levenhuk Ermenrich Zing TC Series Digital Multimeter User Manual

July 31, 2024
levenhuk

levenhuk Ermenrich Zing TC Series Digital Multimeter

Product Specifications:

  • Model : Ermenrich Zing TC21/TC24/TC32 Digital Multimeter
  • Power Source : 2 AA batteries
  • Included Accessories : K-type thermocouple, test leads (red and black), carry bag (TC32 only), user manual, warranty

Product Usage Instructions

Getting Started:

  1. Open the battery compartment cover and insert 2 AA batteries according to the correct polarity.
  2. Close the cover.
  3. Use the knob switch (13) to turn the device on and to select the required operation mode.
  4. To turn the device off, turn the knob switch to the OFF position.

Function Selection:

  1. To select the required function, turn the knob switch (13).
  2. In case there are multiple subfunctions in one function, press the FUNC button (5) and select the needed one.
  3. The names of the functions are displayed on the LCD screen (4).

Data Hold:

  1. During the measurement, press the HOLD button (7) to store the existing readings.
  2. Press again to turn the data hold mode off.

MAX/MIN Mode:

  1. Press the MAX/MIN button (8) to switch between maximum and minimum values.
  2. Press and hold (8) for 2 seconds to exit this function.
  • Backlight : Press the Flashlight/Backlight button (6) to turn the backlight on/off.
  • Flashlight : Press the Flashlight/Backlight button (6) for 3 seconds to turn the flashlight on/off.
  • Input Jack Indicator: When power is on or the function is switching, the corresponding input jack light will flash to indicate that the test lead should be inserted into the corresponding jack.

MODELS

FEATURES

  1. NCV probe
  2. Flashlight
  3. Warning indicator
  4. LCD display
  5. FUNC (Function) button
  6. Flashlight/Backlight button
  7. HOLD (Data hold) button
  8. MAX/MIN (Max./min. values) button
  9. 10A jack
  10. INPUT jack (+)
  11. mA jack
  12. COM jack (-)
  13. Knob switch

Please carefully read the safety instructions and the user manual before using this product. Keep away from children. Use the device only as specified in the user manual.

The kit includes  a digital multimeter, K-type thermocouple, test leads (red and black), carry bag (TC32 only), user manual, and warranty.

Getting started

Open the battery compartment cover and insert 2 AA batteries according to the correct polarity. Close the cover. Use the knob switch (13) to turn the device on and to select the required operation mode. To turn the device off, turn the knob switch to the OFF position.

Function selection

To select the required function, turn the knob switch (13). In case there are multiple subfunctions in one function, press the FUNC button (5) and select the needed one. The names of the functions are displayed on the LCD screen (4).

  • Data hold
    During the measurement, press the HOLD button (7) to store the existing readings. Press again to turn data hold mode off.

  • MAX/MIN mode
    Press the MAX/MIN button (8) to switch between maximum and minimum values. Press and hold (8) for 2 seconds to exit this function.

  • Backlight
    Press the Flashlight/Backlight button (6) to turn the backlight on/off.

  • Flashlight
    Press the Flashlight/Backlight button (6) for 3 seconds to turn the flashlight on/off.

  • Input jack indicator
    When power on or function is switching, the corresponding input jack light will flash to indicate that the test lead should be inserted into the corresponding jack.

Auto power turn-off
After 15 minutes without any key operation, the device will turn off automatically to save battery energy. After automatic shutdown, press any key to return multimeter in operating condition. To cancel the automatic turn-off function, press and hold the FUNC button (5) and turn the knob switch (13) to power on simultaneously. After turning the multimeter off and then turning it on again, the auto power off function will be activated by default.

High voltage/current prompt function
When the measuring voltage is greater than 80V or the measuring current is greater than 1A, the orange backlight will light up, prompting the users to be careful.

