Dynalabs Ltd. provides this publication “as is” without warranty of any kind, express or implied, including but not limited to, the implied warranties of merchantability or fitness for a particular purpose. This document is subject to change without notice, and should not be construed as a commitment or representation by Dynalabs Ltd. This publication may contain inaccuracies or typographical errors. Dynalabs Ltd. will periodically update the material for inclusion in new editions. Changes and improvements to the product described in this manual may be made at any time.

Introduction

Capacitive accelerometers are based on proven micro-electro-mechanical systems (MEMS) technology. These capacitive accelerometers are reliable and long-term stable. These sensors are Differential Ended type DC response sensors. The advantages of these sensors are their outstanding temperature stability, external noise immunity and their lightweight. These sensors feature standard reliable aluminum housing with protection class IP68. Steel housing is also possible.
Dynalabs LN accelerometers provide Low noise – high resolution with an outstanding noise performance from 9 to 680 μg/√Hz. These accelerometers provide a wide frequency range (±5%) from 250 Hz to 1,500 Hz.

DYN-C-1000-LN sensors offer the following options;

  • Custom Cable Length
  • Custom Housing Material
  • Custom Connector
  • Base plate

General Information

Unpacking and Inspection
Dynalabs products provide adequate protection for undamaged products to be transported. Document the damages that occur indirectly during the transport and contact the customer representative.

System Components
The DYN-C-1000-LN has the following components:

  • MEMS Sensor
  • Calibration Certificate
  • Product Manual

Specifications

Table 1: Specifications datasheet

| | 1002LN| 1005LN| 1010LN| 1030LN| 1050LN| 1100LN| 1200LN
---|---|---|---|---|---|---|---|---
Full scale acceceleration| (g)| ±2| ±5| ±10| ±30| ±50| ±100| ±200
Sensitivity| (mV/ g)| 1,350| 540| 270| 90| 54| 27| 13.5
Frequency

range (±5%)

| (Hz)| 700| 1,150| 2,000| 2,300| 2,700| 2,900| 2,500
Non-linearity (full scale)| (%)| 0.15| 0.15| 0.15| 0.15| 0.15| 0.15| 0.15
Noise (in band)| (μg/

√Hz)

| 9| 21| 40| 100| 180| 340| 680
Bias temperature| (mg/

°C)

| ±0.2| ± 0.5| ± 1| ± 3| ± 5| ± 10| ± 20
Shock survivability| (g)| 2,500| 2,500| 2,500| 3,000| 3,000| 3,000| 3,000

Environmental
Table 2: Environmental Specifications datasheet

Protection Level IP 68
Operating Voltage 5 V – 20 V
Operating Temperature -40 °C to +100 °C
Operating Current Consumption mA 7 mA
Isolation Case isolated

Physical
Table 3: Physical Specifications datasheet

Sensing Element MEMS Capacitive
Housing Material Aluminum or Steel
Connector (Optional) D-Sub 9 or 15 pin, Lemo, Binder
Mounting Adhesive or screw mount
Base plate (Optional) Aluminum or Steel

__

Weight (without cable)

| 10 g (aluminum)

20 g (steel)

Outline Drawing
The dimensional properties of DYN-C-1000-LN sensors are given below ;DynaLabs-DYN-C-1000-LN-Uniaxial-Capacitive-Accelerometers-
fig-1

Operation and Installation

General
The general sensor connector configuration is given below; Cable Code/Pin Configuration:

  • Red : V + Power supply voltage +5 to +20 VDC
  • Black : Ground Power GND
  • X : Yellow : Signal(+) Positive, analog output voltage signal for differential mode.
    Blue : Signal(-) Negative, analog output voltage signal for differential mode.

WARNING
Never connect the power supply and/or the power ground to yellow and/or blue cables.
Never connect the power supply to the power ground. Always use a clean power source and check the voltage range.

Sensor Static Calibration Verification

Using gravity, voltage values are measured in the + and – gravity directions, providing a value of ±1 g. The measurement should be made as follows;

  • When the sensitivity value of 1000LN series sensors is used with the data acquisition system, the sensor shows +1 g with the effect of gravity in the direction of the arrow sign.
  • When the sensor is in the opposite direction of the arrow, it shows -1 g with the effect of gravity.

Using gravity, the voltage values that provide 1 g in the + and – directions are measured and compared with the catalog value. The calibration value should be close to the catalog value with 10% tolerance. Sensor catalog sensitivity values are given in Table 1.

Declaration of Conformity

This declaration of conformity is issued under the sole responsibility of the manufacturer. The product(s) are developed, produced and tested according to following EC- directives:

  • 2014/35/EU – Low Voltage Directive (LVD)
  • 2006/42/EU – Machinery Safety Directive
  • 2015/863/EU – RoHS Directive

Applied standards:

  • EN 61010-1:2010
  • EN ISO 12100:2010
  • MIL-STD-810-H-2019 (Test Methods: 501.7 – High Temperature, 502.7 – Low Temperature, 514.8 – Vibration, 516.8 – Shock)