Fisher Scientific 2300 Isotemp Digital Water Bath Instruction Manual

May 15, 2024
Fisher Scientific

Fisher Scientific LOGO Fisher Scientific
lsotemp Water Bath
Installation and Operation Manual 

Visit us online to register your warranty
www.thermoscientific.com/labwarmnty

2300 Isotemp Digital Water Bath

Models covered in this manual

Catalog Number (Model)| Specific Information| Voltage
15-462-2Q (2329)| 2L| 120V
15-462-3Q (2337)| 2L| 240V
15-462-820 (2306)| SS lid| 120V
15-462-S3Q (2314)| SS lid| 230V
15-462-280 (2331)| 2L| 120V
15-462-380 (2338)| 21| 240V
15-462-S2SQ (2307)| SS lid, shallow| 120V
15-462-5380 (2315)| SS lid, shallow| 230V
15-462-50 (23401| 51| 120V
15-462-60 (2341)| 5L| 230V
15-462-S50 (2316)| SS lid| 230V
15-462-S60 (2317)| SS lid| 230V
15-462-100 (2320)| 101| 120V
15-462-110 (2321)| 101| 240V
15-462-S100 (2300)| SS lid| 120V
15-462-S11Q (2301)| Ss lid| 230V
15-462-150 (2322)| 5L/10L, dual chamber| 120V
15-462-160 (2328)| 51/101_, dual chamber| 240V
15-462-5150 (2303)| SS lid| 120V
15-462-S16Q (2304)| SS lid| 230V
15-462-200(2332)| 20L| 120V
15.462-210(2333)| 201| 240V
15-462-S20Q (2308)| SS lid| 120V
15-462-8210 (2309)| SS lid| 230V
15-462-280 (2335)| 28L| 120V
15-462-290 (2336)| 281| 240V
15-462-S28Q (2310)| SS lid| 120V
15-462-S29Q (2312)| SS lid| 230V

MANUAL NUMBER 102370 (7002300)

REV ECR/ECN DATE DESCRIPTION By
0 4/29/10 Transfer to Marietta (was 102370 6/25/10 ccs
1 26775/Sl-10360 2/’11/11 Corrected voltage specs of 15-462-60 (2341) ccs

CAUTION
Contains Parts and Assemblies Susceptible to Damage by Electrostatic Discharge (ESD)
Important Read this instruction manual. Failure to read, understand and follow the instructions in this manual may result in damage to the unit, injury to operating personnel, and poor equipment performance.
Caution All internal adjustments and maintenance must be performed by qualified service personnel.
Material in this manual is for information purposes only. The contents and the product it describes are subject to change without notice. Fisher Scientific makes no representations or warranties with
In no event shall Fisher Scientific be held liable for any damages, direct or incidental, arising respect to this manual. out of or related to the use of this manual.
©2010 Fisher Scientific. All rights reserved.

Important operating and/or maintenance instructions. Read the accompanying text carefully.
Potential electrical hazards. Only qualified persons should perform procedures associated with this symbol.
Equipment being maintained or serviced must be turned off and locked off to prevent possible injury.
Hot surface(s) present which may cause burns to unprotected skin, or to materials which may be damaged by elevated temperatures.
**** Marking of electrical and electronic equipment, which applies to electrical and electronic equipment falling under the Directive 2002/96/EC (WEEE) and the equipment that has been put on the market after 13 August 2005.

This product is required to comply with the European Union’s Waste Electrical & Electronic
Equipment (WEEE) Directive 2002/96/EC. It is marked with the WEEE symbol. Fisher Scientific has contracted with one or more recycling/ disposal companies in each EU Member State European Country, and this product should be disposed of or recycled through them. Further information on Fisher Scientific’s compliance with this directive, the recyclers in your country and information on these products will be available at www.fishersci.com.

Always use the proper protective equipment (clothing, gloves, goggles, etc.)
  Always dissipate extreme cold or heat and wear protective clothing.
Always follow good hygiene practices.
Each individual is responsible for his or her own safety.

Do You Need Information or Assistance on Fisher Scientific Equipment?
The Fisher Scientific Sales Group can provide information on pricing and give you quotations.
We can take your order and provide delivery information on major equipment items or make arrangements to have your local sales representative contact your. Our products are listed on the internet and we can be contacted through our Internet home page.

