Health o meter 565KL,565KG High Resolution Digital Baby Scale Instruction Manual
- September 7, 2024
- Health o meter
Table of Contents
- Health o meter 565KL,565KG High Resolution Digital Baby Scale
- Intended use
- SPECIFICATIONS
- CERTIFICATIONS / CONNECTIVITY / DISPOSAL
- POWERING THE SCALE / SET UP
- Powering the Scale – Inserting Batteries (Batteries not included)
- SET UP INSTRUCTIONS
- OPERATING INSTRUCTIONS
- Maintenance
- CALIBRATION
- Troubleshooting
- WARRANTY
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Health o meter 565KL,565KG High Resolution Digital Baby Scale
PELSTAR, LLC 9500 West 55th St. McCook, IL 60525-7110 USA www.homscales.com
- Health o meter® Professional 565KL Health o meter® Professional 565KG
- Pelstar, LLC 2024
Thank you for purchasing this Health o meter® Professional product. Please read this manual carefully, and keep it for easy reference or training.
Note: This scale has been factory calibrated, and does not require calibration before use.
- Please register your scale for warranty coverage at: www.homscales.com
- For User Instructions updates and revisions please go to: www.homscales.com
Intended use
The Health o meter® Professional 565KL/565KG scales were intended to be used in a professional medical environment by trained medical staff. This product was designed to weigh infants who are laying or sitting on the tray. Do not modify the product or use it for anything other than its intended purpose. To prevent injury and damage to your scale, please follow these instructions very carefully.
- Assemble and operate the scale per the enclosed user instructions.
- For accurate weighing, this scale must be placed on a flat, stable surface.
- For accurate weighing, verify before each use the proper operation according to the procedure described in this manual.
- Do not transport the scale with a patient or object on the tray.
- Do not exceed the weight capacity specified for the scale.
- Do not use it in the presence of flammable or explosive materials.
- Ensure that the scale or AC adapter does not come in contact with liquids, excessive temperature, or excessive humidity.
- If the scale becomes damaged, it should not be operated until properly serviced.
Patient/caregiver safety
- This scale is designed for static weighing of patients only. No scale should be used for patient transfer.
- To prevent patient injury, the patient must be attended to throughout the entire w eight g event.
When operating the scale with batteries:
- When replacing/inserting batteries, be sure to use all new batteries.
When operating the scale with the AC adapter:
- Operate this scale exclusively with the AC adapter provided by Health o met e r ® Professional. Use of an unspecified adapter will void the warranty and can pose a serious safety hazard.
- For use with AC adapter mode l listed on the next page
- Before using this scale, inspect the AC adapter cord for cracking/fraying i n g, or for broken/bent plug prongs.
- Before using this scale, ensure that the AC adapter is plugged into an outlet with the rated voltage appropriate for operation.
- Ensure that the AC adapter outlet is wired to a circuit breaker or other protected power source.
- Unplug the AC adapter, and carefully store both the adapter and adapter cord before moving the scale.
- Operating this scale at voltages and frequencies other than specified can damage the equipment and will void the warranty.
In no event shall Pelstar, LLC be liable for damages or injuries arising out of or connected with the assembly, us e, or misuse of this product
SPECIFICATIONS
General
The Health o meter® Professional 565KL/565KG scales use highly
sophisticated microprocessor technology. Each precision scale is designed to
provide accurate, reliable and repeatable weight measurements. In addition,
each scale is designed to provide the user with features that make the
weighing process simple, fast, and convenient.
Scale Specifications
Capacity and Resolution| 565KL Capacity: 44 lb / 20 kg 565KG Capacity: 20 kg
Resolution: 1g (to 6kg); 2g (to 20kg)
---|---
Power Requirements| Adapter Model No. UES12LCP-120100SPA Input: 100-240V –
50/60Hz, 0.5A
Output:12.0V 1.0A (Included) Or 6 AA Batteries (not included)
Environmental| Operating temperatures: 32°F to 104°F (0°C to 40°C) Storage Specifications: -4°F to 140°F/ -20°C to 60°C
CERTIFICATIONS / CONNECTIVITY / DISPOSAL
Certification Descriptions
Scale Disposal
This Health o meter® Professional scale must be disposed of properly as electronic waste. Follow the national, regional or local regulations which apply to you for disposal of electronic waste or batteries. Do not dispose of this device in the domestic waste stream.
