AEMC INSTRUMENTS F01 Clamp Multimeter User Manual
- June 13, 2024
- AEMC INSTRUMENTS
Table of Contents
AEMC INSTRUMENTS F01 Clamp Multimeter
Product Information
- Product Name: Clamp Multimeter
- Model Number: F01
- Manufacturer: AEMC
- Serial Number: [Serial Number]
- Catalog Number: 2129.51
- Website: www.aemc.com
Statement of Compliance
We guarantee that at the time of shipping, your instrument has met its
published specifications. An NIST traceable certificate may be requested at
the time of purchase or obtained by returning the instrument to our repair and
calibration facility for a nominal charge.
The recommended calibration interval for this instrument is 12 months and
begins on the date of receipt by the customer. For recalibration, please use
our calibration services. Refer to our repair and calibration section at
www.aemc.com.
Table of Contents
Operation
- 4.1 Voltage Measurement –
- 4.2 Audio Continuity Test and Resistance Measurement
- 4.5 Current Measurements –
Maintenance
- 5.1 Changing the Battery
- 5.2 Cleaning
- 5.3 Storage
Introduction
This manual provides instructions for using the Clamp Multimeter Model F01.
Warning: Please adhere to the international electrical symbols and safety precautions mentioned in this manual.
Definition of Measurement Categories
The Clamp Multimeter Model F01 is designed for measurements in the following categories:
- Cat. I: For measurements on circuits not directly connected to the AC supply wall outlet such as protected secondaries, signal level, and limited energy circuits.
- Cat. II: For measurements performed on circuits directly connected to the electrical distribution system. Examples are measurements on household appliances or portable tools.
- Cat. III: For measurements performed in the building installation at the distribution level such as on hardwired equipment in fixed installation and circuit breakers.
- Cat. IV: For measurements performed at the primary electrical supply.
Product Usage Instructions
Operation
-
Voltage Measurement –
Follow the instructions provided in the manual to accurately measure voltage using the Clamp Multimeter. -
Audio Continuity Test and Resistance Measurement
Refer to the manual for step-by-step instructions on performing audio continuity tests and measuring resistance with the Clamp Multimeter. -
Current Measurements –
Learn how to measure current using the Clamp Multimeter by following the instructions provided in the manual.
Maintenance
-
Changing the Battery
Refer to the manual for detailed instructions on how to change the battery of the Clamp Multimeter. -
Cleaning
Follow the provided guidelines in the manual to properly clean the Clamp Multimeter. -
Storage
Learn how to store the Clamp Multimeter correctly by referring to the instructions in the manual.
INTRODUCTION
Warning
- Never use on circuits with a voltage higher than 600V and an overvoltage category higher than Cat. III.
- Use in inside environments with Pollution Degree 2; Temperature 0°C to +50°C; 70% RH.
- Only use accessories compliant with safety standards (NF EN 61010-2-031) 600V min and overvoltage Cat. III.
- Never open the clamp before disconnecting all power sources.
- Never connect to the circuit to be measured if the clamp is not properly closed.
- Before any measurement, check the proper positioning of the cables and switch.
- When measuring current, check for proper alignment of the conductor in relation to the markers and proper closing of the jaws.
- Always disconnect the clamp from any power source before changing the battery.
- Do not perform resistance tests, continuity tests or semi-conductor tests on a circuit under power.
International Electrical Symbols
| This symbol signifies that the instrument is protected by double or
reinforced insulation.
---|---
| This symbol on the instrument indicates a WARNING and that the operator
must refer to the user manual for instructions before operating the
instrument. In this manual, the symbol preceding instructions indicates that
if the instructions are not followed, bodily injury, installation/sample and
product damage may result.
| Risk of electric shock. The voltage at the parts marked with this
symbol may be dangerous.
| This symbol refers to a type A current sensor. This symbol signifies
that application around and removal from HAZARDOUS LIVE conductors is
permitted.
| In conformity with WEEE 2002/96/EC
Definition of Measurement Categories
- Cat. I: For measurements on circuits not directly connected to the AC supply wall outlet such as protected secondaries, signal level, and limited energy circuits.