Customer Service 1-800-766-7000
To order online www.fishersci.com

Fisher Scientific Product Service Support can supply technical information about proper setup, operation or troubleshooting of your equipment. We can fill your needs for replacement parts or provide you with on-site service. We can also provide you with a quotation on our Extended Maintenance Program for our products.
Whatever products you need or use, we will be happy to discuss your applications. If you are experiencing technical problems, working together, we will help you locate the problem and, chances are, correct it yourself … over the telephone without a service call.

When more extensive service is necessary, we will assist you with direct factory trained technicians or a qualified service organization for on-the- spot repair. If your service need is covered by the Fisher Scientific Products warranty, we will arrange for the unit to be repaired at our expense and to your satisfaction.
Regardless of your needs, professional service technicians are available by telephone to assist you concerning our products Monday through Friday from 8:00 a.m. to 6:00 p.m. Eastern Time.
Please call or fax us at:

Customer Service 1-800-766-7000
Orders Fax 1-800-926-1166

International customers, please contact your local Fisher distributor.

Section 1 Introduction

Congratulations on your new Fisher Scientific Isotemp Water Bath.
Isotemp Water Baths offer the versatility needed to handle virtually any clinical laboratory procedure – incubation, inactivation, agglutination -as well as most serological, pharmaceutical, biomedical, and industrial procedures.
These water baths are available with chamber capacities of 2, 5, 10, 20, and 28 liters. The 2 liter unit is available configured with either a standard or shallow chamber. Also, a dual chamber unit is available with 5 and 10 liter chambers, with independent controls for each. All units are available to operate from either 120 or 230 volts.
The baths work equally well with water or bath oil as the medium. All models offer the same temperature ranges: ambient to 60°C without the supplied cover, and ambient to 100°C with the cover in place (within physical limitations).
These baths are microprocessor controlled for precise temperatures and reliable, trouble-free operation. Dual thermostats – one controlling the set point temperature and the other operating as a safety – virtually eliminate the possibility of overheating. Although not designed to operate dry, the bath will not be damaged if it is allowed to run dry.
An energy-saving removable cover helps to reduce evaporation while helping to maintain a uniform, constant temperature. The cover also allows the water bath to reach higher temperatures. The hinged, see­ through gable cover features unique “fins” that keep hands away from hot vapors for easier, safer opening. The cover stays open at the 90° position or, because of its quick-attach hinge, can be lifted off completely to accommodate large glassware.
A plastic rim remains cool to the touch even when operating the bath at maximum temperature.
The unit features a power-on self-test of vital circuitry. This test will automatically be performed each time the unit is turned on. If problems are detected, the system will indicate a warning message on the display.
The calibration of the water bath can be adjusted to optimize its accuracy for the temperature that is used most often in your application. This can be accomplished through a simple procedure accessed from the front panel.
As a convenience, a drain pump is supplied with 10, 20, 28, and dual 5/10 liter models.
All controls are conveniently located on the front panel for easy access. To assure set point security, the temperature can only be changed by entering a specific set menu first. This ensures that the set point cannot be accidentally altered during extended or unattended operation.
To allow you to change the configuration of some units, and to permit easy servicing, your water bath provides front service access to the controller without having to empty the bath.
The water bath features corrosion resistant construction throughout to withstand the rigors of daily lab use.

Section 2 Safety Information

These instructions contain important operating and safety information.
The user must carefully read and understand these instructions before using the water bath.
Your unit has been designed to optimize function, reliability, safety, and ease of use. It is the user’s responsibility to install the bath in conformance with local electrical codes.
To avoid electrical shock, always:

  • Connect the water bath to a properly grounded electrical outlet of the correct voltage and current handling capacity. Check the nameplate on the back of the unit for the voltage and current rating.
  • Disconnect the unit from the power outlet prior to maintenance and servicing. Note that 230V units are double pole, neutral fusing.

To avoid personal injury:

  • Do not use in the presence of flammable or combustible materials; fire or explosion may result. This device contains components which may ignite such materials.

  • Do not remove or modify the grounded power plug. Use only properly grounded outlets to avoid a shock hazard. This unit is not rated for use in hazardous atmospheres.

  • Do not continue to operate the bath if the temperature control fails (displays an “E” code or controls erratically) or the backup control fails (red safety light glows continuously when turned completely clockwise or fails to light when turned completely counterclockwise).

