Biohazard spills should be cleaned using approved liquid promptly. Solvent spills are a fire hazard. Stop the unit immediately, and DO NOT operate until clean up is complete and vapors have dissipated. Multi-Purpose Tube Rotator
DO NOT immerse the unit for cleaning.
DO NOT operate the unit if it shows signs of electrical or mechanical damage.
Note: The rotary plate maybe have a little vibration below 11 rpm, but the vibration will relieve or disapear when add some loads.
Various independent clamps are available with this Multi-Purpose Tube Rotator for convenient combination use. It is applicable for 1.5 to 50 ml microtubes to mix liquid efficiently and gently. It has an angle range of 95 to 180 degrees and can maintain suspension state of biological samples such as blood mixing. Applications include blood coagulation prevention, emulsion diagnosis, immunoprecipitation, and other related areas.