Connectivity Information
To enable reliable transmission of weight this scale is designed to connect to
a computer, monitor, or other electronic device via optional Pelstar® wireless
technology. This connectivity helps minimize a source of potential medical
errors caused by incorrectly copying and then recording patient data. For more
information regarding how to connect the scale to other electronic devices,
please contact Health o meter® Professional Scales Technical Support,
available during normal business hours at
1-800-638-3722.
Note: If this product was purchased as a “BT” model, refer to the Wireless Communication Set Up Instructions included with the packaging.
POWERING THE SCALE / SET UP
Powering the Scale – Inserting Batteries (Batteries not included)
- Locate the battery compartment underneath the scale base and remove battery compartment cover.
- Place 6 new AA batteries into the battery compartment. Make sure the ‘+’ and ‘-’ are positioned correctly. (Use only AA alkaline batteries).
- Replace the battery compartment cover.
Warning: If the scale will not be used for some time, remove the batteries to avoid a safety hazard. Disposal of batteries must be performed in accordance with the prevailing national, regional or local regulations which apply to you.
AC Adapter (included)
Connect the AC adapter into the base of the scale and then into the outlet for
power. Warning: To avoid a safety hazard, use only a Health o meter®
Professional AC adapter.
Measuring Tape
For your convenience a measuring tape is included with the scale. You may
affix the tape to either the bottom or side of the weighing tray depending on
your preference for visually determining the length of the baby, up to 24” or
58 cm.
SET UP INSTRUCTIONS
To enter the Options mode, push and hold the UNIT button while pressing and releasing the button.
Auto Off Function
When powering the scale via the AC adapter, the display can be set to power
off after a period of inactivity. The Auto Off default setting is “disabled”.
Follow this procedure to enable or disable the Auto Off function. Note: When
powering the scale via batteries, the Auto Off function is automatically
enabled.
Enable / Disable Auto Off
-
While the scale is turned off, press and hold the UNIT button then press and release the button. Hold the UNIT button until “AOF=d” or “AOF=E” appears on the display, then release. Press the UNIT button again to change the setting.
AOF=d Disables the Auto Off function
AOF=E Enables the Auto Off function -
Press HOLD/RLS/RECALL to save your selection and exit the Options mode.
-
The scale will restart. The display will show “UEr”, followed by a version number, and then dashes. When the display shows “0.0” (in lb mode) or “0.000” (in kg mode) the scale is ready to use.
SET UP INSTRUCTIONS (CONT.)
Setting Auto Off Time
When powering the scale via batteries or if Auto Off is Enabled for the AC
adapter, the period of inactivity before the scale powers off can be set to a
specific amount of time. The default setting for Auto Off time is 2 minutes.
Follow this procedure to set the number of minutes of inactivity before power
off.
- While the scale is turned off, press and hold the UNIT button then press and release the button. Hold the UNIT button until “AOF=d” or “AOF=E” appears on the display, then release. Use the ZERO/TARE to advance or the button to scroll back the display screen until “AFt” appears. Press the UNIT button to change the Auto Off time.
- AFt = 1 Scale will power off after 1 minute of inactivity
- AFt = 2 Scale will power off after 2 minutes of inactivity
- AFt = 3 Scale will power off after 3 minutes of inactivity
- Choose the setting you prefer and press HOLD/RLS/RECALL to save your selection and exit the Options mode.
- The scale will restart. The display will show “UEr”, followed by a version number, and then dashes. When the display shows “0.0” (in lb mode) or “0.000” (in kg mode) the scale is ready to use.
Auto Hold
The Auto Hold function allows the scale to continue displaying the weight
measurement after the patient is removed from the weighing tray. The Auto Hold
setting can be adjusted to display the weight for 0, 2 minutes, 5 minutes, or
continuously when on AC power. The default setting is 2 minutes. Follow this
procedure to adjust the Auto Hold setting.
- While the scale is turned off, press and hold UNIT button then press and release the button. Hold the UNIT button until “AOF=d” or “AOF=E” appears on the display, then release. Use the ZERO/TARE to advance or the button to scroll back the display screen until “AHd” appears. Press UNIT to change the setting.
- AHd = 0 Weight measurement will not be held on the display after the patient is removed
- AHd = 1 Weight measurement will hold on the display for 2 minutes
- AHd = 2 Weight measurement will hold on the display for 5 minutes
- AHd = C Weight measurement will hold on the display continuously if the scale is on AC power
- Choose the setting you prefer and press HOLD/RLS/RECALL to save your selection and exit the Options mode.