  • Select a fluid that is not corrosive and is not flammable. The following fluids are not recommended and may damage the unit: Chlorides or bleach.
    Strong concentrations of any acid.
    Strong concentrations of any salt.
    Weak concentrations of hydrochloric acid, hydrofluoric acid, hydrobromic acid, hydroiodic acid, sulfuric acid, or chromic acid.
    Weak salt solutions containing sodium chloride, calcium chloride, chromate or chromium compounds.
    High Purity water (deionized water, >lM ohm, etc.) Most photographic solutions.

  • Do not use a flammable liquid. A fire hazard may result. This unit is not explosion proof Unit contains components which may ignite such materials.

  • Use appropriate hand and eye protection when handling hazardous chemicals.

  • The interior of the unit can reach temperatures that can cause burns.
    Avoid contact. The unit can remain hot without visual indication for some time after power is turned off

  • If you will use the water bath with any liquid that will give off fumes, be sure to operate the water bath in a fume hood or with proper ventilation.

  • Use the cover to reduce evaporation and to permit reaching higher temperatures.

  • Hot liquids pose a burn hazard. Be careful not to reach into the bath when it contains hot liquids. Also be careful of steam rising from hot liquids.

  • The unit is intended to be operated with liquid in the chamber.
    However, it will not be damaged if it temporarily runs dry unless ignited by melting samples/ plastic in the bath.

  • Do not use in highly corrosive atmospheres; corrosive fumes and spillage may damage the unit and its internal components, creating a shock hazard.

  • Fumes from acidic solutions cause corrosion of the stainless steel reservoir. Care should be taken to maintain a neutral pH at all times.

  • Refer servicing to qualified personnel.

  • Do not place containers directly on bottom of chamber. Bottom can get extremely hot if no liquid is in the chamber. Always use the diffuser tray.

Section 3 Specifications

These are nominal specifications. Fisher Scientific reserves the right to change specifications or designs at any time without incurring obligation.
Power requirements …………..120VAC or 230VAC ± 10%, 50 or 60 Hz
Ambient conditions …………..+4°C to +35°C, up to 75% relative humidity
Temperature range .. …………..Cover open: Ambient to 60°C;
Cover closed:………….. Ambient to 100°C
Control …………..±O.l°C Between ambient and I00°C
Uniformity …………..±0.2°C at 37°C …………… ±O.S°C at 37°C (shallow pan)
Stability………….. ±0.5°C at 37°C
Max altitude 2000m
“Over Voltage Category II (IEC 664)
Pollution Degree 2 (IEC 664)

*CE Products meet the relevant EC harmonized standards for safety (IECJOJO-l!EN61010 and EMC (EN55014, EN55104, EN6J000-4-2, -4- 4, -4-6, -4-11 and ENV50140.)
“Applies to 230V units only

Catalog Number (Model) Specific Information Voltage
15-462-20 (2329) 2L 120V
15-462-30 (2337) 2L 240V
15-462-S20 (2306) SS lid 120V
15-462-S30 (2314) SS lid 230V
15-462-250(2331) 2L 120V
15-462-380 (2338) 21. 240V
15-462-8280 (2307) SS lid, shallow 120V
15-462-S3S0 (2315) SS lid, shallow 230V
15-462-50 (2340) 5L 120V
15-462-60 (2341) 51 120V
15-462-850(2316) SS lid 230V
15-462-S60123171 SS lid 230V
15-462-100(2320) 10L 120V
15.462-110(2321) 10L 240V
15-462-S100(2300) SS lid 120V
15-462-8110 (23011 Ss lid 230V
15-462-15012322) 5L/101, dual chamber 120V
15.462-16012328) 5L/10L, dual chamber 240V
15-462-8150 (2303) SS lid 120V
15-462-S160 (2304) SS lid 230V
15-462-200 (2332) 20L 120V
15-462-210 (2333) 201. 240V
15-462-S200 (2308) SS lid 120V
15-462-3210 (2309) SS fid 230V
15-462-280 (2335) 28L 120V
15-462-29012336) 28L 240V
15-462-8280 (2310) SS lid 120V
15-462-S29Q (2312) SS lid 230V

Section 4 Unpacking and Installation

The water bath is shipped in a single carton. When unpacking the unit, check each loose item against the packing list below. Should a shortage exist, notify your Fisher Scientific distributor, identifying the part by name and catalog number.