  • Do not measure the voltage higher than 1000V DC or 750V AC, otherwise the device may be damaged.
  • Observe the safety precautions to avoid the risk of electric shock.
  • Put the sensor up to a known power source to test the device prior to using it.

Professional measurement

Voltage measurement (in V)

  • Turn the knob switch (13) to the  position and select the appropriate range (TC21); or
  • Turn the knob switch (13) to the position and press the FUNC button (5) to select the AC or DC voltage measurement function (TC24/TC32).

Plug the black test lead into the COM jack (12) and the red test lead into the INPUT jack (10). Connect the probes to the measuring points (in parallel). The measurement results will appear on the LCD display (4).

  • ТС21:
    When measuring AC current, press the FUNC button (5) to display the frequency.

  • ТС24/ТС32:
    When measuring AC current, the frequency is displayed on LCD simultaneously.

When the voltage is greater than 80V, the orange backlight will light up.

Voltage measurement (in mV) (TC24/TC32 only)
Turn the knob switch (13) to the position and press the FUNC button (5) to select the AC or DC voltage measurement function. Plug the black test lead into the COM jack (12) and the red test lead into the INPUT jack (10). Connect the probes to the measuring points (in parallel). The measurement results will appear on the LCD display (4). When measuring AC voltage, the frequency is displayed on LCD simultaneously.

Frequency/Duty measurement
Turn the knob switch (13) to the ****position and press the FUNC button (5) to select the frequency or duty measurement function. Plug the black test lead into the COM jack (12) and the red test lead into the INPUT jack (10). Connect the probes to the measuring points or in parallel to the circuit to be measured. The measurement results will appear on the LCD display (4).

  • When measuring high current, continuous measurement should be no longer than 15 seconds.
  • To avoid damaging the device, check the fuse before measuring and use the correct jack.

Resistance measurement

  • Turn the knob switch (13) to the position and select the appropriate range (TC21); or
  • Turn the knob switch (13) to the position and press the FUNC button (5) to select the resistance measurement function (TC24/TC32).

Plug the black test lead into the COM jack (12) and the red test lead into the INPUT jack (10). Connect the probes to the resistance measuring points of circuit or resistor (in parallel). The measurement results will appear on the LCD display (4).

When measuring resistance on the line, switch off the power supply and discharge all high-voltage capacitors.

Continuity test

  • Turn the knob switch (13) to the position (TC21); or
  • Turn the knob switch (13) to the position and press the FUNC button (5) to select the continuity measurement function (TC24/TC32).

Plug the black test lead into the COM jack (12) and the red test lead into the INPUT jack (10). Connect the probes to the circuit or component to be tested (in parallel). The measurement results will appear on the LCD display (4). If the resistance is <30Ω, an acoustic signal will be emitted continuously and the green indicator lights up at the same time. If the resistance is between 30Ω to 60Ω, the red indicator lights up.

When measuring continuity on the line, switch off the power supply and discharge all high-voltage capacitors.

Diode test

  • Turn the knob switch (13) to the position (TC21); or
  • Turn the knob switch (13) to the position and press the FUNC button (5) to select the diode measurement function (TC24/TC32).

Plug the black test lead into the COM jack (12) and the red test lead into the INPUT jack (10). Connect the black test probe to the cathode side and the red test probe to the anode side of the diode to be measured. If the polarity of the test probes is reverse to the diode polarity, the OL (Overload) will appear on the screen. The measurement results will appear on the LCD display (4).

When testing diodes on the line, switch off the power supply and discharge all high-voltage capacitors.

Capacitance measurement
Turn the knob switch (13) to the position. Plug the black test lead into the COM jack (12) and the red test lead into the INPUT jack (10). Connect the probes to the capacitance to be measured (in parallel). The measurement results will appear on the LCD display (4).

When measuring capacitance on the line, switch off the power supply and discharge all high-voltage capacitors.

