eyc-tech THS88MAX Dew Point Industrial Transmitter Instruction Manual

June 27, 2024
eyc-tech

eyc-tech THS88MAX Dew Point Industrial Transmitter

eyc-tech THS88MAX Dew Point Industrial Transmitter

Security considerations

Please read this Specification carefully, prior to use of this, and keep the manual properly, for timely reference.
Solemn Statement:
This product can not be used for any explosion-proof area.
Do not use this product in a situation where human life may be affected.
eyc-tech will not bear any responsibility for the results produced by the operators!

Warning!

  • Installation and wiring must be performed by qualified personnel in accordance with all applicable safety standards.
  • This product must be operated under the operating conditions specified in manual to prevent equipment damages.
  •  Please using the product under the ordinary pressure, or it will influence safe problem.
  • This product must be operated under the operating condition specified in this manual to prevent equipment damages.
  • This product must be operated under the normally atmospheric condition to prevent equipment damages.
  • To prevent products damage, always disconnect the power supply from the product before performing any wiring and installation.
  • All wiring must comply with local codes of indoor wiring and electrical installation rules.
  • Please use crimp type terminal.
  • To prevent personal injury, do not touch the moving part of product in operation.
  • It may cause high humidity atmosphere during the product was breakdown. Please take safety strategy

Dimension

Dimension

※ Standard ─ M12-5PIN 2M waterproof cable

Diagram

Diagram

*Please make sure the product and the device which connect with RS-485 are on common ground, avoid damaged product.

Accessories for measuring atmospheric dew point

  • Environment pressure affects the value of dew point. When environment pressure is higher, water vapor condensate easier. Thus, dew point value in high pressure environment is higher than atmospheric dew point.
  • In process line, pressure could change and fluctuate easily, and leads to inconstant reading value. To avoid this problem, eyc-tech suggest user to measure dew point at atmospheric pressure (atmospheric dew point).
  • To help user measuring atmospheric dew point easily, eyc-tech provide related accessories, which including installation base for dew point transmitter, pressure regulator and flowmeter …etc. User can select accessories depends on installation. Please ask our sales personnel for further information.
    Part number| Description
    ---|---
    BASE-THS-001| Installation base(1/2“PT for transmitter), SUS304, Connection for air inlet and outlet: 1/4”PT
    BASE-THS-002| Installation base(1/2“PF for transmitter), SUS304, Connection for air inlet and outlet: 1/4”PT
  • A schematic of installation is shown below. Using pressure regulator to regulate pressure to 1 atm, and adjust airflow to 1~5 LPM to have stable reading of atmospheric pressure.
    Accessories for measuring atmospheric dew point

Software and calibration operation step

Application Program statement

User may download the configuration software on eyc-tech web site. Please decompress the application prior to execute it. Operating System requirements:
above Windows XP. Other application program requirements: above Microsoft Office 2003
Application Program statement

Setting RS-485 connection

  1. Connect product to PC via RS-485 cable

  2. Execute “THS UI”

  3. Click “Interface > Config”
    Setting RS-485 connection

  4. Select the corresponding values of com port as fallowing :
    a. Port:Check Come Port
    b. Baud Rate
    c. Data Frame
    d. Timeout
    e. Retry
    f. Station ID(Default 1)
    Setting RS-485 connection

  5. Click “Apply”

  6. Connect successfully
    a. Show value and trend chart of Dew point temperature
    b. Show “Open Port, Read successful”
    Setting RS-485 connection

Scan RS-485 connection

※Use scan function to connect when forgetting the connection information or having more facilities .

  1. Connect the product to PC via RS-485 cable

  2. Execute “THS UI”

  3. Click “Interface > Config”
    Scan RS-485 connection

  4. Select the corresponding values of com port as fallowing:
    a. Port:
    b. RS-485
    Scan RS-485 connection

  5. Click “Scan” to execute connection facilities

  6. Scan connection facilities and set up
    a. Select Station ID
    b. Click “CLOSE AND EXPORT”
    Scan RS-485 connection

  7. Click “Apply”
    Scan RS-485 connection

  8. Connect successfully
    a. Show values and trend chat Dew point Temperature
    b. Show “Open port, Read successful”
    Scan RS-485 connection

