eti 800801 Dial Thermometer Instruction Manual

June 3, 2024
eti

eti Logo 800801 Dial Thermometer
Instruction Manual
eti 800801 Dial Thermometer

INSTRUMENT OPERATION

Insert the probe into the liquid being measured, the instrument reading will take a few seconds to stabilise.

CALIBRATION ADJUSTMENT – Caution: Do not adjust unless certain the instrument is out of specification. Compare against a reference thermometer/Ice Bath. Using the spanner, twist the lock-nut whilst holding the dial to correct any error. For further instructions, refer to the ETI website: etiltd.com /cal-adjustment.

GUARANTEE

This instrument has a one-year limited guarantee against manufacturing defects. Full details of liability are available at: etiltd.com/terms.

Product code Range
800-801 -40 to 70 °C
800-802 -10 to 110 °C
800-803 0 to 250 °C

Rio Refugio 9648
Parque de Negocios ENEA,
Pudahuel, Santiago ~ CHILE
Tel. +56 2 28988221
www.yalitech.cl

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals