COREMORROW E53.D Series Piezo Controller User Manual

June 3, 2024
COREMORROW

COREMORROW E53.D Series Piezo Controller

COREMORROW-E53.D-Series-Piezo-Controller-product

This document describes the following products:

  • E53.D Servo 1 channel SGS sensor Software

DECLARATION

Declaration!
This user manual is an integrated user manual of the E53.D series piezoelectric controller. Please read this user manual carefully before using this controller. Follow the instructions in the manual during use. If there is any problem, please contact us for technical support. If you do not follow this manual or disassemble and modify the product yourself, the company will not be liable for any consequences arising therefrom.

Please read the following to avoid personal injury and to prevent damage to this product or any other product connected to it. In order to avoid possible hazards, this product can only be used within the specified range.

Notice!
Do not touch any exposed ends of the product and its accessories. There is a high voltage inside. Do not open the case without permission. Do not connect or disconnect input, output, or sensor cables with power on. Please keep the surface of E53.D clean and dry, don’t operate in a humid or static environment. After use, the output voltage should be cleared to zero before turning off the controller switch, such as switching the servo state to the open-loop state.

Danger!
The piezoelectric power amplifier described in this manual is a high-voltage device capable of outputting high currents, which can cause serious or even fatal damage if not used properly. It is strongly recommended that you do not touch any parts that connect to the high-voltage output. Special Note: If you connect it with other products in addition to our company, please follow the general accident prevention procedures.

Operating high-voltage amplification requires training professional operators.

Warning!
If the voltage exceeds the PZT’s tolerable range, it will cause permanent damage to the PZT. Before adding voltage to the PZT poles, it must be ensured that the positive and negative poles of the PZT are connected correctly and the operating voltage is within the allowable range of this PZT.

Cautious!
E53.D housing should be installed on a horizontal surface in an area with a 3CM air flow area to prevent internal convection in the vertical direction. Insufficient airflow can cause equipment to overheat or premature instrument damage.

Security

Introduction

  • Keep the surface of the piezo controller clean and dry.
  • Do not operate in a humid or static environment.
  • Piezo controllers are used to driving capacitive loads (eg piezo actuators).
  • It should not be used in user manuals of other products of the same name.
  • Pay special attention that it cannot be used to drive resistive or inductive loads.
  • Piezo controllers could be used for static and dynamic operating applications.
Safety Instructions

Piezo controller is based on the national safety standard. Improper use may cause personal injury or damage to the piezo controller. The operator is responsible for the correct installation and operation of the piezo controller.

  • Please read the user manual in detail.
  • Please eliminate any faults and potential safety hazards caused by the faults.

If the protective ground wire is not connected or connected incorrectly, there will be a possibility of leakage. If you touch the E53.D piezo controller, it may cause serious or even fatal injuries. If the piezo controller housing is opened without permission, touching the live parts may cause electric shock, resulting in serious or even fatal injury or damage to the piezo controller.

  • Only authorized professional technicians with corresponding qualifications could open the piezo controller.
  • When opening the piezo controller housing, you need to disconnect the power plug.
  • Do not touch any internal parts when operating under bare conditions.

Notes

  • The contents in the user manual are all standard descriptions, and the customized parameters are not explained in detail in this manual.
  • The latest user manual is available for download on CoreMorrow website.
  • When operating the piezo controller, the user manual should be placed near the system for easy reference in time. If the user manual is missing or damaged, please contact CoreMorrow customer service department.
  • Please promptly add all the information given in the manufacturer’s user manual, such as supplements or technical instructions.
  • If your user manual is incomplete, a lot of important information will be missed, causing serious or fatal injury and causing property damage. Please read and understand the content in the user manual before installing and operating the piezo controller.
  • Only professionals who are authorized to meet the technical requirements could install, operate, maintain and clean the piezo controllers.

Introduction

Features
  • 1 channel small size
  • 24V(20~30V)1.5A 36W
  • Peak current 1A
  • Ave current 60mA
  • Unload bandwidth 10KHz
  • Output short circuit protection.
Applications
  • Driving piezo actuators
  • Driving piezo objective scanners

Order information

  • E53.D——Servo SGS sensor Analog/software input control

Accept customized according to requirements:

  1. 12bit gain/-20 ~ 120V output voltage(default)
  2. 15bit gain/-20 ~ 150V output voltage

Driving Principle

COREMORROW-E53.D-Series-Piezo-Controller-fig-1

Driving Principle

  • Average output(Sine wave operation mode)
    • Pa ≈ Upp • Us • f• Cpiezo
  • Pa=Average output[W]
  • Upp=Peak and peak drive voltage [V]
  • Us=Drive voltage[V]((Vs+)-(Vs-))
  • Cpiezo=Piezo actuator capacitance[F]
  • f=Operating frequency of the sine wave[Hz]

Appearance

COREMORROW-E53.D-Series-Piezo-Controller-fig-2

Panel Introduction

COREMORROW-E53.D-Series-Piezo-Controller-fig-3

No. Function Description
Power indicator Green, lights when power’s on
USB port MicroUSB port
RS-232/422 See interface pin definition
Analog input Analog voltage input interface
Servo integral Adjust step response
Sensor monitor 0 ~ 10V

|

Target

| Lights when the controlled displacement deviates from the target value
⑧| Zero| Zero adjustments of the sensor signal
⑨| Sensor connector| Piezo actuator sensor connector
⑩| Drive connector| Piezo driving connector
 | Limit| Over-current indicator

RS-232/422 interface pin definition.COREMORROW-E53.D-Series-Piezo-
Controller-fig-5

No. Function Description
1 Signal source control switch M: Signal source selection by communication

instruction
D: Digital circuit as signal source
A: Using an external analog signal as a signal source
2| Open loop / Servo control switch| M: Open loop / Servo status controlled by communication instructions
OFF: Switch to open loop state
ON: Switch to Servo state

Notes and suggestions

  • E53.D cannot be used to drive inductive loads. If the inductive loads are driven, the product may be damaged.
  • If there is no need, please do not twist the potentiometer easily.

Contact us

Harbin Core Tomorrow Science & Technology Co., Ltd.

CoreMorrow Official and CTO WeChat are below:COREMORROW-E53.D-Series-
Piezo-Controller-fig-6

Documents / Resources