hygiena KIT230203 Maize Detection Kit Instructions

June 25, 2024
Hygiena

hygiena-LOGO

hygiena KIT230203 Maize Detection Kit

hygiena-KIT230203-Maize-Detection-Kit-PRODUCT

Product Information

Specifications

  • Product Name: PCR kit for the qualitative detection of Bt11 DNA
  • Product No.: KIT230203
  • Application: Food testing
  • Method: Real-time PCR
  • Intended Use: For in vitro use only

Product Usage Instructions

Introduction

  • This PCR kit is designed for the qualitative detection of Bt11 DNA in food samples using real-time PCR instruments. It is essential for compliance with regulations regarding genetically modified organisms (GMOs) in food and feed products.

Contents

  • The kit includes all necessary components for conducting the PCR assay, including primers, probes, and enzymes.

Additional Materials Required

  • Users will need access to a real-time PCR instrument, PCR tubes, pipettes, and appropriate laboratory equipment.

General Precautions

  • Follow standard laboratory practices and wear appropriate personal protective equipment when handling the kit components.
  • Ensure proper storage conditions to maintain kit stability.

Data Analysis

  • After running the PCR assay, analyze the results using the software provided with the real-time PCR instrument. Interpret the results based on the presence or absence of Bt11 DNA in the samples.

Troubleshooting

  • If encountering issues during the assay, refer to the troubleshooting section of the user manual for guidance on resolving common problems.

Stability and Storage

  • Store the kit components according to the specified storage conditions mentioned in the user manual to ensure stability and optimal performance.

FAQ

Q: Can this kit be used for GMO detection in crops other than maize?

A: This kit is specifically designed for the detection of Bt11 DNA in maize samples and may not be suitable for other GMO targets.

Q: Is this kit suitable for quantitative analysis?

A: No, this PCR kit is intended for the qualitative detection of Bt11 DNA and is not optimized for quantitative measurements.

Revision A, March 2024

  • PCR kit for the qualitative detection of Bt11 DNA using real-time PCR instruments.

Product No. KIT230203

  • Kit for 50 reactions for a maximum of 48 samples Store at -15 to -25 °C

Introduction

  • Many countries worldwide have implemented legislation for the use, cultivation, and labeling of foodstuffs containing genetically modified organisms (GMOs). These regulations allow the usage of GMOs under certain conditions, often including a defined threshold for labeling or when the import and use of GMOs are prohibited. Thus, reliable methods for the detection and identification of GMOs in food and feed are required.
  • With the foodproof® SL GMO product line, Hygiena® Diagnostics offers a wide range of easy and reliable assays for the detection of GMOs. The foodproof SL GMO Detection Kits allow fast, safe, and easy detection of food and feed samples.

Intended Use

  • The foodproof SL GMO Bt11 Maize Detection Kit is designed to detect the GM-maize Bt11 event-specific gene in various processed foods, raw materials, feed, seeds, etc.
  • This kit provides a real-time PCR Master Mix with enzyme components and the specific primer/probe set for rapid testing by real-time PCR, as well as the Internal Control (IC) system for reliable results.

Principle of PCR detection

  • The foodproof SL GMO Bt11 maize detection assay is a qualitative, duplex real-time PCR test for the detection of the GM maize Bt11 specific gene and the Internal Control (IC) using specific primers and probes labeled with fluorescent dyes. The target sequences are detected through FAM and HEX (VIC) channels, respectively.
  • The primer and probe mixture provided is based on the so-called TaqMan® principle. During PCR amplification, forward and reverse primers hybridize to the target DNA. A fluorogenic probe is included in the same reaction mixture which consists of an oligonucleotide labeled with a 5’-reporter dye and a downstream 3’-quencher. During PCR amplification, the probe is cleaved and the reporter dye and quencher are separated. The resulting increase in fluorescence can be detected through a range of real-time PCR platforms.
  • The monitoring of the fluorescence intensities during the real-time PCR allows the detection of accumulating products without re-opening the reaction tubes after the PCR run.
  • The kit minimizes contamination risk and contains all reagents needed for detection (except for PCR-grade H2O).

Internal Amplification Control

  • This kit contains an Internal Positive Control (IC) as a PCR inhibition Control. The IC allows the user to detect and control possible PCR inhibition.
  • The IC reagents are included in the primer/probe mixture and the IC is co-amplified with target DNA from the sample. The results can be visualized in the HEX (VIC) channel.

Carry-over prevention using UNG systems

  • The foodproof SL GMO Bt11 Maize Detection Kit utilizes the UNG system. Carry-over contamination between PCR reactions can be prevented by including uracil-N-glycosylase (UNG) in the reaction mix.
  • UNG can only prevent carry-over from PCR reactions that include deoxyuridine triphosphate (dUTP) in the original PCR reaction.

Contents

This kit is intended for 50 reactions, including controls.