- The scale will restart. The display will show “UEr”, followed by a version number, and then dashes. When the display shows “0.0” (in lb mode) or “0.000” (in kg mode) the scale is ready to use.
Auto Hold
The Auto Hold function allows the scale to continue displaying the weight
measurement after the patient is removed from the weighing tray. The Auto Hold
setting can be adjusted to display the weight for 0, 2 minutes, 5 minutes, or
continuously when on AC power. The default setting is 2 minutes. Follow this
procedure to adjust the Auto Hold setting.
- While the scale is turned off, press and hold UNIT button then press and release the button. Hold the UNIT button until “AOF=d” or “AOF=E” appears on the display, then release. Use the ZERO/TARE to advance or the button to scroll back to the display screen until “AHd” appears. Press UNIT to change the setting.
- AHd = 0 Weight measurement will not be held on the display after the patient is removed
- AHd = 1 Weight measurement will hold on the display for 2 minutes
- AHd = 2 Weight measurement will hold on the display for 5 minutes
- AHd = C Weight measurement will hold on the display continuously if the scale is on AC power
- Choose the setting you prefer and press HOLD/RLS/RECALL to save your selection and exit the Options mode.
- The scale will restart. The display will show “UEr”, followed by a version number, and then dashes. When the display shows “0.0” (in lb mode) or “0.000” (in kg mode) the scale is ready to use.
Sound Options
The display can be set to stay silent or beep while operating the scale. The
default setting for Sound is “enabled”. Follow this procedure to enable or
disable sound.
-
While the scale is turned off, press and hold the UNIT button then press and release the button. Hold the UNIT button until “AOF=d” or “AOF=E” appears on the display, then release. Use the ZERO/TARE to advance or the button to scroll back the display screen until “Snd” appears. Press UNIT to change the setting.
-
Snd = E Enables sound so the scale beeps when keys are pressed.
Snd = d Disables sound so the scale stays silent when keys are pressed. -
Choose the setting you prefer and press HOLD/RLS/RECALL to save your selection and exit the Options mode.
-
The scale will restart. The display will show “UEr”, followed by a version number, and then dashes. When the display shows “0.0” (in lb mode) or “0.000” (in kg mode) the scale is ready to use.
SET UP INSTRUCTIONS (CONT.)
Locking Unit of Measure (LB or KG) (KL versions only)
The weight measuring unit (pounds / LB or kilograms / KG) can be locked to
display weight only in the selected unit of measure. The unit of measure lock
default setting is “unlocked”. Follow this procedure to lock the UNIT button.
Note: If Everlock® is engaged, this feature will not appear in the Options
mode.
- While the scale is turned off, press and hold the UNIT button then press and release the button. Hold the UNIT button until “AOF=d” or “AOF=E” appears on the display, then release. Use the ZERO/TARE to advance or the button to scroll back the display screen until “Unt” appears. Press the UNIT button to change the setting.
- Unt = U Allows the user to toggle weight measurement between LB and KG.
- Unt = L Locks the unit button so weight is only displayed in pounds (LB)
- Unt = ├ Locks the unit button so weight is only displayed in kilograms (KG)
- Unt = g Locks the unit button so weight is only displayed in grams (g)
- Choose the setting you prefer and press HOLD/RLS/RECALL to save your selection and exit the options mode.
- The scale will restart. The display will show “UEr”, followed by a version number, and then dashes. When the display shows “0.0” (in lb mode) or “0.000” (in kg mode) the scale is ready to use.
Enabling Unit View with Unit of Measure Locked
If the scale is locked per the instructions above, users can still view the
alternate weight measuring unit by enabling the “Unit View” option. Enabling
the “Unit View” option allows the user to press the UNIT button to briefly
display the weight in the alternate weight measuring unit. After releasing the
UNIT button the display will return to the locked weight measuring unit. The
“Unit View” default setting is “Disable”. Follow this procedure to enable
“Unit View” when the weight measuring unit is locked. Note: If Everlock® is
enabled, this feature will not appear in the Options mode.
-
While the scale is turned off, press and hold the UNIT button then press and release the button. Hold the UNIT button until “AOF=d” or “AOF=E” appears on the display, then release. Use the ZERO/TARE button to advance or the button to scroll back the display screen until “UdP” appears. Press the UNIT button to change the setting.
-
UdP = d Disables the UNIT button so it will not display the other weight measuring unit.
UdP = E Allows the user to view the other weight measuring unit by holding down the UNIT button even if the weight measuring unit is locked. Enabling this function will only allow the user to view the other unit while pressing the UNIT button; it will not unlock the weight measuring unit. -
Choose the setting you prefer and press HOLD/RLS/RECALL to save your selection and exit the Options mode.
Everlock® (KL versions only)
The Everlock® feature allows the measuring unit to be permanently locked on
the scale, disabling the UNIT button. The Everlock® default setting is
“disabled”. Follow this procedure to engage Everlock®.
Warning: This lock is permanent and cannot be reversed.
Notes:
- Press HOLD/RLS/RECALLat any time to exit the Everlock® procedure.
- Once Everlock® is engaged the Unt and UdP options will no longer be displayed when scrolling through the Options menu.
- There is no method to undo Everlock® without sending the scale back to the manufacturer.
- While the scale is turned off, press and hold the UNIT button then press and release the button. Hold the UNIT button until “AOF=d” or “AOF=E” appears on the display, then release. Use the ZERO/TARE to advance or the button to scroll back the display screen until “ELC=d” appears. This is the default setting with Everlock disabled.
- Press the UNIT button to start the Everlock procedure. “PASS1” will appear on the display.
- With “PASS1” displayed, press the following keys in this sequence:, , ZERO/TARE, UNIT
- If entered correctly, “ELC=L” (if scale was in LB mode) or “ELC= Ͱ” (if scale was in KG mode) will appear on the display.
- Use the ZERO/TARE button to toggle between (L) Pounds and (Ͱ) Kilograms. When the desired unit is displayed, press UNIT and “PASS2” will appear on the display.
- This will prompt the user to enter the second password. Press the following keys in this sequence:ZERO/TARE, UNIT, (Warning: Entering this password will engage Everlock.)
- If entered correctly, the scale will display “ELC = E”
- Choose the setting you prefer and press HOLD/RLS/RECALL to save your selection and exit the Options mode.
- The scale will restart. The display will show “UEr”, followed by a version number, and then dashes. When the display shows “0.0” (in lb mode) or “0.000” (in kg mode) the scale is ready to use.
Warning: Do not power off the scale until HOLD/RLS/RECALL is pressed to save the Everlock option, the scale has restarted and the display has returned to “0.0”.
- The scale will restart. The display will show “UEr”, followed by a version number, and then dashes. When the display shows “0.0” (in lb mode) or “0.000” (in kg mode) the scale is ready to use.
OPERATING INSTRUCTIONS
Keypad
Screen Icons
IMPORTANT: To prevent falls or injury, place the scale securely far from the edge of the counter, table or other flat surface used to hold the scale. Do not let the scale touch the wall or any other object near or around the scale while it is in use.
Weighing a Patient
Note: When this scale is turned on, it will always default to the settings and units last used (LB or KG).
-
With all weight off the scale, press the button to power on the scale.
-
Place the baby on the scale so that its weight is evenly distributed on the tray. As the scale calculates weight a small ring will flash in the upper left corner of the display along with weight measurements. Depending on the movement of the patient on the scale, it may take several seconds for the scale to lock onto the weight.
-
When a stable weight is calculated, the ring will stay displayed and the weight of the baby will lock. Note: If the display shows “—–” you have exceeded the capacity of the scale.(44 lb / 20 kg)
Caution: To prevent patient injury, the patient must be attended throughout the entire weighing event. -
The weight may be held on the display by pressing the HOLD/RLS/RECALL button. To release, press the HOLD/RLS/RECALL button again.
-
After the weight measurement has been taken and is recorded, remove the patient from the scale. To reweigh the patient without removing them from the scale, press the button.
-
Press and hold the button to power off the scale.
Weight Recall
- To recall the last weight measured, with all weights off the scale and the display showing “0”, press the HOLD/RLS/RECALL button.
- To clear the display of the recalled weight, press the HOLD/RLS/RECALL button again.
Zero Function
The ZERO/TARE button zeroes the scale prior to a weight measurement. The
ZERO/TARE button is also used in the Tare function outlined below.
Tare Function
When using this scale, the weight of an object, such as a blanket, can be
subtracted from the total weight to determine the weight of the patient alone.
