Fisher Scientific 0000853 Fisher Digital Drybath User Manual
- June 8, 2024
- Fisher Scientific
Table of Contents
- Models
- Important Notes
- Safety
- Regulatory Compliance
- Electromagnetic Compatibility
- Introduction
- Features
- Transportation and Storage Conditions
- Preparation Work
- Operation Guidance
- Setting Temperature and Heating Time
- Temperature Measurement and Calibration
- Maintenance of the instrument
- Failure Analysis and Troubleshooting
- Accessories
- Warranty
- Frequently Asked Questions
- WEEE Compliance
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Fisher Scientific 0000853 Fisher Digital Drybath User Manual
Models
This manual covers the models shown below.
NA Model | EU Model | Voltage | Description |
---|---|---|---|
88860021 | N/A | 100-120 Vac | FS Drybath Stdrd 1 blck 100-120 V |
88860022 | N/A | 100-120 Vac | FS Drybath Stdrd 2 blck 100-120 V |
88860023 | N/A | 100-120 Vac | FS Drybath Stdrd 4 blck 100-120 V |
88860024 | 15387928 | 200-240 Vac | FS Drybath Stdrd 1 blck 200-240 V |
88860025 | 15397928 | 200-240 Vac | FS Drybath Stdrd 2 blck 200-240 V |
88860026 | 15317938 | 200-240 Vac | FS Drybath Stdrd 4 blck 200-240 V |
Important Notes
General Practice
CAUTION: Important information is contained in any item and should be
carefully read. Failure to observe the instruction would result in damage or
abnormal function of the instrument.
WARNING: The warning message requires extremely careful operation of a
certain step. Failure to observe the instruction may result in serious
personal injury.
Safety
During operation, maintenance and repair of this instrument, the following basic safety notes should be observed. In case of failure to follow these instructions, the warnings or notes indicated herein, the basic protection provided by the instrument, its safety criteria of design and manufacture, and its predicted use range would be impaired. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
CAUTION: This instrument is for indoor use.
-
The ground connection
To avoid the electric shock, the input power line must be reliably grounded. The instrument is equipped with the three-pin plug that has the third pin (the pin connects the ground), therefore, the plug should be used with the grounded power socket only. This is a kind of safety device. If the plug cannot be plugged into the power socket, please ask the electrician to install a correct power socket, so as to make the grounded plug to work for safety. Fisher Scientific is not responsible for any injury as a result of the user’s failure to observe the following requirements. -
Servicing and Replacement of Parts
The operator should not take the instrument apart without permission. Only qualified professionals are allowed to replace elements or adjust components inside the instrument. Replacement of components while unit is connected to power and/or turned on is prohibited. -
Observe voltage/power requirements
Before the instrument is connected to the AC power source, the voltage of the power source should be the same with the required voltage of the instrument (a deviation of ±10% is allowed). The rated load of the power socket should not be lower than the requirements of the instrument. -
The power cord
The instrument should use the power line cord attached to it. If the power line is destroyed, it must be replaced but not be repaired. The replacement should be carried out with the power line of the same type and same specification (reach out to technical service team for further information). No items should be put on the power cord when the instrument is in operation. -
Connecting and disconnecting to/from
power source The user should hold the plug to remove from power source. When connecting the plug, user should make sure it has been fully plugged in; when removing the plug, do not pull the power line forcefully. Do not position the equipment so that it is difficult to operate the disconnecting device (plug). -
Placement of the instrument
This instrument should be fixed in a low RH (≤ 80%) and low dust place away from water source (e.g. sink or water pipeline) and the room should be well ventilated, and free of corrosive gas or interference of strong magnetic field. The instrument should not be placed in a wet or dusty location.
The openings on this instrument are for ventilation circulation and in order
to avoid over-heat of the instrument body they should not be blocked or
covered. When a single set of instrument is used, the interval between
ventilation opening before and after the instrument and its nearest object
should not be less than 25 cm. Also, don’t use the instrument on loose or soft
surface, or the air inlet of instrument bottom might be blocked. Excessive
temperature will impair performance or result in failure of the instrument.
This instrument should not be used in locations subjected to direct sun light.
The instrument should be kept away from hot gas, oven and all other heat
sources. If the instrument is to be stored for a long time, the power plug
should be withdrawn and the instrument covered with soft cloth or plastic film
to avoid entrance of dust. The product is powered by connecting the mains plug
to a standard socket-outlet. Always place the product in such a way that it is
easy for the operator to disconnect the product from the mains supply
Explanation of symbols
Attention, read user manual before use.