Table 1: Kit Contents

Reagent Cap Label Volume Description
2x real-time PCR Master Mix 2xM 625 µL Buffer containing dNTPs, MgCl2, UNG,

and Taq DNA polymerase
Primer / Probe Mixture| Bt11| 200 µL| Primer/ probe mixture:

·         GM Maize Bt11-specific primer and probe

·         IC-specific primer and probe

·         IC DNA

Control DNA| C| 50 µL| Positive control DNA

Additional Materials, Reagents, and Devices Required

  • Disposable powder-free gloves and laboratory coat
  • Pipettors (0.5 to 10 µL, 2 to 20 µL, 20 to 200 µL, 200 to 1,000 µL)
  • Sterile aerosol-barrier pipette tips
  • Ice or benchtop cooler
  • Vortex mixer
  • Clean bench area or PCR box
  • Tabletop centrifuge with rotor for 2 mL reaction tubes
  • Real-time thermal cycler with FAM and HEX (VIC) detection channels
  • Disposable polypropylene microcentrifuge tubes for PCR
  • PCR-grade H2O
  • For DNA Extraction: foodproof® Sample Preparation Kit III

General Precautions

  • Store extracted positive material (samples, controls, and other amplicons) away from all other reagents and add to the reaction mix in a separate area.
  • Thaw all components thoroughly on ice before starting the experiment.
  • When thawed, mix the components and centrifuge briefly.
  • Do not pipette by mouth.
  • Do not eat, drink, smoke, apply cosmetics, or handle contact lenses in laboratory work areas.
  • Do not use a kit beyond its expiration date.
  • Safety Data Sheets (SDS) can be found at www.hygiena.com/documents.
  • Use disposable gloves, laboratory coats, and eye protection while handling samples and reagents. Thoroughly wash hands afterward.
  • Dispose of all samples and unused reagents in compliance with local regulations.
  • Specimens should be considered potentially infectious and handled in a biological cabinet under Biosafety Level 2 or other appropriate biosafety practices.
  • Clean and disinfect all sample or reagent spills using a disinfectant such as 0.5% sodium hypochlorite or other suitable disinfectant.
  • Avoid contact of specimens and reagents with the skin, eyes, and mucosa. If contact occurs with skin, eyes, or mucosa, immediately flush with water and seek medical attention.
  • Use of this product should be limited to personnel trained in laboratory techniques of DNA amplification.
  • To avoid carry-over contamination with PCR product or control DNA, please note the following points:
  1. Be careful not to contaminate the Primer/Probe Mixture and 2x real-time PCR Master Mix with other PCR products or Control DNA through pipetting. To prevent contamination, the use of aerosol-barrier tips is recommended.
  2. Open and close all sample tubes carefully. Avoid splashing or spraying PCR samples.
  3. It is important to have designated lab areas where PCR reactions are set up, preferentially separated in space from the areas where PCR reactions are analyzed.
  4. The laboratory process must be one-directional; it should begin in the Extraction Area and move to the Amplification and Detection Area. Do not transport samples, equipment, and reagents to the areas where you performed previous steps.

Sampling and handling

Sample Collection

  • Various processed foods, raw materials, feed, and seed samples are routinely examined.

Sample Storage

  • The assay sensitivity can be reduced if you routinely freeze the samples before testing or store them for an extended period. Avoid repeated freezing and thawing of samples, which may lead to DNA degradation and decreased sensitivity.

Nucleic Acid Extraction

  • Carry out DNA isolation according to the extraction kit’s product instructions. For more information, please see www.hygiena.com.

Protocol

DNA Isolation

  • Hygiena Diagnostics provides sample preparation kits suitable for all kinds of foods and raw materials.
  • (See 5. “Additional Required Materials, Reagents, and Devices”)

Preparing the PCR

  • To prevent the risk of contamination with foreign DNA, we recommend that all experiment steps be performed in a PCR cleanroom or separated environment area. Aerosol-barrier pipette tips are recommended for each step.

Thawing the Kit Components

  • The use of ice or a benchtop cooler is recommended during experiments to maintain enzyme activity.

Prepare Reaction Master Mix

  • Each reaction has a total volume of 25 µL; the volume of the DNA sample is 5 µL.

Prepare the reaction mixture according to Table 2 below.

Table 2: PCR reaction mixture

Composition Volume
Primer / Probe Mixture 4 µL
2x real-time PCR Master Mix 12.5 µL
H2O PCR-grade 3.5 µL
Total 20 µL

Add 5 µL of extracted DNA sample into the tube.

Prepare Control Amplification Reactions

  • CONTROL+ Positive control amplification: Add 5 µL of Control DNA instead of sample DNA.
  • CONTROL- Negative control amplification: Add 5 µL of PCR-grade H2O instead of sample DNA

Mixing

  • Mix the reagents in the PCR reaction tubes by tapping a minimum of 5 times. Briefly centrifuge the tubes to remove any air bubbles or drops inside the cap.

Amplification

  • Program your real-time PCR instrument according to the manufacturer’s manual.
  • Create a temperature-time profile on your instrument as follows in Table 3.

