INSTRUMENT OPERATION
Insert the probe into the liquid being measured, the instrument reading will take a few seconds to stabilise.
CALIBRATION ADJUSTMENT – Caution: Do not adjust unless certain the instrument is out of specification. Compare against a reference thermometer/Ice Bath. Using the spanner, twist the lock-nut whilst holding the dial to correct any error. For further instructions, refer to the ETI website: etiltd.com /cal-adjustment.
GUARANTEE
This instrument has a one-year limited guarantee against manufacturing defects. Full details of liability are available at: etiltd.com/terms.
Product code | Range |
---|---|
800-801 | -40 to 70 °C |
800-802 | -10 to 110 °C |
800-803 | 0 to 250 °C |