Weighing a Patient and Taring an Unknown Weight
The minimum weight that can be tared is 0.02 lb / 0.01 kg / 10 g. Note: The
scale must lock onto a weight before it can be tared.
- With all the weight off the scale, press the button to turn on the scale.
- Wait until the “0” appears on the display then place the object on the scale. The display will show a value for the object’s weight.
- Press the ZERO/TARE button. The arrow on the display will appear and point to the Tare icon on the display.
- Keep the object on the platform, and then gently place the patient onto the scale. The scale will automatically deduct the weight of the object and only display the patient’s weight.
- When you remove the patient and the item from the platform, “UndEr” will show on the display. Press the ZERO/TARE button to clear the tared value.
Caution: To prevent patient injury, the patient must be attended throughout the entire weighing event.
Weighing a Patient and Taring a Known Weight (Pre-Tare):
Pre-Tare can be performed for items between 0.2 lb and 5 lb (0.1 kg and 2.26
kg).
- With all the weight off the scale, press the button to turn on the scale.
- Wait until “0” appears on the display. Press and hold the ZERO/TARE button until the arrow on the display points to the Tare icon on the display.
- Use the ZERO/TARE buttons to increase or decrease the weight. Pounds will change in 0.1oz increments and KG will change in 0.001 kg increments.
- Once the desired tare value is set, lock the value by pressing the HOLD/RLS/RECALL button. The display will show “UndEr”.
- Weigh the patient following the weighing procedure outlined on the previous page including the item to be tared. The scale will automatically deduct the tare value and only display the patient’s weight.
- To clear the tare value, press the ZERO/TARE button.
Maintenance
The following pages provide instructions for maintenance, cleaning, calibrating, and troubleshooting of your scale. Maintenance operations other than those described in this manual should be performed by qualified service personnel.
Caution: Before first use, or after long periods of non-use, check the scale for proper operation and function. If the scale does not operate correctly, refer to qualified service personnel.
- Check the overall appearance of the total scale for any obvious damage, wear, and tear. Inspect the AC adapter cord for cracking or fraying, or for broken/bent prongs. *Not included with model 565KL-C or 565KG-C
- Do not store the scale on its side.
- This scale contains electronic parts. Use it carefully. Do not drop, kick, or jump on it.
- Do not store chemicals, heaters, or radiators.
Note: This scale is an extremely sensitive weighing device. If the auto- off function is disabled, battery power will be consumed. Items such as clothing or towels placed or dropped on the scale may cause it to activate. To prevent the run down of the batteries, do not store or leave anything on the scale.
Cleaning and Disinfecting
Proper care and cleaning are essential to ensure a long life of accurate and
effective weighing.
- Health o meter® Professional recommends using one of the following solutions on a soft cloth or disposable wipe:
- mild soap and water solution
- 70% isopropyl alcohol
- solution with 1-5% hydrogen peroxide concentration After cleaning/disinfecting, wipe with a cloth dampened with water and then with a clean dry cloth. To avoid residue build-up or damp surfaces, ensure all scale parts are completely dry after cleaning.
- Never use rough or abrasive materials to clean the scale, as these will damage the scale’s finish.
- Do not submerge the scale in water or any other liquid.
- Do not pour or spray fluids directly on the scale.
CALIBRATION
Your scale has been factory-calibrated and does not require calibration before use. There are no requirements for calibration in the field; users should adhere to their institution’s calibration policies. If required, the scale can be calibrated. A 5kg weight is required for calibration. If the scale is Everlocked into pounds, the calibration procedure will still require the use of a 5kg weight.
- Note: Avoid scale placement near heating/cooling vents, heating appliances, areas of high airflow, or direct sunlight. Calibration must be done in a controlled environment.
- Note: Only weights certified and traceable to national standards should be used for calibration procedures.
- Press and hold HOLD/RLS/RECALL, then press and release the button to power the scale. Remain pressing HOLD/RLS/RECALL until “CAL” appears on the display. Release all buttons.
- Once the buttons are released the following message will display, “CAL59”, “CAL58”etc. until the display counts down to zero. This allows the load cells time to settle before calibration starts. HOLD/RLS/RECALL can be pushed at any time to cancel the countdown, however, this is not recommended.
- Once the time has elapsed, “2Ero SCALE” will scroll horizontally across the screen. With no weight on the scale, press ZERO/TARE.