Table 3: Temperature Time Profile

Temperature Time Cycle
50 °C* 2 min 1
95 °C 10 min 1
95 °C 15 seconds 45
60 °C** 1 min
  • UDG activation step inhibits contamination by PCR product.
  • Detect the fluorescence at this step.

Data analysis

The fluorescence curves are analyzed in FAM and HEX (VIC) fluorescence detection channels (see Table 4). You can predict the presence or absence of the target gene in your samples by analyzing the real-time PCR results.

Table 4: Specific Detection on Fluorescence Channel

Target Gene Fluorophore
Bt11 FAM
IC HEX (VIC)

Interpretation of Results

  • The signal is considered to be positive if the corresponding fluorescence accumulation curve crosses the threshold line.
  • Results are accepted as relevant if both positive and negative amplification controls pass.
  • IC: When amplifying a target sample with a high copy number, the IC may not produce an amplification plot. This does not invalidate the test and should be interpreted as a positive experimental result.

Table 5: Interpretation of Results

 | Positive Control| Negative Control| Bt11| IC| Interpretation
---|---|---|---|---|---


FAM

| HEX (VIC)
Case 1| +| –| +| +| BT11 gene is detected.
Case 2| +| –| +| -*
Case 3| +| –| –| +| BT11 gene is not detected.
Case 4| +| –| –| –| Invalid result; retest
Case 5| +| +| +/-| +/-
Case 6| –| +/-| +/-| +/-

  • Detection of the Internal Amplification Control in the respective channel is not required for a positive result.
  • A high copy number of the target gene can lead to reduced or absent Internal Amplification Control signal.

Troubleshooting

Situation Possible cause Recommendation
Negative control samples are positive. Carry-over contamination ·   Exchange

all critical solutions.

·   Repeat the analysis of all tests with fresh aliquots of all reagents.

·   Take measures to detect and eliminate the source of contamination.

No signal is detected for amplification positive controls.| Incorrect programming of the real-time PCR instrument.| ·   The PCR should be repeated after checking the programming of instruments, storage conditions, and the expiration date.
The kit reagents have expired.
Kit components have not been stored according to the manufacturer’s instructions.
· Incorrect PCR reaction

· Pipetting errors

· Omitted reagents

| ·   The PCR should be repeated after checking for the correct pipetting scheme and reaction setup.
No signal is detected for IC in the HEX (VIC) channel and the Bt11-specific gene in the FAM channel.| PCR inhibitors are present at a high concentration.| ·   DNA extraction should be repeated.

Stability and Storage

Store the kit at -15 to -25 °C through the expiration date printed on the label.

Specifications

  • Sensitivity
    • Limit of detection (LOD) at 0.1%
  • Specificity
    • 100% exclusivity for approximately 100 non-target strains

Quality control

In compliance with the Federal State Institution of Science “Central Research Institute of Epidemiology” ISO 13485 – certified Quality Management System, each lot of foodproof SL GMO Bt11 Maize Detection Kit has been tested against predetermined specifications to ensure consistent product quality.

Ordering information

Product Order No. # Tests
foodproof SL GMO Bt11 Maize Detection Kit KIT230203 50 reactions
foodproof Sample Preparation Kit III KIT230174 50 reactions

Supplementary Information

Ordering Information

  • Hygiena Diagnostics offers a broad range of reagents and services. For a complete overview and more information, please visit our website at www.hygiena.com.

Trademarks

  • foodproof®, microproof®, vetproof®, ShortPrep®, StarPrep®, RoboPrep® and LyoKit® are registered trademarks of Hygiena Diagnostics GmbH.
  • Hygiena® is a registered trademark of Hygiena. Other brand or product names are trademarks of their respective holders.
  • Other brand or product names are trademarks of their respective holders.

Contact and Support

  • If you have questions or experience problems with this or any other product of Hygiena Diagnostics GmbH, please contact our Technical Support staff (www.hygiena.com/support). Our scientists commit themselves to providing rapid and effective help. We also want you to contact us if you have suggestions for enhancing our product performance or using our products in new or specialized ways. Such customer information has repeatedly proven invaluable to us and the worldwide research community.

Reference Number

  • The reference number and original Hygiena Diagnostics GmbH article number: Z 720 01

Change Index

  • Version 1, April 2014
  • The first version of the package insert.
  • Revision A, March 2024
  • Rebranding and new layout.
  • Z 720 01 20 -> INS-KIT230203-RevA
  • Hygiena®
  • Camarillo, CA 93012
  • USA
  • diagnostics.support@hygiena.com
  • Manufactured by
  • Hygiena Diagnostics GmbH Hermannswerder 17
  • 14473 Potsdam
  • Germany
  • www.hygiena.com

Documents / Resources

| hygiena KIT230203 Maize Detection Kit [pdf] Instructions
KIT230203 Maize Detection Kit, KIT230203, Maize Detection Kit, Detection Kit
---|---

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals