COREMORROW E53.D Series Piezo Controller User Manual
- June 3, 2024
- COREMORROW
Table of Contents
COREMORROW E53.D Series Piezo Controller
This document describes the following products:
- E53.D Servo 1 channel SGS sensor Software
DECLARATION
Declaration!
This user manual is an integrated user manual of the E53.D series
piezoelectric controller. Please read this user manual carefully before using
this controller. Follow the instructions in the manual during use. If there is
any problem, please contact us for technical support. If you do not follow
this manual or disassemble and modify the product yourself, the company will
not be liable for any consequences arising therefrom.
Please read the following to avoid personal injury and to prevent damage to this product or any other product connected to it. In order to avoid possible hazards, this product can only be used within the specified range.
Notice!
Do not touch any exposed ends of the product and its accessories. There is a
high voltage inside. Do not open the case without permission. Do not connect
or disconnect input, output, or sensor cables with power on. Please keep the
surface of E53.D clean and dry, don’t operate in a humid or static
environment. After use, the output voltage should be cleared to zero before
turning off the controller switch, such as switching the servo state to the
open-loop state.
Danger!
The piezoelectric power amplifier described in this manual is a high-voltage
device capable of outputting high currents, which can cause serious or even
fatal damage if not used properly. It is strongly recommended that you do not
touch any parts that connect to the high-voltage output. Special Note: If you
connect it with other products in addition to our company, please follow the
general accident prevention procedures.
Operating high-voltage amplification requires training professional operators.
Warning!
If the voltage exceeds the PZT’s tolerable range, it will cause permanent
damage to the PZT. Before adding voltage to the PZT poles, it must be ensured
that the positive and negative poles of the PZT are connected correctly and
the operating voltage is within the allowable range of this PZT.
Cautious!
E53.D housing should be installed on a horizontal surface in an area with a
3CM air flow area to prevent internal convection in the vertical direction.
Insufficient airflow can cause equipment to overheat or premature instrument
damage.
Security
Introduction
- Keep the surface of the piezo controller clean and dry.
- Do not operate in a humid or static environment.
- Piezo controllers are used to driving capacitive loads (eg piezo actuators).
- It should not be used in user manuals of other products of the same name.
- Pay special attention that it cannot be used to drive resistive or inductive loads.
- Piezo controllers could be used for static and dynamic operating applications.
Safety Instructions
Piezo controller is based on the national safety standard. Improper use may cause personal injury or damage to the piezo controller. The operator is responsible for the correct installation and operation of the piezo controller.
- Please read the user manual in detail.
- Please eliminate any faults and potential safety hazards caused by the faults.
If the protective ground wire is not connected or connected incorrectly, there will be a possibility of leakage. If you touch the E53.D piezo controller, it may cause serious or even fatal injuries. If the piezo controller housing is opened without permission, touching the live parts may cause electric shock, resulting in serious or even fatal injury or damage to the piezo controller.
- Only authorized professional technicians with corresponding qualifications could open the piezo controller.
- When opening the piezo controller housing, you need to disconnect the power plug.
- Do not touch any internal parts when operating under bare conditions.
Notes
- The contents in the user manual are all standard descriptions, and the customized parameters are not explained in detail in this manual.
- The latest user manual is available for download on CoreMorrow website.
- When operating the piezo controller, the user manual should be placed near the system for easy reference in time. If the user manual is missing or damaged, please contact CoreMorrow customer service department.
- Please promptly add all the information given in the manufacturer’s user manual, such as supplements or technical instructions.
- If your user manual is incomplete, a lot of important information will be missed, causing serious or fatal injury and causing property damage. Please read and understand the content in the user manual before installing and operating the piezo controller.
- Only professionals who are authorized to meet the technical requirements could install, operate, maintain and clean the piezo controllers.
Introduction
Features
- 1 channel small size
- 24V(20~30V)1.5A 36W
- Peak current 1A
- Ave current 60mA
- Unload bandwidth 10KHz
- Output short circuit protection.
Applications
- Driving piezo actuators
- Driving piezo objective scanners
Order information
- E53.D——Servo SGS sensor Analog/software input control
Accept customized according to requirements:
- 12bit gain/-20 ~ 120V output voltage(default)
- 15bit gain/-20 ~ 150V output voltage
Driving Principle
Driving Principle
- Average output(Sine wave operation mode)
- Pa ≈ Upp • Us • f• Cpiezo
- Pa=Average output[W]
- Upp=Peak and peak drive voltage [V]
- Us=Drive voltage[V]((Vs+)-(Vs-))
- Cpiezo=Piezo actuator capacitance[F]
- f=Operating frequency of the sine wave[Hz]
Appearance
Panel Introduction
No. | Function | Description |
---|---|---|
① | Power indicator | Green, lights when power’s on |
② | USB port | MicroUSB port |
③ | RS-232/422 | See interface pin definition |
④ | Analog input | Analog voltage input interface |
⑤ | Servo integral | Adjust step response |
⑥ | Sensor monitor | 0 ~ 10V |
⑦
|
Target
| Lights when the controlled displacement deviates from the target value
⑧| Zero| Zero adjustments of the sensor signal
⑨| Sensor connector| Piezo actuator sensor connector
⑩| Drive connector| Piezo driving connector
| Limit| Over-current indicator
RS-232/422 interface pin definition.
No. | Function | Description |
---|---|---|
1 | Signal source control switch | M: Signal source selection by communication |
instruction
D: Digital circuit as signal source
A: Using an external analog signal as a signal source
2| Open loop / Servo control switch| M: Open loop / Servo status controlled by
communication instructions
OFF: Switch to open loop state
ON: Switch to Servo state
Notes and suggestions
- E53.D cannot be used to drive inductive loads. If the inductive loads are driven, the product may be damaged.
- If there is no need, please do not twist the potentiometer easily.
Contact us
Harbin Core Tomorrow Science & Technology Co., Ltd.
- Tel: +86-451-86268790
- Email: info@coremorrow.com.
- Website: www.coremorrow.com.
- Address: Building I2, No.191 Xuefu Road, Nangang District, Harbin, HLJ, China.
CoreMorrow Official and CTO WeChat are below:
Documents / Resources
|
COREMORROW E53.D Series Piezo
Controller
[pdf] User Manual
E53.D Series, Piezo Controller, E53.D Series Piezo Controller, Controller
---|---