- Cat. II: For measurements performed on circuits directly connected to the electrical distribution system. Examples are measurements on household appliances or portable tools.
- Cat. III: For measurements performed in the building installation at the distribution level such as on hardwired equipment in fixed installation and circuit breakers.
- Cat. IV: For measurements performed at the primary electrical supply (<1000V) such as on primary overcurrent protection devices, ripple control units, or meters.
Receiving Your Shipment
Upon receiving your shipment, make sure that the contents are consistent with
the packing list. Notify your distributor of any missing items. If the
equipment appears to be damaged, file a claim immediately with the carrier and
notify your distributor at once, giving a detailed description of any damage.
Save the damaged packing container to substantiate your claim.
Ordering Information
Clamp-on Multimeter Model F01 ………………………………..Cat. #2129.51
Includes multimeter, set of red and black leads with probe tips, 9V battery,
carrying pouch and this user manual.
Accessories and Replacement Parts
Replacement set of leads, red and black with probe tips …. Cat. #2118.92
General Canvas Pouch (4.25 x 8.5 x 2″)……………………….. Cat. #2119.75
Only use accessories adapted to the voltage and overvoltage category of the
circuit to be measured (per NF EN 61010).
PRODUCT FEATURES
Description
The Clamp-on Multimeter, Model F01 emphasizes reliability and simplicity of
use to respond to the needs of power professionals.
Features:
- A compact unit, integrating the current sensor for intensity measurements without breaking the test circuit
- Outstanding ergonomic features:
- automatic selection of AC or DC measurement – V only
- automatic selection of measurement ranges
- programmable audio voltage indication (V-Live)
- “over-range” indication
- power auto-off
- Compliance with IEC electrical safety standards and CE markings
- Light and rugged construction for field use
Model F01 Control Functions
- Jaws
- Command Buttons
- 4-way Rotary Switch
- Liquid Crystal Display
R otary Switch Functions
- OFF Deactivation of the clamp, activation is ensured by selection of other functions
- DC and AC voltage measurement (rms value)
- Continuity and resistance measurement
- AC ampere measurement (rms value)
Hold Button Primary Functions
Short Press: Freezes the display. The display is cleared when the button is
pressed again.
Button Held Down: Enables access to secondary functions in conjunction with
the rotary switch.
Hold Button Secondary Functions (with rotary switch)
-
Disable Auto-Off Function
While pressing down the HOLD button, bring the rotary switch from the OFF position to the position. -
The unit emits a double beep, then the symbol flashes.
The selected configuration is put into memory when the button is released (the symbol remains lit continuously). -
Auto-off is reactivated when switch returns to OFF position.
-
Activate The V-Live Function
(Beeper ON when voltage >45V peak)
While pressing down the HOLD button, bring the rotary switch from the OFF position to the V position. The unit emits a double beep, then the V and symbol flashes. The selected configuration is put into memory when the button is released (the V symbol becomes fixed and the symbol flashes).
Proceed in the same way to suppress the V-Live function (the symbol disappears when the button is released). -
Displaying The Internal Software Version
While pressing down the HOLD button, bring the rotary switch from the OFF position to the A position. The unit beeps, the software version is displayed in the form UX.XX for 2 seconds, then all the segments of the display are shown. -
Liquid Crystal Display
The liquid crystal display includes the digital display of the measured values, the related units and symbols.-
Digital Display
4 digits, 9999 counts, 3 decimal points, + and – signs (DC measurement) -
+ OL : Positive value range exceedance (>3999cts)
-
– OL : Negative value range exceedance
-
OL : Unsigned value range exceedance
-
– – – – : Indeterminate value (middle segments)
-
Symbol Display
-
HOLD Function active
-
Constant operation (no power auto-off)
-
Flashing: V-Live function selected
Fixed: Continuity measurement -
AC Measurement in AC mode
-
DC Measurement in DC mode
-
Flashing: power limited to approximately1 hour
Fixed: battery drained, operation and accuracy are no longer guaranteed
-
-
Buzzer
Different sounds are emitted according to the function performed:- Short and medium sound: valid button
- Short and medium sound every 400 ms: voltage measured is higher than the unit’s guaranteed safety voltage
- 5 short and medium recurring sounds: automatic deactivation of the instrument
- Continuous medium sound: continuity value measured below 40Ω
- Modulated medium continuous sound: value measured in volts, higher than 45V peak when the V-Live function is selected
SPECIFICATIONS
Reference Conditions
23°C ±3°K; RH of 45 to 75%; battery power at 8.5V ± 5V; frequency range 45 to
65Hz; position of conductor centered in clamp jaws; conductor diameter .2″
(5mm); no electrical field; no external AC magnetic field.
