AEMC INSTRUMENTS OX 5042 and OX 5042B Handscope Instructions
- September 12, 2024
- AEMC INSTRUMENTS
Table of Contents
OX 5042 and OX 5042B Handscope
“`html
Product Information
Specifications:
- Model: OX 5042 & OX 5042B Handscope
- Measurement Types: AC Voltage, AC Current, Harmonics
- Display: LCD
- Input Channels: Channel A, Channel B
Product Usage Instructions
First-Time Setup:
- Turn the instrument ON.
- Press Memory Mode.
- Use the left/right buttons to highlight the .cfg icon.
AC Voltage Measurement:
- Attach the BNC adapter to the Channel A input.
- Attach the black and red leads to the adapter.
- Select probes and attach them to the leads.
AC Current Measurement:
- Attach the current probe to the Channel B input.
- Clip the probe around the conductor to be measured.
- Turn the instrument ON.
AC Voltage and Current Harmonics:
- Press Harmonics Mode.
- Connect voltage leads and attach the current probe.
- Press Auto Set.
Frequently Asked Questions (FAQ)
Q: What should I do if I encounter an insufficient amplitude
error message during Harmonics measurement?
A: If either channel is not connected, an
insufficient amplitude error message will appear. Pressing Auto Set
may not eliminate the error. Quickly press twice the channel button
for the channel causing the error. This action turns off that
channel, removes the error message, and displays the THD of the
active channel. The inactive channel will be grayed out. To
reactivate a turned-off channel, press its channel button one more
time.
“`
Chauvin Arnoux®, Inc.
15 Faraday Drive · Dover, NH 03820 ·
800-343-1391 ·
603-749-6434 · www.aemc.com
Model OX 5042 & OX 5042B Handscope Instructions for General Electrical
Measurements
First-Time Setup
1. Turn the instrument ON.
2. Press Memory Mode .
3. Use the left/right buttons to highlight the .cfg icon
.
4. Use the up/down buttons to select the Load File icon
and press Enter .
5. Press the left arrow button twice to select files and the up/down buttons
to highlight default.cfg.
6. Press Enter, then Right Arrow, and then Enter again to load the default
configuration.
7. Press Meter Mode then then press Channel B .
8. Press the right navigation button until the Multiplier field is highlighted. By default this is set to x1.
9. Press the down navigation button until the
setting is highlighted.
10. Press Enter to save the setting.
11. Press to highlight the Units field. Press to highlight the A setting.
12. Press to save the setting. The main menu now appears as follows:
13. Press Memory Mode . 14. Use the left/right buttons to highlight the .cfg icon 15. Use the up/down buttons to highlight the Save Mode 16. Use the up button to highlight the Save File icon
. and press Enter . and press Enter .
This completes the First Time Setup.
AEMC® Instruments Technical Support
800-343-1391 (Ext. 351) ·
techsupport@aemc.com
99-MAN 100453 v11 05/24 1
15 Faraday Drive · Dover, NH 03820 ·
800-343-1391 ·
603-749-6434 · www.aemc.com
AC Voltage Measurement
Chauvin Arnoux®, Inc.
1. Attach the BNC adapter (shown on right) to the Channel A input.
2. Attach the 10 ft black and red leads to the adapter.
3. Select the alligator or pencil tip probes and attach them to the leads as needed.
4. Attach the test leads to the measurement point. Turn the instrument ON.
5. Press Meter Mode
then press Channel A . The bottom display should
appear as follows:
NOTE: If this menu screen is not displayed, press the Channel A button again.
6. Verify that AUTO is ON in the menu at the bottom of the display. If AUTO is
not
ON, see Loading a Saved Configuration at the end of this document.
7. The voltage measurement appears on the screen (see the example below):
8. Read the voltage and record:
AEMC® Instruments Technical Support
800-343-1391 (Ext. 351) ·
techsupport@aemc.com
99-MAN 100453 v11 05/24 2
15 Faraday Drive · Dover, NH 03820 · 800-343-1391 · 603-749-6434 · www.aemc.com
Chauvin Arnoux®, Inc.