Non-contact voltage detection (NCV)
Turn the knob switch (13) to the position and press the FUNC button (5) to select the NCV function. NCV (Non-contact voltage detection) is displayed. Slowly bring the NCV probe (1) close to the point to be detected. When the signal of a weak electromagnetic field is detected, the built-in buzzer will produce a slow beep sound, and the LED indicator (3) will glow green. When the signal of a strong electromagnetic field is detected, the built-in buzzer will produce a quick beep sound, and the LED indicator (3) will glow red.

When using this function, remove the test leads from the jacks.

Live detection
Turn the knob switch (13) to the position and press the FUNC button (5) to select the LIVE function. Live (Live detection) appears on the screen. Plug only the red test lead into the INPUT jack (10). Connect the red test probe to the conductor to be measured. When the signal of a weak electromagnetic field is detected, the built-in buzzer will produce a slow beep sound, and the LED indicator (3) will glow green. When the signal of a strong electromagnetic field is detected, the built-in buzzer will produce a quick beep sound, and the LED indicator (3) will glow red.

Temperature measurement
Turn the knob switch (13) to the position. Plug the black test lead of K-type thermocouple into the COM jack (12) and the red test lead into the INPUT jack (10). Connect the thermocouple probe to the object to be measured. The measurement results are shown in Celsius and Fahrenheit simultaneously.

When using the included K-type thermocouple, temperature measurement range is –40… +260°C / –40… +500°F.

Battery test (TC21 only)
The device allows you to check the actual battery voltage under a load of a certain resistance. Such assessment gives a more accurate battery condition status compared to a voltage measurement without a load. Turn the knob switch (13) to the battery test position and select the appropriate range (1.5V or 9V). Plug the black test lead into the COM jack (12) and the red test lead into the mA jack (11). Touch the positive pole with the red test lead and the negative pole with the black test lead. The measurement results (battery voltage) will appear on the display (4).

  • When measuring a battery in the 1.5V range, a 30Ω load resistance is used.
  • When measuring a battery in the 9V range, a 300Ω load resistance is used.

Specifications

  TC21 TC24 TC32
DC voltage, range 600mV / 6V / 60V / 600V / 1000V ±(0.5%+3) 200mV / 2V / 20V

/ 200V /

1000V ±(0.08%+5)

AC voltage, range| 6V / 60V / 600V /

750V ±(0.8%+5)

| 600mV / 6V / 60V / 600V /

750V ±(0.8%+5)

| 200mV / 2V / 20V / 200V /

750V ±(0.5%+25)

DC current, range 1| 60μA / 60mA / 600mA /

10A ±(1.2%+3)

| 600μA / 6000μA / 60mA /

600mA / 10A ±(1.2%+3)

| 200μA / 2000μA / 20mA /

200mA ±(0.5%+5)

DC current, range 2| —| —| 10A ±(0.8%+15)
AC current, range 1| 60mA / 600mA /

10A ±(1.5%+3)

| 600μA / 6000μA / 60mA /

600mA / 10A ±(1.5%+3)

| 200μA / 2000μA / 20mA /

200mA ±(0.8%+25)

AC current, range 2| —| —| 10A ±(1.0%+25)
Resistance, range 1| 600Ω / 6kΩ / 60kΩ / 600kΩ ±(1.0%+3)| 200Ω / 2kΩ / 20kΩ / 200kΩ /

2MΩ / 20MΩ ±(1.0%+15)

Resistance, range 2| 6MΩ / 60MΩ ±(1.5%+3)| 100MΩ ±(3.0%+25)
Capacitance, range 1| 10nF / 100nF / 1μF / 10μF /

100μF / 1mF ±(4.0%+5)

| 10nF / 100nF / 1μF / 10μF /

100μF / 1mF ±(4.0%+5)

| 2nF / 20nF / 200nF / 2μF / 20μF /

200μF / 2mF / 20mF ±(4.0%+50)

Capacitance, range 2| 10mF / 100mF ±(5.0%+5)| —
Frequency, range 1| 10Hz / 100Hz / 1kHz / 10kHz / 100kHz / 1MHz ±(1.0%+3)| 200Hz / 2kHz / 20kHz / 200kHz /

2MHz ±(1.0%+30)

Frequency, range 2| 10MHz ±(3.0%+3)| 10MHz ±(3.0%+30)
Duty, range| 1–99% ±(3.0%+3)| 1–99% ±(3.0%+30)
Temperature measurement range| –40… 1000°C (–40… 1832°F) ±(1.0%+3)
Counts| 6000| 20 000
Operating temperature range| 0… +40°C / 32… +104°F
Operating humidity range| 0–80% RH
Storage temperature range| –10… +60°C / +14… 140°F
Storage humidity range| 0–70% RH
Power supply| 2pcs alkaline AA batteries (1.5V)
Safety rating| EN61010-1, EN61010-2-030, EN61010-2-033, EN61326-1; Cat. III, 1000 V; Cat. IV, 600 V

The manufacturer reserves the right to make changes to the product range and specifications without prior notice.

Care and maintenance

Use the device only within the permitted range. Failure to follow these instructions may result in electric shock, fire and/or personal injury. When measuring voltage, do not exceed the operating conditions (see “Specifications”). When changing functions and ranges, always remove the test probes from the test points. Do not use the device if it is not working properly. Please note that the parameters of the power supply must comply with the technical characteristics of the device. Do not try to disassemble the device on your own for any reason. For repairs and cleaning of any kind, please contact your local specialized service center. Protect the device from sudden impact and excessive mechanical force. Do not use the device in aggressive atmosphere. Store the device in a dry cool place. Only use accessories and spare parts for this device that comply with the technical specifications. Never attempt to operate a damaged device or a device with damaged electrical parts! If a part of the device or battery is swallowed, seek medical attention immediately.

Battery safety instructions
Always purchase the correct size and grade of battery most suitable for the intended use. Always replace the whole set of batteries at one time; taking care not to mix old and new ones, or batteries of different types. Clean the battery contacts and also those of the device prior to battery installation. Make sure the batteries are installed correctly with regard to polarity (+ and –). Remove batteries from equipment that is not to be used for an extended period of time. Remove used batteries promptly. Never short-circuit batteries as this may lead to high temperatures, leakage, or explosion. Never heat batteries in order to revive them. Do not disassemble batteries. Remember to switch off devices after use. Keep batteries out of the reach of children, to avoid risk of ingestion, suffocation, or poisoning. Utilize used batteries as prescribed by your country’s laws.

Ermenrich Warranty

Ermenrich products, except for their accessories, carry a 5-year warranty against defects in materials and workmanship. All Ermenrich accessories are warranted to be free of defects in materials and workmanship for six months from the purchase date. The warranty entitles you to the free repair or replacement of the Ermenrich product in any country where a Levenhuk office is located if all the warranty conditions are met.

For further details, please visit: levenhuk.com/warranty

If warranty problems arise, or if you need assistance in using your product, contact the local Levenhuk branch.

Levenhuk USA 928 E 124th Ave. Ste D, Tampa, FL 33612, USA, +1 813 468-3001, contact_us@levenhuk.com

Levenhuk®, Ermenrich® are registered trademarks of Levenhuk Optics s.r.o. (Europe). 2006–2024 Levenhuk, Inc. All rights reserved.

ermenrich.com

20240724

FAQ

Q: What should I do if the device does not power on?
A: Ensure that the batteries are inserted correctly and try turning the device on again. If the issue persists, contact customer support for assistance.

Q: How do I change the measurement units on the multimeter?
A: Use the knob switch to select different measurement functions and units as needed. Refer to the user manual for specific instructions on unit changes.

Q: Can I use this multimeter to measure temperature?
A: Yes, you can measure temperature using the included K-type thermocouple. Follow the user manual guidelines for accurate temperature measurements.

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>