WARNING: There is a sign of
“CAUTION : HOT SURFACE! or
ATTENTION: SURFACE HOT” on the instrument. The metal part (module) near
the sign should not be touched with any part of the body when the Instrument
is operating in a high temperature state or just finished operation to avoid
burns!
Alternating Current Protective Conductor Terminal
CAUTION: In any of the following cases, immediately cut off the power supply, withdraw the power plug from the power socket and contact the supplier for service:
- Liquid drops into the inside of the instrument.
- The instrument is rained on or water is spilled on it.
- The instrument works abnormally, especially when generating an abnormal sound orodor.
- The instrument is dropped or its casing is damaged.
- The function of instrument obviously changes.
Regulatory Compliance
European Union
The European voltage models of this product meet all the applicable requirements of the European Directives and therefore display the CE Marking. These Directives are captured in the EU Declaration of Conformity which may be obtained from the manufacturer.
Product Safety
This product family has been tested to applicable product safety standards by a Nationally Recognized Test Laboratory (NRTL) and may bear the NRTL’s mark of safety compliance to those applicable standards.
CEEMC EN 61326-1
CE Safety EN 61010-1 CE
Safety EN 61010-2-010
UL 61010-1/CSAC22.2 NO.61010-1
ROHS2.0 2011/65/EU
WARNING : EMC Registration is done on this equipment for business use only. It may cause interference when the product would be used in home. This warning statement applies a product for business use.
Electromagnetic Compatibility
This device complies with Part 15 Subpart B of the FCC Rules. Operation is subject to the following two conditions:
- this device may not cause harmful interference, and
- this device must accept any interference received, including interference that may cause undesired operation.
Any changes or modifications not expressly approved by the manufacturer could void the user’s authority to operate the equipment.
Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Canadian ISED IC Notice
- This ISM digital apparatus complies with Canadian ICES-001, Class A
- Cet appareil ISM est conforme á la norme NMB-001 du Canada, Classe A
End of Life Care
Some considerations and suggestions are listed below for proper disposal of
this product. While addressing these actions for safe recycling and disposal,
please follow all guidelines, Safety Data Sheets (SDS), or regulations
applicable to your country and region.
- This product has materials and components that may be recycled or reused according to local guidelines and regulations.
- Remove any batteries present before disposal. Batteries, battery packs, and accumulators should not be disposed of as unsorted household waste. Please use the public collection system to return, recycle, or treat them in compliance with the local regulations.
- Remove all samples and items before defrosting a unit to room ambient temperatures.
- Clean up any chemical or biological safety hazards using appropriate methods.
- Have a certified technician remove the Refractory Ceramic insulation from the unit then dispose properly.
Introduction
The product is a dry bath with advanced microcomputer control, which can be widely applied to sample reservation, enzyme reservation and reaction, DNA amplification, electrophoresis degeneration and serum coagulation, etc. The characteristics of the product are as follows:
- The digital display for control of the temperature.
- Heat blocks help avoid possible contamination while containing the sample inside a tube.
- The heat blocks are easy to replace, clean and disinfect and are suitable for various tubes sizes.
- Built-in over-temperature protection device warrants sample and user safety.
- Temperature deviation adjustment.
Features
This chapter introduces the usage, transportation, storage conditions of the instrument, as well as its basic parameters, performance and functions.
Normal Working Conditions
Ambient temperature 10°C ~ 30°C
Relative humidity ≤ 80%
Power supply 100-120 V ~ 50/60 Hz or 200-240 V ~ 50/60 Hz
Operation Up to 2000 m
Pollution Degree 2
Overvoltage category II Indoor use
Mains supply voltage fluctuations ±10%
CAUTION : Before using the instrument, please make sure the working condition meets the above requirements. Especially that the power line is reliably grounded.
Transportation and Storage Conditions
Ambient temperature: -20°C~+55°C
Relative humidity: ≤ 80%
Basic Parameters
Table 2. 100-120 V
NA Catalog Number | 88860021 | 88860022 | 88860023 |
---|---|---|---|
Model Name | Digital 1 block 100-120 V | Digital 2 block | |
100-120 V | Digital 4 block 100-120 V | ||
Temperature Range | Ambient Temperature +5°C ~ 130°C (Ambient Temperature 25°C) | ||
Temperature Uniformity | ±1.0°C@37°C | ||
Temperature Accuracy | ±0.5°C@37°C | ||
Temperature Fluctuation | ±0.5°C | ||
Temperature Rise Time | ≤ 20 min (rise from 30°C to 130°C) | ≤ 25 min (rise from |
30°C to 130°C)
**** Size (L × W × H)| 288 × 200 × 100 mm11.3 × 7.87 × 3.94 in| 318 × 200 ×
100 mm12.5 × 7.87 × 3.94 in| 450 × 200 × 100 mm17.7 × 7.87 × 3.94 in
Weight| 2.9 kg (6.39 lbs)| 3.3 kg (7.28 lbs)| 4.7 kg (10.36 lbs)
Fuse Protector (ø5×20)| F 250 V 2.5 A| F 250 V 5 A| F 250 V 8 A
Table 3. 200-240 V
NA Catalog Number | 88860024 | 88860025 | 88860026 |
---|---|---|---|
EU Catalog Number | 15387928 | 15397928 | 15317938 |
Model Name | Digital 1 block 200-240 V | Digital 2 block | |
200-240 V | Digital 4 block 200-240 V | ||
Temperature Range | Ambient Temperature +5°C ~ 13°C (Ambient Temperature 25°C) | ||
Temperature Uniformity | ±1.0°C@37°C | ||
Temperature Accuracy | ±0.5°C@37°C | ||
Temperature Fluctuation | ±0.5°C | ||
Temperature Rise Time | ≤ 20 min (rise from 30°C to 130°C) | ≤ 25 min (rise from |
30°C to 130°C)
**** Size (L × W × H)| 288 × 200 × 100 mm11.3 × 7.87 × 3.94 in| 318 × 200 ×
100 mm12.5 × 7.87 × 3.94 in| 450 × 200 × 100 mm17.7 × 7.87 × 3.94 in
Weight| 2.9 kg (6.39 lbs)| 3.3 kg (7.28 lbs)| 4.7 kg (10.36 lbs)
Fuse Protector (ø5×20)| F 250 V 2.5 A| F 250 V 2.5 A| F 250 V 3.15 A
CAUTION: The instrument is only used for heating operation: from low temperature to high temperature.
Preparation Work
This chapter introduces structure of the instrument, user interface and functions of all buttons and preparation before startup. Read the content in the chapter carefully before using the instrument for the first time.
Structure Diagram
Display Panel Diagram
Instructions for Buttons
| Setting or adjusting button:
Press the button to display present setting temperature and press the button
again to display present time. After the unit has reached set point and
temperature is stable, calibration can be set by pressing the button.
---|---
****| ** Under the status of setting, press “ ” or “ ” to modify
the flashing figure. The instrument will enter temperature calibration mode if
user press the two buttons simultaneously under the status of stopping.
| When setting or adjusting,
pressing the button can move the flashing figure. During the operation, the
display can be switched between the temperature and the time.
| ** For starting up or
stopping program.
Operation Guidance
This chapter introduces the displaying window and the operation of buttons of
the dry bath with constant temperature.
Note : The illumination pattern of the ‘Heat Light’ indicates the
Operation status.
- Heat Light OFF (Unit Stopped) = “Set/Cal” button used for Setpoint adjustment.
- Heat Light blinks/flashes as Block is heating or cooling toward Setpoint.
- Heat Light ON (Unit Started and stable at setpoint) = “Set/Cal” used for Calibration.
Examination Before Startup
Please confirm the following content before plugging the plug into the power socket.
- The power source is in accordance with the voltage requirement of the instrument (please refer to the chapter II for the requirement of the voltage).
- Make sure the plug has been fully plugged into the power socket.
- The power line is reliably grounded.
CAUTION : If the display of the instrument is abnormal after startup, please turn off the power source immediately and contact with the supplier.
Startup
Press the power switch and the instrument will power on. All indicator lights and display serial ports will be turned on. About 3 ~ 5 seconds later, temperature display window will show the real-time temperature. All indicator lights will be turned off.
Setting Temperature and Heating Time
Set Temperature and Hold-Time Setpoint:
-
Connect power cord to appropriate electrical supply.
-
Make Power switch, located near power cord on rear, to the ‘ON’ position.
-
Confirm the Heat-indicator is ‘off’. If ‘ON’ or ‘blinking’, press “Stop”.
-
Set-up Temperature:
a. Press the “SET/CAL” button. The instrument is in the Temperature Setpoint mode with the display window showing the previous temperature set value, and the left digit flashing.
b. Press “ ” to increase or “ ” to decrease the flashing\ digit of the Setpoint.
c. Press “SHIFT” to move the flashing right, to the next digit.
d. Repeat Steps 4b and 4c to complete Temperature Setpoint selection. -
Setting Hold-Time Setpoint:
a. After the temperature is selected, press “SET/CAL” button. The instrument is in the Hold-time Setpoint mode with the display window showing the previous Hold-time set value, and the left digit flashing.
b. Press “ ” to increase or “ ” to decrease the flashing digit of the Hold-Time. Choose a time from 00:01 to 99:59, hours and minutes.
c. Press “SHIFT” to move the flashing right, to the next digit.
d. Repeat Steps 5b and 5c to complete Hold-Time Setpoint selection. For Continuous operation, Choose 99:59 then increase to display “- -:- -”.
e. After the Time setting is complete, Wait for 5 seconds and the instrument will automatically exit the Setpoint mode. The display will now show the current measured temperature. -
Press “Start” button to begin heating the Block. The green “Heat Light” will blink as Block is heating toward set point.
-
Allow 30-minutes for temperature to stabilize. The heat indicator stops blinking and remains constantly “ON” to indicate the temperature is stable, and instrument is ready to use
Instruction: There is limitation for temperature and time setting. The setting temperature range is: 15 ~ 135°C; the time range is: 0 ~ 99 h 59 min and continuous mode “- -:- -“.
Start and Stop Instrument
START: Press “START/STOP” to start operation, the indicator light “HEAT”
will be flashing. When the instrument achieves the setting temperature, the
“HEAT” light will stop flashing. This indicates that the unit will enter hold
temperature status. The digital display window will display the real-time
temperature and the timing function by pressing “SHIFT” button. When the
holding time equals to the setting time, the instrument will alarm 5 beeps.
STOP: Under the working status, press “START/STOP” to stop working. At
that time, corresponding working indicator light “HEAT” will be turned off.
The digital display window displays the real-time temperature.
Temperature Measurement and Calibration
-
The instrument has been operational for minimum of 30-minutes at desired setpoint.
-
An independent temperature measurement device is in the designed 5 mm hole in Block, along with paraffin oil, or similar heat-transfer oil.
-
Confirm the Heat-indicator is ‘ON’ steady, not ‘blinking’.
-
Compare the Display and independent measurements. And check if the temperature gap is > 0.2°C or not.
-
If less than 0.2°C, no calibration should be made.
-
If more than 0.2°C, calibrate the Temperature by these steps:
a. Press and hold the “SET/CAL” button. The instrument is in the Temperature Calibration mode with the display window showing the measured temperature value, and the left digit flashing.
b. Press “ ” to increase, or “ ” to decrease, the flashing digit to match the same digit of the independent measurement.
c. Press “SHIFT” to move the flashing right, to the next digit.
d. Repeat Steps 6b and 6C to complete Temperature caibration.
e. After the temperature calibration is completed, press “SET/CAL” button to exit the calibration mode and continue the Start mode. -
The Heat-indicator may start ‘blinking’ as the temperature is stabilized to the calibrated temperature.
-
When the Heat-indicator is ‘ON’ steady, the temperature is stable. After the instrument stable operated for additional 10 minutes, compare the display temperature and measurement again. If the temperature gap is > 0.2°C, repeat all steps of 6.
CAUTION:
- To ensure the accuracy of temperature, the instrument shall be calibrated after temperature stabilizes for 30 minutes.
- Calibrate the instrument with qualified Grade II standard filled thermometer.
- Calibration point: Fill the middle hole of the heat block with paraffin oil and soak the thermometer bulb.
Maintenance of the instrument
The holes of the blocks should be regularly cleaned with the damp cloth to ensure the test tube be well contacted with the wall of the holes so as to have good heat conduction. If the surface of the instrument is polluted, it can be cleaned with a slightly damp soft cloth. Generally unit falls under protective system of IP20 pursuant to IEC 60529.
WARNING: When cleaning the instrument, the power supply should be shut off and unplugged. The instrument surface should be cleaned with a non- corrosive cleaning agent.
Failure Analysis and Troubleshooting
This chaper introduces possible failures of this instrument, explanation and troubleshooting
CAUTION: The user is not allowed to open thehousing of the instrument for inspection during the warranty period. If any failures that require opening the housing for inspection occur, the supplier or manufacturer should be contacted timely.
Table 4. Failure Analysis and Troubleshooting
No. | Error | Cause | Solution |
---|
1
| ****
Display window is not turned on after switching on power supply
| Power supply fails to be connected| Check power supply and connect it.
Fuse burned out| Replace fuse
Damage of the switch| Replace the switch.
Others| Contact Fisher Scientific.
2| Temperature display window shows “001” and the buzzer alarm in a
di…di…sound| Short circuit of sensor| Check if the connecting line of
sensor is damaged, and replace it.
3| Temperature display window shows “002” and the buzzer alarm in a
di…di…sound| Open circuit of sensor| Check if the connecting line of
sensor is damaged, and replace it.
4| Temperature display does not conform to actual temperature| Sensor is
damaged or in poor contact| Contact Fisher Scientific.
Accessories
Table 5. Modular Block Accessories
| Catalog No.| Heating Blocks| No. of Holes| Block
Dimensions W x D x H (mm/in)
---|---|---|---|---
| 88860101| For 6 mm test tubes| 46| 124 x 76 x 39 /4.9
x 3.0 x 1.5
| 88860102| For 10 mm test tubes| 28| 124 x 76 x 39
/4.9 x 3.0 x 1.5
| 88860103| For 1.5 ml test tubes| 28| 124 x 76 x 39
/4.9 x 3.0 x 1.5
| 88860104| For 2.0 ml test tubes| 28| 124 x 76 x 39
/4.9 x 3.0 x 1.5
| 88860105| For 12 mm & 13 mm test tubes| 24| 124 x 76
x 39 /4.9 x 3.0 x 1.5
| 88860106| For 15 mm & 16 mm test tubes| 15| 124 x 76
x 39 /4.9 x 3.0 x 1.5
| 88860107| For 17 mm & 18 mm test tubes| 12| 124 x 76
x 39 /4.9 x 3.0 x 1.5
| 88860108| For 20 mm test tubes| 8| 124 x 76 x 39 /4.9
x 3.0 x 1.5
| 88860109| For 25 mm test tubes| 6| 124 x 76 x 39 /4.9
x 3.0 x 1.5
| 88860110| For 0.5 ml test tubes| 40| 124 x 76 x 39
/4.9 x 3.0 x 1.5
| 88860111| For 96 well Elisa plate| 1| 124 x 76 x 39
/4.9 x 3.0 x 1.5
| 88860112| For mixed size test tubes 32 holes for 6 mm test tubes
21 holes for 10 mm test tubes| 32 + 21| 124 x 76 x 39 /4.9 x 3.0 x 1.5
| 88860113| For mixed size test tubes 18 holes for 1.5 ml test
tubes 10 holes for 2.0 ml test tubes| 18 +10| 124 x 76 x 39 /4.9 x 3.0 x
1.5
| 88860114| For mixed size test tubes 3 holes for 25 mm test tubes
12 holes for 13 mm test tubes 6 holes for 6 mm test tubes| **** 3+12+6| 124 x
76 x 39 /4.9 x 3.0 x 1.5
Table 5. Modular Block Accessories
| Catalog No.| Heating Blocks| No. of Holes| Block
Dimensions W x D x H (mm/in)
---|---|---|---|---
| 88860115| For mixed size test tubes 30 holes for 0.5 ml test
tubes 20 holes for 0.2 ml test tubes| 30+20| 124 x 76 x 39 /4.9 x 3.0 x
1.5
| 88860116| For 15 ml flat bottom test tubes| 15| 124 x
76 x 113 /4.9 x 3.0 x 4.55
| 88860117| For 50 ml flat bottom test tubes| 4| 124 x
76 x 113 /4.9 x 3.0 x 4.55
| 88860118| For 15 ml conical bottom test tubes| 15|
124 x 76 x 113 /4.9 x 3.0 x 4.55
| 88860119| For 50 ml conical bottom test tubes| 4| 124
x 76 x 113 /4.9 x 3.0 x 4.55
| 88860120| For 96 well Non-skirted PCR plate| 1| 124 x
76 x 39 /4.9 x 3.0 x 1.5
| 88860121| For 96 well half/full skirt| |
Warranty
When used in laboratory conditions and according to these operation
instructions and maintenance, this product is warranted for 24 months against
defective materials or workmanship. The 24 month warranty period begins from
the delivery date of this product.
For product quality or performance issues, contact Fisher Scientific Customer
Service.
Frequently Asked Questions
Scenario 1
Does unit turn off once the program time is complete:
Example- If unit is set for one program to run at 45°C for 2 hrs.
Q-Would the unit turn off upon completion of this 2 hrs?
A-The instrument doesn’t turn off by itself, it just stops running.
Q-If not- would the unit hold the set temperature beyond this 2 hrs?
A-After this 2 hrs, the unit doesn’t hold the set temperature. It will cool to
ambient temp.
Q-Would unit return to ambient temp?
A-The unit will return to ambient temp.
Scenario 2
Suppose the unit is running a set program and there is power outage. Once the
power resumes
Q-Would unit restart automatically once power is restored
A-The instrument will automatically turn on.
Q-Would unit resume the program from where it was left?
A-It can’t.
Q-Would customer need to start the program manually?
A-Yes.
Q-Would the unit keep the saved program in memory?
A-The unit has no memory function.
Q-Would customer need to create new program?
A-Need to reset the program.
WEEE Compliance
WEEE Compliance. This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2012/19/EU. It is marked with the following symbol. Fisher Scientific has contracted with one or more recycling/disposal companies in each EU Member State, and this product should be disposed of or recycled through them. Further information on our compliance with these Directives, the recyclers in your country, and information on Fisher Scientific products which may assist the detection of substances subject to the RoHS Directive are availableat fisher.com/ WEEERoHS.rrrrrrrrrt
IF YOU NEED ASSISTANCE:
Fisher Scientific products are backed by a global technical support team ready to support your applications. We offer cold storage accessories, including remote alarms, temperature recorders, and validation services. Visit fishersci.com or call:
United States
+1-800-766-7000
fishersci.com
Canada
+1-800-234-7437
fishersci.ca
Austria
+43(0)800-20 88 40
at.fishersci.com
Netherlands
+31(0)20 4887 70 00
nl.fishersci.com
Belgium
+32(0)56 260 260
be.fishersci.com
Norway
+47 22 95 59 59
fishersci.no
Denmark
+45 70 27 99 20
fishersci.dk
Portugal:
+351 21 425 33 50
pt.fishersci.com
Germany
+49(0)180 5258221
de.fishersci.com
Spain
+34 002 239 303
es.fishersci.com
Ireland
+959(0)1 885 5854
ie.fishersci.com
Sweden
+46 31- 68 94 30
fishersci.se
Italy
+39 02 950 59 478
it.fishersci.com
Switzerland
+41(0)56 618 41 11
ch.fishersci.com
Finland
+358(0)9 8027 6280
fishersci.fi
UK
+44(1)1509 555 500
fisher.co.uk
France
+33(0)388 67 14 14
fishersci.fr
References
- Lab Equipment and Lab Supplies | Fisher Scientific
- Lab Equipment and Lab Supplies | Fisher Scientific
- Lab Equipment and Lab Supplies | Fisher Scientific
- Lab Equipment and Lab Supplies | Fisher Scientific
- Lab Equipment and Lab Supplies | Fisher Scientific
- Lab Equipment and Lab Supplies | Fisher Scientific
- Fisher | Emerson US
- Lab Equipment and Lab Supplies | Fisher Scientific
- Lab Equipment and Lab Supplies | Fisher Scientific
- Lab Equipment and Lab Supplies | Fisher Scientific
- Lab Equipment and Lab Supplies | Fisher Scientific
- Lab Equipment and Lab Supplies | Fisher Scientific
- Lab Equipment and Lab Supplies | Fisher Scientific
- Lab Equipment and Lab Supplies | Fisher Scientific
- Lab Equipment and Lab Supplies | Fisher Scientific
- Lab Equipment and Lab Supplies | Fisher Scientific
- Lab Equipment and Lab Supplies | Fisher Scientific
- Lab Equipment and Lab Supplies | Fisher Scientific
- Lab Equipment and Lab Supplies | Fisher Scientific
- EPM Weee Compliance | Thermo Fisher Scientific - US
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>