COREMORROW E70.D4S Series Piezo Controller User Manual
- June 3, 2024
- COREMORROW
Table of Contents
COREMORROW E70.D4S Series Piezo Controller
This document describes the following products:
- E70.D4S Piezo Controller SGS sensor 4 channels
DECLARATION
This user manual is an integrated user manual of the E70.D4S series piezoelectric controller. Please read this user manual carefully before using this controller. Follow the instructions in the manual during use. If there is any problem, please contact us for technical support. If you do not follow this manual or disassemble and modify the product yourself, the company will not be liable for any consequences arising therefrom. Please read the following to avoid personal injury and to prevent damage to this product or any other product connected to it. In order to avoid possible hazards, this product can only be used within the specified range.
Notice!
Do not touch any exposed ends of the product and its accessories. There is
a high voltage inside. Do not open the case without permission. Do not connect
or disconnect input, output, or sensor cables with power on. Please keep the
surface of E70.D4S clean and dry, don’t operate in a humid or static
environment. After use, the output voltage should be cleared to zero before
turning off the controller switch, such as switching the servo state to the
open-loop state.
Danger!
The piezoelectric power amplifier described in this manual is a high-
voltage device capable of outputting high currents, which can cause serious or
even fatal damage if not used properly. It is strongly recommended that you do
not touch any parts that connect to the high-voltage output. Special Note: If
you connect it with other products in addition to our company, please follow
the general accident prevention procedures. Operating high-voltage
amplification requires training professional operators.
Warning!
If the voltage exceeds the PZT’s tolerable range, it will cause permanent
damage to the PZT. Before adding voltage to the PZT poles, it must be ensured
that the positive and negative poles of the PZT are connected correctly and
the operating voltage is within the allowable range of this PZT.
Cautious!
E70.D4S housing should be installed on a horizontal surface in an area with
a 3CM air flow area to prevent internal convection in the vertical direction.
Insufficient airflow can cause equipment to overheat or premature instrument
damage.
Security
Features
- Keep the surface of E70.D4S clean and dry, don’t operate in a humid or static environment. E70.D4S is used to drive capacitive loads (such as piezoelectric ceramic actuators).
- E70.D4S cannot be used in the user manuals of other products with the same name. Pay attention to E70.D4S cannot be used to drive inductive loads.
- E70.D4S can be used for static and dynamic operating applications
- E70.D4S with SGS sensor can use servo operation mode.
Safety instructions
E70.D4S is based on nationally recognized safety standards. Improper use may
cause personal injury or damage it. The operator is responsible for the
correct installation and operation of it.
- Please read the user manual in detail.
- Please immediately eliminate any malfunctions and safety hazards caused by malfunctions. If the protective grounding wire is not connected or connected incorrectly, there will be the possibility of electric leakage. If you touch E70.D4S piezo controller, it can cause serious or even fatal injuries.
- Only authorized professionals could open the E70.D4S series controller. When opening the E70.D4S series controller, please disconnect the power plug. Please do not touch any internal parts when operating in an exposed state.
- If open E70.D4S privately, touching live parts may cause electric shock, resulting in serious or even fatal injury or damage to the E70.D4S series controller.
User Manual Notes
- The contents described in the user manual are standard product descriptions, special product parameters are not described in detail in this manual.
- When using the E70.D4S series piezo controller, the user manual should be placed near the system for easy reference in time. If the user manual is lost or damaged, please contact our customer service department.
- Please timely add all the information given in the manufacturer’s user manual, such as supplements or technical descriptions.
- If your user manual is incomplete, it will miss a lot of important information, cause serious or fatal injuries, and cause property damage. You have read and understood the contents of the user manual before installing and operating the E70.D4S series piezoelectric controller.
- Only authorized professionals who meet the technical requirements can install, operate, maintain and clean the E70.D4S series of digital piezoelectric controllers.
Features and Applications
E70.D4S series piezo controller is a 4-channel piezoelectric ceramic controller with a small size, high power, low power consumption, high bandwidth, and low ripple noise. The controller uses a dedicated operational amplifier circuit to ensure high-voltage and high-current output capability. By optimizing the sensing servo module, the accuracy and stability of control are improved. The reliable anti-interference design ensures the high-frequency response of the controller.
Series
Model
|
Description
---|---
E70.D4S
| Piezo controller, 4channels, SGS sensor, Software control and analog input control
Front Pane
Symbol | Function | Description |
---|
RS-232/422
|
D-SUB 9
| Set DIP switch/software to select control mode. Connect the computer with the controller interface module through RS-232/422 port access terminal to realize computer control.
USB
|
Micro- female
| Set DIP switch/software to select control mode. Connect the computer to the controller interface module through the USB port access terminal to realize computer control.
Power supply
|
SF812-S2
| Power connector socket. Connect via power adapter or DC power supply.
Switch| KCD1-102| Control the power on and off of the piezo controller.
M OFF ON 1
|
MSK-13C01
| Toggle switch position to switch open loop/servo control. Function description is shown in table below.
M OFF ON 2
|
MSK-13C01
| Toggle switch position to switch open loop/servo control. Function description is shown in table below.
M OFF ON 3
|
MSK-13C01
| Toggle switch position to switch open loop/servo control. Function description is shown in table below.
M OFF ON 4
|
MSK-13C01
| Toggle switch position to switch open loop/servo control. Function description is shown in table below.
M D A
|
MSK-13C01
| Toggle switch position to switch digital/analog control. The function
description is shown in the table below.
Switch| Position| Function
---|---|---
1
| ON| CH1 output control servo
OFF| CH1 output control open loop
M| Software control optional open loop/servo
2
| ON| CH2 output control servo
OFF| CH2 output control open loop
M| Software control optional open loop/servo
3
| ON| CH3 output control servo
OFF| CH3 output control open loop
M| Software control optional open loop/servo
4
| ON| CH3 output control servo
OFF| CH3 output control open loop
M| Software control optional open loop/servo
5
| A| Output control analog input
D| Output control digital input
M| Software control optional digital/analog input
Checking
E70.D4S controller has been carefully checked for electrical and mechanical
aspects before shipment. When you receive the device, unpack and inspect the
surface of the system for any obvious signs of damage. If it is damaged, it
may be damaged during transportation, please contact our customer service
department in time. Check whether the accessories are complete according to
the packing list. Please keep the original packaging materials for subsequent
maintenance and use.
installment
Installation Precautions
Note! Improper installation of the E70.D4S piezoelectric controller may cause
personal injury or damage to the E70.D4S piezoelectric controller!
Installation and use of E70.D4S should be close to the power source so that
the power plug can be easily and quickly disconnected from the main power
source. Use included power cord to connect the E70.D4S piezoelectric
controller system. If the power cord provided by our company must be replaced,
please use a power cord with a large enough size and effective grounding.
Ensure ventilation
Note! Overheating of equipment due to high temperature may damage the E70.D4S
controller!
- Ensure that the cooling area of the controller is sufficiently cooled.
- Ensure that there is adequate ventilation equipment.
- Keep ambient temperature to a non-critical level(<50℃ ).
- The temperature of cooling surface of the controller is>50 ℃, it is recommended to take external heat dissipation measures to improve the stability of the controller.
Connect power
Use the power adapter (output range +20V~+30V/3A) to connect to the power
supply interface of the E70.D4S power supply.
Cable connection
When the power supply is disconnected, connect the PZT&Sensor cable to the
E70. D4S controller interface. Note that the number on the piezoelectric
actuator corresponds to the number of the controller. In analog control mode,
when the signal source (signal generator, analog signal source, DA control
card) output is 0, connect the SMB cable to the SMB interface of the E70.D4S
controller. Connect to the computer control mode of the PC, connect to the PC
through the cable connection USB interface or RS-232/422 interface socket.
Operating
Control mode optional(Output control)
Toggle Switches | Setting |
---|---|
M D A | A |
Digital mode
Toggle Switches | Setting |
---|---|
M D A | D |
Servo mode optional(Output control)
Toggle Switches | Servo mode | Setting |
---|
1/2/3/4
| Open loop| OFF
Servo| ON
Software mode optional(Software control)
Toggle Switches | Setting |
---|---|
1 | M |
2 | M |
3 | M |
4 | M |
5 | M |
Parameter
Environmental conditions
The using environment of the E70.D4S controller:
Environmental conditions | Condition description |
---|---|
Application | For room use only |
Environment humidity
| Highest relative humidity 80%, temperature can reach 30℃ Highest relative
humidity 50%, temperature can reach 40℃
Operating temperature| 0 ℃ — +50 ℃
Storage temperature| –10 ℃ — +85 ℃
Driving Principle
Pin Definition
No.| Pin Definition|
---|---|---
1| +10V
2| Sensor signal 1+
3| Sensor signal 1-
4| GND
5| PZT GND
6| PZT OUTPUT
Monitor Out
No.| Pin Definition|
---|---|---
1| Sensor output 1
2| Sensor output 2
3| Sensor output 3
4| Sensor output 4
5| GND
RS-232/422
No.| Pin Definition|
---|---|---
1| not used
2| RS-232 TxD
3| RS-232 RxD
4| not used
5| GND
6| RS-422 RxD+
7| RS-422 RxD-
8| RS-422 TxD+
9| RS-422 TxD-
Power Calculation
- Average output(Sine wave operation mode)
- Pa ≈ Upp • Us • f• Cpiezo
- Pa=Average output[W]
- Upp=Peak and peak drive voltage
- [V] Us=Drive voltage[V]
- ((Vs+)-(Vs-))
- Cpiezo=Piezo actuator capacitance[F]
- f=Operating frequency of the sine wave[Hz]
Maintenance, storage, transportation
Cleaning measures
Note! The PCB board of the function module in the E07.D4S system is an ESD
(electrostatic discharge) sensitive device. Take precautions against any
static build-up of these devices before use to avoid contact with circuit
component leads and PCB wiring. Before touching any electronic components, the
body first touches the grounding conductor to discharge static electricity,
ensuring that any type of conductive particles (metal, dust or debris, pencil
lead, screws) enter the device. Be careful not to drop the equipment when
cleaning, to avoid any form of mechanical shock! Disconnect the power plug of
the E07.D4S system before cleaning. Prevent cleaning fluid and any liquid from
entering the system module to avoid short circuits. For the surface of the
system chassis and the front panel of the module, please do not use an organic
solvent for surface wiping.
Transportation and storage
This product is packed in a carton. Transportation must be carried out under
product packaging conditions, and direct rain and snow, direct contact with
corrosive gases and strong vibrations should be avoided during transportation.
The instrument can be transported under various conditions of normal
transportation and should avoid dampness, load, collision, extrusion,
irregular placement and other adverse conditions during transportation.If the
instrument is not used for a long time, the instrument should be packaged and
stored. The instrument should be stored in a non-corrosive atmosphere and in a
well-ventilated, clean room. In the process of transportation, storage and
use, attention should be paid to fire prevention, shockproofing,
waterproofing, and moisture proof.
Contact us
Harbin Core Tomorrow Science & Technology Co., Ltd.
Tel: +86-451-86268790
Email: info@coremorrow.com
Website: www.coremorrow.com
Address: Building I2, No.191 Xuefu Road, Nangang District, Harbin, HLJ, China
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>