Electrical Specifications
Voltage (V)
Range | 40V | 400V | **600V*** |
---|---|---|---|
Measuring Range** | 0.2V to 39.99V | 40.0V to 399.9V | 400 to 600V |
Accuracy | 1% of Reading |
+ 5cts
| 1% of Reading
+ 2cts
| 1% of Reading
+ 2cts
Resolution| 10mV| 0.1V| 1V
Input Impedance| 1MW
Overload Protection| 600VAC/DC
*In DC, the display indicates +OL above +600V and -OL above -600 V.
In AC, the display indicates OL over 600Vrms.
In AC if the value of the voltage measured is <0.15V the display indicates
0.00.
Audio Continuity ( ) / Resistance Measurement (Ω)**
Range | 400 W |
---|---|
Measuring Range | 0.0 to 399.9W |
Accuracy* | 1% of Reading + 2cts |
Resolution | 0.1W |
Open Circuit Voltage | £3.2V |
Measuring Current | 320µA |
Overload Protection | 500VAC or 750VDC or peak |
*with compensation for measurement lead resistance
Current (A)
Display Range | 40A | 400A | **600A*** |
---|---|---|---|
Measuring Range** | 0.20 to 39.99A | 40.0 to 399.9A | 400 to 600A peak |
Accuracy | 1.5% of Reading + 10cts | 1.5% of Reading + 2cts | |
Resolution | 10mA | 100mA | 1A |
*The display indicates OL over 400Arms.
**In AC, if the value of the current measured is <0.15A, the display shows
0.00.
- Battery: 9V alkaline battery (type IEC 6LF22, 6LR61 or NEDA 1604)
- Battery Life: 100 hrs approx
- Auto-off: After 10 minutes of no activity
Mechanical Specifications
Temperature:
- Reference Range
- Operating Range
- Storage Range (without battery)
-
Operating Temperature: 32 to 122°F (0 to 50°C); 90% RH
-
Storage Temperature: -40 to 158°F (-40 to 70°C); 90% RH
-
Altitude:
Operation: ≤2000m
Storage: ≤12,000m -
Dimensions: 2.76 x 7.6 x 1.46″ (70 x 193 x 37mm)
-
Weight: 9.17 oz (260g)
-
Clamp Tightening Capacity: ≤1.00” (≤26mm)
Safety Specifications
-
Electrical Safety
(as per EN 61010-1 ed. 95 and 61010-2-032, ed. 93)- Double Insulation
- Category III
- Pollution Degree 2
- Rated Voltage 600V (RMS or DC)
-
Electric Shocks (test as per IEC 1000-4-5)
- 6kV in RCD mode on the voltmeter function, aptitude criterion B
- 2kV induced on the current measurement cable, aptitude criterion B
-
Electromagnetic Compatibility (as per EN 61326-1 ed. 97 + A1)
-
Emission: class B
Immunity: -
Electrostatic discharges:
4kV on contact, aptitude criterion B
8kV in the air, aptitude criterion B -
Radiated field: 10V/m, aptitude criterion B
-
Fast Transients: 1kV, aptitude criterion B
-
Conduit interference: 3V, aptitude criterion A
-
-
Mechanical Resistance
- Free fall 1m (test as per IEC 68-2-32)
- Impact: 0.5 J (test as per IEC 68-2-27)
- Vibration: 0.75mm (test as per IEC 68-2-6)
-
Auto-extinction (per UL94)
- Housing V0
- Jaws V0
- Display window V2
Variations in Operating Range
Influence
Quantities
| Meas. Range Quantities| Quantity Influenced| Influence
Typical Max
---|---|---|---
Battery Voltage| 7.5 to 10V| All| –| 0.2% R + 1ct
Temperature| 32 to 122°F| V A
W
| 0.05% R/50°F
0.1% R/50°F
0.1% R/50°F
| 0.2% R/50°F + 2cts
0.2% R/50°F + 2cts
0.2% R/50°F + 2cts
Relative Humidity| 10 to 90% RH| V A
W
| ≤1ct 0.2% R
≤1ct
| 0.1% R + 1ct 0.3% R + 2cts 0.3% R + 2cts
Frequency
| 40Hz to 1kHz 1kHz to 5kHz 40Hz to 400Hz 400Hz to 5kHz| V
A
| see curve
see curve
| 1% R + 1ct
6% R + 1ct
1% R + 1ct
5% R + 1ct
Position of conductor in the jaws
(f ≤ 400Hz)
| Position on internal perimeter of jaws| ****
A
| ****
1% R
| ****
1.5% R + 1ct
Adjacent conductor with AC current (50Hz) running through| Conductor in contact with external perimeter of jaws| ****
A
| ****
40 dB
| ****
35 dB
Conductor clamped| 0 to 400VDC or rms| V| <1ct| 1ct
Application of voltage to the clamp| 0 to 600VDC or rms| A| <1ct| 1ct
Peak factor
| 1.4 to 3.5 limited to 600A peak 900V peak| A V| 1% R
1% R
| 3% R + 1ct
3% R + 1ct
Rejection of series mode in DC| 0 to 600V/50Hz| V| 50 dB| 40 dB
Rejection of series mode in AC| 0 to 600VDC
0 to 400ADC
| V A| <1ct
<1ct
| 60 dB
60 dB
Rejection of common mode| 0 to 600V/50Hz| V A| <1ct 0.08A/100V| 60 dB
0.12A/100V
Influence of external magnetic field| 0 to 400A/m (50Hz)| A| 85 dB| 60 dB
Number of jaw open- ing movements| 50000| A| 0.1% R| 0.2% R + 1ct
Typical Frequency Response Curves
-
– V = f (f)
-
– I = f (f)
OPERATION
Voltage Measurement – ( )
- Connect the measurement leads to the instrument’s terminals, complying with the polarities indicated: red lead on the “+” terminal and black lead on the “COM” terminal.
- Set the rotary switch to the “ ” position.
- Connect the unit to the voltage source to be measured, making sure that the voltage does not exceed the maximum acceptable limits (see § 3.2.1).
- Range switching and AC/DC selection are automatic
If the signal measured is >45V peak, the audio indication is activated if the V-Live function is selected (see § 2.6.2).
For voltages ≥600Vdc or rms, a repetitive beep indicates that the measured voltageis higher than the acceptable safety voltage (OL).
- Range switching and AC/DC selection are automatic
Audio Continuity Test – ( ) and
Resistance Measurement – (Ω)
- Connect the measurement leads to the terminals.
- Set the rotary switch to the “ ” position.
- Connect the unit to the circuit to be tested. The buzzer is continuously active as soon as contact is established (circuit closed) and if the resistance value measured is less than 40Ω.
NOTE : Above 400Ω, the display indicates OL.
Current Measurements – ( )
- Set the rotary switch to position “ ” position.
- Clamp the conductor carrying the current to be measured, checking for proper closing of the jaws and for foreign matter in the gap.
Range switching and AC/DC selection are automatic.
MAINTENANCE
Use only factory specified replacement parts. AEMC® will not be held
responsible for any accident, incident, or malfunction following a repair done
other than by its service center or by an approved repair center.