Packing List

Qty Item Supplied
1 Water Bath
1 Cover Assembly
1 Diffuser Tray
1 Thermometer Clip and Grommet
1 Siphon Pump (10, 20, 28, & dual 5/10 only)
1 Instructions
1 Warranty Card

Note If there is shipping damage, keep entire shipment intact – retaining the carton and all packing material – and file a claim with the final carrier.
Usually the firm will send an investigator to ascertain liability . .

Assembly and Installation

Follow these steps to assemble and install the water bath before operation:

  1. The location must:
    • Be indoors.
    • Provide an adequate source of power. Check the label on the back of the unit for voltage and current requirements.
    • Provide adequate clearance to insert samples.
    • Be level, fixed, and capable of supporting the weight of the unit when filled with liquid and samples.
    • Be free of drafts and wide ambient temperature variations such as near a heater or air conditioning vents.
    • Provide a fume hood if hazardous fumes are anticipated when using the water bath.
    • Be convenient to a sink for filling and draining the water bath.

  2. Install the cover by inserting the hinge into the lip at the rear of the bath. Angle the cover up while pushing down to lock the hinge in place under the stainless steel tank lip. Remove the cover by lifting the lid and pushing the hinge towards the back to release it from under the tank lip. Covers on the 2 liter shallow, 5 liter, and 10 liter model baths can also be installed and removed by sliding the hinge into place under the rim and tank lips from either side.

  3. Connect the line cord to a suitable grounded electrical outlet.

Repositioning the Control Panel
With 20 and 28 liter units, the control panel can be positioned on the side or the front of the unit in order to achieve the best use of available bench space. To change the position of the control panel, follow these steps:

  1. Turn off the unit and disconnect from power source.

  2. Allow to cool.

  3. Remove thermometer and dip. Remove the cover.

  4. Assuming that the water bath is empty, turn it over on a protected surface.

  5. Remove two screws retaining the control panel and two screws retaining the blank panel.

  6. Remove the blank panel.

  7. Carefully move the control panel to the new location and reinstall.
    Note Do not disconnect internal wiring or sensors from control panel.

  8.  Install the blank panel at the other location.

  9. Return unit to location of use.

  10. Reconnect power, turn on and test unit.
    Note This procedure applies to 20 and 28 liter units only.

Filling the Unit
Follow these steps to fill the unit in preparation for use:

  1. Fill the water bath with tap water or oil so that the liquid level is approximately 1 1/2 inches (3/4 inch with 2 liter shallow model) from the top of the tank when full anticipated load is placed in bath.
  2. If a thermometer is used, use the dip provided to hold the thermometer to the side of the bath. Slide the ****-ring on the thermometer to position the thermometer to the proper depth.
  3. When dosing the cover, place the thermometer along the front edge. A notch along the front of the cover dears the thermometer and dip with the cover dosed.
  4. For optimum results, the same fluid level should be maintained throughout the operating period.
  5. If using water, algicide may be added to reduce algae formation. Follow the instructions supplied with the algicide.

Note If the bath has been used previously, the tank should be thoroughly deaned before refilling .

Controls and Indicators
All controls and indicators are located on the front panel for ease of operation.
Note The water bath, full of water (within 1- 1.5” of top of tank), with lid, is designed to reach 37°C in less than 60 minutes or 90°C in less than 180 minutes .
Power on/ off:
Controls line power to unit. Set to 1 position to turn on power, 0 posi­ tion to turn off power.
°For °C:
Shows whether temperature is being displayed in Fahrenheit or Centigrade.
Display:
During standby shows actual temperature; during setups, shows menu modes and values being selected.
Alarm:
Lights to show that the backup controller is regulating the temperature rather than the primary controller.
Program:
Lights to show that unit is in Program mode rather than Run mode; see menu below.
Heat:
Lights to show when heat is being supplied to the chamber; cycles on and off as heat is requested by the controller.
Menu:
Allows the mode of operation to be selected. Each time the Menu button is pressed, the mode changes in this sequence:

  • Select temperature units in °C or F°(display shows _C_F).
  • Calibrate the unit (display shows CAL).
  • Display the actual temperature of the chamber (normal mode ofoperation).
  • The unit will return to normal operation, displaying the actual temperature, if no controls are touched for a few seconds.