AC Current Measurement
1. Select the current probe based on the current to be measured: · MN251T (if in your kit) for measuring from (1 to 200) A · MN379T (if in your kit) for measuring from (0.05 to 120) A · MiniFlex® sensor for measuring (10 to 3000) A.
2. Attach the current probe to the Channel B input.
3. Clip the current probe around the conductor to be measured.
4. Turn the instrument ON.
7. Press Meter Mode
then press Channel B .
8. The bottom display should appear as follows:
NOTE: If this menu screen is not displayed, press the B Channel button again.
9. Verify that AUTO is ON and X1000 is in the menu at the bottom of the
display. If AUTO is not ON, see Loading a saved configuration at the end of
this document.
10. The current measurement appears on the Channel B display (see the example
below):
11. Read the current and record:
AEMC® Instruments Technical Support
800-343-1391 (Ext. 351) ·
techsupport@aemc.com
99-MAN 100453 v11 05/24 3
15 Faraday Drive · Dover, NH 03820 · 800-343-1391 · 603-749-6434 · www.aemc.com
Chauvin Arnoux®, Inc.
AC Voltage and Current Harmonics
1. Press Harmonics Mode . Connect the voltage leads and attached the current
probe. Press Auto Set .
2. The Voltage and Current THD (Total Harmonics Distortion) percentages
appear on the top of the display (see the example below):
Typical Display
Error Message (see note below)
Record the Voltage THD percentage:
%
Record the Current THD percentage:
%
NOTE: If either channel is not connected, an insufficient amplitude error message, as shown above, appears. If pressing Auto Set does not eliminate the error, quickly press twice the channel button for the channel causing the error. This turns OFF that channel, removes the error message, and displays the THD of the active channel. The channel that is OFF is grayed out. To re- activate a channel that is turned OFF, press its channel button one more time.
AEMC® Instruments Technical Support 800-343-1391 (Ext. 351) · techsupport@aemc.com
99-MAN 100453 v11 05/24 4
15 Faraday Drive · Dover, NH 03820 · 800-343-1391 · 603-749-6434 · www.aemc.com
Chauvin Arnoux®, Inc.
Millivolt Drop Current Harmonics
1. Press Meter Mode
and then press Channel A .
2. Verify the following menu appears.
NOTE: If this menu screen is not displayed, press the Channel A button again.
Verify that AUTO is ON in the menu at the bottom of the display. If AUTO is
not ON, see Loading a Saved Configuration at the end of this document.
3. Move the test leads to each measurement point, and record each millivolt
drop:
If applicable, record the completed measurements in your report software.
AEMC® Instruments Technical Support 800-343-1391 (Ext. 351) · techsupport@aemc.com
99-MAN 100453 v11 05/24 5
15 Faraday Drive · Dover, NH 03820 · 800-343-1391 · 603-749-6434 · www.aemc.com
Chauvin Arnoux®, Inc.
Three most IMPORTANT things to remember:
- Make sure you are in AUTO for Meter mode. 2) Turn Channels OFF when they are not being used. 3) Scale factor must be set correctly for each mode.
Loading a Saved Configuration
1. Press Memory Mode .
2. Use the left/right buttons to highlight the .cfg icon
.
3. Use the up/down buttons to select the Load File icon
and press Enter .
4. Press the left arrow button twice to select files and the up/down buttons
to highlight setup_00.cfg.
5. Press Enter, then right arrow, and then Enter again to load the saved
configuration.
Saving a Configuration
1. Press Memory Mode .
2. Use the left/right buttons to highlight the .cfg icon
.
3. Use the up/down buttons to highlight the Save Mode
and press Enter .
4. Use the up button to highlight the Save File icon
and press Enter .