- The scale will be zero and “PLACE 5 ON SCALE PRESS HOLD” will scroll across the screen.
- Place a certified 5kg weight on the center of the weighing tray, then pressHOLD/RLS/RECALL. “CAL” will flash on the screen while the scale is calibrating.
- A. If the weiis with in the hehet calibration range, a four-digit number will be displayed for approximately 5 seconds. Remove the weight from the scale and the scale will restart.
- B. If the weight placed on the scale does not match the correct calibration standard(5 kg), “Out OF RANGE” will appear on the display. Calibration is not performed. Press ZERO/TARE to exit calibration mode. Obtain the correct amount of calibration weight and begin the calibration process again.
Troubleshooting
Before contacting service personnel, refer to the following instructions to check and to correct any failures.
Symptom | Possible Cause | Corrective Action |
---|---|---|
Scale does not turn on | 1. Dead battery |
2. Faulty electrical outlet
3. Bad power supply
| 1. Replace batteries
2. Use a different outlet
3. Replace AC adapter
Questionable weight or the scale does not zero| 1. External object
interfering with the scale| 1. Remove interfering object from the scale
2. The display did not show “0.0” before weighing| 2. Remove the patient
from the weighing tray, zero the
scale and begin weighing process again
3. Scale is not placed on a
level surface
| 3. Place the scale on a level
surface and begin weighing process again
4. Scale is out of
calibration
| 4. Check weight with known
weight value
The display shows “UNDER”| Negative weight, indicating a tare was present. Load on the scale is below initial
zero point.
| Clear scale of any weight and re-zero the scale by pressing ZERO/TARE
button.
WARRANTY
Limited Warranty
What does the Warranty Cover?
This Health o meter® Professional scale is warranted from date of purchase
against defects of materials or in workmanship for two (2) years. If the
product fails to function properly, return the product, freight prepaid, and
properly packed to Pelstar, LLC (see “To Get Warranty Service”, below, for
instructions). If the manufacturer determines that a defect of material or in
workmanship exists, the customer’s sole remedy will be the replacement of the
scale at no charge. Replacement will be made with a new or manufactured
product or component. If the product is no longer available, a replacement may
be made with a similar product of equal or greater value. All replaced parts
are covered only for the original warranty period.
Who is Covered?
The original purchaser of the product must have proof of purchase to receive
warranty service. Please save your invoice or receipt. Pelstar dealers or
retail stores selling Pelstar products do not have the right to alter, modify
or in any way change the terms and conditions of this warranty.
What is Excluded?
Your warranty does not cover normal wear of parts or damage resulting from any
of the following: negligent use or misuse of the product, use on improper
voltage or current, use contrary to the operating instructions, abuse
including tampering, damage in transit, or unauthorized repair or
alternations. Further, the warranty does not cover natural disasters, such as
fire, flood, hurricanes, and tornadoes. This warranty gives you specific legal
rights, and you may also have other rights that vary from country to country,
state to state, province to province or jurisdiction to jurisdiction. To get
Warranty Service make sure you keep your sales receipt or document showing
proof of purchase. Call (+1)
800-638-3722 or (+1)
708-377-0600 to receive a return
authorization (RA) number, which must be included on the return label. Attach
your proof of purchase to your defective product along with your name,
address, daytime telephone number, and description of the problem. Carefully
package the product and send it with shipping and insurance prepaid to:
- Pelstar, LLC Attention R/A#_____Return Department9500 West 55th StreetMcCook, IL 60525
Extended Warranty Available
This scale is eligible for Health o meter® Professional ScaleSurance Extended
Warranty Program. ScaleSurance extends the warranty period for an additional
two years. This extension to the standard Limited Warranty can be purchased
with new scales or for a facility’s existing scale before its current warranty
has expired. To learn more, visit www.homscales.com/scalesurance/ or contact
your medical supply distributor.
- Not available in all countries
PELSTAR, LLC
9500 West 55th St. McCook, IL 60525-7110 USA
1-800-638-3722 or
1-708-377-0600 PLEASE REGISTER YOUR
SCALE FOR WARRANTY COVERAGE AT: www.homscales.com
Health o meter® is a registered trademark of Sunbeam Products, Inc. used under
license. Health o meter® Professional products are manufactured, designed,
and owned by Pelstar, LLC. We reserve the right to improve, enhance, or modify
Health o meter® Professional product features or specifications without
notice. © Pelstar, LLC 2024