Set:
When in the normal mode of operation, allows the temperature setpoint to be changed. When in the _C_F mode, allows the temperature units displayed to be changed. When in the CALibration mode, allows the temperature offset to calibrate the unit to be changed.
Increase:
After entering other than normal mode, increases the displayed value.
Decrease:
After entering other than normal mode, decreases the displayed value.
Safety controller:
Sets the Limit temperature above to the primary setpoint; should be set at a higher temperature than the setpoint to give the safety control in the event that the primary control fails.
Note Safety control is not as accurate a control as digital control.

Section 5 Operation

Follow these procedures for the operation of the unit, including:

  • Power up and initial indications.
  • Setting the temperature units in °F or °C
  • Setting the temperature setpoint.
  • Setting the backup temperature.
  • Calibrating the unit.
  • Emptying and deaning the unit.

Power Up
Follow these steps to turn on the unit to prepare it for use:

  1. Set the power switch to the ON position. The unit will go through a power-on self test that will take several seconds. During this time the display will show the unit’s capacity in liters.
  2. When the unit has completed its self-test, it will maintain the bath at the last setpoint temperature. The Heat indicator wiU light when the unit is applying heat to the bath. This indicator will cycle on and off during normal operation.
  3. When the display shows normal operation after completing the self­ test, continue by setting the temperature setpoint and backup temperature.
  4. If the display shows any message after completing the self-test, do not use the unit. Refer to the Troubleshooting section to determine what is causing the message to appear on the display.

Setting Temperature Units to °F or °C
Note which indicator is lit to the left of the display, either F for Fahrenheit or C for Centigrade. If the desired indicator is lit, omit this procedure. To change the temperature units, follow this procedure:

  1. Press Menu button so display shows _C_F. Press and hold the Set while simultaneously pressing the Increase or Decrease button until the display shows the desired temperature units, C for Centigrade or F for Fahrenheit.
  2. Release all controls.Within a few seconds, the display will return to normal operation. The desired indicator (F or C) should be lit to the left of the display.

Setting the Temperature Setpoint

Follow these steps  to set the controller so that it maintains the desired temperature:

  1. Turn Safety controller knob fully clockwise.

  2. Press and hold the Set button to see the current setpoint. The Program indicator illuminates to show that this is the setpoint. If this setpoint is correct, there is no need to change the setting.
    The display will return to normal operation when the Set button is released, displaying the actual temperature.

  3. To change the setpoint, press and hold the Set button while simultaneously pressing the Increase or Decrease button. The display shows the temperature setpoint as you change it. Release the Set button when the desired setting is achieved. The display will return to normal operation within a few seconds, displaying the actual temperature.

  4. To check the setpoint without affecting normal operation, press the Set button at any time. To change the setpoint, repeat Step 3 above.

  5. Set the Safety temperature according to the procedure below.

Note If Alarm Light comes on during normal cycling, turn backup controller knob slightly clockwise and monitor (if backup and digital controls overlap, the bath will have a wider temperature control band).

Setting the Safety Temperature
Follow these steps to set the water bath Safety temperature controller.

  1. Set the Temperature set point 2 to 5° higher than desired temperature.Wait until bath temperature stabilizes as shown by displayed actual temperature.
  2. Turn backup controller knob counterclockwise until Alarm light just comes on. Turn the knob clockwise slightly until the light goes off.
  3. Set the digital set point 5° higher and verify the Safety is controlling 2 to 5° above desired set point.
  4. Set the Temperature control down to desired temperature, allow the temperature to drift down, and verify the alarm light does not come on. If the light does come on, go back to Step 1, setting the Safety higher.

Calibrating the Unit
To check the accuracy of the unit, clip a certified thermometer, not supplied, to the side of the bath as described in the installation instructions. After allowing sufficient time for the temperature to stabilize, compare the thermometer reading to the actual temperature displayed on the unit. If the displayed temperature does not agree with the thermometer, calibrate the controller as follows:

  1. Note the difference between the displayed temperature and the thermometer. For example: If the displayed temperature is 37 and the thermometer reads 36, the difference is -1, meaning that the display should read 1 lower than it now shows.
  2. Press Menu button until display shows CAL.
  3. Press and hold the Set button while simultaneously pressing the Increase or Decrease button to set the desired temperature offset.
    Release the Set button when the desired setting is achieved. The display will return to normal operation within a few seconds, displaying the actual temperature.

In the example above, set the display to read -1.
Note The temperature offset can be set up to a range of 5.5°C or 10.0°F.

Emptying and Cleaning the Unit
Please be advised that stainless steel can and will rust if not regularly deaned and properly maintained. It is recommended that the bath be deaned at least on a monthly basis for moderate or continuous use applications. Follow these steps to empty and dean the unit after using it:

  1. Turn off power, unplug the unit, and allow to cool completely.
  2. If you have a water bath of 2 or 5 liter capacity, pour the liquid from the unit into an appropriate disposal container.
  3. If you have any other size water bath, allow the unit to cool and use the hand pump to empty the unit into an appropriate disposal container.
  4. Clean the inside of the water bath with mild detergent (such as Joy dishwashing detergent) and warm water. Do not scrub any surface with steel wool. (Steel wool leaves small metal particles behind that will rust, causing the pan to look rusty). In instances where a heavy coating of residue has accumulated inside the tank or where there is evidence that corrosion is beginning, the use of a stainless steel cleaner (such as 3M Stainless Steel Cleaner and Polish) is recommended.
  5. Rinse and wipe all tank surfaces with distilled water or equivalent.
  6. Unit is now ready for next use.

Section 6 Troubleshooting

If you have problems using the unit, follow these general procedures to track down the cause from the symptoms you are experiencing. If an error message appears on the display, refer to the specific error in this chart.
If the problem is not resolved using the table below, contact Technical Services.

Problem Possible Cause
No power indication Unit not plugged in; no power at the outlet; incorrect

power (make sure outlet matches label on back of unit); defective power cord; defective power switch or fuse (s) (call Technical Services).
Na heating| Setpoint or backup not set properly; defective heater or controller (call Technical Services).
Always heating| Defective controller (call Technical Services).
Very slow heating of samples| Empty tank or extremely low liquid level in tank (add liquid to the tank to improve heating).
Unit heating up slower than normal| Defective controller (call Technical Services).
Alarm light on constantly| Setpoint or backup not set properly; defective primary controller (call Technical Services).
Display not correct| Temperature F or C set in error; temperature not calibrated (perform calibration procedure with thermometer); defective controller (call Technical Services).
Erratic operation| Broken internal electrical connection; defective heater or controller (call Technical Services).
Error Message: El| Failure of temperature sensor or controller (call Technical Services).
Error Message: E2| Failure of controller (call Technical Services).
Display shows at power-up| Wrong voltage; controller not set up properly (call Technical Services).

Section 7 Replacement Parts

Item Pert No.
Cover, 2 liter Plastic Lid SPN102304
Stainless Steel Lid 15-462-32Q
Cover, 2 liter shallow Plastic Lid SPN102305
Stainless Steel Lid 15-462-330
Cover, 5 liter Plastic Lid SPN 1 02305
Stainless Steel Lid SPNCVM1598X1
Cover, 10 liter Plastic Lid SPN102306
Stainless Steel Lid 15-462-340
Cover, 20 and 28 liter Plastic Lid SPN102307
Stainless Steel Lid 15-462-350
Hinge, lid (for 2 liter) 102383
Hinge, rim (for 2 liter) 102384
Hinge, lid (for 10, 20, & 28 liter) 102386
Hinge, lid (2 liter shallow & 5 liter) Plastic Lid 102386
Stainless Steel Lid 106702
Hinge, rim (28L and Dual Models) 102387
Hinge, rim (for 10 & 20 liter) 102459
Hinge, rim (for 2 liter shallow & 5 liter) HGMX1
Clip, thermometer (for 2L shallow) 102423
Clip, thermometer (all except 2LS) 102424
Pump, Siphon 102391
Controller PC board (120V) 102419
Controller PC Board (230V) 104068
Display PC board 102420
Controller, safety 102499
Sensor, RTD 102331
Diffuser Tray, 2 liter 102352
Diffuser Tray, 2L shallow or 5 liter 102353
Diffuser Tray, 10 liter 102354
Diffuser Tray, 20 or 28 liter 102355
Fuse, Type 3AG 1.25x.25, 10 Amp 45920
Fuse, 5x2Omm, F5A, 250V 102487

Fisher Scientific 2300 Isotemp Digital Water Bath - Schematic
Unit

102370(7002300)Rev. 1
© 2010 Fisher Scientific Inc.
All rights reserved
Printed in U.S.A.

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Fisher Scientific User Manuals

Related Manuals