Setting RS-485 Mod Bus Protocol

  1. Setting RS-485 connection step as step 5.1

  2. Click “Setting”
    Setting RS-485 ModBus Protocol

  3. Select Modbus Protocol parameter
    a. Station ID:1~247
    b. Baud Rate:9600, 19200, 38400, 57600, 115200
    c. Data Frame:None-8Bit-1Stop, None-8Bit-2Stop, Even-8Bit-1Stop, Even- 8Bit-2Stop, Odd-8Bit-1Stop, Odd-8Bit-1Stop
    Setting RS-485 ModBus Protocol

  4. Click “Apply”

  5. Execute connection as step 5.2 or 5.3 again

Display and save data

  1. Show Data: Click ”Display”
    Display and save data

  2. Icon function statements
    | Clear the chart records| | Save the data measuring when the system start connecting before clinking the Export icon
    ---|---|---|---
    Display and save data| Change the chart style| | Show/Not show the statement of Y axis
    | Select the OUTPUT channel| | Show/Not show the statement of X axis
    | Set line color chosen from OUTPUT| | Show/Not show the statement of Y secondary axis
    | Snap chart| | Show/ Not show chart
    | Show/Not show measuring data
    | Adjust time range of X axis
    | Adjust time range of Y axis

  3. Setting time interval of record
    a. File > Log Interval
    b. Select time interval of record
    Display and save data

  4. Save/Log measuring data
    a. Log measuring range: Save the data measuring when the system start connecting before clinking the Export icon
    a -1. Click Display > Export
    Display and save data
    a-2. Appoint path and Key in file name > save
    1. If file name is some as the path name, the original file will be over write.
    Display and save data
    b. Log measuring data: Log the data which is on from start or off
    b-1. Display > Log(OFF)
    Display and save data
    b-2. Appoint path and Key in file name > save > Log (ON)
    ※1. If file name is some as the path name, the original file will be over write.
    Display and save data

Choose parameter of Output

  1. Click “Output”
    Choose parameter of Output

  2. Select relative parameters of Output1
    a. Measures
    b. Responding Rate
    c. Voltage or current analog range
    d. Upper and Lower point of Output
    Choose parameter of Output

  3. Click “Apply”
    Choose parameter of Output

  4. Select relative parameters of Output2
    a. Measures
    b. Responding Rate
    c. Upper and Lower point of Output
    Choose parameter of Output

  5. Click “Apply”

Convert 4-20mA to 0-10V

  1. Select Current to 0-20mA in UI.
    Choose parameter of Output

  2. Please Connect 500 ohms in parallel at the output.
    Diagram

Inspection and maintenance

  1. Maintenance
    Since this product is inspected and calibrated for high accuracy at the factory before shipment, no calibration on the installation site is necessary when this product is installed. For inspection and maintenance follow the instructions below:

    • Periodic inspection
      Periodically inspect this product for its sensing accuracy, and clean the cover. Set the period between inspections based on atmospheric dust and other contaminants in the installation environment.
  2. Troubleshooting

    • Sensor maintenance
      Do not damage sensor surface during the maintenance process.

    • Troubleshooting
      If any problem occurs during operation, refer to the table below for appropriate solutions

Problem Cleck items Soluations
  • No output
  • Unstable output

|

  • Disconnected wiring
  • Loose wiring
  • Power supply voltage
  • Sensor damages

|

  • Re-perform wiring

  • Crew on terminal tightly or replace
    wires

  • Replace the sensor

  • Slow response to
    output

  • Errow in output

|

  • Moisture /condensation on
    the product

  • Check installed location

  • Check installed angle

  • Check dust and contamination on the sensor

|

  • Remove the sensor and filter. Dry
    power-off state sensor in clean air
    seasoning

  • Refer to the section

  • Align measurement head with flow
    direction

  • Cleaning the filter

  • Changing the filter

  • Calibrate

  • Replace the sensor

CUSTOMER SUPPORT

eyc-tech Measuring Specialist
enhance your capability with sensor technology
Air flow | Humidity | Dew point | Differential pressure | Liquid flow
Temp. | Pressure | Level | Air quality | Signal meter
Tel.: 886-2-8221-2958
Web: www.eyc-tech.com
e-mail: info@eyc-tech.com
www.eyc-tech.comLogo