Changing the Battery
Disconnect the instrument from any source of electricity.
- Set the switch to OFF.
- Slide a screwdriver into the slot at the top of the battery cover (rear of the clamp) and push the battery cover upwards.
- Replace the used battery with a 9V battery (type LF22), observing the polarities.
- Install the battery in its housing, then reattach the battery cover.
Cleaning
Disconnect the instrument from any source of electricity.
- Use a soft cloth lightly dampened with soapy water.
- Rinse with a damp cloth and then dry with a dry cloth.
- Do not splash water directly on the clamp.
- Do not use alcohol, solvents or hydrocarbons.
- Make sure the gap between the jaws is kept clean and free from debris at all times, to help ensure accurate readings.
Storage
If the instrument is not used for a period of more than 60 days, remove the battery and store it separately.
Repair and Calibration
To ensure that your instrument meets factory specifications, we recommend that
it be submitted to our factory Service Center at one-year intervals for
recalibration, or as required by other standards or internal procedures.
For instrument repair and calibration:
You must contact our Service Center for a Customer Service Authorization
Number (CSA#). This will ensure that when your instrument arrives, it will be
tracked and processed promptly. Please write the CSA# on the outside of the
shipping container. If the instrument is returned for calibration, we need to
know if you want a standard calibration, or a calibration traceable to
N.I.S.T. (Includes calibration certificate plus recorded calibration data).
Chauvin Arnoux®, Inc.
d.b.a. AEMC® Instruments
15 Faraday Drive
Dover, NH 03820 USA
Tel: 800-945-2362 (Ext. 360)
603-749-6434 (Ext. 360)
Fax: 603-742-2346 or
603-749-6309
repair@aemc.com
(Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to N.I.S.T.
are available.
NOTE: All customers must obtain a CSA# before returning any instrument.
Technical and Sales Assistance
If you are experiencing any technical problems, or require any assistance with
the proper operation or application of your instrument, please call, mail, fax
or e-mail our technical support hotline:
Chauvin Arnoux®, Inc.
d.b.a. AEMC® Instruments 200 Foxborough Boulevard Foxborough, MA 02035, USA
Phone: 800-343-1391
508-698-2115
Fax : 508-698-2118
techsupport@aemc.com
www.aemc.com
NOTE: Do not ship Instruments to our Foxborough, MA address.
Limited Warranty
The Model F01 is warranted to the owner for a period of one year from the date
of original purchase against defects in manufacture. This limited warranty is
given by AEMC® Instruments, not by the distributor from whom it was purchased.
This warranty is void if the unit has been tampered with, abused or if the
defect is related to service not performed by AEMC® Instruments.
For full and detailed warranty coverage, please read the Warranty Coverage
Information, which is attached to the Warranty Registration Card (if enclosed)
or is available at www.aemc.com. Please keep the
Warranty Coverage Information with your records.
What AEMC® Instruments will do:
If a malfunction occurs within the one-year period, you may return the
instrument to us for repair, provided we have your warranty registration
information on file or a proof of purchase. AEMC® Instruments will, at its
option, repair or replace the faulty material.
YOU CAN NOW REGISTER ONLINE AT: www.aemc.com
Warranty Repairs
What you must do to return an Instrument for Warranty Repair:
First, request a Customer Service Authorization Number (CSA#) by phone or by
fax from our Service Department (see address below), then return the
instrument along with the signed CSA Form. Please write the CSA# on the
outside of the shipping container. Return the instrument, postage or shipment
pre-paid to:
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
Service Department
15 Faraday Drive • Dover, NH 03820 USA
Tel: 800-945-2362 (Ext. 360)
603-749-6434 (Ext. 360)
Fax: 603-742-2346 or
603-749-6309
repair@aemc.com
Caution: To protect yourself against in-transit loss, we recommend you
insure your returned material.
NOTE: All customers must obtain a CSA# before returning any instrument.
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
15 Faraday Drive
Dover, NH 03820 USA
Phone : 603-749-6434
Fax: 603-742-2346
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>